Equipment qualification and calibration in FDA Form 483 observations
Showing that equipment and instruments work as intended: installation and operational qualification, calibration and preventive maintenance. FDAgent finds 153 such observations in 115 Form 483s issued to 108 facilities, for inspections that ended between October 2008 and September 2026. Each is quoted below from the 483 FDA posted.
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.
- Observations
- 153
- Each quoted below
- Form 483s
- 115
- Posted by FDA
- Facilities
- 108
- Distinct FEIs
- Most recent
- Sep 9, 2026
- Inspection end date
Observations by year
These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).
Peak: 50 in 2026
| Year | Observations |
|---|---|
| 2008 | 2 |
| 2009 | 0 |
| 2010 | 0 |
| 2011 | 0 |
| 2012 | 1 |
| 2013 | 1 |
| 2014 | 0 |
| 2015 | 0 |
| 2016 | 11 |
| 2017 | 9 |
| 2018 | 9 |
| 2019 | 9 |
| 2020 | 0 |
| 2021 | 4 |
| 2022 | 9 |
| 2023 | 16 |
| 2024 | 9 |
| 2025 | 23 |
| 2026 | 50 |
How FDAgent sorts observations into this theme
Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.
It is tied to one of these 21 CFR entries
21 CFR 211.6321 CFR 211.68(a)21 CFR 211.160(b)(4)21 CFR 820.70(g)21 CFR 820.72
A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.
Or its text contains one of these phrases
“installation qualification” · “operational qualification” · “equipment qualification” · “IQ” · “OQ” · “PQ” · “IOQ” · “calibrat*” · “preventive maintenance” · “preventative maintenance” · “requalif*”
* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.
The 153 observations, newest first
Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.
Fresenius Kabi USA, LLCInspection ended May 14, 2026 · FEI 3001833549 · Observation 2
… For example, Test (b) (4) ((b) (4) vial), which was used to requalify visual inspector on 1/21/2026, contains (b) (4) accept units and [illegible] defect units. Of the (b) (4) defect units, [illegible] contain foreign material particulates of the following types and sizes: (b) (4). Your threshold study established a PoD of [illegible] (b) (4)% for (b) (4) ((b) (4) stopper) particulates; however, defects at this size are not represented in Test …
Contains “requalif*”Read the Form 483 →
Galaxy Medical Products LLCInspection ended Apr 28, 2026 · FEI 3010888324 · Observation 1
The quality control unit lacks authority to review production records to assure that no errors have occurred and fully investigate errors that have occurred. Specifically, Your firm failed to document and investigate the use of an out-of-calibration thermometer during the filling of (b) (4) cylinders. On 04/27/2026, an out-of-calibration thermometer (Serial # (b) (4)) that had exceeded its calibration due date by approximately four months was …
Contains “calibrat*”Read the Form 483 →
Jiangsu Hansoh Pharmaceutical Group Co Ltd.Inspection ended Apr 24, 2026 · FEI 3008316144 · Observation 6
Routine calibration, inspection, checking of automatic, mechanical, electronic equipment is not performed according to a written program designed to assure proper performance. The following discrepancies were noted during the review of the (b) (4) weight check systems on aseptic filling lines in workshops (b) (4) A. Workshop (b) (4) The weight checks performed on the (b) (4) balances prior to use, do not bracket the amount being weighed. The …
Contains “calibrat*”Read the Form 483 →
USV Private LimitedInspection ended Apr 17, 2026 · FEI 3004086192 · Observation 6
… for sterility, lacks particle shedding evaluations for cleanroom garments and wipes, and has not conducted supplier audits or established a qualification and requalification program. For example, sterile gowns from (b) (4) were accepted without independent sterility testing, and sterile (b) (4) from (b) (4) were tested using statistically unjustified single-unit sampling, despite direct contact with surfaces in aseptic processing areas.
