The 8 observations on this Form 483
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.
- 1
Inadequate validation of aseptic and sterilization process
Procedures designed to prevent microbiological contamination of drug products purporting to be sterile did not include adequate validation of the aseptic and sterilization process. Specifically, A. Your firm conducted Airflow Visualization (AFV) studies (smoke studies) for ISO 5 Laminar Flow Hood (LFH) and Laminar Flow Hood (LFH) under both static and dynamic conditions in August 2025 and February 2026. No formal study reports or supporting documentation were available for review for either study. The records lacked the following critical study elements: the type of smoke sources, the stated study purpose, targeted test locations, and personnel activities performed during dynamic conditions, real-time observations, and conclusions. Although video recordings of the AFV studies were captured, they were not accompanied by written protocols, contemporaneous documentation, or summary reports sufficient to contextualize, interpret, or substantiate the visual findings. B. Discrepancies were identified in the Airflow Visualization (AFV) records (videos) for ISO 5 Laminar Flow Hood (LFH) and Laminar Flow Hood (LFH) from studies completed in August 2025 and February 2026. The following deficiencies were observed: i. Incomplete Coverage of Critical Production Steps. Across both August 2025 and February 2026 AFV studies, the dynamic airflow visualization did not encompass all critical steps performed during commercial production operations. For example, the following steps were not captured in either study period: needle, syringe preparation, cap preparation, opening of syringe and cap boxes (trays of (b)(4) units), drawing drug product from vials into a mL syringe using (b)(4), picking up and attaching caps to filled syringes, and the capping process. Only the filling of each mL syringe to the 0.07 mL mark was consistently captured. This deficiency was observed across multiple video records. ii. Stationary Fogger Wand and Absence of Airflow Visualization Over Critical Work Zones. In multiple video records from February 2026, the fogger wand was held stationary rather than maneuvered to cover all critical work surfaces and product-contact areas. For example: In the provided video (LFH Repacking, dynamic), no airflow visualization was observed over the syringe to be filled or cap containers from T=[illegible] to T=[illegible] and no AFV was observed over the filling area at T=[illegible]. In the provided video (LFH Sterile, dynamic), the fogger wand was stationary, and no airflow visualization was captured over the initial collection of drug product from the (b)(4) container into the syringe (T=[illegible] to T=[illegible]). In the provided video (LFH, Repacking, dynamic), the fogger wand remained stationary throughout the entire video, with no airflow visualization captured over the syringe and cap containers at any point during the study. iii. Absence of AFV During Dynamic (Operational) Conditions. In the provided video (LFH) (February 2026), airflow visualization was observed only under static conditions (T=[illegible] to T=[illegible]). Once the operator began preparation for production, no AFV was conducted during the dynamic phase (T=[illegible] to T=[illegible]), despite dynamic studies being required to reflect actual operational conditions. iv. Absence of Active Air Sampling During AFV Studies. In all three August 2025 video records covering both LFH and LFH, no active air sampling was performed concurrent with the AFV studies. Active air sampling during AFV studies is necessary to assess and document the microbiological quality of the environment under simulated or actual operational conditions. The AFV studies for ISO 5 LFH and LFH conducted in both August 2025 and February 2026, were incomplete and did not adequately simulate or represent actual commercial production operations, including critical manual steps and interventions. These deficiencies indicate that unidirectional airflow protection over open product and product-con
Shortened here; the PDF has the full text.
