21 CFR 640.11: General requirements
FDA's inspection data lists 16 Form 483 citations of 21 CFR 640.11 at 14 facilities, from March 2011 to March 22, 2019.
Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.
- Form 483 citations
- 16
- At 15 inspections
- Facilities cited
- 14
- Distinct FEIs
- Most recent citation
- Mar 22, 2019
- Blood Bank of Alaska, Inc.
- Warning letters
- 0
- State a violation of it
The regulation
The current text of 21 CFR 640.11 from the Electronic Code of Federal Regulations.
§ 640.11 General requirements.
(a) Storage. Immediately after processing, the Red Blood Cells shall be placed in storage and maintained at a temperature between 1 and 6 °C.
(b) Inspection. The product shall be inspected immediately after separation of the plasma, periodically during storage, and at the time of issue. The product shall not be issued if there is any abnormality in color or physical appearance or if there is any indication of microbial contamination.
21 CFR Part 640: Additional Standards for Human Blood and Blood Products, Subpart BRead on eCFR
Citations by year
Form 483 citations of 21 CFR 640.11 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).
Peak: 4 in 2011
| Year | Form 483 citations |
|---|---|
| 2011 | 4 |
| 2012 | 3 |
| 2013 | 1 |
| 2014 | 2 |
| 2015 | 3 |
| 2018 | 2 |
| 2019 | 1 |
How FDA words this citation
FDA's short descriptions and citation text on these 16 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.
Short descriptions
- General requirements - storage13 citations
- General requirements - inspection3 citations
Citation text
- “Failure to store and maintain the Red Blood Cells between 1 and 6 degrees Celsius immediately after processing.”7 citations
- “Failure to store the Red Blood Cells between 1 and 6 degrees Celsius immediately after processing.”5 citations
- “Failure to inspect Red Blood Cells immediately after separation of the plasma, periodically during storage, and at the time of issue.”2 citations
- “Failure to inspect Red Blood Cells immediately after separation of the plasma, periodically during storage, and at the time of issue and prevent issuance if Red Blood Cells are abnormal in color, physical appearance, or indicative of microbial contamination.”1 citations
Show 1 more
- “Failure to maintain the Red Blood Cells between 1 and 6 degrees Celsius immediately after processing.”1 citations
Facilities cited most recently
The latest inspections where FDA cited 21 CFR 640.11, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.
Show 4 more
Often cited at the same inspection
Of the 15 inspections where FDA cited 21 CFR 640.11, the share where it also cited each of these sections.
- 21 CFR 606.100Standard operating procedures10 (67%)
- 21 CFR 606.160Records8 (53%)
- 21 CFR 640.4Collection of the blood5 (33%)
- 21 CFR 606.60Equipment4 (27%)
- 21 CFR 640.25General requirements4 (27%)
- 21 CFR 640.33 (20%)
Ask FDAgent about 21 CFR 640.11
FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.
Related
About this data
- Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
- Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
- A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 640.11; FDAgent quotes that sentence exactly.