FDA has 8 inspections on file for USV Private Limited in Dabhel, Daman, India (FEI 3004086192), from 2009 to 2026. The latest ended on and was classified OAI, official action indicated. FDA has issued 2 warning letters under this FEI, the latest on (CDER).
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA posted the Form 483 from that inspection. FDAgent read its 9 observations from the PDF; they are listed below. Three of the four FOIA requests for this site have asked FDA for this inspection's 483.
No court action or import refusal is filed under this FEI.
Four regulations were cited at more than one inspection; they are listed below.
As of Nov 21, 2026, the site is registered with FDA as a drug establishment for manufacturing, packing, labeling, analysis and sterilization, through Dec 31, 2027. Its registrant, USV Private Limited (D-U-N-S 650611585), has registered 2 other sites.
Sources:FDA firm profile ↗Form 483 PDF ↗Warning letter ↗FDA drug establishment registrations ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 8, 2009–2026
- 3 OAI · 5 VAI · 0 NAI
- Inspections with 483 observations shown
- 4 of 8
- Form 483s posted by FDA
- ↗
- Requested under FOIA
- 4 requests, latest Sep 10, 2026
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
- Drug establishment registration
- Through
Inspection history
8 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDrugs · Drug Quality AssurancePosted 483 · 9 observations
Read this Form 483 on its own page →Form 483 PDF on fda.gov
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.
- Observation 1
Poor aseptic practices and sanitization failures
Procedures designed to prevent microbiological contamination of drug products purporting to be sterile are not established and followed. Specifically, A. During filling of (b) (4) on the (b) (4) line for protocol DMF/MSP/26/016 "Protocol for filling of (b) (4) in (b) (4) mL (b) (4) Bottle in (b) (4) Bottle)" Filling line, poor aseptic practices were observed: 1. An operator was observed breaching first air by reaching over open sterile bottles near the (b) (4) (b) (4) during the replacement of viable air sampling plates. The bottles exposed during this breach were not removed from the line. During this intervention, the RABS (b) (4) was (b) (4) a Grade B environment and (b) (4) operator, gowned to Grade B standards, performed a sanitization procedure on the (b) (4) which was (b) (4) into the Grade B zone. This procedure involved wiping the sterilized (b) (4) with (b) (4), which removes the (b) (4) validated sterile state, rendering it only sanitized. This sanitization was further compromised when the operator's forearm and hand passed directly over the (b) (4), obstructing first air. This was not recreated in either smoke study in (b) (4) 2. An operator placed a forceps adjacent to the fill line in a manner that caused the forceps to lean over open bottles containing product, thereby obstructing first air. The bottles exposed during this breach were not removed from the line. 3. First air was breached over open product containers when the operator made multiple adjustments to the (b) (4) during active filling operations. The bottles exposed during these breaches were not removed from the line. 4. During execution of the qualified intervention "Bottle Fall on Conveyor Belt Removed by a Technician with Forceps," an operator successfully removed the first fallen bottle using forceps but when he was pushing the line of vials forward towards the (b) (4), a second bottle fell during the intervention process. Instead of removing the second fallen bottle, the operator stood the bottle upright and pushed it toward the (b) (4) using forceps. B. During manufacturing of (b) (4) Injection (b) (4) mcg/mL, batch (b) (4), an aseptic connection was performed between the (b) (4) tank and (b) (4) tubing leading to the (b) (4) tank. Despite this representing the highest risk operation for microbial contamination, the batch manufacturing record does not document when the connection was performed, its duration, or which personnel performed it. Additionally, no finger dabs are collected after this aseptic connection, and the operator performing the connection inside the Grade A area is gowned only to Grade B standards. C. During set-up and filling operations, the following equipment and materials were transferred between Grade B and Grade A areas without adequate sanitization: 1. For (b) (4) Injection (b) (4) mcg/mL, batches (b) (4) (b) (4) containers: (b) (4) used for sanitization were observed being transferred between Grade B and Grade A areas without sanitization of the exterior surfaces of the containers. These containers were used held directly over the Doctor's hands. The Doctor is the person responsible for all assembly practices within the RABS (b) (4) 2. For (b) (4) Injection (b) (4) mcg/mL, batch (b) (4) a (b) (4) used to transport environmental monitoring settle plates was moved between Grade B and Grade A areas. This (b) (4) was not sanitized before entering the Grade A environment.
