Equipment qualification and calibration in FDA Form 483 observations
Showing that equipment and instruments work as intended: installation and operational qualification, calibration and preventive maintenance. FDAgent finds 153 such observations in 115 Form 483s issued to 108 facilities, for inspections that ended between October 2008 and September 2026. Each is quoted below from the 483 FDA posted.
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.
- Observations
- 153
- Each quoted below
- Form 483s
- 115
- Posted by FDA
- Facilities
- 108
- Distinct FEIs
- Most recent
- Sep 9, 2026
- Inspection end date
Observations by year
These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).
Peak: 50 in 2026
| Year | Observations |
|---|---|
| 2008 | 2 |
| 2009 | 0 |
| 2010 | 0 |
| 2011 | 0 |
| 2012 | 1 |
| 2013 | 1 |
| 2014 | 0 |
| 2015 | 0 |
| 2016 | 11 |
| 2017 | 9 |
| 2018 | 9 |
| 2019 | 9 |
| 2020 | 0 |
| 2021 | 4 |
| 2022 | 9 |
| 2023 | 16 |
| 2024 | 9 |
| 2025 | 23 |
| 2026 | 50 |
How FDAgent sorts observations into this theme
Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.
It is tied to one of these 21 CFR entries
21 CFR 211.6321 CFR 211.68(a)21 CFR 211.160(b)(4)21 CFR 820.70(g)21 CFR 820.72
A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.
Or its text contains one of these phrases
“installation qualification” · “operational qualification” · “equipment qualification” · “IQ” · “OQ” · “PQ” · “IOQ” · “calibrat*” · “preventive maintenance” · “preventative maintenance” · “requalif*”
* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.
The 153 observations, newest first
Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.
Farmakeio Outsourcing LLCInspection ended Sep 5, 2025 · FEI 3014982757 · Observation 5
Routine calibration and checking of automatic and electronic equipment is not performed according to a written program designed to assure proper performance. Specifically, during a review of your firm's batch record for Testosterone/Triamcinolone Acetonide Pellet 200mg/40mcg, Lot 3468, Expiry 7/23/2026, Sections 2.3.1.3 - 2.3.1.5, document pellets being processed through the [(b)(4)] Systems, [(6)(4)] Metal detection System. Section 2.3.1.5, …
Opens with FDA's citation wording for 21 CFR 211.68(a)Read the Form 483 →
Elite Orthopedics, LLCInspection ended Jul 15, 2025 · FEI 3011803392 · Observation 2
Documentation of equipment maintenance, cleaning, sanitization, and calibration was not maintained. Specifically, Since 06/09/2023, your firm did not document pre-installation information, preventative maintenance, and control programming for the (b) (4) use to store (b) (4). As instructed in the (b) (4) Instructions User Manual for the following elements, • Unit requirements • As needed preventative maintenance, such as battery alarm check and …
Contains “calibrat*”Read the Form 483 →
Elite Orthopedics, LLCInspection ended Jul 15, 2025 · FEI 3011803392 · Observation 3
Storage temperatures of HCT/Ps were not recorded and maintained. Specifically, From (b) (4), your firm distributed (b) (4) tissue products and failed to record (b) (4) temperature of storage areas and perform a (b) (4) review of the (b) (4) Temperature Log. Also, did not maintained (b) (4) calibrations records for the (b) (4) thermometer.
Contains “calibrat*”Read the Form 483 →
Foremount Enterprise Co., LtdInspection ended Jun 19, 2025 · FEI 3010402691 · Observation 4
… Product Code CBP exported and intended for the US market; however, your firm was not able to provide protocols and objective evidence to support your related Installation Qualification (IQ), Operational Qualification (OQ), and Performance Qualification (PQ) Reports FM-RN-01 and FM-RN-02 dated October 22, 2024.
Contains “installation qualification”Read the Form 483 →
Medi-Fare Drug Pharmaceutical Compounding LLCInspection ended Jun 6, 2025 · FEI 3009925820 · Observation 9
Employees engaged in the manufacture and processing of a drug product lack the training required to perform their assigned functions. A. Your visual inspection qualification training is inadequate regarding both your initial qualification and your lack of requalification. 1. You lack an annual requalification process. 2. Your initial qualification is not scientifically established nor standardized. Your initial visual inspection qualification is …
Contains “requalif*”Read the Form 483 →
AVEM Labs LLCInspection ended May 21, 2025 · FEI 3017596057 · Observation 3
Equipment used for measuring was not calibrated according to established schedules. Specifically, you have not calibrated your temperature monitoring devices, (b) (4) , S/N (b) (4) and S/N (b) (4) . These temperature monitoring devices are used to monitor temperatures of finished and quarantined lots of amniotic membrane (b) (4) BioVair DL.
