Protherics UK Ltd. Form 483, June 2022
FDA issued this Form 483 to Protherics UK Ltd. in Llandysul, United Kingdom (FEI 3001417487) at the end of a biologics inspection that ended June 23, 2022. It lists 3 observations. FDA classified the inspection Voluntary Action Indicated (VAI).
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: FDA's Electronic Reading Room, posted Oct 26, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.
- Inspection ended
- Jun 23, 2022
- Biologics · Blood and Blood Products
- FDA's outcome
- VAI
- Voluntary Action Indicated
- Posted by FDA
- Oct 26, 2026
- 1,586 days after the inspection
The 3 observations on this Form 483
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.
- 1
Deficient QC testing procedures and controls
QC testing procedures are deficient. Specifically, A. There is a lack of study to support the (b)(4) of sample (b)(4) for endotoxin analysis per (b)(4)-13.160, v7, section 6.12.3. B. (b)(4) media are not being compared to previously tested and approved batch of the same media type. Instead, they are being compared to (b)(4) obtained from (b)(4) per (b)(4) (b)(4)-13.015, v15, section 6.7.1. C. There is a lack of (b)(4) under (b)(4) conditions despite (b)(4) qualification requires the assessment of a (b)(4) recovery per (b)(4)-13.002, v7, section 2. D. (b)(4) prepared (b)(4) cultures are not qualified prior to release for use. (b)(4) cultures are being used as (b)(4) for (b)(4) and (b)(4). E. Drug samples are not (b)(4) in analytical sequences to verify (b)(4) control. A (b)(4) is analyzed (b)(4) sample and does not confirm that the (b)(4) instrument is in a calibrated state upon the completion of the analysis. As a result, drug product results were invalidated and retested, due to a (b)(4) calibration test failure. For example, i). Quality Event 55795 - The (b)(4) calibration for the Quality Control (b)(4) was conducted per SOP (b)(4) entitled, "(b)(4) (b)(4)" on June 29, 2020, and failed the (b)(4) accuracy test during the calibration. Consequently, all CroFab and DigiFab (b)(4) drug product release tests that were performed (b)(4) (b)(4) were invalidated and retested. All sample retests had passing results. F. (b)(4) are not included at the end of analytical sequences to ensure (b)(4) control for the following (b)(4) analyzes. i). 13.059, entitled, "Determination of Drug Product Protein content by (b)(4)" ii). (b)(4) 13.0028, entitled, "Product speciation - (b)(4)" iii) (b)(4)-13.110, entitled, "Measurement of DigiFab biological activity (Potency) & specific content by"
- 2
Deficient facility qualification testing in Grade area
(b)(4) testing (b)(4) conducted in - Grade (b)(4) area in September 2021 as part of the facility qualification was deficient. Specifically, A. There was a lack of consistent supply of (b)(4) for an adequate assessment of (b)(4) HEPA (b)(4) or the absence of (b)(4) as required per (b)(4)-04.013, v1, section 6.2. B. Aseptic operations carried out during routine processing including worst-case activities which cause the most potential for product contamination were not adequately simulated under dynamic conditions as required per (b)(4)-04.013, v1, section 6.4.
- 3
Deficient environmental monitoring system for classified areas
Classified (b)(4) areas are deficient regarding the system for monitoring environmental conditions. Specifically, A. There is a lack of documented risk based assessments for identifying the environmental monitoring (EM) sample locations. B. (b)(4)-13.009, entitled "Viable Environmental Monitoring Procedures", v27 is deficient in that the above procedures do not define specific EM sample locations when no processing is scheduled.
Other Form 483s FDA posted for this facility
Every Form 483 FDA posts under FEI 3001417487, newest first.
- Form 483, August 2024Inspection ended Aug 13, 2024 · Posted Oct 26, 2026VAI
Ask FDAgent about this Form 483
FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.
Related
About this data
- The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
- The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
- FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
- FDAgent is independent of FDA.