Catalent Indiana LLC
1300 S Patterson Dr, Bloomington, Indiana, United States
FEI 3005949964DUNS 119085807
FDA has 13 inspections on file for Catalent Indiana LLC in Bloomington, Indiana (FEI 3005949964), from 2011 to 2026. The latest ended on and was classified OAI, official action indicated.
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA posted the Form 483 from that inspection. FDAgent read its 8 observations from the PDF; they are listed below. One of the 53 FOIA requests for this site has asked FDA for this inspection's 483, and one more for the 483 from a later inspection, ending May 22, 2026, that FDA's Data Dashboard doesn't list yet.
FDA issued a warning letter on (CDER · Office of Manufacturing Quality). No court action or import refusal is filed under this FEI.
As of Nov 21, 2026, the site is registered with FDA as a drug establishment for manufacturing, packing and labeling, through Dec 31, 2027. Its registrant is Catalent Indiana, LLC (D-U-N-S 172209277).
Sources:FDA firm profile ↗Form 483 PDF ↗Warning letter ↗FDA drug establishment registrations ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 13, 2011–2026
- 2 OAI · 7 VAI · 4 NAI
- Inspections with 483 observations shown
- 4 of 13
- 3 posted 483s not matched to an inspection
- Requested under FOIA
- 53 requests, latest FY2023
- Warning letters
- · CDER · Office of Manufacturing Quality ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
- Drug establishment registration
- Through
Inspection history
13 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC · FDA reading room, last synced Nov 23, 2026, 04:33 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDrugs · Drug Quality AssurancePosted 483 · 8 observations
Read this Form 483 on its own page →Form 483 PDF on fda.gov
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.
- Observation 1
Inadequate investigation records
Written records of investigations into the failure of a batch or any of its components to meet specifications do not always include the conclusions and follow-up. Specifically, A. Extrinsic biological particulates (mammalian hair) have continued to be identified in finished sterile drug products. However, investigations into these events have not been completed in a timely manner and do not demonstrate that the root cause has been adequately identified and corrected. From approximately 08/30/2025 to present, you have identified mammalian hair in finished units in at least (b)(4) batches, including three identified in 2026, representing a recurring contamination issue (approximately (b)(4) batches out of (b)(4) inspected in 2026). Probable root cause has been attributed to stopper components; however, this conclusion is based on an ongoing, aggregated trend investigation (REC 1093464), that was opened on 08/28/2025, and has not been substantiated through complete, batch specific investigations. All (b)(1) deviations are currently open. You stated that closure of those deviations is pending completion of this trend investigation. Despite implementation of multiple corrective and preventative actions, you continue to observe mammalian hair in finished drug products. B. Investigations into repeated water intrusion events in classified manufacturing areas were not adequate to ensure timely correction of the identified root cause or to evaluate the impact of these events on aseptic operations. On or around 01/05/2026, water intrusion (roof leak) into the aseptic processing area resulted in the cancellation of a media fill lot (b)(4), Line (b)(4), Room (b)(4), with the root cause attributed to a roof leak and water saturation in the interstitial spaces above manufacturing suites in Building (b)(4) (REC 1149054). On or around 02/15/2026, a second roof leak resulted in water intrusion impacting Vial Line (b)(4), Room (b)(4), and (b)(4), Room (b)(4), during execution of media fill lot (b)(4), which was subsequently aborted (REC 1166977). On or around 03/04/2026, a third water intrusion event occurred in Room (b)(4) during execution of media fill lot (b)(4) on (b)(4), which was also aborted (REC 1175609). These three events occurred within a period of approximately two months, while corrective actions from the previous deviations remained open and prior to completion of the roof repair. Despite identification of the roof as the source of the leaks, you did not implement effective interim controls or complete corrective actions prior to resuming manufacturing in the affected areas. Additionally, you did not evaluate the impact of repeated leaks on commercial batches manufactured in the affected areas prior to the repair of the roof in Building (b)(4) on or around 04/14/2026 ((b)(4) batches made on lines (b)(4) and Syringe Line (b)(4) and did not implement enhanced environmental monitoring during the period of elevated risk. Review of maintenance records identified multiple prior roof leak events in Building (b)(4) in 2025 (approximately seven work orders), further demonstrating a recurring condition that was not effectively corrected. C. The defect classification system for visual inspection does not ensure that particulate matter (PM) related defects are consistently evaluated and investigated. For example, you classify PM such as atypical particulates, adhered matter, light/dark particles, and dark/light fibers as intrinsic, unless determined to be extrinsic per A-WI-22-01-002-001, Work Instruction for Particle and Matter Identification, Evaluation and Trending, Version: 8, Effective Date: 2026-03-30, and do not require an investigation unless the PM is determined to be extrinsic or an action limit is exceeded. As a result, PM identified during visual inspection are not consistently assessed to determine their origin, significance, or potential impact to product quality. From 07/01/2025 to present, you have identified approximately 84
Shortened here; the PDF has the full text.