Contains “calibrat*”Read the Form 483 →
Nanjing Cuccess Pharmaceutical Co., LtdInspection ended Mar 21, 2025 · FEI 3027129265 · Observation 2
Routine checking of automatic and electronic equipment is not performed according to a written program designed to assure proper performance. Specifically, I reviewed qualification, mapping, and routine monitoring records for the long-term (25°C/60% RH) stability chamber, ID 911C49003. I observed the following deficiencies: A. Stability chamber requalification protocol VP1026-24-01-00 and report VR1026-24-01-00 indicate the chamber must be in a …
Opens with FDA's citation wording for 21 CFR 211.68(a)Read the Form 483 →
Emcure Pharmaceuticals LimitedInspection ended Feb 25, 2025 · FEI 3005782879 · Observation 1
… Maintenance and Equipment Integrity Issues: (b)(4) showed physical deterioration (paint chips, rust, peeling) and [illegible] maintenance "KEM-008/06, Preventive Maintenance for [illegible] (b)(4) Effective Date: 06 Aug 2024" did not require inspection for such damage. (b)(4) with numerous [illegible] and (b)(4) #1402 had (b)(4) dents; cleaning validation did not address potential residue in hard-to-clean areas nor incorporate dent study results.
Contains “preventive maintenance”Read the Form 483 →
Shouguang Fukang Pharmaceutical Co., Ltd.Inspection ended Feb 21, 2025 · FEI 3003941038 · Observation 2
… The last preventive maintenance was performed on 9/22/2024 and no issues were noted regarding the current condition of the damaged bottom of this vessel. C) Specifically, on 2/17/2025, inspection of the firm's [illegible] water system located in Building [illegible] that supplies [illegible] water to Workshop # [illegible] where the API [illegible] is manufactured noted that the distribution [illegible] that supplies [illegible] water to this …
Contains “preventive maintenance”Read the Form 483 →
Ajinomoto Company Inc.Inspection ended Feb 21, 2025 · FEI 3002806336 · Observation 1
… The SOP required a calibration of the balance after power is restored, but the employees did not record a calibration after power was restored. The investigation determined that the record in use was not sufficient to allow the employees to properly and reliably record a check and the form was revised; however, there is no discussion of why the form was not designed correctly originally and whether or not any changes need to be made to prevent a …
Contains “calibrat*”Read the Form 483 →
Piramal, Piramal Pharma LimitedInspection ended Feb 17, 2025 · FEI 3008763768 · Observation 1
… In addition, the equipment alarms have not been challenged as part of Installation, Operation, and Performance Qualification (IQ/OQ/PQ). D. The software used to control and monitor the [illegible] process step for APIs had the following alarms reported throughout Batch # [illegible] (API batch processed between February 08, 2025, and February 11, 2025): (b)(4) 1. Approximately six (6) (b)(4) alarms; 2. One (1) (b)(4) alarm; 3. Approximately …
Contains “IQ”Read the Form 483 →
Laurus Labs LimitedInspection ended Jan 31, 2025 · FEI 3014255847 · Observation 1
Written procedures, including changes thereof, are not drafted, reviewed and approved by appropriate organization units, and approved by the quality control unit. Specifically, You failed to ensure that equipment qualification protocols are reviewed and approved by the quality unit. During the inspection, we observed multiple equipment used for the manufacturing of active pharmaceutical ingredients (i.e., [illegible]) qualified without an …
Contains “equipment qualification”Read the Form 483 →
Zhejiang Huahai Pharmaceutical Co., Ltd.Inspection ended Jan 24, 2025 · FEI 3003999190 · Observation 2
… Calibration of Scale prior to filling ii. Calibration of Scale after being filled iii. Adjustment o~ (b)C1 . . (b) (4 iv. Adjustment of .~ 4 (bTC, v 6 (bHl Removing, • v·,aIs at . . (b)(4) vii adjustment leading to l j 3. Smoke studies for Workshop were performed using [ ] generated from here has been no evaluation of whether this produces neutrally buoyant smoke.
