Equipment qualification and calibration in FDA Form 483 observations

Showing that equipment and instruments work as intended: installation and operational qualification, calibration and preventive maintenance. FDAgent finds 153 such observations in 115 Form 483s issued to 108 facilities, for inspections that ended between October 2008 and September 2026. Each is quoted below from the 483 FDA posted.

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.

Observations
153
Each quoted below
Form 483s
115
Posted by FDA
Facilities
108
Distinct FEIs
Most recent
Sep 9, 2026
Inspection end date

Observations by year

These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).

Peak: 50 in 2026

Equipment qualification and calibration observations in Form 483s by inspection year
YearObservations
20082
20090
20100
20110
20121
20131
20140
20150
201611
20179
20189
20199
20200
20214
20229
202316
20249
202523
202650

How FDAgent sorts observations into this theme

Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.

It is tied to one of these 21 CFR entries

21 CFR 211.6321 CFR 211.68(a)21 CFR 211.160(b)(4)21 CFR 820.70(g)21 CFR 820.72

A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.

Or its text contains one of these phrases

“installation qualification” · “operational qualification” · “equipment qualification” · “IQ” · “OQ” · “PQ” · “IOQ” · “calibrat*” · “preventive maintenance” · “preventative maintenance” · “requalif*”

* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.

The 153 observations, newest first

Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.

  1. MicroBioPharm Japan Co., LtdInspection ended Feb 13, 2026 · FEI 3003992082 · Observation 3

    Your process validations are inadequate. Specifically, You have performed a process validation for the manufacturing of the API (b)(4) (DMF# (b)(4)) using production equipment that have not been qualified. For example, the process verification protocols, (b)(4) have used major production equipment that have not been adequately qualified. (b)(4) The lack of equipment qualification undermines the validity of the process validation study, as there …

    Contains “equipment qualification”Read the Form 483 →

  2. Almon Healthcare Private LimitedInspection ended Feb 13, 2026 · FEI 3030509366 · Observation 8

    Failure to monitor the (b)(4) water system to ensure that water is of appropriate quality. Your firm lacks both preventive maintenance and sanitization programs for the (b)(4) water system, failing to ensure appropriate water quality for API manufacturing and cleaning equipment/utensils that directly come in contact with APIs. Water testing is limited to collection from the (b)(4) water collection at production point-of-use locations, with …

    Contains “preventive maintenance”Read the Form 483 →

  3. Biomylz Private LimitedInspection ended Feb 9, 2026 · FEI 3013697420 · Observation 1

    … Laboratory logbooks used to document media preparation, pH meter equipment usage, incubator usage, sterilization (b)(4) activities, conductivity meter calibration, and pH meter calibration were observed with pages that were not numbered in advance. As a result, there is no assurance that pages cannot be removed, replaced, or altered without detection, and that laboratory records are complete and reliable.

    Contains “calibrat*”Read the Form 483 →

  4. SNJ Labs Private Ltd.Inspection ended Feb 6, 2026 · FEI 3026587575 · Observation 1

    … However, no preventive maintenance was performed on the instrument since 2020. Your firm retested the samples and invalidated the original OOS results (OOS/24/001 (b) (4) %], OOS/24/006 (b) (4) %], OOS/24/007 (b) (4) %] and OOS/24/010 (b) (4) %]; specification limit: (b) (4) %). Your firm's corrective actions and preventive actions to the assignable root cause were inadequate and that resulted in repeated sample failures. During April-September …

    Contains “preventive maintenance”Read the Form 483 →

  5. Apotek Production & Laboratorier ABInspection ended Jan 29, 2026 · FEI 3003766921 · Observation 5

    Equipment used in the manufacture, processing, packing or holding of drug products is not of appropriate design and suitably located to facilitate operations for its intended use. Specifically, 1. The ge is located above the end of the filling line ge at the ceiling level. HEPA filtered air is drawn up and directly off the line when a ge on the end is removed for the duration of machine set up. ga are located on each side of the line and extend …

    Opens with FDA's citation wording for 21 CFR 211.63Read the Form 483 →

  6. Centrient Pharmaceuticals India Private LimitedInspection ended Jan 27, 2026 · FEI 3004497364 · Observation 2

    … This condition demonstrates inadequate preventive maintenance and poses a risk of (b)(4) contamination to the API. The use of equipment in this degraded state may adversely affect the quality of the drug substance. On 27-Jan-2026, your firm informed us that approximately (b)(4) had been shaved off from the inner surface of the (b)(4) ID #(b)(4) 302 based on the preliminary investigation.