- Observation 2
Inadequate Grade A separation and airflow protection
Separate or defined areas to prevent contamination or mix-ups are deficient regarding operations related to aseptic processing of drug products. Specifically, Aseptic processing operations fail to provide adequate protection of Grade A (ISO Class 5) environments and first air over critical exposure points. For example: A. Rooms (b) (4) (Vial Filling and Sealing) and (b) (4) Filling): Grade A (b) (4) laminar airflow areas are separated from Grade B areas solely by an incomplete (b) (4) (approximately (b) (4) (b) (4) barrier (b) (4). This barrier design is fundamentally inadequate to prevent air migration and cross-contamination from Grade B into Grade A areas because: • The (b) (4) barrier does not provide complete physical separation between Grade A and Grade B areas • Both Grade A and Grade B areas are maintained at the same pressure (b) (4) pascals), with no pressure differential to ensure unidirectional airflow from the more critical Grade A area to the less critical Grade B area • The combination of incomplete physical separation and absence of pressure cascade allows unrestricted air exchange between classified areas, failing to protect Grade A environments from Grade B contamination • The design provides no assurance that particulates, microorganisms, or other contaminants from the Grade B environment cannot migrate into the Grade A area where sterile products and components are exposed B. Room (b) (4) Filling Line): After leaving the sterile bottle bowl, sterile open bottles travel approximately (b) (4) on a (b) (4) transfer belt that is only sanitized (not sterilized). This conveyor passes directly under a conveyor power box containing electrical connections. The bottles then enter a (b) (4) bottle chute where they travel approximately (b) (4) before entering a (b) (4) which moves them to another conveyor track and on to the filling station. This design prevents unidirectional laminar airflow from reaching the open bottle tops throughout the (b) (4) transport pathway, exposes sterile containers to sanitized (not sterile) equipment for approximately (b) (4) (approximately (b) (4) places non-sterile equipment directly above open sterile containers, and traps any particles generated by moving conveyor mechanisms in (b) (4) spaces with the sterile bottles. Your firm has not validated through airflow visualization studies that this design provides adequate first air protection or that sanitization of the (b) (4) track and conveyor systems is sufficient to prevent contamination of sterile bottles during this (b) (4) transport where continuous Grade A protection is blocked. These design deficiencies fail to ensure that aseptic processing operations adequately protect sterile drug products from environmental contamination during critical manufacturing steps.
Show observations 3 to 9
- Observation 3
Inadequate smoke studies and media fills
Procedures designed to prevent microbiological contamination of drug products purporting to be sterile did not include adequate validation of the aseptic process. Specifically, airflow visualization studies (smoke studies) do not demonstrate appropriate airflow patterns. Additionally, airflow visualization studies and media fills fail to represent actual commercial manufacturing conditions: A. Your airflow visualization studies do not include adequate evaluation of airflow patterns during transitions of personnel, materials, and equipment between Grade B and Grade A environments within the manufacturing suite: 1. During the manufacturing of (b) (4) Injection (b) (4) mcg/mL, Batches (b) (4) (b) (4) in room (b) (4) Vial Filling & Capping Room (b) (4) operators were observed moving materials and equipment between the Grade B and Grade A sides of this barrier. The airflow visualization studies did not evaluate whether these transitions across the barrier maintain unidirectional airflow to prevent contamination of the Grade A area from the adjacent Grade B environment. 