Contains “calibrat*”Read the Form 483 →
Janssen Vaccines CorporationInspection ended Nov 19, 2024 · FEI 3012637764 · Observation 3
… UDA Attaching the (b) (4) to filling line, right hand is blocked by left hand Tightening the (b) (4) assembly, not enough smoke to verify UDA (b) (4) (b) (4) Calibration Adjusting the IPC (b) (4) not adequate smoke to verify UDA Adjusting the (b) (4) with both hands: not enough smoke to UDA (b) (4) (b) (4) Sample Bag Replacement and (b) (4) Removing the sample bag, not enough smoke to verify UDA Replacing the sample bag, not enough smoke to …
Contains “calibrat*”Read the Form 483 →
Janssen Vaccines CorporationInspection ended Nov 19, 2024 · FEI 3012637764 · Observation 5
Equipment used in the manufacture, processing, packing, or holding of drug products is not of appropriate design to facilitate operations for its intended use. Specifically, Your firm's (b) (4) integrity test (b) (4) machines are not adequately qualified such that the (b) (4) (b) (4), and operating ranges (b) (4) associated with your drug products manufacturing are not adequately qualified. You carried out Operation Qualification in two parts; …
Opens with FDA's citation wording for 21 CFR 211.63Read the Form 483 →
Staska Pharmaceuticals IncInspection ended Sep 18, 2024 · FEI 3017473850 · Observation 4
… Additionally, viable air monitoring was conducted during the calibration of the repeater pump and not during routine filling.
Contains “calibrat*”Read the Form 483 →
Serum Institute of India Pvt. Limited Manjari SiteInspection ended Jul 2, 2024 · FEI 3020223359 · Observation 1
Requalification of the (b) (4) used to (b) (4) the filling (b) (4) filling line and (b) (4)) where COVID vaccine is filled into syringes is inadequate. Specifically, the maximum load runs performed during re-qualification of the (b) (4) failed your acceptance criterion for observing positive (b) (4) in 2022 and during (b) (4) requalification run in 2023. In addition, positive (b) (4) were found in the same nearby locations during requalification …
Contains “requalif*”Read the Form 483 →
PCAS Finland OyInspection ended Jun 7, 2024 · FEI 3003469991 · Observation 2
… There is no written procedure for inspection, component replacement, calibration or other maintenance frequency/schedule for this equipment.
Contains “calibrat*”Read the Form 483 →
Swati Spentose Private LimitedInspection ended May 31, 2024 · FEI 3009765202 · Observation 2
… Over one hundred OOS/Incidents investigations have been concluded as human errors since Jan 2021 despite ongoing retraining and requalification of analysts. Human errors in the QC lab such as incorrect entry of the sample details into LIMS system, dilution error, preparation error, wrong weight error, weight transfer error etc. have occurred repeatedly since January 2021 (as shown in the table below) despite of corrective actions including …
Contains “requalif*”Read the Form 483 →
Alkem Laboratories LimitedInspection ended Mar 27, 2024 · FEI 3006370533 · Observation 1
QUALITY SYSTEM OBSERVATION 1 There is a failure to thoroughly review any unexplained discrepancy and the failure of a batch or any of its components to meet any of its specifications whether or not the batch has been already distributed. Specifically, A. During the preventive maintenance of Perkin Elmer UV Spectrophotometer ID #G/QC/225 connected to software and computer (UV WinLab ES 6.5.0.103 and computer ID #ABA/EQ/QC/COM/01187) on 06/28/2023, …
Contains “preventive maintenance”Read the Form 483 →
21 CFR sections these observations open with
The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.
- 21 CFR 211.63Equipment design, size, and locationIn this theme's rule15
- 21 CFR 211.22Responsibilities of quality control unit10
- 21 CFR 211.113Control of microbiological contamination9
- 21 CFR 211.68Automatic, mechanical, and electronic equipmentIn this theme's rule8
- 21 CFR 211.160General requirementsIn this theme's rule6
- 21 CFR 211.192Production record review6
- 21 CFR 211.100Written procedures; deviations5
- 21 CFR 211.42Design and construction features3
- 21 CFR 820.75Process validation2
- 21 CFR 1271.200Equipment2
Ask FDAgent about equipment qualification and calibration in Form 483s
FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.
Related
About this data
- The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
- Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
- A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.
- FDAgent is independent of FDA.