    Contains “preventive maintenance”Read the Form 483 →

  7. Neeyaan LLCInspection ended Jan 27, 2026 · FEI 3024366553 · Observation 9

    Equipment used in the manufacture, processing, packing or holding of drug products is not of appropriate design and of adequate size to facilitate operations for its intended use. Specifically, A) No Installation Qualification (IQ), Operational Qualification (OQ) and Performance Qualification (PQ) have been conducted for the Tablet Press (b) (4) (Room (b) (4) Room) which was used in the compression of OTC drug product (b) (4), USP ((b) (4) …

    Opens with FDA's citation wording for 21 CFR 211.63Read the Form 483 →

  8. Neeyaan LLCInspection ended Jan 27, 2026 · FEI 3024366553 · Observation 11

    … No preventative maintenance program exists for the HEPA filters.

    Contains “preventative maintenance”Read the Form 483 →

  9. Neeyaan LLCInspection ended Jan 27, 2026 · FEI 3024366553 · Observation 13

    … SOP-20 lacks specific instructions regarding the number of material lots to be evaluated during vendor qualification/requalification and the types of testing required to verify vendor adherence to specifications. Additionally, at least one specific identity test for API (b) (4), USP is not performed routinely; according to the firm's R&D Specialist, certificates of analysis from the vendor are accepted in lieu of identity testing for some …

    Contains “requalif*”Read the Form 483 →

  10. US Specialty Formulations LLCInspection ended Jan 14, 2026 · FEI 3010680515 · Observation 4

    … Th~(l:5) (4 ~run 2025 PQ identified locatio accumulated lethality. Although all performed. (iv) {6} ~ b This represents a repeat observation from previous

    Contains “PQ”Read the Form 483 →

  11. Dr. Reddy’s Laboratories LimitedInspection ended Dec 12, 2025 · FEI 3009193040 · Observation 2

    … (b)(4) As of 12/12/2025, you have not requalified your Tablet Compression Machine ID # PR034 and (b)(4)(b)(4)(b)(4) ID # PR182 assembly in Room . The make and model of the (b)(4)(b)(4) (b)(4)(b)(4)(b)(6) ID’s # are different. Your Head of Production stated that (b)(4) the working principles for are same thus requalification is not needed. (b)(4) You manufacture multiple drug products on Tablet Compression Machine ID # PR034 and (b)(4)(b)(4)(b)(4) …

    Contains “requalif*”Read the Form 483 →

  12. Eugia Pharma Specialities LimitedInspection ended Nov 14, 2025 · FEI 3009883410 · Observation 4

    Equipment used in the manufacture, processing, packing or holding of drug products is not of appropriate design to facilitate operations for its intended use. A. Aseptic processing operations conducted in filling line (b) (4) are deficient in design and controls to prevent microbiological contamination of sterile drug products. On November 05, 2025, we evaluated the filling operation in (b) (4) Line (b) (4) for (b) (4) Injection (b) g; Batch (b) …

    Opens with FDA's citation wording for 21 CFR 211.63Read the Form 483 →

  13. Lonza AGInspection ended Oct 28, 2025 · FEI 3002930340 · Observation 4

    … The (b)(4) instrument used to seal unique tracking number, and is not maintained or calibrated. c. The (b)(4) used for the (b)(4) do not have a preventative maintenance program. The product contact part (b)(4) and replaced b c. tth is "failure-based" (they are used until they fail, and then they are replaced). The facility has a single (b)(4) used for (b)(4) at the time (b)(4) compatible with the (b)(4) failure and was being repaired since on …