2. During the filling of (b) (4) Bottle Collection Protocol for (b) (4) in (b) mL (b) (4) Bottle on (b) (4) Filling Line (DMF/MSP/26/016) in room (b) (4) Filling (b) (4) (b) (4), operators were observed moving materials and equipment between the Grade B and Grade A sides of this barrier. The airflow visualization studies did not evaluate whether these transitions across the (b) (4) barrier maintain unidirectional airflow to prevent contamination of the Grade A area from the adjacent Grade B environment. B. Your airflow visualization studies fail to represent the most challenging conditions that occur during routine commercial production. Specifically: 1. During commercial manufacturing of (b) (4) Injection (b) (4) mcg/mL, Batch (b) (4) multiple operators were observed working simultaneously in the Grade B and Grade A areas performing aseptic connections, material transfers, and equipment adjustments. However, your airflow visualization studies only simulated these activities using (b) (4) operators, failing to evaluate the impact of multiple personnel on airflow patterns and potential air disruption in critical zones. 2. Your airflow visualization studies demonstrate inconsistencies in the evaluation of aseptic connections between different study versions, and fail to include all critical aseptic connections performed during every commercial batch: a. During your (b) (4) airflow visualization study, the aseptic connection between the sterile tubing and the (b) (4) tank attached to the manifold was performed and evaluated as an intervention; however, during your (b) (4) airflow visualization study, this same aseptic connection was not performed or evaluated, despite this connection being made during every commercial batch, including (b) (4) Injection (b) (4) mcg/mL, Batches (b) (4) This omission eliminates assessment of airflow patterns during a critical sterile operation that involves removal of sterile wrapping from connection ports, manipulation of sterile tubing, and direct exposure of product contact surfaces in the Grade A environment. b. During your (b) (4) airflow visualization study, portions of the aseptic connection of the (b) (4) tubing connecting the (b) (4) tank to the product (b) (4) tank were evaluated; however, the complete intervention was not simulated. Specifically, the passing of the sterile tubing under the filling line was not included in the smoke study evaluation. During your (b) (4) airflow visualization study, this aseptic connection was not simulated or evaluated at all, despite this complete aseptic connection being performed during commercial manufacturing of all batches on the vial filling line including (b) (4) Injection (b) (4) mcg/mL, Batches (b) (4) including passing sterile tubing under the filling line. C. Turbulent and non-unidirectional airflow observed during smoke studies: Review of airflow visualization study videos revealed i
Shortened here; the PDF has the full text.
- Observation 4
Inadequate visual inspection program for steriles
Control procedures are not established which monitor the output and validate the performance of those manufacturing processes that may be responsible for causing variability in the characteristics of in-process material and the drug product. Specifically, your visual inspection process for sterile products is inadequate: A. Your firm manufactures sterile injectable products including (b) (4) (b) (4) Injection. Review of your visual inspection procedures, statistical evaluation protocols, and trending reports revealed the following deficiencies: Specific Findings 1. No Action Limits Established - Your statistical evaluation protocols calculate only alert limits using Mean (b) (4) Standard Deviations. No action limits are established for any critical defect type. 2. No alert or action limits established for: Black particles: * No limits established for (b) (4) (all strengths: (b) (4) mcg/mL) Limits established only for (b) (4) mg/(b)(4) mL Glass particles: * No limits established for (b) (4) mg/(b)(4) mL Limits established only for (b) (4) mcg/mL Broken vials: * No limits established for any product Cracked vials: * No limits established for any product Fibers (when tracked separately from particles): * No limits established for any product Fibers are combined with particles as "particle/fiber" in trending data, preventing separate monitoring and control SOP DMF-SOP-0350 Section 5.8.11 classifies these as "Critical defects" that "may cause serious adverse reactions or death of the patient if the product is used." Your trending reports document occurrences of these defects: • Broken/cracked vials: Documented in trending data but no limits established • Fibers: Detected in multiple batches but no separate limits when not combined with particles (b) (4) Although an alert limit of (b) vials is established for glass particles in (b) (4) mcg/mL, Batch (b) (4) mcg/mL, US product, manufactured February 2023) contained (b) (4) glass particles double the alert limit yet no investigation was performed. 