- 2
Inadequate sterility failure investigations and batch review
There is a failure to thoroughly review any unexplained discrepancy and the failure of a batch or any of its components to meet any of its specifications whether or not the batch has been already distributed. Specifically, Sterility test failure investigations for aseptically re-packaged Avastin 1.25 mg/0.05 mL were inadequate and failed to identify probable root causes. Since October 2025, the firm has rejected four batches due to sterility failures. Despite the rejection of these batches, other batches manufactured on the same day and in the same laminar flow hood in the table below were released without a scientifically sound justification or supporting data demonstrating adequate risk mitigation. Fill date: 10/17/2025 Rejected lot # 20251017-6D87A8 Released lot manufactured in the 2nd hood 20251017-01A42E Released lots manufactured in the same hood 20251017-0FFE8B, 20251017-03F240, 20251017-235C2A. Fill date: 10/21/2025 Rejected lot 20251021-01F7A1 Released lot manufactured in the 2nd hood 20251021-8CC0D9 Released lots manufactured in the same hood 20251021-1BEBE6, 20251021-5C30B4, 20251021-817D6D. Fill date: 3/09/2026 Rejected lot 20260309-8FE1A0 Released lot manufactured in the 2nd hood 20260309-141A33 Released lots manufactured in the same hood 20260309-3645AC, 20260309-80AC52. Fill date: 6/15/2026 Rejected lot 20260615-4D607D Released lot manufactured in the 2nd hood 20260615-812D57 Released lots manufactured in the same hood 20260615-7E9720, 20260615-42D2D3. Your firm's failure to identify probable root causes for sterility test failures raises significant concerns regarding the robustness of the aseptic filling operations and the integrity of the contamination control strategy. The release of concurrent batches — manufactured under the same conditions as the rejected lots without adequate scientific justification or documented risk mitigation poses a direct risk to patient safety, as those products may not meet sterility assurance requirements.
- 3
Deficient separate areas to prevent contamination
Separate or defined areas to prevent contamination or mix-ups are deficient regarding the manufacturing and processing operations. Specifically, A. Personnel and material flow into and out of the ISO 7 and ISO 8 classified areas occurs through a single shared pathway that passes directly through the visual inspection and labeling room. Gowning, degowning, and material transfer — including finished product exiting the cleanroom and incoming components entering — all traverse the same corridor used for labeling and visual inspection of previously released product, creating a convergence of traffic that is inconsistent with unidirectional flow principles. Specifically, operators completing shoe covers, head covers, masks, and hand washing in the non-classified area must traverse this shared labeling room pathway to reach the ISO 8 entry door. Upon degowning, operators must similarly walk from ISO 7, through ISO 8, and through this same labeling room, where other staff are actively performing packaging, labeling, and visual inspection activities, to reach an outside corridor. No designated degowning area exists within the ISO 8 anteroom. Additionally, no physical or spatial separation exists within the ISO 8 anteroom to support or delineate a gowning zone, and operators were observed leaning against ISO 8 walls and tables for stability during gowning, introducing a potential source of contamination to sterile garments. B. Within this same access route, a door handle providing passage between the ISO 7 and ISO 8 classified areas and the labeling room was observed to be visibly degraded due to repeated exposure to cleaning and disinfecting agents. Furthermore, the curved geometry of the handle prevents adequate cleaning and disinfection of all contact surfaces. C. Airflow vents in the ISO 7 cleanroom and ISO 8 anteroom are obstructed. For example, multiple items — including storage containers, a table, and (b)(4) bottles — were observed blocking the air vents in the ISO 8 anteroom. The shared personnel and material pathway through the visual inspection and labeling room, the absence of a designated gowning and degowning zone, a deteriorating and inadequately cleanable door handle, and obstructed airflow vents collectively represent significant deficiencies in contamination control infrastructure. These conditions compromise the environmental integrity of the ISO 7 and ISO 8 classified areas and present a risk of cross-contamination to finished product, released materials, and ongoing sterile production operations.
- 4
Lack of adequate written sanitation procedures
There is a lack of written procedures assigning responsibility, providing cleaning schedules and describing in sufficient detail the methods, equipment and materials to be used for sanitation. Specifically, A. Your firm's cleaning procedure is inadequate. i. Your firm stated that only the ISO 5 laminar flow hood receives cleaning (b)(4). However, your firm failed to clean the ISO 7 classified area (b)(4) leaving cleaning to the end of production for the day. ii. Your firm lacks scientifically established and data-supported cleaning intervals for the cleanroom. A review of cleaning records revealed a 49-day gap between the deep cleaning performed on 02/09/2026 and the subsequent cleaning on 03/30/2026. Furthermore, your firm acknowledged that scheduling cleaning on a (b)(4) basis — rather than at fixed intervals — could permit up to (b)(4) between cleaning events. Cleaning intervals of this duration, without scientific justification or validated hold-time data, present an unacceptable risk of microbial proliferation and environmental contamination in a classified cleanroom, potentially compromising the sterility and quality of drug products manufactured therein.
