Skip to content

GSK Vaccines GmbH Form 483, March 2024

FDA issued this Form 483 to GSK Vaccines GmbH in Marburg, Germany (FEI 3001314775) at the end of a biologics inspection that ended March 6, 2024. It lists 4 observations. FDA classified the inspection Voluntary Action Indicated (VAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Oct 26, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Mar 6, 2024
Biologics · Vaccines and Allergenic Products
FDA's outcome
VAI
Voluntary Action Indicated
Posted by FDA
Oct 26, 2026
964 days after the inspection
Facility
FEI 3001314775
Marburg, Germany

The 4 observations on this Form 483

Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.

  1. 1

    Aseptic controls not established and followed

    Procedures designed to prevent microbiological contamination of drug products purporting to be sterile are not established and followed. Specifically, a. On 2/26/2024 during (b) (4) operation of MenA Vaccine batch (b) (4) using (b) (4) unit, and after exchanging (b) (4) located approximately (b) (4) vials, the grade(b)(4) operators did not reject the (b) (4) product vials located directly below the intervention. As per your SOP instruction# 9000085445-02, before carrying out a (b) (4) change, the line must be (b) (4) so that there are no vials in the (b) (4). However, the operators did not ensure that the (b) (4) line was (b) (4) to the intervention. In addition, there is not documentation available in the executed batch records to ensure that the operators (b) (4) the (b) (4) lines or reject the (b) (4) product vials (b) (4) the interventions. The firm has used the (b) (4) unit to (b)(4) RabAvert (batch (b) (4), Expiry (b) (4)) and MenA Vaccines (batch (b) (4) and Expiry (b) (4)) for US market. b. On 3/6/2024, during set up operations of (b) (4) unit for MenA batch (b) (4), I observed the shoulder of Grade(b)(4) operator touch the Grade(b)(4) side of the (b) (4). No sanitization was performed on the Grade (b)(4) side of the (b) (4) before closing the (b) (4).

  2. 2

    Inadequate investigation of Aluminum Hydroxide discrepancy

    There is a failure to thoroughly review any unexplained discrepancy whether or not the batch has been already distributed. Specifically, investigation related to (b) (4) of Aluminum Hydroxide (b) (4) is not adequate. Aluminum Hydroxide is an adjuvant used for Infanrix, (b) (4), Pediarix, and Boostrix Vaccine products for US market. On 06Apr2023, (b) (4) and (b) (4) were observed on the (b) (4) of (b) (4), which is used for the (b) (4) of Aluminum Hydroxide bulk material (Deviation # 200991257). The root cause was identified as (b) (4) of (b) (4) due to (b) (4) contact with Aluminum Hydroxide suspension. Based on the investigation, you have identified that a contact time of more than (b) (4) can lead to (b) (4) parts. However, the investigation was not extended to other (b) (4) surfaces to re-evaluate the established dirty hold times of 72 hours for fixed equipment, such as but not limited to (b) (4), and (b) (4) for (b) (4) parts, such as not limited to (b) (4) utilized for Aluminum Hydroxide(b) (4) production.

  3. 3

    Deficient segregation of expired and quarantine materials

    Separate or defined areas to prevent mix-ups are deficient. Specifically, your firm failed to discard and/or transfer (b) (4) materials to a physical quarantine location to prevent the materials being used beyond (b) (4) dates. (b) (4) materials were observed to be stored in your (b) (4) on 2/27/2024. There was no clear demarcation between (b) (4) materials and (b) (4) or (b) (4) materials in your (b) (4). More than (b) (4) material batches (material lots) had been identified that are currently stored at the same location with your (b) (4) or (b) (4) materials in your (b) (4). In addition, your written procedure LSOP-9000056691-12, "Execution of Destructions" requires that, "materials to be destroyed are (b) (4) in room (b) (4)" until physical destruction. There are more than (b) (4) material batches (material lots) identified with status "(b) (4)" that are currently stored in your (b) (4).

  4. 4

    Incorrect expiry on labels and shipping documents

    Documents and (b) (4) used for distribution of the (b) (4) drug products are deficient. Specifically, your firm failed to ensure correct (b) (4) were applied to (b) (4) drug product containers before release. The final (b) (4) on your containers that hold (b) (4) drug products contain incorrect expiration date. The expiration date on the (b) (4) was not based on the associated stability studies. Furthermore, your documentations that accompanied with the (b) (4) drug product shipment, including the packing lists, contain incorrect information. The documents do not bear an expiration date determined by the associated stability studies and do not match with the expiration date on the container (b) (4) that hold (b) (4) drug products.

Other Form 483s FDA posted for this facility

Every Form 483 FDA posts under FEI 3001314775, newest first.

All inspections at this facility →

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.