RayzeBio, Inc Form 483, February 2026
FDA issued this Form 483 to RayzeBio, Inc in Indianapolis, Indiana (FEI 3030530655) at the end of a drug inspection that ended February 24, 2026. It lists 11 observations. FDA classified the inspection Voluntary Action Indicated (VAI).
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: FDA's Electronic Reading Room, posted Oct 20, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.
- Inspection ended
- Feb 24, 2026
- Drugs · Drug Quality Assurance
- FDA's outcome
- VAI
- Voluntary Action Indicated
- Posted by FDA
- Oct 20, 2026
- 238 days after the inspection
The 11 observations on this Form 483
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.
- 1
Quality control procedures inadequate and not followed
The responsibilities and procedures applicable to the quality control unit are not in writing and fully followed. Specifically, A. Your firm's procedure SOP-00013 ("Batch Review and Release Process," v. 5, effective 10/31/2025) for shipping drug products under quarantine is inadequate. The procedure allows shipment of sterile injectable drug products to end customers before the completion of analytical tests and quality review of batch manufacturing records, but it does not specify the steps to be taken if the product fails to meet finished product quality specification(s). This procedure does not ensure the quality of products prior to distribution. B. Your firm lacked a procedure for recalls that details the steps to follow, within the shelf life of the products, in the case of batch failures due to analytical test failures or manufacturing deviations. C. Per SOP-00012 ("Corrective and Preventive Action (CAPA) Handling," v. 2, effective 04/14/2025), Quality Assurance (QA) is responsible for tracking CAPA quality events, and CAPA due dates can be extended during the Implementation phase with appropriate justification. A review of the firm's list of CAPAs revealed that many remain open past their due dates without justification. For example: 1. CAPA-00217 was initiated to address missed pipettes calibration. CAPA-00217 was opened on 08/11/2025 to verify that pipettes were either calibrated or retired and to assess the need for a process change surrounding the calibration of pipettes. Neither action was performed, and CAPA-00217 remains open without a justification with a past due date of 11/14/2025. 2. CAPA-00218 was initiated 08/13/2025 to address overdue calibration work orders, have overdue equipment calibrated, and the calibration work order documentation completed. CAPA-00218 remains open without justification and has a past due date of 10/31/2025. 3. CAPA-00222 was initiated on 08/18/2025 to address overdue maintenance work orders, including overdue schedules for (b)(4), and (b)(4) preventive maintenance (PM) for HVAC equipment PM work orders; (b)(4) system ((b)(4) in Suite (b)(4) and Suite (b)(4) refrigerators, freezers, incubators, and stability chambers in micro and analytical labs; and BSCs and fume hoods in QC micro/sterility and production rooms. CAPA-00222 remains open without justification and has a past due date of 11/30/2025. 4. CAPA-00215 was initiated on 08/08/2025 to activate all freezer and refrigerator assets and their associated PM schedules active in the (b)(4) ((b)(4)). This CAPA was initiated following the failure to perform preventative maintenance for freezers and refrigerators per the recommended frequency. CAPA-00215 remains open without justification and has a past due date of 09/12/2025. 5. CAPA-00214 was initiated on 08/08/2025 to have all HVAC assets and their associated PM schedules active in the (b)(4). CAPA-00214 remains open without justification and has a past due date of 09/12/2025.
