Equipment qualification and calibration in FDA Form 483 observations

Showing that equipment and instruments work as intended: installation and operational qualification, calibration and preventive maintenance. FDAgent finds 153 such observations in 115 Form 483s issued to 108 facilities, for inspections that ended between October 2008 and September 2026. Each is quoted below from the 483 FDA posted.

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.

Observations
153
Each quoted below
Form 483s
115
Posted by FDA
Facilities
108
Distinct FEIs
Most recent
Sep 9, 2026
Inspection end date

Observations by year

These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).

Peak: 50 in 2026

Equipment qualification and calibration observations in Form 483s by inspection year
YearObservations
20082
20090
20100
20110
20121
20131
20140
20150
201611
20179
20189
20199
20200
20214
20229
202316
20249
202523
202650

How FDAgent sorts observations into this theme

Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.

It is tied to one of these 21 CFR entries

21 CFR 211.6321 CFR 211.68(a)21 CFR 211.160(b)(4)21 CFR 820.70(g)21 CFR 820.72

A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.

Or its text contains one of these phrases

“installation qualification” · “operational qualification” · “equipment qualification” · “IQ” · “OQ” · “PQ” · “IOQ” · “calibrat*” · “preventive maintenance” · “preventative maintenance” · “requalif*”

* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.

The 153 observations, newest first

Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.

  1. Greenwich RxInspection ended Sep 9, 2026 · FEI 3043120213 · Observation 10

    Inadequate testing on (b)(4) used to sterilize drug products. Specifically, during a review of your firm's (b)(4) test of (b)(4) used in the compounding of finished drug sterile products, your firm failed to use calibrated gauge(s) and equipment to ensure adequate testing and (b)(4) were capable of meeting their intended use.

    Contains “calibrat*”Read the Form 483 →

  2. RC Outsourcing LLCInspection ended Sep 3, 2026 · FEI 3011967886 · Observation 5

    Routine calibration, inspection and checking of automatic, mechanical and electronic equipment is not performed according to a written program designed to assure proper performance. Specifically, Your firm has not performed temperature mapping under the worst-case conditions for your critical equipment during equipment qualification. For example, your firm operates incubators (b)(4); and (b)(4) refrigerators (b)(4), all of which are used for …

    Opens with FDA's citation wording for 21 CFR 211.68(a)Read the Form 483 →

  3. TAILSTORM HEALTH INCInspection ended Aug 28, 2026 · FEI 3015929581 · Observation 2

    … The most recent (b) (4) qualification is documented in (b) (4).PQ.0004, Performance Qualification-(b) (4) Report, (b) (4) PE001, report approved 06/19/26. C. Your procedures only require a (b) (4) air removal test during (b) (4) qualification of the (b) (4). Technicians do not perform (b) (4) tests before any routine (b) (4) cycles. There is inadequate written justification for the frequency of air removal tests. Technicians use (b) (4) PE001 to …

    Contains “PQ”Read the Form 483 →

  4. TAILSTORM HEALTH INCInspection ended Aug 28, 2026 · FEI 3015929581 · Observation 3

    Equipment used in the manufacture, processing, packing or holding of drug products is not of appropriate design to facilitate operations for its intended use and cleaning and maintenance. Specifically, [illegible] System (b) (4) A. We observed a leaking diaphragm valve after the final (b) (4) step on the (b) (4) skid used to supply (b) (4) for all sterile injectable drugs that contain (b) (4). Examples of drugs that contain (b) (4) from this …

    Opens with FDA's citation wording for 21 CFR 211.63Read the Form 483 →

  5. TAILSTORM HEALTH INCInspection ended Aug 28, 2026 · FEI 3015929581 · Observation 4

    … Dispensing room Balance PE026 was found out-of-tolerance at the (b) (4) and (b) (4) weights during the 10/08/25 calibration. Balance PE026 was found out-of-tolerance at the (b) (4) weight during the 04/15/25 calibration. There was no investigation documented. E. Dispensing room Balance PE027 was found out-of-tolerance at the (b) (4) and (b) (4) weights during the 10/08/25 calibration. Balance PE027 was found out-of-tolerance at the (b) (4) and …