- Observation 2
Unqualified visual inspection controls
Control procedures are not established which monitor the output and validate the performance of those manufacturing processes that may be responsible for causing variability in the characteristics of in-process material and the drug product. Specifically, A. Visual inspection methods have been implemented without adequately qualifying inspection sensitivity, defect detectability, or comparability between methods. For example: Your threshold (b)(4) studies failed to generate (b)(4) across the full visible range of defects and failed to establish (b)(4) based on defect size at which defects become reliably detectable. Without establishing (b)(4) supported by (b)(4), you have not demonstrated that inspection sensitivity has been adequately qualified or that the visual inspection process is capable of consistently detecting and rejecting critical defects. Additionally, MVI studies were conducted using aggregated "particulate" defect categories that combined multiple particulate defect types (ex: glass, metal, rubber, fibers). Further, defect units were characterized by using broad size ranges rather than actual measured particle sizes, which does not allow for evaluation of detection capability as a function of particle size. Review of your threshold studies supporting the manual baseline demonstrated that certain particulate defect types, such as glass, black fiber, and hair, exhibited low or negligible probabilities of detection. However, particulate inspection performance is characterized by using (b)(4), which masks this variability and does not reflect detection capability for specific defect types or sizes. B. The visual inspection program is not adequately designed or implemented to ensure that in-process control procedures provide consistent detection, classification, and evaluation of product defects. For example: Fiber defects (ex: light and dark fibers) identified during visual inspection are classified as Major defects and are described as particulate matter likely originating from the manufacturing process under A-WI-SOP-03-04-001-002, Parenteral Inspection Defect Criteria List, Version: 39, Effective Date: 2025-12-17. Your procedures require that particles be identified to determine whether they are intrinsic or extrinsic; however, fiber materials such as polyester fibers and lint from manufacturing materials (such as (b)(4)) are considered intrinsic following identification. As a result, fibers that may originate from the manufacturing environment or personnel are categorized as process related material, rather than extrinsic contamination. This classification does not ensure that fiber particulates are appropriately evaluated within the context of extrinsic contamination risk.
