Skip to content

Blood Bank of Alaska, Inc.

1215 Airport Heights Dr, Anchorage, Alaska, United States

FEI 3071403

FDA has 7 inspections on file for Blood Bank of Alaska, Inc. in Anchorage, Alaska (FEI 3071403), from 2010 to 2024. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

No warning letter, court action or import refusal is filed under this FEI.

Six regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
7, 2010–2024
0 OAI · 6 VAI · 1 NAI
Inspections with 483 observations shown
5 of 7
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
  2. VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. General requirements - storage

      Failure to store and maintain the Red Blood Cells between 1 and 6 degrees Celsius immediately after processing.

    2. Establish, maintain and follow manufacturing SOPs

      Written standard operating procedures including all steps to be followed in the collection of blood and blood components for allogeneic transfusion, autologous transfusion and further manufacturing purposes were not always followed.

  3. VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Maintain and clean equipment

      Failure to maintain equipment used in the storage of blood and blood products.

    2. Establish, maintain and follow manufacturing SOPs

      Written standard operating procedures including all steps to be followed in the collection , processing , storage and distribution of blood and blood components for homologous transfusion are not always maintained and maintained on the premises.

    3. Adverse Reaction- Reports of Investigations

      Written reports of investigations of adverse reactions, including conclusions and follow up, are not prepared and maintained.

    4. Thorough investigations

      Failure to perform a thorough investigation and make a record of the conclusions and follow-up of an unexplained discrepancy.

    5. Required records

      Failure to maintain processing records.

  4. VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products10 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Concurrent documentation

      Records are not concurrently maintained with the performance of each significant step in the collection, processing, storage, and distribution of each unit of blood and blood components so that all steps can be clearly traced.

    2. Final product - evidence of thawing

      Failure to store the final plasma product in a manner to show evidence of thawing.

    3. Provide space for examination

      Failure to provide adequate space for private examinations of individuals to determine their suitability as blood donors.

    4. Adverse Reaction - Investigations

      A thorough investigation of each reported adverse reaction was not made.

    5. Consideration of all results

      Failure to consider all test results in determining whether the test results meet the test objectives.

    Show 5 more citations
    1. Establish, maintain and follow manufacturing SOPs

      Written standard operating procedures including all steps to be followed in the collection, processing, storage, and distribution of blood and blood components for homologous transfusion, autologous transfusion, and further manufacturing purposes are not always maintained and followed.

    2. Notification w/in 8 weeks

      Failure to make reasonable attempts to notify the donor within 8 weeks after determining that the donor is deferred or determined not to be suitable for donation.

    3. Follow up and counseling

      Failure to provide appropriate information concerning medical follow-up and counseling.

    4. Provisions to monitor lab test procedures & instruments

      Failure to establish adequate provisions for monitoring the reliability, accuracy, precision, and performance of laboratory test procedures and instruments.

    5. Thorough investigations

      Failure to perform a thorough investigation and make a record of the conclusions and follow-up of an unexplained discrepancy.

  5. VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Person performing, test results, interpretation

      Records fail to be as detailed as necessary so as to provide a complete history of the work performed.

    2. Equipment calibration frequency

      Equipment used in the processing and storage and distribution of blood and blood components is not observed, standardized and calibrated with at least the frequency required.

    3. Establish, maintain and follow manufacturing SOPs

      Written standard operating procedures including all steps to be followed in the collection, processing, compatibility testing, storage, and distribution of blood and blood components for homologous transfusion and further manufacturing purposes are not always followed.

  6. VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Provide space for examination

      Failure to provide adequate space for private examinations of individuals to determine their suitability as blood donors.

    2. Adverse Reaction - Investigations

      A thorough investigation of each reported adverse reaction was not made.

    3. Thorough investigations

      Failure to perform a thorough investigation and make a record of the conclusions and follow-up of an unexplained discrepancy.

    4. Follow up and counseling

      Failure to provide appropriate information concerning medical follow-up and counseling.

    5. Record review prior to release

      All records pertinent to a lot or unit were not reviewed before the release or distribution of a lot or unit of final product.

  7. VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood ProductsCitations not published

Regulations FDA cited here

The 10 sections of 21 CFR in FDA’s published citations for 5 of this facility’s 7 inspections, with how many of them cited each.

Show all 10

Other FEIs recorded as Blood Bank of Alaska, Inc.

Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3071403 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3071403. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.