Equipment qualification and calibration in FDA Form 483 observations

Showing that equipment and instruments work as intended: installation and operational qualification, calibration and preventive maintenance. FDAgent finds 153 such observations in 115 Form 483s issued to 108 facilities, for inspections that ended between October 2008 and September 2026. Each is quoted below from the 483 FDA posted.

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.

Observations
153
Each quoted below
Form 483s
115
Posted by FDA
Facilities
108
Distinct FEIs
Most recent
Sep 9, 2026
Inspection end date

Observations by year

These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).

Peak: 50 in 2026

Equipment qualification and calibration observations in Form 483s by inspection year
YearObservations
20082
20090
20100
20110
20121
20131
20140
20150
201611
20179
20189
20199
20200
20214
20229
202316
20249
202523
202650

How FDAgent sorts observations into this theme

Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.

It is tied to one of these 21 CFR entries

21 CFR 211.6321 CFR 211.68(a)21 CFR 211.160(b)(4)21 CFR 820.70(g)21 CFR 820.72

A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.

Or its text contains one of these phrases

“installation qualification” · “operational qualification” · “equipment qualification” · “IQ” · “OQ” · “PQ” · “IOQ” · “calibrat*” · “preventive maintenance” · “preventative maintenance” · “requalif*”

* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.

The 153 observations, newest first

Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.

  1. Aurobindo Pharma Limited Unit 4Inspection ended Sep 16, 2016 · FEI 3008461619 · Observation 3

    … The Line #(b) (4) stoppering equipment qualification document # SH1026OQ Rev 1.0 and FU4-PN-(b) (4)-001-EQP-PQ-002 does not include a quantitative evaluation of the (b) (4) for products intended for (b) (4). During the inspection on September 7, I observed a large number (b) (4) of product vials with dislodged and variable (b) (4) (b) (4) during production of Injection (batch (b) (4). B. The (b) (4) 002 re-qualification document FU4-PN (b) (4) …

    Contains “equipment qualification”Read the Form 483 →

  2. Aurobindo Pharma Limited Unit 4Inspection ended Sep 16, 2016 · FEI 3008461619 · Observation 4

    The use of instruments and recording devices not meeting established specifications was observed. Specifically, the "Indu Softweb Studio" SCADA electronic recording system for temperatures found in the micro lab rooms for incubating Environmental Monitoring, Reference Cultures, and other media incubation shows multiple out-of-specification temperatures and loss of communication errors through June, July, and August 2016. As evidenced by repeated …

    Opens with FDA's citation wording for 21 CFR 211.160(b)(4)Read the Form 483 →

  3. Indoco Remedies LimitedInspection ended Sep 4, 2016 · FEI 3005124189 · Observation 5

    Equipment used in the manufacture, processing, packing or holding of drug products is not of appropriate design and suitably located to facilitate operations for its intended use and cleaning and maintenance. Specifically. a. The vial washer that will be used for injectable products filled on Line has surrounding ®® that keep the outfeed for washed vials under Grade A laminar air flow, and is surrounded by a Grade D area. The space is …

    Opens with FDA's citation wording for 21 CFR 211.63Read the Form 483 →

  4. Cipla LimitedInspection ended Aug 30, 2016 · FEI 3004081307 · Observation 3

    Equipment used in the manufacture, processing, packing or holding of drug products is not of appropriate design to facilitate operations for its intended use. Specifically, at the) of each of | “IE ‘machines, the process includes a ®® system, but it is only validated to ®® holes that may occur at the ® of each vial and not thronghont the container itself This is nerformed . (4) © @ (0) (4) following | by which a 2 There is no routine vial …

    Opens with FDA's citation wording for 21 CFR 211.63Read the Form 483 →

  5. Claris Injectables LimitedInspection ended Aug 26, 2016 · FEI 3004610460 · Observation 3

    OBSERVATION 3 The use of apparatus not meeting established specifications was observed. Specifically, the ™* used during sterility testing has not been qualified for that purpose.