- 2
Aseptic contamination prevention procedures not validated
Procedures designed to prevent microbiological contamination of drug products purporting to be sterile are not established, written, and followed. Such procedures shall include validation of all aseptic and sterilization processes. Specifically, A. Procedures designed to prevent microbiological contamination of drug products purporting to be sterile do not include adequate validation of the aseptic process. Your firm failed to provide supportive data for the initial validation of the process confirmation batches. For example: 1. Per the firm's procedure SOP-00155 ("Aseptic Process Simulations (Media Fills)," v. 1, effective 05/28/2025), for initial validation, a minimum of (b)(4), successful media fills are required to consider the line aseptically validated. Your firm claims the process is validated in Suite (b)(4) however, your firm failed to validate a minimum of (b)(4), successful media fills in Suite (b)(4) as described in the following table: Media Fill Batch Number - Date Performed - Media Fill Result: (b)(4) - Invalidated; Pass; Pass; Invalidated; Pass; Pass; Invalidated; Pass; Pass; Invalidated. 2. The firm's validation master plan (Document# TD-00001, "Indianapolis Site Validation Master Plan," version 4, effective 05/13/2025) claims validated media fill parameters for worst-case conditions that lack supporting data from media fill runs in manufacturing Suites (b)(4). The Validation Master Plan stated the following parameters: (b)(4) batch size, (b)(4) mL vial size, (b)(4) mL fill volume, (b)(4) hour (b)(4) clean hold time, (b)(4) hour filling time, and (b)(4) hour aseptic processing time. However, these parameters were not demonstrated in the actual media fill runs performed as described below: a. For Suite (b)(4) (Room 1012) process confirmation batches: Media fill (b)(4), performed on (b)(4) had a (b)(4) fill time. Media fill (b)(4) performed on (b)(4) had a batch size of (b)(4) vials, a (b)(4) hours (b)(4) min (b)(4) clean hold time, a (b)(4) hour (b)(4) min filling (dispensing) time, and a (b)(4) hour (b)(4) min aseptic processing time. Media fill (b)(4), performed on (b)(4), had a (b)(4) mL final product vial size, a (b)(4) mL fill volume, and a (b)(4) hours (b)(4) min (b)(4) clean hold time. b. For Suite (b)(4) (Room 1013) process confirmation batches: Media fill (b)(4), performed on (b)(4), had a (b)(4) mL final product vial size, a (b)(4) mL and (b)(4) mL fill volume, a (b)(4) hours (b)(4) min (b)(4) clean hold time, a (b)(4) hours (b)(4) min filling (dispensing) time, and a (b)(4) hours (b)(4) min aseptic processing time. Media fill (b)(4), performed on (b)(4) had a (b)(4) hours (b)(4) min (b)(4) clean hold time. Media fill (b)(4), performed on (b)(4), had a (b)(4) mL final product vial size, and a (b)(4) mL fill volume. B. The following observations were made during review of the most current smoke study (air flow visualization): 1. The following activities were not evaluated during the smoke studies for vial filling in Suite (b)(4): the drug product bulk transfer from the (b)(4) ((b)(4)) to the (b)(4) ((b)(4)) and the final product vial transfer from (b)(4) to (b)(4). 2. The smoke study was not performed in (b)(4) and (b)(4), both of which have an ISO 5 classification. 3. Operators did not move slowly and deliberately. Rapid operator movements within the aseptic area were observed in both the smoke study videos and routine production operations. 4. Assessment of the inside surfaces of the isolators in Suite (b)(4) and Suite (b)(4) could not be made because your firm placed black plastic sheets on the interior walls, preventing evaluation of the (b)(4) walls' state of repair, presence of scratches, and air flow assessment at the (b)(4). C. Your firm lacks post-batch cleaning procedures to prevent contamination. SOP-00098 ("Cleaning Procedure: (b)(4) ISO 5 (b)(4)," v2, effective 04/11/2025) requires only (b)(4) cleaning of the (b)(4) (ISO 5) and production room (ISO 7). Batch records confirm no post-batch c
Shortened here; the PDF has the full text.
