Laboratorio Banco Vida
Upr Comprehensive Cancer Center Grounds Of Medical Center, San Juan, Puerto Rico, United States
FEI 3015767533
FDA has 2 inspections on file for Laboratorio Banco Vida in San Juan, Puerto Rico (FEI 3015767533), from 2019 to 2021. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 3 citations from this inspection, all under 21 CFR Part 1271, the rules for human cell and tissue products. They are listed below.
The most recent OAI inspection ended on Sep 18, 2019. FDA issued a warning letter on (CBER). No court action or import refusal is filed under this FEI.
One regulation was cited at more than one inspection: 21 CFR 1271.160(b)(6).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 2, 2019–2021
- 1 OAI · 1 VAI · 0 NAI
- Inspections with 483 observations shown
- 2 of 2
- Form 483s posted by FDA
- ↗
- Requested under FOIA
- 1 request, latest Sep 3, 2025
- Warning letters
- · CBER ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Deviations--evaluation, cause, corrective action
Investigation of deviations related to core CGTP requirements did not include corrective action(s) to address the deviation and prevent recurrence.
Reduced risk representation
Written representation that processing methods reduce the risk of transmission of communicable disease by HCT/Ps was not based on a fully validated or verified process.
Procedures to meet core CTGP
Procedures appropriate to meet core CGTP requirements for all steps that you perform in the manufacture of HCT/Ps were not documented.
- OAISignificant objectionable conditions; FDA recommended regulatory actionBiologics · Human Cellular, Tissue, and Gene TherapiesPosted 483 · 6 citations
Read this Form 483 on its own page →Form 483 PDF on fda.gov
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Records incomplete
Records were not detailed as necessary to provide a complete history of work performed.
Validation & approval--established procedures
Processes with results which could not be fully verified by inspection and tests, were not validated and approved according to established procedures.
Deviations--Investigation, documenting, trending
The quality program does not include the investigation of HCT/P deviations relating to core CTGP requirements.
Monitoring-microorganisms where appropriate
Environmental conditions were not monitored for microorganisms.
Size, construction, location--prevent mix-ups etc.
Facilities were not of suitable size and construction to prevent contamination of HCT/Ps with communicable disease agents and ensure the orderly handling of HCT/Ps without mix-ups.
Show 1 more citation
Adequate control, proper conditions
Environmental conditions existed in which contamination or cross contamination of HCT/Ps or equipment could occur, and environmental conditions were not adequately controlled and proper conditions for operations were not provided.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Sep 3, 2025. It asked for a Form 483.
Inspection ending Sep 18, 2019OAI
1 request. In the inspection history above. FDA has posted its Form 483.
- Sep 3, 2025FDA-FOIA-2025-10513Form 483· 1 of 2 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 6 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.
- 21 CFR 1271.160Establishment and maintenance of a quality program2 of 2
- 21 CFR 1271.230Process validation2 of 2
- 21 CFR 1271.180Procedures1 of 2
- 21 CFR 1271.190Facilities1 of 2
- 21 CFR 1271.195Environmental control and monitoring1 of 2
- 21 CFR 1271.270Records1 of 2
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCBER · BiologicsFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3015767533 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3015767533. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.