Equipment qualification and calibration in FDA Form 483 observations

Showing that equipment and instruments work as intended: installation and operational qualification, calibration and preventive maintenance. FDAgent finds 153 such observations in 115 Form 483s issued to 108 facilities, for inspections that ended between October 2008 and September 2026. Each is quoted below from the 483 FDA posted.

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.

Observations
153
Each quoted below
Form 483s
115
Posted by FDA
Facilities
108
Distinct FEIs
Most recent
Sep 9, 2026
Inspection end date

Observations by year

These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).

Peak: 50 in 2026

Equipment qualification and calibration observations in Form 483s by inspection year
YearObservations
20082
20090
20100
20110
20121
20131
20140
20150
201611
20179
20189
20199
20200
20214
20229
202316
20249
202523
202650

How FDAgent sorts observations into this theme

Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.

It is tied to one of these 21 CFR entries

21 CFR 211.6321 CFR 211.68(a)21 CFR 211.160(b)(4)21 CFR 820.70(g)21 CFR 820.72

A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.

Or its text contains one of these phrases

“installation qualification” · “operational qualification” · “equipment qualification” · “IQ” · “OQ” · “PQ” · “IOQ” · “calibrat*” · “preventive maintenance” · “preventative maintenance” · “requalif*”

* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.

The 153 observations, newest first

Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.

  1. Catalent Indiana LLCInspection ended Sep 1, 2022 · FEI 3005949964 · Observation 7

    Routine checking of equipment is not performed according to a written program designed to assure proper performance. Specifically, A. Your firm does not assure that the (b)(4) is working within expected parameters during set up for a run. Your firm uses a challenge set with gross defects only prior to running in order to check (b)(4) operability. The challenge set does not include vials that are at a known visible range of detection. According to …

    Opens with FDA's citation wording for 21 CFR 211.68(a)Read the Form 483 →

  2. Catalent Indiana LLCInspection ended Sep 1, 2022 · FEI 3005949964 · Observation 11

    Records are not maintained so that data therein can be reviewed at least annually to evaluate the quality standards of each drug product to determine the need for changes in specifications or manufacturing or control procedures. Your firm did not establish criteria during supplier qualification and requalification based on recurring supplier complaints in A-SOP-03-03-003, Supplier Qualification. In addition, your firm has not been tracking and …

    Contains “requalif*”Read the Form 483 →

  3. Carolina InfusionInspection ended Aug 12, 2022 · FEI 3015826782 · Observation 4

    … The compound room was measured as requiring a [illegible] reading using calibrated equipment as 0.0140. On the same day, viable air sampling was performed; samples were processed by (b)(4) under Sample IDs [illegible]. The room, which is pressure neutral to the hallway adjacent to the sterile compounding cleanroom, recovered both highly pathogenic organisms and CFU counts that exceeded the action limit for the sampling area. Recovered organisms …

    Contains “calibrat*”Read the Form 483 →

  4. Protherics UK Ltd.Inspection ended Jun 23, 2022 · FEI 3001417487 · Observation 1

    … A (b)(4) is analyzed (b)(4) sample and does not confirm that the (b)(4) instrument is in a calibrated state upon the completion of the analysis. As a result, drug product results were invalidated and retested, due to a (b)(4) calibration test failure. For example, i). Quality Event 55795 - The (b)(4) calibration for the Quality Control (b)(4) was conducted per SOP (b)(4) entitled, "(b)(4) (b)(4)" on June 29, 2020, and failed the (b)(4) accuracy …

    Contains “calibrat*”Read the Form 483 →

  5. ID Biomedical Corporation of QuebecInspection ended May 20, 2022 · FEI 3004835647 · Observation 4

    Preventative maintenance activities were incomplete. Specifically, the (b)(4) in the (b)(4) on the (b)(4) are not part of a routine preventative maintenance program to include visual inspection. These (b)(4) (b)(4) are used during [illegible]

    Contains “preventative maintenance”Read the Form 483 →

  6. Serum Institute of India Pvt. Ltd.Inspection ended Apr 19, 2022 · FEI 3020223359 · Observation 8

    Equipment cleaning following preventive maintenance was not performed on the [illegible] used in manufacturing. As a result, (b)(4) batch failed endotoxin testing.