- 5
No temperature mapping under worst-case conditions
Routine calibration, inspection and checking of automatic, mechanical and electronic equipment is not performed according to a written program designed to assure proper performance. Specifically, Your firm has not performed temperature mapping under the worst-case conditions for your critical equipment during equipment qualification. For example, your firm operates incubators (b)(4); and (b)(4) refrigerators (b)(4), all of which are used for product storage and environmental monitoring (EM) sample incubation. A review of the initial qualification records for these units revealed that temperature mapping was performed exclusively under empty-state conditions. When asked whether temperature mapping under operationally loaded conditions had been performed to reflect the units' current operational use, your firm confirmed that only the initial empty-state mapping had been completed and no subsequent studies have been initiated.
- 6
Unjustified subvisible particle specifications for Avastin
Laboratory controls do not include the establishment of scientifically sound and appropriate specifications designed to assure that drug products conform to appropriate standards of identity, strength, quality and purity. Specifically, Your firm continues to fail to provide a scientific rationale to support the established method and limits for subvisible particle testing of Avastin (bevacizumab) 1.75 mg/0.07 mL for intravitreal injection. Your firm releases this product using a subvisible particulate matter specification of (b)(4) particles/container at (b)(4) um and (b)(4) particles/container at (b)(4) um. All lots associated with endophthalmitis adverse events were released using this particulate matter specification. This observation is a repeat observation from the previous FDA 483, List of Observations, issued on 11/04/2024.
- 7
Employees not given required GMP training
Employees are not given training in the particular operations they perform as part of their function, current good manufacturing practices and written procedures required by current good manufacturing practice regulations. Specifically, A. Your firm does not conduct periodic CGMP training for employees. B. Not all personnel receive basic training on the theory and practice of the Quality Management System and Good Manufacturing Practices, as required, per SOP-6 Training and Personnel, v2.0, effective September 30, 2025, Section 5.2.1. C. Your firm has not established training or qualification requirements to ensure that personnel responsible for reading and interpreting EM plate results are able to accurately enumerate and differentiate microbial growth and surface colonization.
- 8
Expiration dates unsupported by stability studies
Drug product expiration dates are not related to the storage conditions stated on the labeling, as determined by stability studies. Specifically, Your firm's product labels do not specify required storage temperature conditions, and the assigned beyond-use date (BUD) is not supported by stability testing. For example: your firm has failed to specify required storage temperature conditions on product labels, despite assigning a beyond-use date (BUD). For example, your firm stated that stability testing is not conducted for its non-sterile compounded products on the basis that the water activity content is less than [illegible]%, with a BUD assigned at a controlled room temperature of [illegible]. However, all three non-sterile product labels reviewed during the inspection indicated only "room temperature," without further qualification or specification of the required temperature range, which is inconsistent with your firm's stated storage conditions.
Other Form 483s FDA posted for this facility
Every Form 483 FDA posts under FEI 3011967886, newest first.
- Form 483, August 2025Inspection ended Aug 28, 2025 · Posted Sep 22, 2025OAI
- Form 483, September 2024Inspection ended Sep 27, 2024 · Posted Dec 12, 2024OAI
- Form 483, September 2022Inspection ended Sep 23, 2022 · Posted Feb 3, 2023OAI
- Form 483, February 2019Inspection ended Feb 8, 2019 · Posted May 5, 2020VAI
- Form 483, February 2019Inspection ended Feb 8, 2019 · Posted Jul 9, 2020VAI
- Form 483, March 2017Inspection ended Mar 31, 2017 · Posted Jan 17, 2024OAI
- Form 483, March 2016FDA record date Mar 18, 2016 · Posted Jan 17, 2024
- Form 483, February 2016Inspection ended Feb 9, 2016 · Posted Jan 17, 2024VAI
Ask FDAgent about this Form 483
FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.
Related
About this data
- The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
- The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
- FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
- FDAgent is independent of FDA.