Grifols Diagnostic Solutions, Inc.
4560 Horton St, Emeryville, California, United States
FEI 1000124057
FDA has 12 inspections on file for Grifols Diagnostic Solutions, Inc. in Emeryville, California (FEI 1000124057), from 2010 to 2025. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA's Data Dashboard shows Form 483 citations for this inspection, but FDAgent does not hold their text. One of the two FOIA requests for this site has asked FDA for this inspection's 483.
The most recent OAI inspection ended on May 24, 2012. No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 12, 2010–2025
- 1 OAI · 8 VAI · 3 NAI
- Inspections with 483 observations shown
- 2 of 12
- 1 with citations published, text not held
- Form 483s posted by FDA
- ↗
- Requested under FOIA
- 2 requests, latest Feb 10, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
12 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Bioresearch MonitoringCitations published; text not held
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood ProductsPosted 483 · 1 observation
Read this Form 483 on its own page →Form 483 PDF on fda.gov
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.
- Observation 1
Inadequate process validation and continued verification
Licensed Antigen Component Manufacturer for IVDs Your firm has not appropriately ruled out process validation for manufacturing of licensed antigen HCV c100-3 instead of using a verification approach, and your firm has not analyzed process data across all antigen types using your firm's CPV (b) (4) Reports to assess that manufacturing processes are in a qualified/validated state. A. Process Validation Master Plan, GDSE-VAL-VMP-000005, V21 & 22, indicates in part in section 9.3 that a verification approach is an alternative to re-validation in which manufacturing process performance is monitored and evaluated. Your firm used a process verification approach to address (b) (4) for HCV c100-3 antigen (b) (4) manufacturing (yeast (b) (4), and yeast (b) (4) (b) (4), as indicated by Process Verification Report GDSE-VAL-REP-000973. The Master Plan also indicates in part that a risk assessment will determine if a verification is appropriate. This was done as per Risk Assessment for (b) (4) HCV c100-3 (b) (4) in (b) (4) GDSE-MFG-RSKA-000652. The risk assessment indicates that "Run (b) (4) of (b) (4) runs used for evaluation of risk) did not pass certain assays. This apparent failure was not addressed to substantiate the decision to perform a process verification (i.e., (b) (4) run) instead of a full re-validation (i.e., PPQ) of the HCV c100-3 process (b) (4) B. Your firm's SOP, Process Validation Stage 3 - Continued Process Verification, GDSE-MFG-SOP-000071, Ver 17 and 18 requires, in part, the trend analysis of process output parameters and critical quality attributes across all your firm's licensed and unlicensed antigen products, as part of (b) (4) process verification reviews to ensure that manufacturing processes (b) (4) are in a qualified/validated state. The last (b) (4) review was documented to have occurred in the (b) (4) of 2023.
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood ProductsCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood ProductsCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood ProductsCitations not published
- NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
- OAISignificant objectionable conditions; FDA recommended regulatory actionBiologics · Blood and Blood ProductsCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
- NAINo objectionable conditions that need follow-upDrugs · Drug Quality AssuranceNone
- NAINo objectionable conditions that need follow-upDrugs · Drug Quality AssuranceNone
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood ProductsCitations not published
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on Feb 10, 2026. Each one asked for a Form 483.
Inspection ending Oct 31, 2025VAI
1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Apr 30, 2026. FDAgent holds no copy that FDA posted.
- Feb 10, 2026FDA-FOIA-2026-1389Form 483· 1 of 5 sitesRecords released
Inspection ending Dec 13, 2024VAI
1 request. In the inspection history above. FDA has posted its Form 483.
- Apr 7, 2025FDA-FOIA-2025-3166Form 483· 1 of 4 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 1 section of 21 CFR in FDA’s published citations for 1 of this facility’s 12 inspections.
- 21 CFR 820.25Personnel1 of 1
Observations read from posted 483 PDFs carry no CFR section of FDA’s, so they are not counted here.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1000124057 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1000124057. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.