Skip to content

JOINN Laboratories (Suzhou) Co. Ltd Form 483, May 2026

FDA issued this Form 483 to JOINN Laboratories (Suzhou) Co. Ltd in China (FEI 3012611386) at the end of an inspection that ended May 15, 2026. It lists 2 observations.

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Nov 12, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
May 15, 2026
—
FDA's outcome
Not classified
Not classified yet
Posted by FDA
Nov 12, 2026
181 days after the inspection
Facility
FEI 3012611386
China

The 2 observations on this Form 483

Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.

  1. 1

    Management failed to ensure personnel understood functions

    21 CFR 58.83

    Testing facility management failed to assure that all personnel clearly understood the functions they were to perform. Specifically, a. Section 4.3.1 of SOP ADM-027 Version 16, "Archive Administration," effective July 5, 2023, states, "The study number, archive date, specimen type, storage location, amount and other relevant information shall be documented in database (the Ziguang Archive management software)." For study (b)(4), the Ziguang archive management software entry listed 63 wet tissue bags on May 11, 2026; however, 62 wet tissue bags were physically verified as archived on shelves 52-1-7 and 51-3-7 in Building 8, room 8103. b. Section 4.3.1 (Specimen Storage) of SOP ADM-027 Version 23.0, "Archive Administration," effective March 23, 2026, states, "When Document Management System (DMS) is adopted, the archive administrator shall enter the information such as the specific placement position and quantity of the samples." Thirty-four trays of (b)(4) blocks (e.g. (b)(4)) were temporarily stored in the Building 27 document archives (room 27104) due to space limitations in the Building 27 specimen archives (room 27102); however, DMS was not updated to reflect this temporary storage location when observed on May 11, 2026. c. Section 4.2.5 of SOP PATH-020 Version 3, "Microtomy of (b)(4) Sections," effective February 2, 2023, states, "Verify the block code and record the block code on the glass slide (or verify the block code if the glass slide is printed with a block code)." For study (b)(4), upon comparing the (b)(4) blocks to the corresponding slides, the slide generated from (b)(4) block (b)(4) code 3 (brain section 3, dorsal) was erroneously labeled as (b)(4) code 3B (brain section 3, ventral), and the slide generated from (b)(4) block (b)(4) code 3B (brain section 3, ventral) was erroneously labeled as (b)(4) code 3 (brain section 3, dorsal). d. 21 CFR 58.83 states, "All reagents and solutions in the laboratory areas shall be labeled to indicate identity, titer or concentration, storage requirements, expiration date." (b)(4) Mobile phase for GLP study found in room 8232 was not labeled with an expiration date or (b)(4) Wrong number storage conditions. e. Section 4.5.1.2 of SOP ADM-029 Version 15.0, "Use and Management of Equipment," effective April 2, 2026, states, "When using the following externally calibrated/verified equipment, their corrected values or error values should be introduced ***. Hygrometers used for monitoring environmental temperature and humidity (excluding EBl system monitoring probes)." The Room Temperature and Humidity Record (Form BA-041-F03) for Room 8237 recorded temperature and humidity readings twice daily without applying the corrected values from April 2, 2026, through May 12, 2026.

  2. 2

    Facility size and construction not suitable

    The testing facility is not of suitable size and construction to facilitate the proper conduct of nonclinical laboratory studies. Specifically a. In Building 7, where reserve animals were being held, the floor in one area of animal room 7176 did not slope toward the drain. We observed discoloration and pooling liquid in one part of this area. b. In Building 7, in the corridor near room 7178, we observed bubbling and cracking in the surface of the PVC floor. c. In Building 5, near animal room 5264, the vertical locking mechanism for the door separating the silent area corridor would not stay in place. We observed non-human primates pushing this door open while undergoing observations.

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.