3. Inadequate Particle Characterization Procedures - Your procedures do not require characterization of particles and fibers to determine whether they are extrinsic (contamination) or intrinsic (process-inherent) unless an alert limit is exceeded. Your current procedure states that extrinsic particulates indicating filth, sterility assurance issues, or cGMP violations are unacceptable even if statistical sampling criteria are met. However, without characterization, you cannot determine if particles are extrinsic. For defect types without established alert limits (glass particles, fibers when tracked separately, broken/cracked vials), no investigation is triggered and particles are never characterized. This means batches may be released containing uncharacterized particles of unknown origin and unknown patient safety risk, and you cannot apply your own batch disposition criteria requiring identification of extrinsic particulates. B. For (b) (4) products packaged in (b) (4) bottles, including (b) (4) (b) (4) your visual inspection process does not include adequate destructive testing methods to ensure product quality. Your current procedure only allows for external visual inspection of (b) (4) containers, which cannot detect: 1. Particulate matter suspended in the product solution. 2. Contamination, discoloration, or phase separation of the product. 3. Foreign matter or defects that may be present inside the container but not visible through (b) (4) packaging materials. (b) (4) Although your firm initiated destructive testing on (b) (4) this practice has not been adequately implemented or validated. Specifically: 1. Lack of Operator Training: Personnel performing destructive testing have not received formal training on the procedure, acceptance criteria, or defect recognition. 2. Absence of Training Materials: No defect kit or reference standards have been established to train operators on identifying unacceptable product attributes during vis
Shortened here; the PDF has the full text.
- Observation 5
Inadequate personnel finger dab monitoring technique
Procedures designed to prevent microbiological contamination of drug products purporting to be sterile are not established, written, or followed. Specifically, personnel monitoring procedures for aseptic operations are inadequate. The finger dabbing method consisting of touching the plate with four fingers together instead of rolling each finger separately was observed on the following occasions: A. April 9, 2026: During the setup for sterile filling operations for (b) (4) Injection (b) (4) mcg, Batch (b) (4), operators rolled four fingers together as a group and the thumb separately, rather than rolling each finger individually across the (b) (4) plate surface. This technique does not ensure adequate surface area contact and sample collection from each individual finger. B. April 14, 2026: During the filling of (b) (4) on the (b) (4) line for protocol DMF/MSP/26/016 "Protocol for filling of (b) (4) in (b) mL (b) (4) Bottle in (b) (4) Bottle)", (b) (4) operators continued to demonstrate the same inadequate finger dabbing technique observed on April 9, 2026. Without proper contact between each finger and the media plate, accurate results cannot be obtained to detect microbial contamination on operators' (b) (4)
- Observation 6
No supplier qualification for aseptic consumables
Written procedures for qualifying, controlling, and evaluating suppliers and materials used in aseptic processing areas are not established, written or followed. Specifically, your firm lacks a comprehensive, risk-based program for qualifying and monitoring suppliers of critical consumables used in ISO 5 (Grade A) areas, including garments, (b) (4), wipes, and disinfectants. Your firm relies solely on supplier Certificates of Analysis without performing independent verification testing for sterility, lacks particle shedding evaluations for cleanroom garments and wipes, and has not conducted supplier audits or established a qualification and requalification program. For example, sterile gowns from (b) (4) were accepted without independent sterility testing, and sterile (b) (4) from (b) (4) were tested using statistically unjustified single-unit sampling, despite direct contact with surfaces in aseptic processing areas.
