Skip to content

21 CFR 606.60: Equipment

FDA's inspection data lists 317 Form 483 citations of 21 CFR 606.60 at 255 facilities, from December 2008 to August 26, 2026.

Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.

Form 483 citations
317
At 299 inspections
Facilities cited
255
Distinct FEIs
Most recent citation
Aug 26, 2026
Oklahoma Blood Institute
Warning letters
0
State a violation of it

The regulation

The current text of 21 CFR 606.60 from the Electronic Code of Federal Regulations.

§ 606.60 Equipment.

(a) Equipment used in the collection, processing, compatibility testing, storage and distribution of blood and blood components shall be maintained in a clean and orderly manner and located so as to facilitate cleaning and maintenance. The equipment shall be observed, standardized and calibrated on a regularly scheduled basis as prescribed in the Standard Operating Procedures Manual and shall perform in the manner for which it was designed so as to assure compliance with the official requirements prescribed in this chapter for blood and blood products.

(b) Equipment that shall be observed, standardized and calibrated with at least the following frequency, include but are not limited to:

Equipment | Performance check | Frequency | Frequency of calibration

Temperature recorder | Compare against thermometer | Daily | As necessary.

Refrigerated centrifuge | Observe speed and temperature | Each day of use | Do.

Hematocrit centrifuge | | | Standardize before initial use, after repairs or adjustments, and annually. Timer every 3 mo.

General lab centrifuge | | | Tachometer every 6 mo.

Automated blood-typing machine | Observe controls for correct results | Each day of use |

Hemoglobinometer | Standardize against cyanmethemoglobin standard | ......do |

Refractometer | Standardize against distilled water | ......do |

Blood container scale | Standardize against container of known weight | ......do | As necessary.

Water bath | Observe temperature | ......do | Do.

Rh view box | ......do | ......do | Do.

Autoclave | ......do | Each time of use | Do.

Serologic rotators | Observe controls for correct results | Each day of use | Speed as necessary.

Laboratory thermometers | | | Before initial use.

Electronic thermometers | | | Monthly.

Vacuum blood agitator | Observe weight of the first container of blood filled for correct results | Each day of use | Standardize with container of known mass or volume before initial use, and after repairs or adjustments.

(c) Equipment employed in the sterilization of materials used in blood collection or for disposition of contaminated products shall be designed, maintained and utilized to ensure the destruction of contaminating microorganisms. The effectiveness of the sterilization procedure shall be no less than that achieved by an attained temperature of 121.5 °C (251 °F) maintained for 20 minutes by saturated steam or by an attained temperature of 170 °C (338 °F) maintained for 2 hours with dry heat.

21 CFR Part 606: Current Good Manufacturing Practice for Blood and Blood Components, Subpart DRead on eCFR

Citations by year

Form 483 citations of 21 CFR 606.60 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).

Peak: 45 in 2011

Form 483 citations of 21 CFR 606.60 by inspection year
YearForm 483 citations
20082
200927
201035
201145
201231
201324
201422
201513
201618
201715
201810
201916
20203
20213
20229
20235
20248
202516
202615

How FDA words this citation

FDA's short descriptions and citation text on these 317 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.

Short descriptions

  • Equipment observed, standardized, calibrated134 citations
  • Equipment calibration frequency79 citations
  • Maintain and clean equipment67 citations
  • Provide proper equipment to meet requirements37 citations

Citation text

  • “Failure of equipment to perform in the manner for which it was designed so as to assure compliance with the official requirements prescribed in 21 CFR 606.”37 citations
  • “Equipment used in the collection of blood and blood components is not observed, standardized and calibrated with at least the frequency required.”24 citations
  • “Equipment used in the processing of blood and blood components is not observed, standardized and calibrated with at least the frequency required.”19 citations
  • “Equipment used in the storage and distribution of blood and blood components is not observed, standardized and calibrated with at least the frequency required.”12 citations
Show 2 more
  • “Equipment used in the storage and distribution of blood and blood components is not observed on a regularly scheduled basis as prescribed in the SOP Manual.”9 citations
  • “Failure to maintain equipment used in the collection of blood and blood products in a clean and orderly manner.”8 citations

Facilities cited most recently

The latest inspections where FDA cited 21 CFR 606.60, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.

Show 4 more

Often cited at the same inspection

Of the 299 inspections where FDA cited 21 CFR 606.60, the share where it also cited each of these sections.

Ask FDAgent about 21 CFR 606.60

FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
  • Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
  • A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 606.60; FDAgent quotes that sentence exactly.
  • FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.