Contains “requalif*”Read the Form 483 →
Jiangsu Hengrui PharmaceuticalsInspection ended Apr 15, 2026 · FEI 3006689263 · Observation 1
… Software Qualification: Your firm failed to adequately qualify the RABs (b)(4) filling equipment, equipment #(b)(4) The OQ/PQ protocol, V-F-P 59 203P, is inadequate this protocol only tested the equipment's ability to reject overfills and underfills but failed to demonstrate the equipment's ability to accurately and consistently confirm fill weights within acceptable limits. Additionally, the equipment lacks data storage capability, lacks …
Contains “OQ”Read the Form 483 →
Jiangsu Hengrui PharmaceuticalsInspection ended Apr 15, 2026 · FEI 3006689263 · Observation 2
Procedures designed to prevent microbiological contamination of drug products purporting to be sterile are not established. A. Visual Inspection 1. Your firm uses only one defect kit for operator training, ongoing requalification, and performance testing. This kit contains only one example of each defect type. Some of the defects have been in use for over four years, allowing operators to repeatedly view, and are tested on identical defects …
Contains “requalif*”Read the Form 483 →
AstraZeneca ABInspection ended Apr 10, 2026 · FEI 3003342394 · Observation 2
… However, the preventative maintenance program was not updated to address the equipment wear condition that caused this deviation.
Contains “preventative maintenance”Read the Form 483 →
Conroe Welding Supply, IncInspection ended Apr 3, 2026 · FEI 1625647 · Observation 1
… Your firm failed to establish a preventative maintenance SOP for (b) (4) Pumping System that used for production and the (b) (4) used for testing and failed to establish maintenance logs. N. Your firm failed to establish a SOP for receiving bulk medical (b) (4) gas. O. Your firm failed to establish a SOP for label control, insurance and inspection of labels.
Contains “preventative maintenance”Read the Form 483 →
Adimmune CorporationInspection ended Apr 2, 2026 · FEI 3008919998 · Observation 3
Your firm failed to establish, follow and document preventive maintenance procedures, specifically, a. There is no documentation to support the replacement of (b) (4) utilized at the (b) (4) supply source for drug product overlay after (b) (4) uses according to Standard Operating Procedure (SOP) (b) (4) 140: BLD (b) (4) Filling & Packaging Line (b) (4) The Standard of Integrity Test in BLD (b) (4) Filling & Packaging Plant Line (b) (4), effective …
Contains “preventive maintenance”Read the Form 483 →
Hospira, Inc.Inspection ended Mar 10, 2026 · FEI 1925262 · Observation 5
… to replace ge The Quality Unit did not ensure that the current PM aligned with the QRMR, demonstrating inconsistency between the risk assessment and actual preventative maintenance practices. ii. Change Control Oversight: Preventative Maintenance Program Change Request (PMCR) Task Plan #M-P5490QT, PM Schedule #MCP-5490, dated 7 August 2025, stated "cancel PM due to frequency changes in SOP-95715" and “gy are no longer changed gy after SOP 95715 …
Contains “preventative maintenance”Read the Form 483 →
Unilab Chemicals and Pharmaceuticals Private Ltd.Inspection ended Feb 27, 2026 · FEI 3004985767 · Observation 3
… (b) (4) Vessel No-1 (UNI(b) (4)/01) Preventive Maintenance Record Logbook page intended for (b) (4) was already pre-filled with observations for machine parts, including: (b) (4) There is no documented reasoning as to why the observation page intended for (b) (4) were already prefilled without any signature or date. E. The Microbiology Department's Burning of (b) (4) Records Equipment Log No. F/MIC/005 was missing multiple entries of Laminar Air …
Contains “preventive maintenance”Read the Form 483 →
Eugia Pharma Specialities LimitedInspection ended Feb 27, 2026 · FEI 3011960448 · Observation 3
… products work under (b)(14); however, the qualification of visual inspection the first time the employee is qualified is not performed again at subsequent requalifications. Of the visual inspection staff, (b)(4) were first qualified in (b)(4), (b)(4) were first qualified in (b)(4) with the remainder first qualified more recently, and none were challenged after (b)(4) since their first qualification.