Show observations 3 to 8
- Observation 3
Inadequate aseptic/sterilization validation
Procedures designed to prevent microbiological contamination of drug products purporting to be sterile did not include adequate validation of the aseptic and sterilization processes. Specifically, Your sterility assurance program(s) are not adequately established which include airflow studies, intervention analysis to include occluded surfaces and your media fill program. A. Airflow studies performed on your (b)(4) lines including but not necessarily limited to (b)(4) Line (b)(4), Syringe Line (b)(4) and Vial Line (b)(4) do not adequately demonstrate the performance of challenging conditions, zones of impact, rejection requirements and adherence to first air principles. Interventions include but are not necessarily limited to the following: 1. High risk interventions change (b)(4), (b)(4), (b)(4) 2. Moderate and low risk interventions (b)(4) (b)(4) or load (b)(4) (b)(4) (b)(4) (b)(4) (b)(4) B. Intervention risk assessments and written SOPs do not adequately and comprehensively characterize and document requirements during the performance of interventions: 1. Your intervention risk assessment leverages (b)(4) input parameters as (b)(4) and (b)(4). In part (b)(4) lacks adequate assessment as operators are allowed to break first air with their (b)(4) as long as they comply with your written records to include aseptic observer requirements. Interventions impacted include but are not limited to high risk interventions (b)(4) and moderate low risk interventions include (b)(4). a. (b)(4) design impacts your risk assessment(s) to specifically include high risk intervention (b)(4) as this intervention inherently requires first air blockage during its performance. Other interventions lacking adequate space to routinely perform them adequately includes but is not limited to moderate low interventions (b)(4). 2. Your intervention risk assessment continues to allow for interventions which may include occluded surfaces such as moderate low intervention (b)(4), as well as atypical interventions. C. Media fills are not designed to ensure they provide worst case conditions to the aseptic filling process including but not limited to worst case interventions, are not necessarily included. Intervention variation such as line location (to include within and across locations) or zones as well as all (b)(4) used, vial sizes and open vial stress condition(s). Intervention counts performed during approximate period 6/2025 - 12/2025 is included as: (b)(4) 7
- Observation 4
Insufficient stopper sampling plan
The number of containers and amount of material taken from each container is not based upon appropriate criteria. Specifically, Your firm's sampling of incoming stoppers is not based on appropriate statistical criteria and does not ensure that samples are representative of the lot. Specifically, your firm samples (b)(4) of stoppers from each received lot for testing and release, without documented justification that this sampling approach is representative of the entire lot. This practice does not ensure variability within the lot is adequately assessed.
- Observation 5
Environmental monitoring failures
Aseptic processing areas are deficient regarding the system for monitoring environmental conditions. Specifically, Environmental monitoring is not always performed in response to interventions located and/or impacting critical surfaces. Examples include but are not limited to: (b)(4) materials, tools to include (b)(4).
- Observation 6
Quality unit responsibilities lacking
The responsibilities and procedures applicable to the quality unit are not in writing and fully followed. Specifically, You lack comprehensive trending analysis regarding your environmental monitoring program as well as your (b)(4) and (b)(4) decontamination (b)(4) program. Both programs lack adequate spatial analysis such as intra and inter location requirements over justified periods.
- Observation 7
Incomplete master production records
Master production and control records lack complete manufacturing and control instructions. Specifically, The work instruction A-WI-22-06-001-001 titled "Packaging and Labeling Defect Criteria List" that is utilized to assist the Quality Unit in identifying drug product defects during AQL inspections of finished sterile drug product batches is void of detailed critical and major (b)(4) defect descriptions, such as but not limited to the categories '(b)(4)' and (b)(4).
- Observation 8
Equipment design not adequate
Equipment used in the manufacture, processing, packing, or holding of drug products is not of appropriate design to facilitate operations for its intended use. Specifically, The (b)(4) Assembly Machine (Equipment ID #(b)(4)) is utilized to assemble and label sterile drug product (b)(4) into (b)(4). The firm assembles and labels the sterile drug products (b)(4) on the (b)(4) assembly machine. The equipment performance qualification studies titled '(b)(4) Assembly Machine (System (b)(4)) Performance Qualification Protocol Summary Report - VL-087JUN21, VPQ-(b)(4)-00019-S-VL-087JUN21, Client (b)(4),' effective: 01/23/2023, '(b)(4) Assembly Machine System (b)(4) Performance Qualification Protocol Summary Report - VL-087JUN21, Client (b)(4) Addendum 11, VPQ-(b)(4)-00019-S-VL-087JUN21-ADD11,' effective: 01/27/2026, and '(b)(4) Assembly and Labeling Machine (System (b)(4)) Performance Qualification Protocol, A-VPQ-00061-P,' effective: 04/08/2025, do not assess the assembly and labeling functionality as indicated in the production of (b)(4) and (b)(4) for the production batch sizes of (b)(4) and (b)(4) finished drug product (b)(4).
