North Shore University Hospital Northwell Health
300 Community Dr, Manhasset, New York, United States
FEI 2475795
FDA has 17 inspections on file for North Shore University Hospital Northwell Health in Manhasset, New York (FEI 2475795), from 2009 to 2026. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA posted the Form 483 from that inspection. FDA published 3 citations from this inspection, all under 21 CFR Part 606, the manufacturing rules for blood and blood components. They are listed below. Two of the six FOIA requests for this site have asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
10 regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Form 483 PDF ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 17, 2009–2026
- 0 OAI · 12 VAI · 5 NAI
- Inspections with 483 observations shown
- 12 of 17
- Form 483s posted by FDA
- ↗
- Requested under FOIA
- 6 requests, latest FY2023
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
17 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood ProductsPosted 483 · 3 citations
Read this Form 483 on its own page →Form 483 PDF on fda.gov
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Person performing, test results, interpretation
Records fail to identify the person performing the work so as to provide a complete history of the work performed.
Establish, maintain and follow manufacturing SOPs
Written standard operating procedures including all steps to be followed in the compatibility testing and distribution of blood and blood components for allogeneic transfusion were not always followed.
Required records
Failure to maintain quality control and general records.
- NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene TherapiesNone
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Who must report - manufacturer
Failure to submit a biological deviation report.
Establish, maintain and follow manufacturing SOPs
Written standard operating procedures including all steps to be followed in the storage and distribution of blood and blood components for allogeneic transfusion were not always followed.
Required records
Failure to maintain quality control and general records.
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Human Cellular, Tissue, and Gene Therapies3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Determination based on screening and testing
HCT/P donors were not determined to be eligible based on the results of donor screening and testing.
Testing, screening, donor eligibility procedures
You did not establish and maintain procedures for screening and determining the eligibility of HCT/P donors.
Risk factors, clinical evidence
Donors were not screened by a review of relevant medical records for risk factors of communicable disease agents and diseases.
- NAINo objectionable conditions that need follow-upBiologics · Vaccines and Allergenic Products1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Risk factors, clinical evidence
Donors were not screened by a review of relevant medical records for risk factors of communicable disease agents and diseases.
- NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
- NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Person performing, test results, interpretation
Records fail to be as detailed as necessary so as to provide a complete history of the work performed.
Required records
Failure to maintain general records.
Establish, maintain and follow manufacturing SOPs
Written standard operating procedures including all steps to be followed in the collection, processing, storage and distribution of blood and blood components for allogeneic transfusion were not always maintained and followed.
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Human Cellular, Tissue, and Gene Therapies3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Reproductive cells/ tissues- general
Donors of reproductive cells or tissues not recovered by a method that ensures freedom from contamination were not screened by a review of relevant medical records for of communicable diseases of the genitourinary tract.
Procedures for all other requirements
Procedures for all steps performed in the screening and determining of donor eligibility of HCT/Ps were not established , maintained , defined , documented , implemented and followed .
Accurate, indelible, legible
Donor eligibility records are not accurate , indelible and legible .
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Arm preparation
The phlebotomy site is not prepared by a method that gives maximum assurance of a sterile container of Whole Blood.
Calibration/Inspection/Checking not done
Routine checking of automatic equipment is not performed according to a written program designed to assure proper performance.
Establish, maintain and follow manufacturing SOPs
Written standard operating procedures including all steps to be followed in the collection, processing, compatibility testing, storage, and distribution of blood and blood components for homologous transfusion and autologous transfusion are not always maintained and followed.
Required records
Failure to maintain quality control and general records.
- NAINo objectionable conditions that need follow-upDrugs · Bioresearch MonitoringNone
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Human Cellular, Tissue, and Gene Therapies3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Testing, screening, donor eligibility procedures
Procedures for all steps performed in the screening and determining of donor eligibility of HCT/Ps were not established, maintained, defined, documented, implemented, and followed.
Reproductive cells/ tissues- general
Donors of reproductive cells or tissues not recovered by a method that ensures freedom from contamination were not screened by a review of relevant medical records for risk factors of communicable diseases of the genitourinary tract.
Accurate, indelible, legible
Donor eligibility records are not accurate, indelible, and legible.
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Person performing, test results, interpretation
Records fail to identify the person performing the work so as to provide a complete history of the work performed.
Adequate identification and handling of test samples
Failure to adequately provide for identification and handling of all test samples so that they are accurately related to its donor.
Calculations performed are in the records
Laboratory records do not include a record of all calculations performed in connection with the test.
Concurrent documentation
Records are not concurrently maintained with the performance of each significant step in the collection and processing of each unit of blood and blood components so that all steps can be clearly traced.
