Equipment qualification and calibration in FDA Form 483 observations
Showing that equipment and instruments work as intended: installation and operational qualification, calibration and preventive maintenance. FDAgent finds 153 such observations in 115 Form 483s issued to 108 facilities, for inspections that ended between October 2008 and September 2026. Each is quoted below from the 483 FDA posted.
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.
- Observations
- 153
- Each quoted below
- Form 483s
- 115
- Posted by FDA
- Facilities
- 108
- Distinct FEIs
- Most recent
- Sep 9, 2026
- Inspection end date
Observations by year
These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).
Peak: 50 in 2026
| Year | Observations |
|---|---|
| 2008 | 2 |
| 2009 | 0 |
| 2010 | 0 |
| 2011 | 0 |
| 2012 | 1 |
| 2013 | 1 |
| 2014 | 0 |
| 2015 | 0 |
| 2016 | 11 |
| 2017 | 9 |
| 2018 | 9 |
| 2019 | 9 |
| 2020 | 0 |
| 2021 | 4 |
| 2022 | 9 |
| 2023 | 16 |
| 2024 | 9 |
| 2025 | 23 |
| 2026 | 50 |
How FDAgent sorts observations into this theme
Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.
It is tied to one of these 21 CFR entries
21 CFR 211.6321 CFR 211.68(a)21 CFR 211.160(b)(4)21 CFR 820.70(g)21 CFR 820.72
A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.
Or its text contains one of these phrases
“installation qualification” · “operational qualification” · “equipment qualification” · “IQ” · “OQ” · “PQ” · “IOQ” · “calibrat*” · “preventive maintenance” · “preventative maintenance” · “requalif*”
* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.
The 153 observations, newest first
Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.
Jiangsu Hengrui Pharmceuticals Co., LtdInspection ended Jan 16, 2024 · FEI 3007373503 · Observation 1
… D) Production operator requalification procedures with respect to performing routine and non-routine interventions are inadequate. During our review of six (6) media fill studies across all manufacturing workshops, we observed operators who performed the bulk of interventions, operators who performed several interventions, and operators who performed no interventions during media fill studies, who were all considered to be requalified as a result …
Contains “requalif*”Read the Form 483 →
Jiangsu Hengrui Pharmceuticals Co., LtdInspection ended Jan 16, 2024 · FEI 3007373503 · Observation 6
… Subsequently, our review found that equipment qualification activities for these equipment have never been performed by your firm. Additionally, during the inspection on 01/11/2024, we requested your firm to take us to the viewing window of your workshop (b)(4) in order to observe your MET ONE 3400 NVP equipment located in the Grade B cleanroom. Upon arrival, we were told to wait until personnel gowned and entered the cleanroom to show us the …
Contains “equipment qualification”Read the Form 483 →
Laurus Synthesis Private LimitedInspection ended Dec 12, 2023 · FEI 3011524794 · Observation 1
… The damage was not present during preventive maintenance done on November 21, 2021. A total of (b) (4) batches were manufactured since the last Preventive Maintenance, including (b) (4) batches of (b) (4) (b) (4) and (b) (4) batches of (b) (4) (b) (4). The investigation did not include screening all (b) (4) batches for (b) (4) particles. Only (b) (4) lots of product (b) (4) manufactured after product changeover were sampled and inspected for (b) …
Contains “preventive maintenance”Read the Form 483 →
Laurus Synthesis Private LimitedInspection ended Dec 12, 2023 · FEI 3011524794 · Observation 3
… This (b) (4) was stated to have been inspected as clean on December 02, 2023 during the (b) (4) preventive maintenance. C. Product contact transfer hoses used to transfer (b) (4) during production and non-product contact transfer hoses are stored in the accessory room on a shelf with the hose ends covered with bags. These hoses were not stored in a way to ensure that any residual solvent or moisture after use and cleaning does not remain within …
Contains “preventive maintenance”Read the Form 483 →
Antaria Pty Ltd.Inspection ended Nov 17, 2023 · FEI 3025978151 · Observation 3
… b) The "calibration standard" listed in the template for the assay of (b) (4) via (b) (4) chromatography shows a date of preparation as 14FEB23. The bottle of working standard currently being used is dated 12SEP23. Per your technician, the date of 12SEP23 is also incorrect. The technician stated they had prepared the working standard several days earlier but did not change the date on the bottle. c) There is no documentation of the (b) (4) of the …
Contains “calibrat*”Read the Form 483 →
Antaria Pty Ltd.Inspection ended Nov 17, 2023 · FEI 3025978151 · Observation 4
… SOP 200.02 Shelf Life of Laboratory Reagents and Samples, rev 03 issued 07JUL2021 defines recommended expiration dates for laboratory prepared reagents and calibration standards however, your firm has no documentation of any validation work to justify these recommended expiration dates.
