Skip to content

Grifols Diagnostic Solutions Inc. Form 483, December 2024

FDA issued this Form 483 to Grifols Diagnostic Solutions Inc. in Emeryville, California (FEI 1000124057) at the end of a biologics inspection that ended December 13, 2024. It lists 1 observation. FDA classified the inspection Voluntary Action Indicated (VAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Oct 26, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Dec 13, 2024
Biologics · Blood and Blood Products
FDA's outcome
VAI
Voluntary Action Indicated
Posted by FDA
Oct 26, 2026
682 days after the inspection
Facility
FEI 1000124057
Emeryville, California

The 1 observation on this Form 483

Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.

  1. 1

    Inadequate process validation and continued verification

    Licensed Antigen Component Manufacturer for IVDs Your firm has not appropriately ruled out process validation for manufacturing of licensed antigen HCV c100-3 instead of using a verification approach, and your firm has not analyzed process data across all antigen types using your firm's CPV (b) (4) Reports to assess that manufacturing processes are in a qualified/validated state. A. Process Validation Master Plan, GDSE-VAL-VMP-000005, V21 & 22, indicates in part in section 9.3 that a verification approach is an alternative to re-validation in which manufacturing process performance is monitored and evaluated. Your firm used a process verification approach to address (b) (4) for HCV c100-3 antigen (b) (4) manufacturing (yeast (b) (4), and yeast (b) (4) (b) (4), as indicated by Process Verification Report GDSE-VAL-REP-000973. The Master Plan also indicates in part that a risk assessment will determine if a verification is appropriate. This was done as per Risk Assessment for (b) (4) HCV c100-3 (b) (4) in (b) (4) GDSE-MFG-RSKA-000652. The risk assessment indicates that "Run (b) (4) of (b) (4) runs used for evaluation of risk) did not pass certain assays. This apparent failure was not addressed to substantiate the decision to perform a process verification (i.e., (b) (4) run) instead of a full re-validation (i.e., PPQ) of the HCV c100-3 process (b) (4) B. Your firm's SOP, Process Validation Stage 3 - Continued Process Verification, GDSE-MFG-SOP-000071, Ver 17 and 18 requires, in part, the trend analysis of process output parameters and critical quality attributes across all your firm's licensed and unlicensed antigen products, as part of (b) (4) process verification reviews to ensure that manufacturing processes (b) (4) are in a qualified/validated state. The last (b) (4) review was documented to have occurred in the (b) (4) of 2023.

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.