- 3
Environmental monitoring system deficient for aseptic areas
Aseptic processing areas are deficient regarding the system for monitoring environmental conditions. Specifically, A. The firm's environmental monitoring and personnel monitoring procedures are inadequate. For example: 1. Your firm failed to analyze raw Environmental Monitoring data for patterns (e.g., increasing CFUs) that indicate a loss of control in aseptic areas. Data trending is not performed and compiled (b)(4) in accordance with procedures SOP-00044 ("Environmental Monitoring," v1, effective 06/24/2025) and SOP-00191 ("(b)(4) - Trending and Reporting"). 2. Environmental monitoring points are not scientifically justified or located in critical areas, failing to detect potential sources of contamination. No risk assessment study was performed to define sampling locations and frequencies of monitoring in ISO 5 and ISO 7 areas. 3. The (b)(4) and (b)(4) were qualified with ISO 5 classification during Installation and Operational Qualification for Suite (b)(4) and Suite (b)(4) (b)(4) ISO 5 (b)(4). However, your environmental monitoring program monitors these two (b)(4) as ISO 6 areas without adequate justification. 4. On February 2, 2026, during aseptic fill operations for lot (b)(4) in Suite (b)(4) (Room 1012), an operator touched four locations when transferring material between the (b)(4) and the (b)(4): the (b)(4) door button (monitored), the door handle, the door's (b)(4)-facing side, and the tray (b)(4) where the operator pulled the tray (all three unmonitored). This practice was also observed during smoke studies. 5. There is no personnel monitoring occurring above the shoulders, although aseptic operators were observed during set up activities leaning inside the aseptic (b)(4) and (b)(4) from their waist up to their head. 6. On February 2, 2026, during aseptic fill operations for lot (b)(4) in Suite (b)(4) (Room 1012), the operators' fingertips were sampled without (b)(4) the finger as required by the firm's procedure SOP-00043 ("Viable Surface Testing", v 1, effective 06/24/2025). 7. The firm does not monitor the inside of ISO 7 cleanroom transfer carts used for carrying Transfer Materials (MT-(b)(4), MT-(b)(4), MT-(b)(4)) that are loaded into the (b)(4) and Transfer Materials (MT-(b)(4) and MT-(b)(4)) that are loaded into the (b)(4). B. Your firm failed to perform periodic review of quality impactful alarms. Between 08/26/2024 and 03/21/2025, your facility reported 19,463 differential pressure, temperature and humidity alarms. Below are alarms from 03/10/2025 to 03/21/2025 only: Alarm Type # of alarms Area: Differential Pressure (DP) 227 - Rooms 1024-1026A (ISO7-ISO8), Rooms 1011A-1011B (ISO8-CNC); Humidity and Temperature 38 - (b)(4) Suite (b)(4), Suite (b)(4); Humidity 38 - QC Micro Lab, 38 - QC Analytical Lab, 38 - Packaging and Labeling Room, Receiving Warehouse East Room, Receiving Warehouse West Room, Shipping Warehouse East Room, Shipping Warehouse West Room. DEV-00117 (initiated on January 29, 2025, and closed on March 31, 2025) was created to revise Alarm Management SOP-00209 to include a Quality Impactful Alarm Periodic Review requirement. Deviation DEV-00183 (initiated on May 20, 2025, and closed on August 1, 2025) was created for failure to perform the (b)(4) alarm management review. SOP-00209 was not updated until December 16, 2025. Additionally, CAPA-00106 and CAPA-00201 were initiated on December 9, 2024, and July 22, 2025, respectively, to address environmental monitoring alarms and procedural gaps. Both CAPAs remain open. To date, no periodic review of quality impactful alarms has been performed. C. The firm's procedure SOP-00195, "Cleanroom Certification," v 0, effective 04/11/2025, defines the (b)(4) Acceptance Criteria for ISO 7 rooms as (b)(4) and for ISO 8 rooms as (b)(4) and defines the Differential Pressure (Water Column) Acceptance Criteria for all cleanrooms as (b)(4) inches water column. Upon review of the 06/19/2025 cleanroom certification, the following four classified rooms were found
Shortened here; the PDF has the full text.