    Contains “preventive maintenance”Read the Form 483 →

  7. Serum Institute of India Pvt. Ltd.Inspection ended Apr 19, 2022 · FEI 3020223359 · Observation 9

    Corrective actions are not completed in timely manner. For example: a) (b)(4) cycles were aborted three times due to low concentration of (b)(4) Preventive maintenance to be completed to [illegible] after cleaning of [illegible] b) On 1/17/2022, leakage was observed in [illegible] equipment during [illegible] of (b)(4) was used for [illegible] This same (b)(4) SOP (#1012-0054) to be revised to include instruction to perform leak test and …

    Contains “preventive maintenance”Read the Form 483 →

  8. Dr. Reddy's Laboratories Ltd.Inspection ended Oct 29, 2021 · FEI 3006549835 · Observation 4

    … Additionally, conditions of the r calibration dat of calibration date 6) You failed to adequately challenge you utilized to items used on your distribution studies wer 002-00'. During 7) risk assessment was conducted, however, you material and

    Contains “calibrat*”Read the Form 483 →

  9. Fujifilm Diosynth Biotechnologies USAInspection ended Apr 21, 2021 · FEI 3000718852 · Observation 4

    … Validation and Re-validation of (b) (4) Procedures and Media Holds, Revision 08, Effective Date 19 Nov 2020, states that there is no requirement to requalify Process (b) (4) Cycles.

    Contains “requalif*”Read the Form 483 →

  10. Wuxi BiologicsInspection ended Apr 2, 2021 · FEI 3010606982 · Observation 3

    Adequate schedules have not been established for the preventative maintenance of equipment. Specifically, your firm failed to establish adequate preventative maintenance schedules for the following equipment: (b) (4) (MFG-UOE-3290), (b) (4) (MFG-DFC-3016), and (b) (4) (MFG-DFC-3017). For example, WX-SOP-00171, entitled Operation and Cleaning of Biologic (b) (4) Pump, states the (b) (4) of biologic (b) (4) pumps will be replaced (b) (4) In January …

    Contains “preventative maintenance”Read the Form 483 →

  11. Imprimis NJOF, -LLCInspection ended Jan 29, 2021 · FEI 3013024146 · Observation 6

    … In addition, there is no established frequency for preventative maintenance and no records are available for cleaning of air ducts on periodic basis. The firm's Preventative Maintenance Program for facilities and manufacturing is incorporated in (b) (4) Management Software. It was noted that (b) (4) Management Software Installation Qualification (IQ) was completed in (b) (4), however, post implementation of Software, a PQ was not performed. The …

    Contains “preventative maintenance”Read the Form 483 →

  12. Nanobots Healthcare, LLC dba Assurance InfusionInspection ended Dec 3, 2019 · FEI 3015617746 · Observation 10

    Your integrity testing device (Manufacturer: (b) (4) ) has not been calibrated since 2015. The manufacturer's recommendation for re-calibration is annually.

    Contains “calibrat*”Read the Form 483 →

  13. MPRX, Inc. dba Medical Park PharmacyInspection ended Nov 22, 2019 · FEI 3004102002 · Observation 2

    Calibration of mechanical equipment is not performed daily or on a routine schedule designed to assure proper performance. Specifically, a) You have not calibrated the scale used to weigh bulk drug substances/chemicals since you moved to this location. b) You stated, and I observed you do not verify the weighing of ingredients/components for each lot made. c) Formula Worksheets do not list ingredients and the quantity of each in the order of the …

    Contains “calibrat*”Read the Form 483 →

  14. Aurobindo Pharma Ltd. Unit 5Inspection ended Oct 28, 2019 · FEI 3004051616 · Observation 4

    … Also, temperature, pressure and time used for sealing during packaging processes are not recorded/documented, nor are the sealing machines used calibrated for temperature, pressure and time. b) Concerning (b)(4) manufacturing (used in API manufacturing and the source; (b)<> for (b)(1) production, (b)(1) and used in cleaning), operating ranges for process parameters are not fully established, such as determining and monitoring acceptable …