    Contains “calibrat*”Read the Form 483 →

  6. Carolina Infusion LLCInspection ended Aug 18, 2026 · FEI 3015826782 · Observation 3

    Inadequate (b)(4) on (b)(4) used to sterilize drug products. Specifically, on (b)(4) after the manufacture of tirzepatide for injection 26 mg/ml lot 08052026(b)(4) and bulk tirzepatide/cyanocobalamin for injection 26 mg/2 mg per ml lot 08052026(b)(4) we observed that the (b)(4) was not performed since the gauge used to test the (b)(4) appeared to be out of calibration and stuck on the pressure reading of approximately 18 PSI. This resulted in the …

    Contains “calibrat*”Read the Form 483 →

  7. IntegraDose Compounding Services LLCInspection ended Aug 12, 2026 · FEI 3014483112 · Observation 2

    … Following the start of operations at your new facility on February 2, 2026, your firm did not perform requalification of visual inspectors or AQL inspectors for the new facility. Instead, your firm continued to use qualifications performed at your previous facility. No assessment or justification for carrying over previous facility qualifications to the new facility was identified during the inspection. Visual inspection qualification is specific …

    Contains “requalif*”Read the Form 483 →

  8. Rubicon Research LimitedInspection ended Jul 3, 2026 · FEI 3016480976 · Observation 1

    Equipment used in the manufacture, processing, packing or holding of drug products is not of appropriate design, of adequate size and suitably located to facilitate operations for its intended use. Specifically, (b)(4) used in the [illegible] 122, which is listed as the equipment [illegible] for manufacturing for (b)(4) mg strengths of (b)(4) Tablets, is not qualified. Only an installation qualification has been completed. B) (b)(4) Machine …

    Opens with FDA's citation wording for 21 CFR 211.63Read the Form 483 →

  9. Eugia Steriles Private LimitedInspection ended Jun 19, 2026 · FEI 3027357163 · Observation 2

    Procedures designed to prevent microbiological contamination of drug products purporting to be sterile did not include adequate validation of the aseptic process. Specifically, A. According to the Quality Head, the following (b)(4) sensors, identified as part of the vial (b)(4) process, have not been calibrated or verified: System Component P&ID (Sensor Names) (b)(4) (b)(4) System (b)(4). There is no calibration record for such sensors involving …

    Contains “calibrat*”Read the Form 483 →

  10. Eugia Steriles Private LimitedInspection ended Jun 19, 2026 · FEI 3027357163 · Observation 4

    … The manufacturing investigation attributed the most probable root cause to potential issues with connections to vessels during (b)(4) calibration that may not have been accounted for during manufacturing final volume make-up. Additionally, (b)(4) during calibration present in the vessel (b)(4) may have been drained prior to final volume make-up. The operator stated that he likely forgot to add the vessel lid for (b)(4) of the batch during the …

    Contains “calibrat*”Read the Form 483 →

  11. Jubilant HollisterStier LLCInspection ended Jun 17, 2026 · FEI 3010477 · Observation 4

    Equipment used in the manufacture, processing, packing or holding of drug products is not of appropriate design to facilitate operations for its intended use. Specifically, the quality unit failed to requalify the existing (b)(4) system following the addition of an (b)(4) storage tank and distribution loop in 2011. The (b)(4) system supplies the (b)(4) system, which is used in the manufacture of (b)(4) drug products. A) No Installation …

    Opens with FDA's citation wording for 21 CFR 211.63Read the Form 483 →

  12. Lyophilization Services of New England, Inc. dba PCI Pharma ServicesInspection ended Jun 10, 2026 · FEI 3008546445 · Observation 1

    … 7.1 General Requirements, Subsection 7.1.1 specifies (b)(4) and/or (b)(4) intervals is based on visual inspection of each system for discoloration during preventative maintenance activities as performed at a minimum every (b)(4) years. You fail to follow the procedure with routine preventative maintenance failing to visually inspect stainless steel surfaces for absence of (b)(4). k. On 28 May 2026, observed were (b)(4) sterilized rotary pumps …

    Contains “preventative maintenance”Read the Form 483 →

  13. Lyophilization Services of New England, Inc. dba PCI Pharma ServicesInspection ended Jun 10, 2026 · FEI 3008546445 · Observation 3

    … PQ-J0001-001, Environmental Monitoring Performance Qualification (EMPQ) of Clean Rooms at the LSNE (b)(4), issue date 28 August 2023 identifies a limit of personnel to be present in gowning (b)(4) 120-C at any given time. SOP 07593, "Gowning Procedure and Personnel Flows", Revision 6, Effective date 20 February 2026 fails to prevent the potential for contamination in personnel gowning (b)(4) 120-C by not restricting access based on the EMPQ. You …