    Contains “calibrat*”Read the Form 483 →

  14. Implantica CE Reflux LimitedInspection ended Oct 24, 2025 · FEI 3015217460 · Observation 1

    … is no record showing that the speed, rpm, and other parameters employed on the RS-300 devices match the values specified in the original validation reports (IQ/OQ/PQ). 2. Corrective Action - Incomplete Batch-Record Amendment The corrective-action portion of the CAPA fails to provide documented proof that all affected batch records, including previously released batch records (job-router forms, batch-release sheets, and any downstream …

    Contains “IQ”Read the Form 483 →

  15. UBI Pharma Inc.Inspection ended Oct 14, 2025 · FEI 3004713402 · Observation 1

    Equipment used in the manufacture, processing, packing or holding of drug products is not of appropriate design to facilitate operations for its [illegible]. Specifically, the visual inspection of your vials is carried out in [illegible] then undergo Acceptance Quality Limit (AQL) testing before releasing. You (6)(4) R eVI (b)(4) 41 confirmed that the [illegible] was not qualified (6)(4) mcg/mL ~~ mL and ~~ mL vials. The qualification is based on …

    Opens with FDA's citation wording for 21 CFR 211.63Read the Form 483 →

  16. UBI Pharma Inc.Inspection ended Oct 14, 2025 · FEI 3004713402 · Observation 2

    … On 10/8/2025 upon a reassessment, your operator needed training and requalification. It was also noted that the operator did not set up the (li)(4) [illegible] correctly and measure [illegible] as described in Section [illegible]. The intensity set up was also [illegible]. You recorded at least 15 complaints for [illegible] empty vial, and crack/leak (Table 1). You investigated only three (3) (Batch No. (b)(4) Expiry date (b)(4) date [illegible] …

    Contains “requalif*”Read the Form 483 →

  17. UBI Pharma Inc.Inspection ended Oct 14, 2025 · FEI 3004713402 · Observation 3

    Procedures designed to prevent microbiological contamination are not established and/or followed. Specifically, Protocol PQ/GT2.3.224 Version [illegible] Hood for [illegible] Machine and the Version 02 (Effective date 4/12/2024) provide procedure for performing and evaluating the air flow of [illegible] room [illegible]. Air flow studies performed [illegible]. A. In multiple intervention simulation [illegible] (b)(4) clearly not defined. E.g., …

    Contains “PQ”Read the Form 483 →

  18. Sun Pharmaceuticals Medicare Ltd.Inspection ended Sep 19, 2025 · FEI 3002957541 · Observation 3

    … The Block (b)(4) vial filling line (b)(4) is used to aseptically fill products for (b)(4) (b)(4) APS requalification Batch (b)(4) was filled on 07/05/2024. After (b)(4) of the incubation, positive growth was observed in (b)(4) container units. After (b)(4) of the incubation, growth was seen in (b)(4) container units. Your investigation identified that incorrect size of (b)(4) stoppers being the root cause for contamination. The contaminating …

    Contains “requalif*”Read the Form 483 →

  19. Farmakeio Outsourcing LLCInspection ended Sep 5, 2025 · FEI 3014982757 · Observation 1

    … Additionally, your firm's metal detection system for finished pellets was not operational until May 19, 2025, metal detection performance qualification (PQ) quality unit approval date. Before this date, no finished pellets underwent metal detection as part of its manufacturing process. The [(D)(4)] intermediate product, which is (b)(4) and then processed into pellets, was visually inspected, packaged, [(6)(4)] sterilized, and distributed without …

    Contains “PQ”Read the Form 483 →

  20. Farmakeio Outsourcing LLCInspection ended Sep 5, 2025 · FEI 3014982757 · Observation 4

    Equipment used in the manufacture, processing, packing or holding of drug products is not of appropriate design and of adequate size to facilitate operations for its intended use. Specifically, A. Your firm's [(b)(4)] Systems, [(6)(4)] Metal detection System qualification failed to show the equipment is adequately designed and sized to facilitate operations for its intended use. During a review of your firm's [(b)(4)], Pharmaceutical Metal …

    Opens with FDA's citation wording for 21 CFR 211.63Read the Form 483 →

21 CFR sections these observations open with

The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.

Ask FDAgent about equipment qualification and calibration in Form 483s

FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
  • Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
  • A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.
  • FDAgent is independent of FDA.