- 4
Failure to investigate discrepancies and batch failures
There is a failure to thoroughly review any unexplained discrepancy and the failure of a batch or any of its components to meet any of its specifications. Specifically, A. A review of the EMPQ (REC-00703, "Environmental Performance Qualification Summary Report," v0, approved 08/23/2024) showed that protocol deviations were inadequately investigated. Protocol Deviation Reports 11-16 identified 14 samples exceeding CFU action limits, totaling 315 CFUs. The Quality Unit failed to adequately investigate and implement corrective actions after identifying the CFUs, determine microorganism sources or trace them to production operations, and evaluate potential quality impact on sterile drug product manufacturing. B. A review of the firm's environmental monitoring excursions list (EME (b)(4)) for the period from 05/20/2025 to 01/20/2026 revealed that approximately 55 of 113 EME investigations remain open (37 in ISO 5 areas) without adequate investigation, with many investigations for excursions dating back to 10/06/2025 (n=18) not yet initiated. Multiple batches were released to market with unresolved excursions. For example: EME-2025-046: On 11/11/2025, an excursion notification was submitted in response to a personnel monitoring excursion of one (1) CFU (Action limit (b)(4) CFU), sampling date 11/04/2025, following the sterility testing in the Biological Safety Cabinet (ISO 5) in Room 1010 for lot (b)(4) and lot (b)(4). To date, no investigation has been initiated in response to this excursion. Both batches, lot (b)(4) and lot (b)(4), were released to the U.S. market. EME-2025-053: On 12/22/2025, an excursion notification was submitted in response to a right glove monitoring excursion during the formulation activities of lot (b)(4) in the (b)(4) (ISO 5) in Room 1012 of one (1) CFU (Action limit (b)(4) CFU), sampling date 12/16/2025. To date, no investigation has been initiated in response to this excursion. Lot (b)(4) was released to the U.S. market. EME-2025-055: On 12/26/2025, an excursion notification was submitted in response to a right glove monitoring excursion during the formulation activities of lot (b)(4) in the (b)(4) (ISO 5) in Room 1012 of one (1) CFU (Action limit (b)(4) CFU), sampling date 12/17/2025. To date, no investigation has been initiated in response to this excursion. Lot (b)(4) was released to the U.S. market. EME-2025-056: On 12/26/2025, an excursion notification was submitted in response to a right glove monitoring excursion during the filling activities of lot (b)(4) in the (b)(4) (ISO 5) in Room 1013 of one (1) CFU (Action limit (b)(4) CFU), sampling date 12/17/2025. To date, no investigation has been initiated in response to this excursion. Lot (b)(4) was released to the U.S. market. EME-2026-002: On 01/14/2026, an excursion notification was submitted in response to a viable active air monitoring excursion during the formulation activities of lot (b)(4) in the (b)(4) (ISO 5) in Room 1012 of two (2) CFU (Action limit (b)(4) CFU), sampling date 01/05/2026. Lot (b)(4) was rejected; however, no investigation has been initiated in response to this excursion.
- 5
Isolator decontamination method not validated
Equipment are not cleaned, maintained, sanitized or sterilized at appropriate intervals to prevent malfunctions or contamination that would alter the safety, identity, strength, quality, or purity of the drug product. Specifically, Your firm uses manual cleaning for (b)(4) ISO 5 Isolators in Suite (b)(4) (Room 1012) and Suite (b)(4) (Room 1013) but has not established or validated a decontamination method (e.g., (b)(4)). The firm's Validation Master Plan (TD-00001, v4, effective 05/13/2025) requires Installation and Operational Qualification (IQ/OQ) for the (b)(4) generator and (b)(4) referencing change control PC-00089, which was initiated on 03/25/2025 for (b)(4) implementation in ISO 5 spaces. No progress has been made since initiation, and your firm has provided no explanation or justification for failing to implement a validated sterilization method for the isolators.