    Contains “calibrat*”Read the Form 483 →

  15. Rapid Equine Solutions, LLCInspection ended Aug 9, 2019 · FEI 3006708880 · Observation 8

    Procedures describing the calibration of instruments, apparatus, gauges and recording devices are not written or followed. Specifically, review of the Certificate of Analysis for (b) (4) Cartridges lot# (b) (4) revealed that the firm is not conducting cartridge calibration per SOP P-604 Testing Via the (b) (4) System, v0 effective 2/25/2015, section 8.3.1.2. The following drug products Folic Acid 10 mg/mL (b) (4) Lot#20190722-1, BUD: 01/18/2020, …

    Opens with FDA's citation wording for 21 CFR 211.160(b)(4)Read the Form 483 →

  16. Hieber’s PharmacyInspection ended Jul 17, 2019 · FEI 1000121499 · Observation 1

    … Instead personnel perform a (b) (4) visual inspection of the pressures on the (b)(4) and the clean room gauge, and record the pressures on the Cleaning and Calibration Log. Additionally, there is no pressure gauge to monitor the pressure or pressure differential between the ISO 7 clean room and the uncontrolled environment in the vestibule. This is a repeat observation from the 2017 inspection.

    Contains “calibrat*”Read the Form 483 →

  17. Hubei Gedian Humanwell Pharmaceutical Co., Ltd.Inspection ended Jun 28, 2019 · FEI 3004117486 · Observation 1

    The preventive maintenance of the (b) (4) tanks is inadequate. Specifically, A. On June 24, 2019, (b) (4) tanks (b) (4) were observed to have loose plastic tape pieces attached to the seal inside the (b) (4) tanks. Your Workshop Director stated that the tape pieces were used to hold the seal of the tank in place during the maintenance process. The last maintenance for these (b) (4) tanks occurred on August 14, 2018. The seal is replaced on an (b) …

    Contains “preventive maintenance”Read the Form 483 →

  18. Sun Pharmaceutical Industries Ltd.Inspection ended Jun 11, 2019 · FEI 3002809586 · Observation 4

    OBSERVATION 4 A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures. Specifically, the requalification of your [illegible] under Protocol No.: QUA-S/0253 Requalification Protocol/Report for [illegible] (equipment ID: 240) does not demonstrate that the air supply, assembly air, and clamp air pressures required per your reference document were …

    Contains “requalif*”Read the Form 483 →

  19. Torrent Pharmaceuticals LimitedInspection ended Apr 16, 2019 · FEI 3005029956 · Observation 2

    … It was also discovered that the firm has no procedural requirement during the tablet coating process to calibrate the (b)(4) before the start of a new batch. Without a (b)(4) calibration step, there is no assurance that the volumetric (b)(4) is uniformly maintained (across all (b)(4)) during the entire coating process. Lamotrigine ER Tablets 200 mg is an example of a functionally coated tablet product that does not include a (b)(4) calibration …

    Contains “calibrat*”Read the Form 483 →

  20. Torrent Pharmaceuticals LimitedInspection ended Apr 16, 2019 · FEI 3005029956 · Observation 3

    Routine calibration of automatic and electronic equipment is not performed according to a written program designed to assure proper performance. Specifically, The qualification of Tablet (b)(4) machine (equipment ID# (b)(4)-03) is deficient. This equipment is proposed for the commercial manufacturing of (b)(4) tablets ((b)(4) and (b) mg) for (b)(4) #(b)(4) and (b)(4) Tablets ((b)(4) and (b) mg) for (b)(4) #(b)(4). The following issues were …

    Opens with FDA's citation wording for 21 CFR 211.68(a)Read the Form 483 →

21 CFR sections these observations open with

The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.

Ask FDAgent about equipment qualification and calibration in Form 483s

FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
  • Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
  • A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.
  • FDAgent is independent of FDA.