- OAISignificant objectionable conditions; FDA recommended regulatory actionDrugs · Drug Quality AssurancePosted 483 · 1 observation
Read this Form 483 on its own page →Form 483 PDF on fda.gov
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.
- Observation 1
Inadequate investigations and CAPA
OBSERVATION 1 Written records of investigations into the failure of a batch or any of its components to meet specifications do not always include the conclusions and follow-up. Specifically, A. REC 831826 was initiated on January 31, 2024, in response to Lot (D) (4), Project Code [illegible] being released without identification of two atypical extrinsic particles found during 100% manual visual inspection. One of the particles was identified as cat hair, while the second was not evaluated. You stated that only one representative particle is required to be identified if other similar defects are found, meaning no determination into the source or impact of the second particle was evaluated. Additionally, your investigation failed to determine a root cause for the contamination, assess the potential impact to the rest of the lot, or evaluate whether similar issues may have occurred in upstream batches. The review of related deviations was limited to a one year time frame, and no complaints were assessed. Although components were considered a potential root cause, no corrective or preventative actions were taken to address this. Since this time, approximately 20 additional deviations related to hair contamination during 100% manual visual inspection have occurred. In addition, approximately 14 other deviations were opened as a result of a gap analysis conducted in response to this deviation. There are approximately [illegible] batches associated with these deviations. No commercial lots of drug products have been rejected due to hair. These batches span across approximately [illegible] project codes (including, but not limited to: (b) (4)). One example of a recently released lot is project code (b) (4), lot (b) (4), released on or around 29Apr25. B. You failed to conduct adequate historical reviews during investigations by limiting searches for deviations to one year and by not consistently querying (b) (4) with an appropriate scope. For example, REC 1035680 was opened on April 21, 2025, in response to an extrinsic particle observed during 100% manual inspection of lot (b) (4), project code (b) (4). The particle was identified as "proteinaceous mammalian hair." The recurrence analysis was limited to deviations related to stoppers and only covered the date range of April 14, 2024, to April 14, 2025, rather than encompassing all relevant (b) (4) records (e.g., LIRs, CAPAs, change controls, complaints) involving stoppers. Additionally, your investigations do not consistently include a review of complaints when evaluating deviations. While remediation efforts have begun, this gap has not yet been fully addressed. C. REC 691329 was initiated on April 9, 2023, for Lot (b) (4), Project Code (b) (4), due to a failed AQL for a critical defect, plunger not seated. You identified root cause as a lack of defined procedures or instructions for inserting the syringes into the trays. However, operators interviewed during this investigation stated that they have long experienced this issue and developed informal methods to package the product in a way that avoids the defect. You failed to capture this feedback in a separate deviation and no action was taken to formally evaluate or address this tribal knowledge. D. You failed to adequately investigate and implement effective corrective actions to address the presence of foreign material, such as hair, in your sterile drug products. On or around July 2024 you initiated [illegible] supplier complaint records (SCRs) for your stopper manufacturer, (b) (4), approximately seven months after hair was found during 100% visual inspection of finished product. Since then, approximately 20 additional deviations have been initiated for hair contamination in finished product batches. Despite an ongoing trend of hair related to stoppers, you have not fully evaluated or implemented additional corrective actions to mitigate this risk. You currently have an open protocol for opening one additional bag of stopp
Shortened here; the PDF has the full text.
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Vaccines and Allergenic ProductsPosted 483 · 6 observations
Read this Form 483 on its own page →Form 483 PDF on fda.gov
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.
