SNJ Labs Private Ltd.
Plot No 5 To 16, Survey No. 137 Ta. Kotda Sangani, Dist. Rajkot, Padavala, India
FEI 3026587575DUNS 932580580
FDA has one inspection on file for SNJ Labs Private Ltd. in Padavala, India (FEI 3026587575). It ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA posted the Form 483 from that inspection. FDAgent read its 6 observations from the PDF; they are listed below. One FOIA request has asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
As of Nov 21, 2026, the site is registered with FDA as a drug establishment for API manufacturing, through Dec 31, 2027. Its registrant is SNJ Labs Pvt. Ltd. (D-U-N-S 932580580).
Sources:FDA firm profile ↗Form 483 PDF ↗FDA drug establishment registrations ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 1, in 2026
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- ↗
- Requested under FOIA
- 1 request, latest Apr 17, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
- Drug establishment registration
- Through
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssurancePosted 483 · 6 observations
Read this Form 483 on its own page →Form 483 PDF on fda.gov
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.
- Observation 1
Inadequate investigation of critical OOS deviations
Failure of your quality unit to ensure that critical deviations are investigated and resolved. a. Your firm had four OOS results for the (b) (4) content in the finished active pharmaceutical ingredient (API) (b) (4) from April-September 2024. Root causes identified included loose wire connections to the (b) (4) instrument, black coating on the (b) (4) poor (b) (4) performance, and instrument malfunction. However, no preventive maintenance was performed on the instrument since 2020. Your firm retested the samples and invalidated the original OOS results (OOS/24/001 (b) (4) %], OOS/24/006 (b) (4) %], OOS/24/007 (b) (4) %] and OOS/24/010 (b) (4) %]; specification limit: (b) (4) %). Your firm's corrective actions and preventive actions to the assignable root cause were inadequate and that resulted in repeated sample failures. During April-September 2024, your firm analyzed and released (b) (4) finished API batches from Plant (b) (4) and (b) (4) batches of (b) (4) from Plant (b) (4) without performing any risk assessments. b. Your firm did not thoroughly investigate the OOS (OOS/24/005) results for (b) (4) assay by HPLC in (b) (4) (b) (4) API intended for the US market. The assay results were (b) (4) % (first replicate) and (b) (4) % (second replicate); specification limit: (b) (4) % to (b) (4) % w/w. The identified root cause was due to a (b) (4) failure at (b) (4) in the HPLC system; however, the chromatographic data acquisition was documented at (b) (4) The investigation did not verify these details or conduct further investigation for an assignable root cause. Your firm invalidated the initial OOS results, retested the samples with passing results, and implemented no preventive actions.
- Observation 2
Non-validated non-stability-indicating test methods
Your firm failed to use validated and stability-indicating test procedures. Your firm failed to use validated, stability-indicating methods to test active pharmaceutical ingredients (API) (b) (4) (b) (4) intended for the US market when assigning retest dates of (b) (4) The analytical method used for testing stability samples is not stability indicating and cannot identify degradation products. Furthermore, your firm started stress studies for (b) (4) API on (b) (4) intended for the US market, however your firm is currently assigning a (b) (4) retest date to this product. Your firm manufactured (b) (4) batches of (b) (4) and (b) (4) batches of (b) (4) in 2025.
Show observations 3 to 6
- Observation 3
Inadequate computerized systems data integrity controls
Your firm failed to exercise sufficient controls over computerized systems to prevent unauthorized access or changes to data and failure to have adequate controls to prevent omission of data. a. The finished product testing data for (b) (4) intended for the US market using stand-alone instruments, atomic absorption spectrophotometer AAS (SNJ/QC/AAS/37), UV/VIS Spectrophotomer (SNJ/QC/UV/39), and voltameter (SNJ/QC/VM/45) in the computer "C or E" drive were backed up to (b) (4) external hard drives (b) (4) (b) (4) according to your IT administrator. Analysts have unrestricted privileges to cut, copy, paste, delete and rename data files on computer hard drives and there is no assurance that data are original/unaltered before backup. Your firm has the capability to save HPLC data on the server, however, your firm has not conducted gap assessments for GMP data files saved on computers connected to stand alone instruments. b. Your firm's analysts have privileges of changing date and time settings on computers attached to the atomic absorption spectrophotometer and UV-Vis spectrophotometer. c. Your firm failed to validate the Excel software used for laboratory data calculations. Your QC Manager confirmed this on February 4, 2026. The results calculated for (b) (4) assay for (b) (4) finished product batch (b) (4) using Excel software were not available during the inspection and your QC manager stated that your firm's practice is to perform calculations using the software, take a printout without saving or locking the files for review.