Contains “calibrat*”Read the Form 483 →
Antaria Pty Ltd.Inspection ended Nov 17, 2023 · FEI 3025978151 · Observation 6
Calibration of balances used in the QC Laboratory are not calibrated appropriately. For example, the balances are not calibrated over the operating range and there is a failure to demonstrate that the balance meets the requirements defined for repeatability, accuracy, and linearity.
Contains “calibrat*”Read the Form 483 →
Antaria Pty Ltd.Inspection ended Nov 17, 2023 · FEI 3025978151 · Observation 9
Your firm has not fully qualified the stability chambers used for long term (30C/65%RH - Oven ID O-0007) and accelerated stability studies. The chambers were installed in April 2022. The Operational Qualification (OQ) report approved on August 15, 2022 by QA contains no information about the actual operation of the chambers including no temperature and humidity readings, no verification documentation that the digital read out of the temperature …
Contains “operational qualification”Read the Form 483 →
Catalent Indiana, LLCInspection ended Nov 15, 2023 · FEI 3005949964 · Observation 6
… to contain any information documenting the incidences nor log statements as required by the written procedure, A-SOP-07-01-015: Performing Maintenance and Calibration, Version 44, Effective 2023-08-28. Corrective actions were not performed to address these WOs and subsequently, on 31 OCT 2023, recurrences of this same issue of (e) (4) malfunctioning were apparent. During walkthroughs on 31 OCT 2023, this malfunction was observed in real time …
Contains “calibrat*”Read the Form 483 →
Sumitomo Osaka Cement Co., LtdInspection ended Oct 27, 2023 · FEI 3003943727 · Observation 4
… However, the calibration of the old (b)(4) was performed between (b)(4) to (b)(4) (Certificate# FK 610003-1-8) and between (b)(4) to (b)(4) (Certificate# FK 600446-1-8) in July 2022 for the new (b)(4) in September 2022, thus not covering the entire operating range. 7. The firm does not monitor appearance and assay of (b)(4) (critical quality attributes) on stability samples of all (b)(4) API for US instead monitors residue on ignition, …
Contains “calibrat*”Read the Form 483 →
Central Glass Co., Ltd.Inspection ended Sep 15, 2023 · FEI 3002806649 · Observation 4
… In addition on 9/15/23 I observed that there is no calibrated device to monitor the temperature of the room where the media is stored.
Contains “calibrat*”Read the Form 483 →
Jiangsu Hengrui Pharmaceuticals CO. LTD.Inspection ended Aug 29, 2023 · FEI 3003404148 · Observation 9
… (b) (4) The (b) (4) does not have automatic documentation of (b) (4) are documented (b) (4) Qualification by way of (b) (4) mapping is performed (b) (4) and calibration of (b) (4) probe is conducted (b) (4)
Contains “calibrat*”Read the Form 483 →
Qilu Antibiotics Pharmaceutical Co., LTD.Inspection ended May 30, 2023 · FEI 3009831352 · Observation 5
The responsibilities and procedure applicable to the quality control unit are not in writing. Specifically, Your firm does not have established training procedures to oversee the initial and requalification of aseptic filling operators and visual inspection personnel for workshop (b) (4) units (b) (4) A. Your firm has one SOP titled QUAL116.P "Training for Manufacturing and Quality System Personnel" governing the training program in workshop (b) …
Contains “requalif*”Read the Form 483 →
Seqirus Vaccines LimitedInspection ended May 26, 2023 · FEI 3002806949 · Observation 1
Failure to conduct investigations into unexplained discrepancies. Specifically, during the review of the calibration certificate for the (b)(4) used for the testing of the (b)(4) (b)(4) for the bulk formulated (b)(4) of Fluad. We observed that the September 2021 Calibration Certificate "As Found" data had not met the acceptance criteria for the (b)(4) measurement (b)(4). Your firm failed to follow procedure as documented in DOC-000134669 v8.0 …
Contains “calibrat*”Read the Form 483 →
Apicore Pharmaceuticals Private Ltd.Inspection ended May 5, 2023 · FEI 3006767695 · Observation 2
Routine calibration of electronic equipment is not performed to assure proper performance. Specifically, Analytical balance QCE #684 with a weight range of 20 mg to 200 g is used in your firm's QC chemistry laboratory to weigh test samples. Your firm's verification of this balance includes only the 50 mg to 200 g weight range, and does not include the lower weight range of 20 mg to 50 mg.