- Observation 7
Missing component identity testing for impurity
Each component is not tested for conformity with all written specifications for purity, strength, and quality. Specifically, your firm lacks appropriate testing of components used in the manufacture of your drug product. For example, your firm lacks a specific identity test to detect (b) (4) in all shipments, containers, and lots of (b) (4) before use in the manufacturing of (b) (4) Tablets. In the last two years your firm has manufactured over (b) batches of (b) (4) Tablets for distribution into the US market including but not limited to: batches (b) (4) (b) (4) The product has an expiration date of (b) (4)
- Observation 8
Inadequate OOS investigation and sampling justification
There is a failure to thoroughly review any unexplained discrepancy and the failure of a batch or any of its components to meet any of its specifications whether or not the batch has been already distributed. Specifically, your firm's investigation into the out-of-specification (b) (4) results for (b) (4) (b) (4) Tablets, USP (b) (4) mg (Batch (b) (4) OOS DEV-08261) is inadequate because it fails to provide sufficient scientific justification for the proposed change from (b) (4) sampling (b) (4) mg) to (b) (4) sampling (b) (4) mg) and does not correlate (b) (4) (b) (4) results with (b) (4) dosage unit (b) (4). Your investigation concludes that the (b) (4) sampling plan is "susceptible to variations in sample (b) (4) resulting in variability of the amount of collected (b) (4) sample and impacting the (b) (4) result" due to "(b) (4)" in this (b) (4) (b) (4) however, your firm did not explain why this sampling plan was acceptable for (b) (4) previously manufactured and released batches between 2023-2025, nor does it provide data demonstrating that these released batches exhibited acceptable (b) (4) in the (b) (4) dosage units despite allegedly inadequate (b) (4) sampling. Furthermore, the firm proposes that larger sample sizes (b) (4) minimum (b) (4) mg) are necessary to achieve representative (b) (4) (b) (4) results, yet each individual tablet contains only (b) (4) mg of (b) (4) significantly less than the minimum (b) (4) sampling size (b) (4) mg). Your firm failed to explain how the (b) (4) process, which utilizes an even smaller (b) (4) per unit (b) (4) mg) than the allegedly inadequate (b) (4) sampling (b) (4), can consistently produce tablets with acceptable (b) (4) if the (b) (4) does exhibit the degree of (b) (4) claimed in the investigation.
- Observation 9
Incomplete lab data from Empower processing changes
Laboratory records do not include complete data derived from all tests, examinations, and assay necessary to assure compliance with established specifications and standards. Specifically, Your analysts have been given permission within the Empower software to view numerical results in the review window, allowing them to see area counts of peaks while changing the processing methods, without requiring the chromatograms to be saved. Using this method your firm have tested over (b) (4) batches of finished drug products for assay by HPLC in the last two years. Using this method your firm has tested over (b) (4) batches of finished drug products for assay by HPLC in the last two years including but not limited to (b) (4) (b) (4)
- OAISignificant objectionable conditions; FDA recommended regulatory actionDrugs · Drug Quality Assurance10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Procedure Deviations Recorded and Justified
Deviations from written production and process control procedures are not recorded and justified.
Validation lacking for sterile drug products
Procedures designed to prevent microbiological contamination of drug products purporting to be sterile did not include validation of the aseptic process.
Absence of Written Procedures
Your firm failed to establish adequate written procedures for production and process controls designed to assure that the drug products have the identity, strength, purity, and quality that they are purported or represented to possess.
Investigations of discrepancies, failures
There is a failure to thoroughly review any unexplained discrepancy and the failure of a batch or any of its components to meet any of its specifications whether or not the batch has been already distributed.
Procedures for sterile drug products
Procedures designed to prevent microbiological contamination of drug products purporting to be sterile are not established and followed.
Show 5 more citations
Acceptance of drug products
Determinations of conformance to appropriate written specifications for acceptance are deficient for drug products.
Training , Education , Experience overall
Employees engaged in the manufacture, processing, packing and holding of a drug product lack the training and experience required to perform their assigned functions.
Lab controls established, including changes
The establishment of laboratory control mechanisms including any changes thereto, are not drafted by the appropriate organizational unit.
In-process materials specifications testing
Examination and testing of samples is not done to assure that in-process materials conform to specifications.
Written program not followed
The written stability testing program is not followed.
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Establish reliability of supplier's C of A
Establishment of the reliability of the closure supplier's report of analyses is deficient in that the test results are not appropriately validated at appropriate intervals.
Training , Education , Experience overall
Employees engaged in the manufacture and processing of a drug product lack the training required to perform their assigned functions.