- Observation 4
Failure to maintain clean manufacturing facilities
Failure to properly maintain buildings and facilities used in the manufacture of API. Your firm failed to maintain facility and equipment in clean status for manufacturing and processing of (b) (4) (b) (4) (Plant (b) and (b) (4) API (Plant (b) (4) a. Clean room (b) (4) Plant (b) showed (b) (4) material spots on the ceiling (approximately (b) (4) above the (b) (4) Floors had stains and peeling seals between the floor and wall behind the (b) (4) Room corners near the ceiling showed dust-like material and cracks on February 2, 2026. The (b) (4) cleaning record showed that the room was last cleaned on (b) (4) and there was no documentation on the logbook for (b) (4) cleaning of ceiling. b. Clean room (b) (4) Plant (b) used for (b) (4) API (b) (4) was not in clean condition on February 2, 2026. Floors showed (b) (4) liquid like substance leaking from (b) (4) (b) (4) of (b) (4) materials scattered on the floor next to (b) (4) and (b) (4) liquid stains on floors and walls. The (b) (4) cleaning record showed that the room was cleaned on (b) (4) equipment (b) (4) inside room (b) (4) showed apparent (b) (4) deposits throughout, despite cleaning logs showing the last cleaning on (b) (4) c. (b) (4) contained (b) (4) inside despite status boards showing "cleaned" on February 2, 2026. Equipment cleaning logs indicated that cleaning was performed on (b) (4) d. Clean room (b) (4) area in Plant (b) showed (b) (4) deposits on the floor, (b) (4) runover stains on walls, rusted (b) (4) pipelines, rust like deposits on the (b) (4) to the (b) (4) and (b) (4) stains on the (b) (4) of the (b) (4) on February 4, 2026. (b) (4) API is exposed in this room before loading and unloading from the (b) (4)
- Observation 5
Inadequate raw material sampling procedures
Failure to have adequate written procedures for the sampling of raw materials. Your firm has inadequately written SOPs for raw material sample collection and in-process control samples that fail to provide sufficient instructions for QC personnel and operators. SOP QCG/006-05 (Sampling of Raw materials and packing materials) lacks detailed information on how to collect samples from raw material bags. SOP PDG/032-02 (SOP for sampling procedure of in-process and intermediate sample) does not provide adequate details for collecting in-process samples for (b) (4)
- Observation 6
Failure to follow laboratory controls procedures
Failure to follow and document laboratory controls at the time of performance and failure to document and explain any departures from laboratory procedures. a. Your QC microbiology laboratory analysts failed to document original inoculation counts in growth promotion testing (GPT) data sheets for (b) (4) media for (b) (4) Vendor lot (b) (4) in house lot (b) (4) Additionally, growth promotion results for (b) (4) media (b) (4) (b) (4) media (b) (4) vendor lot (b) (4) in house lot (b) (4) were documented as "positive" on (b) (4) exceeding the <(b) (4) timeframe specified in SOP QMG/009-03 "SOP for procedure of Media efficacy testing-growth promotion and inhibitory property test." b. Your QC microbiology laboratory analysts confirmed on February 4, 2026, that growth promotion testing (GPT) was not performed for (b) (4) (lot (b) (4) and (b) (4) (vendor lot (b) (4) lots that were received and currently used for microbiological testing of finished products, environmental monitoring and water testing. Furthermore, your QC Manager confirmed that method verification was not performed for growth promotion studies. c. Your firm failed to initiate laboratory incidents when analysts did not correctly follow HPLC processing methods to analyze (b) (4) assay data for (b) (4) API intended for the US market by HPLC. Additionally, your firm has not established an auto-integration processing method for analyzing HPLC chromatography data for (b) (4) intended for the US market.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Apr 17, 2026. It asked for a Form 483.
Inspection ending Feb 6, 2026VAI
1 request. In the inspection history above. FDA has posted its Form 483.
- Apr 17, 2026FDA-FOIA-2026-3942Form 483· 1 of 2 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
FDA drug establishment registration
FDA registration search- Registered operations
- API manufacturing
- Registration runs through
- Dec 31, 2027
- Registrant
- SNJ Labs Pvt. Ltd.D-U-N-S 932580580
From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3026587575 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3026587575. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.