Establish, maintain and follow manufacturing SOPs
Written standard operating procedures including all steps to be followed in the collection, processing, and distribution of blood and blood components for homologous transfusion are not always maintained and followed.
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Human Cellular, Tissue, and Gene Therapies3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Risk factors, clinical evidence
Donors were not screened by a review of relevant medical records for risk factors of communicable disease agents and diseases.
Risks associated with xenotransplantation
Donors were not screened by a review of relevant medical records for disease risks associated with xenotransplantation.
Procedures for all other requirements
Procedures for all steps performed in the testing, screening, and determining of donor eligibility of HCT/Ps were not established, maintained, defined, documented, implemented, and followed.
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Storage temps./agitation
Failure to store platelets immediately after resuspension at 20 to 24 degrees Celsius with continuous gentle agitation.
Establish, maintain and follow manufacturing SOPs
Written standard operating procedures including all steps to be followed in the collection, processing, storage, and distribution of blood and blood components for homologous transfusion and autologous transfusion are not always maintained and followed.
General requirements - storage
Failure to store and maintain the Red Blood Cells between 1 and 6 degrees Celsius immediately after processing.
Procedure Deviations Recorded and Justified
Deviations from written production and process control procedures are not justified.
Required records
Failure to maintain quality control and general records.
Show 2 more citations
Calculations performed are in the records
Laboratory records do not include a record of all calculations performed in connection with the test.
Arm preparation
The phlebotomy site is not prepared by a method that gives maximum assurance of a sterile container of Whole Blood.
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Corrective action
Failure to maintain a record of corrective action when the quality control testing for platelets does not meet the prescribed requirements.
Legibility and indelibility
Records are illegible and not indelible.
Event files -- failure to establish and maintain
MDR event files have not been established and maintained.
SOPs not followed / documented
Written production and process control procedures are not followed in the execution of production and process control functions and documented at the time of performance.
Reports and information documentation
The written MDR procedure does not include documentation and recordkeeping requirements for all Medical Device Reports and information submitted to FDA and device manufacturers.
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Human Cellular, Tissue, and Gene TherapiesCitations not published
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Six requests to FDA under the Freedom of Information Act named this site, most recently on FY2023. Each one asked for a Form 483.
Inspection ending Mar 11, 2026VAI
2 requests. In the inspection history above. FDA has posted its Form 483.
- Aug 15, 2026FDA-FOIA-2026-7903Form 483· 1 of 4 sitesRecords released
- May 12, 2026FDA-FOIA-2026-4870Form 483· 1 of 4 sitesRecords released
Inspection ending Apr 26, 2023VAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- FY2023FDA-FOIA-2023-4325Form 483Closed
Inspection ending Jul 20, 2022VAI
2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Nov 10, 2025. FDAgent holds no copy that FDA posted.
- Jul 22, 2025FDA-FOIA-2025-9057Form 483· 1 of 2 sitesClosed
- Dec 12, 2024FDA-FOIA-2024-10943Form 483· 1 of 2 sitesRecords released
Inspection ending Mar 12, 2020NAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Jan 8, 2025FDA-FOIA-2025-275Form 483· 1 of 3 sitesClosed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 16 sections of 21 CFR in FDA’s published citations for 12 of this facility’s 17 inspections, with how many of them cited each.
- 21 CFR 606.160Records7 of 12
- 21 CFR 606.100Standard operating procedures6 of 12
- 21 CFR 1271.75How do I screen a donor?5 of 12
- 21 CFR 1271.47What procedures must I establish and maintain?4 of 12
- 21 CFR 211.100Written procedures; deviations2 of 12
- 21 CFR 211.194Laboratory records2 of 12
- 21 CFR 640.4Collection of the blood2 of 12
- 21 CFR 640.25General requirements2 of 12
Show all 16
- 21 CFR 1271.55What records must accompany an HCT/P after the donor-eligibility determination is complete; and what records must I retain?2 of 12
- 21 CFR 211.68Automatic, mechanical, and electronic equipment1 of 12
- 21 CFR 600.14Reporting of biological product deviations by licensed manufacturers1 of 12
- 21 CFR 606.140Laboratory controls1 of 12
- 21 CFR 640.11General requirements1 of 12
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 12
- 21 CFR 803.18What are the requirements for establishing and maintaining MDR files or records that apply to me?1 of 12
- 21 CFR 1271.50How do I determine whether a donor is eligible?1 of 12
Other FEIs recorded as North Shore University Hospital Northwell Health
Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.
North Shore University Hospital Northwell HealthFEI 3004011569Manhasset, New York, United States
Last inspected 2026-03-04NAI
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 2475795 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 2475795. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.