The 4 observations on this Form 483
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.
- 1
Drug substance shelf‑life nonconformance
OBSERVATION 1 The information provided by the firm does not conform (b)(4) the (b)(4) drug substance (DS) shelf life shelf life applied to DS Process Performance Qualification (PPQ) batches (Batch (b)(4) PPQ batches (Batch (b)(4) was (b)(4)
- 2
Integrity test procedure deficiencies
OBSERVATION 2 The (b)(4) (b)(4) integrity test procedure used for follow the vendor recommendations (document in 'Operation of (b)(4) (b)(4). For example: a. (b)(4) version: (b)(4) indicates to ___. However the (b)(4) or test set-up Figure (b)(4) the (b)(4) Manufacturing Executing System (MES) guidance b. (b)(4) version (b)(4) indicates that the include d j n SOP CHVI-346656 v2.0 or in c. (b)(4) version (b)(4) indicates to observe and disregard for: (b)(4) (observation time). i. The (b)(4) MES instructions indicate to start the timer immediately after the (b)(4) and conduct the test for ~ c (MES waiting time instruction parts) ii. SOP CHVI-346656 v2.0 step (b)(4) without mentioning any waiting time. d. (b)(4) version (b)(4) indicates that the integrity test passes (b)(4) (b)(4). The (b)(4) MES guidance for the operators indicate that the (b)(4)
- 3
PP/CPP deviations and OOS results
OBSERVATION 3 Process Parameters (PPs) are not adequately controlled within established ranges manufacturing process. Specifically, PP deviations occurred during Batch (b)(4) manufacturing, including deviations for Critical Process Parameters (CPPs). In addition, Batch (b)(4) and Batch (b)(4) had out-of-specification (OOS) and borderline results, respectively, assay at release. The CPP deviations observed include: a. (b)(4) batch (b)(4) outside proven acceptable range during (b)(4) of production (b)(4). e PAR of n, (b)(4) and NO~ of h (b)(4) and (b)(4) as per CHVI-626612): at c % (b)(4) exceeded the (b) limit, and r)(4) limit. (b)(4) for (b)(4) c. (b)(4) Batch (b)(4) Excursion during the (b)(4) step for Batch b (b)(4). the acceptable range is (b)(4) (b)(4) Batch (b)(4) had OOS (b)(4) result of (b)(4). Batch (b)(4) had (b)(4) result of (b)(4) the lower specification limit (b)(4)
- 4
Critical materials and equipment maintenance issues
OBSERVATION 4 Critical material used to manufacture (b)(4) lifetime. SRer.i:6r.::iJ1v._·______..., a. The (b)(4) integrity Tester used (b)(4) DSC (b)(4) between life _____, entering and damaging the instrument. and is replaced only when damaged. During the FDA inspection execution on October 24, 2025, a (b)(4) ______ through the (b)(4) in the results of the (b)(4). b. The (b)(4) instrument used to seal unique tracking number, and is not maintained or calibrated. c. The (b)(4) used for the (b)(4) do not have a preventative maintenance program. The product contact part (b)(4) and replaced b c. tth is "failure-based" (they are used until they fail, and then they are replaced). The facility has a single (b)(4) used for (b)(4) at the time (b)(4) compatible with the (b)(4) failure and was being repaired since on October 28, 2025.
Other Form 483s FDA posted for this facility
Every Form 483 FDA posts under FEI 3002930340, newest first.
- Form 483, March 2023Inspection ended Mar 28, 2023 · Posted Jul 15, 2024VAI
Ask FDAgent about this Form 483
FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.
Related
About this data
- The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
- The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
- FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
- FDAgent is independent of FDA.