Sun Pharmaceutical Industries Ltd. Form 483, June 2019
FDA issued this Form 483 to Sun Pharmaceutical Industries Ltd. in Halol, India (FEI 3002809586) at the end of a device inspection that ended June 11, 2019. It lists 4 observations. FDA classified the inspection Voluntary Action Indicated (VAI).
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: FDA's Electronic Reading Room, posted Jun 21, 2019. Form 483 PDF on fda.gov. FDAgent is independent of FDA.
- Inspection ended
- Jun 11, 2019
- Devices · Compliance: Devices
- FDA's outcome
- VAI
- Voluntary Action Indicated
- Posted by FDA
- Jun 21, 2019
- 10 days after the inspection
The 4 observations on this Form 483
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.
- 1
Aseptic process procedures not followed
OBSERVATION 1 Procedures designed to prevent microbiological contamination of drug products purporting to be sterile are not followed. The Aseptic Process Validation Master Plan was not followed during execution of media fill run [illegible]. Specifically, [illegible] media fill failure on the [illegible] resulted in media fill failure with over [illegible] media batch contamination in the filled, incubated units. An investigation into the media fill failure determined that several 'new' operations occurred during this media fill run. These included the following that were either not recorded in detail as performed or not performed per written batch instructions: - Removal of the '<b><>' to the aseptically sealed tank containing sterile '<b><>' media. - Maintenance Activity to repair/reset the <b><> of the filling station and bevel gear due to reversal of <b><> movement resulting in intake of air instead of delivery of media. - Quality Assurance Impact Evaluation as part of the 'Breakdown Maintenance Report' regarding the filling station reversal (Sr. No. [<b><>]).
- 2
Laboratory data review not documented
OBSERVATION 2 Established laboratory control mechanisms are not documented at the time of performance. The assurance of the data from the all analytical testing data is not provided. Specifically, Environmental monitoring of plates on June 5, 2019 included review of more than [illegible] sample sets. In my review of the data in support of the EM review, it was noted that in selected examples, the primary reviewer performed the plate analysis almost 2 hours prior to the witness review. As explained by your EM QC staff and detailed in written procedure SOP QCM-042, the witness is to perform the witness review of the reading of the EM plates at the same time as the primary review. For example, and as noted in the MODA system for EM data: (a) sample plate initial read/results entry: 10:09:05, review: 12:05:17 (b) sample plate initial read/results entry: 10:09:05, review: 12:05:17
- 3
Design plans inadequately established
OBSERVATION 3 Design plans that describe or reference the design and development activities and define responsibility for implementation have not been adequately established. Specifically, your design plan for ~b)() __(Injection, E_J<> pcg/mL, tbH > mL & tbH > mL) associated to <b><> was approved on May 20[illegible]. However, your firm began the design activities associated to this product on August 2012. Additionally, your firm did not have an established device history file that demonstrated conformance to an established design plan.
- 4
Process validation/qualification inadequate
OBSERVATION 4 A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures. Specifically, the requalification of your [illegible] under Protocol No.: QUA-S/0253 Requalification Protocol/Report for [illegible] (equipment ID: 240) does not demonstrate that the air supply, assembly air, and clamp air pressures required per your reference document were challenged to determine optimal production parameters. According to your reference document, [illegible] Instruction approved on 12/07/2019, the pneumatic assembly machine requires an air supply of [illegible]; assembly air cylinders require [KbH> ± '<b><> si'] and clamp air cylinders require [bH> ± '<b><> si']. During the review of Protocol No.: QUA-S/0253, it was noted that the air supply pressure was recorded as ~bH> (approximately ~b><> psi). However, no other data was documented for the assembly air cylinders, the clamp air cylinders, and for the tolerances provided by your supplier. Moreover, the pneumatic assembly machine (ID: 240) was used to assemble exhibit lots associated to '(b)(4) _Jib)(4) (b)(4)'.
Other Form 483s FDA posted for this facility
Every Form 483 FDA posts under FEI 3002809586, newest first.
- Form 483, June 2025Inspection ended Jun 13, 2025 · Posted Jul 9, 2025OAI
- Form 483, May 2022Inspection ended May 9, 2022 · Posted May 13, 2022OAI
- Form 483, December 2019Inspection ended Dec 13, 2019 · Posted Dec 20, 2019OAI
- Form 483, August 2018Inspection ended Aug 31, 2018 · Posted Sep 10, 2018NAI
- Form 483, February 2018Inspection ended Feb 23, 2018 · Posted Mar 6, 2018VAI
- Form 483, December 2016Inspection ended Dec 1, 2016 · Posted Dec 13, 2016OAI
Ask FDAgent about this Form 483
FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.
Related
About this data
- The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
- The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
- FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
- FDAgent is independent of FDA.