Procedures for sterile drug products
Procedures designed to prevent microbiological contamination of drug products purporting to be sterile are not established, written and followed.
Scientifically sound laboratory controls
Laboratory controls do not include the establishment of scientifically sound and appropriate test procedures designed to assure that components, in-process materials and drug products conform to appropriate standards of identity, strength, quality and purity.
Investigations of discrepancies, failures
There is a failure to thoroughly review any unexplained discrepancy and the failure of a batch or any of its components to meet any of its specifications whether or not the batch has been already distributed.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDrugs · Drug Quality AssuranceCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Cleaning / Sanitizing / Maintenance
Equipment and utensils are not cleaned and maintained at appropriate intervals to prevent contamination that would alter the safety, identity, strength, quality or purity of the drug product.
Extent of discrepancy, failure investigations
Investigations of an unexplained discrepancy and a failure of a batch or any of its components to meet any of its specifications did not extend to other batches of the same drug product and other drug products that may have been associated with the specific failure or discrepancy.
Absence of Written Procedures
There are no written procedures for production and process controls designed to assure that the drug products have the identity, strength, quality, and purity they purport or are represented to possess.
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Four requests to FDA under the Freedom of Information Act named this site, most recently on Sep 10, 2026. Each one asked for a Form 483.
Inspection ending Apr 17, 2026OAI
3 requests. In the inspection history above. FDA has posted its Form 483.
- Sep 10, 2026FDA-FOIA-2026-8795Form 483Records released
- Aug 12, 2026FDA-FOIA-2026-7754Form 483· 1 of 3 sitesClosed
- Jul 20, 2026FDA-FOIA-2026-6891Form 483· 1 of 2 sitesRecords released
Inspection date not given
1 request. None of FDA’s log entries for these requests say records were released.
- Dec 16, 2020FDA-FOIA-2020-8764Form 483 + EIRWithdrawn
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 9 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 8 inspections, with how many of them cited each.
- 21 CFR 211.192Production record review3 of 3
- 21 CFR 211.25Personnel qualifications2 of 3
- 21 CFR 211.100Written procedures; deviations2 of 3
- 21 CFR 211.113Control of microbiological contamination2 of 3
- 21 CFR 211.160General requirements2 of 3
- 21 CFR 211.67Equipment cleaning and maintenance1 of 3
- 21 CFR 211.84Testing and approval or rejection of components, drug product containers, and closures1 of 3
- 21 CFR 211.110Sampling and testing of in-process materials and drug products1 of 3
Show all 9
- 21 CFR 211.166Stability testing1 of 3
Observations read from posted 483 PDFs carry no CFR section of FDA’s, so they are not counted here.
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDER · DrugsLetter on fda.govFDA record
- Warning letterFormal notice of significant violationsCDER · DrugsFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
FDA drug establishment registration
FDA registration search- Registered operations
- Manufacturing, packing, labeling, analysis, sterilization
- Registration runs through
- Dec 31, 2027
- Registrant
- USV Private LimitedD-U-N-S 650611585
From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.
Other sites with the same registrant
FDA’s registration file names USV Private Limited (D-U-N-S 650611585) as the registrant for this site and the 2 others below. That exact D-U-N-S number is the only link. A registrant is the firm that filed the registration, which is not always the owner.
- USV Private LimitedTakula Khed, Dist. Ratnagiri, India · FEI 30029777986 inspections6latestSep 12, 2025VAINone on file
- USV Private LimitedGujarat, India · FEI 3027308441Registered; no FDA inspection on file
Other FEIs recorded as USV Private Limited
Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.
USV Private LimitedFEI 3003255171Govandi, India
Last inspected 2022-08-19VAI
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. This site’s records are filed under FEI 3004086192 in FDA’s data; other sites under its registrant are listed in summary only. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every inspection and enforcement record on this page is filed under FEI 3004086192. Other sites filed under the same registrant D-U-N-S are listed in summary only. Records filed under another FEI or name are otherwise not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.