- 6
Inadequate cleaning and disinfection system
Your firm failed to provide adequate system for cleaning and disinfecting the room and equipment to produce aseptic condition. Specifically, Aseptic processing areas are deficient regarding the system for cleaning and disinfecting the room and equipment to produce aseptic conditions. For example: A. Your firm has not conducted disinfectant efficacy studies to demonstrate that the disinfectants used ((b)(4) sterile (b)(4), sterile (b)(4), and (b)(4)) can adequately reduce bioburden from Suite (b)(4) and Suite (b)(4) cleanroom surfaces. The firm failed to perform studies showing these disinfectants effectively inactivate or remove microorganisms (bacteria, fungi, yeast, molds, and bacterial spores) from Grade A surfaces composed of various materials including acrylonitrile butadiene styrene, polycarbonate, polypropylene, acetal homopolymer, silicone, and polyvinyl chloride. B. Per SOP-00037 ("Aseptic Technique," v0, effective 06/28/2024, section 5.1.4), operators must perform handwashing with antimicrobial soap and water before gowning. However, on 02/02/2026 during gowning observation for lot (b)(4), no sink was present in the Production Gowning Room (Room 1006), and handwashing was not required for ISO 7 area entry. The Senior QA Specialist confirmed that handwashing is not required upon cleanroom entry; operators only apply (b)(4) to hands before donning gloves. Sinks are only available in the kitchen and restrooms.
- 7
Routine calibration not performed per program
Routine calibration of automatic and electronic equipment is not performed according to a written program designed to assure proper performance. Specifically, On 02/18/2026, I performed a review of your (b)(4) system used for the workflows and activities associated with preventive maintenance (PM) and calibration of equipment in the production and QC laboratory. The review revealed a total of (b)(4) pipettes ((b)(4) pipettes owned by Production and (b)(4) pipettes owned by QC Analytical) that were overdue for calibration and had passed the allowable grace period of (b)(4) days. No work orders had been generated in (b)(4) for the calibration of this equipment. Additionally, no deviation was created for missing the calibration due date (01/31/2026).
- 8
Inadequate training for visual inspection operators
Employees engaged in the manufacture, processing, packing and holding of a drug product lack the training required to perform their assigned functions. Specifically, Operators performing 100% manual visual inspection of (b)(4) finished vials lack adequate training per SOP-00134 (Visual Inspection Guidelines, v3, effective 11/21/2025). While the SOP defines three defect classifications (Critical, Major, Minor) covering 17 total defect types including (b)(4) glass container defects, (b)(4) container closure defects, and (b)(4) content defects (with inherent, extrinsic, and intrinsic particulates), the qualification test kit only includes particulate defects. Operators are not qualified to identify the full range of defects or differentiate between the three particulate types.
- 9
Incomplete reference standard testing records
Laboratory records do not include complete records of any testing and standardization of laboratory reference standards. Specifically, Your firm did not perform reference standard characterizations of the (b)(4) reference standard used in the final release testing of (b)(4). Your firm uses the reference standard (b)(4) ((b)(4)) lot# (b)(4) to conduct the radiochemical purity, identity, and peptide concentration determination by HPLC of (b)(4).
- 10
Failure to test components and validate suppliers
Your firm failed to test samples of each component for identity and conformity with all appropriate written specifications for purity, strength, and quality. Your firm also failed to validate and establish the reliability of your component supplier's test analyses at appropriate intervals. Specifically, 1. Your firm does not conduct identity testing of raw materials used in the production of (b)(4). 2. Your firm does not conduct testing on conformity of containers and closures to ensure that the manufacturer's results are valid and reliable, including sterility of these containers and closures. 3. Your firm has never performed supplier audits on any of the raw materials received at your facility.
- 11
Master records lack theoretical yield statement
Master production and control records lack a statement of theoretical yield including the maximum and minimum percentages of theoretical yield beyond which investigation is required. Specifically, The master batch record for the products (b)(4) Solution for infusion does not have theoretical yield specifications at final dilution stage of the drug product.
Other Form 483s FDA posted for this facility
Every Form 483 FDA posts under FEI 3030530655, newest first.
- Form 483, February 2026Inspection ended Feb 24, 2026 · Posted Jul 13, 2026VAI
Ask FDAgent about this Form 483
FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.
Related
About this data
- The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
- The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
- FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
- FDAgent is independent of FDA.