    Contains “PQ”Read the Form 483 →

  14. Lyophilization Services of New England, Inc. dba PCI Pharma ServicesInspection ended Jun 10, 2026 · FEI 3008546445 · Observation 5

    Equipment used in the manufacture, processing, packing or holding of drug products is not of appropriate design to facilitate operations for its intended use. Specifically, your firm is unable to use some (b)(4) (b)(4) without breaking first air to open vials. For example, (b)(4) on the (b)(4) (b)(4) which your firm uses for routine interventions during the filling process for sterile drug product (b)(4), cannot be introduced into the (b)(4) …

    Opens with FDA's citation wording for 21 CFR 211.63Read the Form 483 →

  15. PQ Pharmacy LLCInspection ended Jun 1, 2026 · FEI 3017374013 · Observation 2

    Equipment used in the manufacture, processing, packing or holding of drug products is not of appropriate design to facilitate operations for its intended use. Specifically, your firm has not established adequate procedures and controls for the maintenance, certification, and qualification of HEPA filters, laminar air flow hoods (LAFH), biological safety cabinets (BSC), and cleanroom HVAC systems used for sterile drug production. HVAC failures …

    Opens with FDA's citation wording for 21 CFR 211.63Read the Form 483 →

  16. PQ Pharmacy LLCInspection ended Jun 1, 2026 · FEI 3017374013 · Observation 3

    … DEV2025053 08/28/2025 Complete failure of the cleanroom AC unit, attributed to advanced coil corrosion, and inadequate preventative maintenance, specifically a lapse in routine inspection and servicing of the HVAC coils. DEV2026005 04/13/2026 inspection the unit where he observed low coolant level. DEV2026006 04/17/2025 [Technician arrived at the facility at 20APR2020 and inspected the unit where he observed a leak in the evaporator coil.] E) …

    Contains “preventative maintenance”Read the Form 483 →

  17. PQ Pharmacy LLCInspection ended Jun 1, 2026 · FEI 3017374013 · Observation 4

    Records of the inspections of automatic, mechanical or electronic equipment, including computers or related systems are not maintained. Specifically, A) During post-use (b)(4) of the sterilizing (b)(4) used for prednisolone sodium phosphate 1% / moxifloxacin HCl 0.5% / bromfenac 0.09% PF, Lot PMB121525CH BUD 12/15/2026, two (2) consecutive tests resulted in the error message "Error: (b)(4) Not Obtainable"; neither were recorded in the batch …

    Opens with FDA's citation wording for 21 CFR 211.68(a)Read the Form 483 →

  18. Shilpa Medicare LimitedInspection ended May 29, 2026 · FEI 3018694237 · Observation 2

    … On 10/22/2025, you recorded a calibration failure in test parameter wavelength accuracy during the calibration of Agilent HPLC ID #QC-E080 where peak was not eluted in the standard (b) (4) solution. You notified your service provider on 10/28/2025 and visited your firm on 12/26/2025 for repair. You attributed the root cause to deterioration of DAD lamp and source of lens after replacing the DAD lamp and recalibrated the equipment. You then stated …

    Contains “calibrat*”Read the Form 483 →

  19. Shanghai AmbioPharm, Inc.Inspection ended May 22, 2026 · FEI 3006494603 · Observation 3

    … During the review of the qualification document, [illegible] Requalification Report", Effective Date: 06/27/2025, I observed the assessed [illegible] during the qualification were [illegible] However, during the review of the batch record for Batch No: [illegible] I observed the actual (b)(4) and (b)(4) parameters are not recorded.

    Contains “requalif*”Read the Form 483 →

  20. JOINN Laboratories (Suzhou) Co. LtdInspection ended May 15, 2026 · FEI 3012611386 · Observation 1

    … Section 4.5.1.2 of SOP ADM-029 Version 15.0, "Use and Management of Equipment," effective April 2, 2026, states, "When using the following externally calibrated/verified equipment, their corrected values or error values should be introduced ***. Hygrometers used for monitoring environmental temperature and humidity (excluding EBl system monitoring probes)." The Room Temperature and Humidity Record (Form BA-041-F03) for Room 8237 recorded …

    Contains “calibrat*”Read the Form 483 →

21 CFR sections these observations open with

The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.

Ask FDAgent about equipment qualification and calibration in Form 483s

FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
  • Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
  • A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.
  • FDAgent is independent of FDA.