Skip to content

21 CFR 606.65: Supplies and reagents

FDA's inspection data lists 146 Form 483 citations of 21 CFR 606.65 at 129 facilities, from October 2008 to July 16, 2026.

Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.

Form 483 citations
146
At 144 inspections
Facilities cited
129
Distinct FEIs
Most recent citation
Jul 16, 2026
UC Health Poudre Valley Hospital
Warning letters
0
State a violation of it

The regulation

The current text of 21 CFR 606.65 from the Electronic Code of Federal Regulations.

§ 606.65 Supplies and reagents.

All supplies and reagents used in the collection, processing, compatibility testing, storage and distribution of blood and blood components shall be stored in a safe, sanitary and orderly manner.

(a) All surfaces coming in contact with blood and blood components intended for transfusion shall be sterile, pyrogen-free, and shall not interact with the product in such a manner as to have an adverse effect upon the safety, purity, potency or effectiveness of the product. All final containers and closures for blood and blood components not intended for transfusion shall be clean and free of surface solids and other contaminants.

(b) Each blood collecting container and its satellite container(s), if any, shall be examined visually for damage or evidence of contamination prior to its use and immediately after filling. Such examination shall include inspection for breakage of seals, when indicated, and abnormal discoloration. Where any defect is observed, the container shall not be used, or, if detected after filling, shall be properly discarded.

(c) Representative samples of each lot of the following reagents or solutions shall be tested on a regularly scheduled basis by methods described in the Standard Operating Procedures Manual to determine their capacity to perform as required:

Reagent or solution | Frequency of testing

Anti-human globulin | Each day of use.

Blood grouping reagents | Do.

Lectins | Do.

Antibody screening and reverse grouping cells | Do.

Hepatitis test reagents | Each run.

Syphilis serology reagents | Do.

Enzymes | Each day of use.

(d) Supplies and reagents that do not bear an expiration date shall be stored in such a manner that the oldest is used first.

(e) Supplies and reagents shall be used in a manner consistent with instructions provided by the manufacturer.

(f) Items that are required to be sterile and come into contact with blood should be disposable whenever possible.

21 CFR Part 606: Current Good Manufacturing Practice for Blood and Blood Components, Subpart DRead on eCFR

Citations by year

Form 483 citations of 21 CFR 606.65 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).

Peak: 22 in 2011

Form 483 citations of 21 CFR 606.65 by inspection year
YearForm 483 citations
20081
200919
201016
201122
201215
20139
201410
20154
20167
20175
20186
20191
20201
20211
20225
20233
20241
202514
20266

How FDA words this citation

FDA's short descriptions and citation text on these 146 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.

Short descriptions

  • Following manufacturer's instructions126 citations
  • Safe, sanitary, orderly storage14 citations
  • Testing of representative samples of reagent lots6 citations

Citation text

  • “Failure to use supplies and reagents in a manner consistent with instructions provided by the manufacturer.”126 citations
  • “Failure to store all supplies and reagents used in the collection of blood and blood components in a safe, sanitary and orderly manner.”5 citations
  • “Representative samples of each lot of Blood grouping reagents are not tested on a regularly scheduled basis as described in the SOP Manual to determine their capacity to perform as required by the regulations.”3 citations
  • “Failure to store all supplies and reagents used in the collection and processing of blood and blood components in a safe, sanitary and orderly manner.”2 citations
Show 2 more
  • “Failure to store all supplies and reagents used in the compatibility testing of blood and blood components in a safe, sanitary and orderly manner.”2 citations
  • “Failure to store all supplies and reagents used in the processing of blood and blood components in a safe, sanitary and orderly manner.”2 citations

Facilities cited most recently

The latest inspections where FDA cited 21 CFR 606.65, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.

Show 4 more

Often cited at the same inspection

Of the 144 inspections where FDA cited 21 CFR 606.65, the share where it also cited each of these sections.

Ask FDAgent about 21 CFR 606.65

FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
  • Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
  • A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 606.65; FDAgent quotes that sentence exactly.
  • FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.