Contains “calibrat*”Read the Form 483 →
Zen Chemicals Pvt. LtdInspection ended Mar 28, 2023 · FEI 3015345351 · Observation 2
… The first set of samples consisted of (b) (4) injections which were analyzed on December 7-8, 2021 for (b) (4) calibration. The injections were rerun due to wrong vial number. 2. The second set of samples consisted of (b) (4) injections and were analyzed on June 2, 2022 for (b) (4) stability of (b) (4). The injections were rerun due to wrong preparation of sample. 3. The third set of samples consisted of (b) (4) injections which were analyzed on …
Contains “calibrat*”Read the Form 483 →
Zen Chemicals Pvt. LtdInspection ended Mar 28, 2023 · FEI 3015345351 · Observation 4
Qualification of equipment were not performed. Specifically, Your firm failed to perform Performance Qualification (PQ) for equipment but not limited to (b) (4) equipment # ZEN/MC/P-56, (b) (4) equipment # (b) (4) equipment # ZEN/MC/P-44, (b) (4) (equipment # (b) (4). According to section 7.8.4 of your SOP, QA/SOP/GEN/39, document titled, 'Standard Operating Procedure for Qualification of Facility / Equipment's / System' states, that QA in …
Contains “PQ”Read the Form 483 →
SciVac LTDInspection ended Mar 27, 2023 · FEI 3012695367 · Observation 3
… Your visual inspection operator failed multiple requalification efforts during the approximate period of 7/3/2022 through 7/28/2022 and no deviation was opened. ii. Your gowning qualification SOP does not require investigations if the operator fails (b)(4) gowning qualification.
Contains “requalif*”Read the Form 483 →
Lupin Limited Unit 1Inspection ended Nov 23, 2022 · FEI 3002807511 · Observation 6
Routine calibration of automatic and mechanical equipment is not performed according to a written program design to assure proper performance. Specifically, The Metal Detection 2022 requalification protocol & report, document number MDP16/RQ/CFM-02/22-018 equipment tag number CFM-02 & MET-04, includes a challenge with (b) (4) (b) (4) challenge (b) (4) (approximately (b) (4) mm diameter). The Deputy General Manager of Quality Assurance confirmed …
Contains “calibrat*”Read the Form 483 →
Catalent Indiana LLCInspection ended Sep 1, 2022 · FEI 3005949964 · Observation 5
… firm did not consider individual defect vial performance when establishing acceptance criteria during (b)(4) ((b)(4) System 2414) Performance Qualification (PQ) (b)(4) mL, VPQ-258-000-0002-P, recipe (b)(4) Rev 1.0. 2. Additionally, instances where defects were not caught by the (b)(4) were not evaluated because they were averaged out based on the acceptance criteria in part A. Your firm runs each defect type and size on the (b)(4) in (b)(4). …
Contains “PQ”Read the Form 483 →
21 CFR sections these observations open with
The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.
- 21 CFR 211.63Equipment design, size, and locationIn this theme's rule15
- 21 CFR 211.22Responsibilities of quality control unit10
- 21 CFR 211.113Control of microbiological contamination9
- 21 CFR 211.68Automatic, mechanical, and electronic equipmentIn this theme's rule8
- 21 CFR 211.160General requirementsIn this theme's rule6
- 21 CFR 211.192Production record review6
- 21 CFR 211.100Written procedures; deviations5
- 21 CFR 211.42Design and construction features3
- 21 CFR 820.75Process validation2
- 21 CFR 1271.200Equipment2
Ask FDAgent about equipment qualification and calibration in Form 483s
FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.
Related
About this data
- The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
- Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
- A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.
- FDAgent is independent of FDA.