Contains “requalif*”Read the Form 483 →
Eugia Pharma Specialities LimitedInspection ended Feb 27, 2026 · FEI 3011960448 · Observation 4
… Corrective actions in the investigation included adding new preventative maintenance to the (b)(4), adding a review of (b)(4) inspection challenge vial for larger glass pieces, and requalification of the visual inspection employees; however, there is no discussion of whether the glass vials used for the product are still acceptable or why the visual inspection was unable to see the glass in the time and conditions of the visual inspection …
Contains “preventative maintenance”Read the Form 483 →
RayzeBio, IncInspection ended Feb 24, 2026 · FEI 3030530655 · Observation 1
… CAPA-00217 was initiated to address missed pipettes calibration. CAPA-00217 was opened on 08/11/2025 to verify that pipettes were either calibrated or retired and to assess the need for a process change surrounding the calibration of pipettes. Neither action was performed, and CAPA-00217 remains open without a justification with a past due date of 11/14/2025. 2. CAPA-00218 was initiated 08/13/2025 to address overdue calibration work orders, have …
Contains “calibrat*”Read the Form 483 →
RayzeBio, IncInspection ended Feb 24, 2026 · FEI 3030530655 · Observation 3
… The (b)(4) and (b)(4) were qualified with ISO 5 classification during Installation and Operational Qualification for Suite (b)(4) and Suite (b)(4) (b)(4) ISO 5 (b)(4). However, your environmental monitoring program monitors these two (b)(4) as ISO 6 areas without adequate justification. 4. On February 2, 2026, during aseptic fill operations for lot (b)(4) in Suite (b)(4) (Room 1012), an operator touched four locations when transferring material …
Contains “operational qualification”Read the Form 483 →
RayzeBio, IncInspection ended Feb 24, 2026 · FEI 3030530655 · Observation 5
… The firm's Validation Master Plan (TD-00001, v4, effective 05/13/2025) requires Installation and Operational Qualification (IQ/OQ) for the (b)(4) generator and (b)(4) referencing change control PC-00089, which was initiated on 03/25/2025 for (b)(4) implementation in ISO 5 spaces. No progress has been made since initiation, and your firm has provided no explanation or justification for failing to implement a validated sterilization method for the …
Contains “operational qualification”Read the Form 483 →
RayzeBio, IncInspection ended Feb 24, 2026 · FEI 3030530655 · Observation 7
Routine calibration of automatic and electronic equipment is not performed according to a written program designed to assure proper performance. Specifically, On 02/18/2026, I performed a review of your (b)(4) system used for the workflows and activities associated with preventive maintenance (PM) and calibration of equipment in the production and QC laboratory. The review revealed a total of (b)(4) pipettes ((b)(4) pipettes owned by Production …
Opens with FDA's citation wording for 21 CFR 211.68(a)Read the Form 483 →
Orean Personal Care LtdInspection ended Feb 20, 2026 · FEI 3010999645 · Observation 2
… In your production lab, loose sheets of the pH calibration test data (for 2024 and 2025, 2026) were stored. The data were not reviewed by the quality assurance. Your Quality Head (b)(6) was not sure why old data were not bounded and archived. b. The Cleaning Validation Protocols and Reports were not reviewed and signed by quality assurance. c. The Process Validation Protocol and Reports were not reviewed and signed by quality assurance.
Contains “calibrat*”Read the Form 483 →
Orean Personal Care LtdInspection ended Feb 20, 2026 · FEI 3010999645 · Observation 3
… The (b)(4) system was not qualified when installed in 2022 and complete preventative maintenance records are not kept of the (b)(4) system, including the cleaning and sanitization of the system. You do not have a written procedure that guides the sampling, sanitizing, cleaning and maintenance of the (b)(4) system. C. On 2/16/2026, an expired standard of pH 7.0 buffer (Lot #(b)(4)) was found in your production lab. The label on the bottle read, …
Contains “preventative maintenance”Read the Form 483 →
MicroBioPharm Japan Co., LtdInspection ended Feb 13, 2026 · FEI 3003992082 · Observation 2
Your firm failed to establish and follow appropriate written procedures for equipment qualification. Specifically, Your firm has failed to establish and follow appropriate written procedures designed to ensure that equipment used in the manufacture of active pharmaceutical ingredients (APIs) and API intermediates is qualified for its intended use. The equipment that lacks adequate qualification or requalification include, but are not limited to, …
Contains “equipment qualification”Read the Form 483 →
21 CFR sections these observations open with
The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.
- 21 CFR 211.63Equipment design, size, and locationIn this theme's rule15
- 21 CFR 211.22Responsibilities of quality control unit10
- 21 CFR 211.113Control of microbiological contamination9
- 21 CFR 211.68Automatic, mechanical, and electronic equipmentIn this theme's rule8
- 21 CFR 211.160General requirementsIn this theme's rule6
- 21 CFR 211.192Production record review6
- 21 CFR 211.100Written procedures; deviations5
- 21 CFR 211.42Design and construction features3
- 21 CFR 820.75Process validation2
- 21 CFR 1271.200Equipment2
Ask FDAgent about equipment qualification and calibration in Form 483s
FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.
Related
About this data
- The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
- Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
- A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.
- FDAgent is independent of FDA.