    Opens with FDA's citation wording for 21 CFR 211.160(b)(4)Read the Form 483 →

  6. Qingdao Kangyuan Pharmaceutical Co., LTDInspection ended Apr 22, 2016 · FEI 3006512854 · Observation 7

    The scale, ID # SC-DZC-17-06, which is used in the (b)(4) room, is (b)(4) calibrated using a (b)(4) standard weight. However the scale weighing range is between (b)(4) to (b)(4) In addition the vendor calibration certificate dated (b)(4) did not indicate that the scale had been calibrated within the scale weighing range. The scale (b)(4) calibration does not cover the stated scale weighing range.

    Contains “calibrat*”Read the Form 483 →

  7. Qingdao Kangyuan Pharmaceutical Co., LTDInspection ended Apr 22, 2016 · FEI 3006512854 · Observation 10

    The written procedure, Gauge and Calibration Management, SOP 03-103, version 02, effective date 7/20/15, did not include adequate instructions regarding where a balance should be placed for weighing and measuring operations. This procedure only states to place the instrument in a clean and dry environment. The procedure also does not have instructions regarding re-calibration of a balance after it is moved.

    Contains “calibrat*”Read the Form 483 →

  8. Drug Depot Inc. dba APS PharmacyInspection ended Jan 22, 2016 · FEI 3010621916 · Observation 1

    … In addition, there are no written calibration procedures or actual calibration documentation available for your [illegible] and This includes lack of documentation for [illegible] is for glassware and finished product while preparing sterile drug products. 2. [illegible] of finished drug products (i.e., testosterone pellets), finished product vials, rubber vial stoppers, and utensils used in the filling of sterile drug products. 3. (b)(4) …

    Contains “calibrat*”Read the Form 483 →

  9. Cosma SpAInspection ended Jan 15, 2016 · FEI 3002806754 · Observation 10

    Observation 10 The following discrepancies were noted during the review of the maintenance and calibration program: 1. Schedules of maintenance and calibration are not managed through change control. The frequency of the maintenance and calibration of equipment is tracked using Database Maintenance software. Changes to this program are not tracked and can be performed by the Safety and Maintenance Manager. 2. Raw data is not retained during the …

    Contains “calibrat*”Read the Form 483 →

  10. Qingdao Kangyuan Pharmaceutical Co., LTDInspection ended Jan 25, 2013 · FEI 3006512854 · Observation 15

    There was no identification, calibration record, or usage log on a scale found in the raw material weighing and dispensing room. Reportedly, this scale is used to weigh (b)(4) used in the product formulation and recorded in the BPR.

    Contains “calibrat*”Read the Form 483 →

  11. Cangene CorporationInspection ended Jun 21, 2012 · FEI 3003153579 · Observation 4

    … Deviation Report #122799, dated August 11, 2010, documents the most current revision of pH calibration document not being used in the QC Raw Materials laboratory. The investigation states that the (b)(4) (b)(4) did not replace the old document with a newer version of the procedure. B. Deviation Report #197739, dated October 21, 1011 [illegible], documents that an incorrect version of a form was being used for raw material testing in the QC …

    Contains “calibrat*”Read the Form 483 →

  12. Genzyme CorporationInspection ended Oct 10, 2008 · FEI 1000305672 · Observation 6

    [illegible] studies executed August 2007 during the operational qualification of the HVAC system for fill suite FF2-16, did not demonstrate the following: Critical aseptic connections [illegible] Routine functions of aseptic core operators, for example: a. manually placing stoppers or reorienting stoppers using forceps for filled vials b. withdrawing unfilled vials from the filling line for weight checks c. redirecting filled vials [illegible] …

    Contains “operational qualification”Read the Form 483 →

  13. Genzyme CorporationInspection ended Oct 10, 2008 · FEI 1000305672 · Observation 10

    For the Cryoshippers which are used to transport master cell banks and working cell banks between manufacturing facilities: The use of these cryoshippers has not been validated. The manual for these shippers lists preventative maintenance steps for both continuous maintenance and annual maintenance. The firm has not conducted any maintenance on any of [illegible] shippers currently in use. The manual for these shippers states that the life …

    Contains “preventative maintenance”Read the Form 483 →

21 CFR sections these observations open with

The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.

Ask FDAgent about equipment qualification and calibration in Form 483s

FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
  • Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
  • A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.
  • FDAgent is independent of FDA.