- Observation 1
Quality unit responsibilities not followed
The responsibilities and procedures applicable to the quality control unit are not fully followed. Specifically, We observed numerous instances in which the Quality Unit did not appropriately exercise its responsibility to ensure that drug products manufactured meet applicable good manufacturing practices and meet established specifications of identity, strength, quality, and purity as is required by A-POL-03-01-001, Quality Manual. The observations that follow demonstrate ways in which the Quality Unit: • Did not always ensure that "if errors have occurred, that errors have been fully investigated" (§5.5.1); • Did not always ensure that "validation and revalidation activities are appropriately executed and documented" (§5.5.1); • Did not always ensure operations designed to monitor the output of manufacturing processes that may cause variability in the quality characteristics of pharmaceutical products were performed by appropriately qualified production personnel (namely, qualification of personnel performing manual visual inspection) (§7.5.1); and • Did not always ensure that "all equipment used in the manufacturing, processing, packaging, or holding of Drug Substance, Drug Product, or Medical Devices is maintained in a good state of repair" (§6.3.2.2).
- Observation 2
Inadequate investigation of deviations
There is a failure to thoroughly review any unexplained discrepancy and the failure of a batch or any of its components to meet any of its specifications whether or not the batch has been already distributed. Specifically, A. Between 31 OCT 2021 and 31 OCT 2023, there have been approximately 194 deviations l6f<5>.<6fC1xc (b) (4) discussin failures. These failures are reported across the =(b�)�(�4)- 4 systems of all bl< filling lines b 4 ). Of these deviations, nine were identified as requiring new action(s) to prevent recurrence of the issue. Approximately 31 actions were initiated directly from these nine deviations. Six of the nine deviations were identified as requiring checks to ensure the action(s) taken were effective. 171 of these deviations do not identify a definitive root cause; in these cases, for major deviations, governing procedure A-SOP-21-01-054, Drug Product 15) 4 Program, instructs investigators to indicate a possible/probable root cause as 15 4 in the deviation's root cause grid. 155 deviations list a possible/probable root cause as (b) (4) ______. There are approximately (b) <41 batches associated with these deviations, with approximately bf<.J batches pertaining to commercial drug and biological drug products. No commercial lots of drug and biological drug products have been rejected due to (15) (4) failures. These ffJ batches span acrosf project codes ti) 4) In the same time period, there have been approximately 668 corrective maintenance (CM) actions associated with D 4 that failed D 4; between the close of the previous FDA inspection on 12 MAY 2023 through 30 OCT 2023, there have been approximately 85 corrective maintenance actions associated with (ti) (4). Document TEC-000-300-164, Drug Product Primary Trending Report Building A, B, and C - Q4FY23 (t> (4) Program, includes the following statistics: . . . of Failures . . . of Batches atch Impact ate vg. Days in se On 31 OCT 2023 during filling o (o) (4) Lot 6 4 , approximately four D 4 (6) (4) faile 6 4 ; Investigators Tan and Kuo observed three of these failures occur in real time. B. Since 01 NOV 2021, there have been approximately 170 deviations associated with particles. There are approximately 170 batches associated with these deviations. Of these 170 deviations, approximately 38 were identified as requiring new action(s) to prevent recurrence of the issue; approximately 28 deviations were identified as requiring checks to ensure the action(s) taken were effective. Trends are tracked in the existing process for trends associated with deviations; there is not currently a trending program for particles in ti) (4) similar to what's done for trending (o) (4) failures. C. REC 723587, Short Description "Pest found in 6 System Room 880, MBR (6 4 -) (4) 6)", WO: 5519936/5519935, discusses the discovery of a pest inside [illegible] on 14 JUN 2023 during teardown activities of product. At (•) (4) was underway: lot # 1;] was [illegible] filled [illegible] lots were rejected. When discussing possible scenarios by which the pest may have entered the (o) (4) this investigation (REC 723587) did not consider nor discuss a (6)(4) failure for [illegible], which was observed to have a (1:>) (4) (REC 723577). D. Deviation #730619, opened 29 JUN 2023, was opened in response to retain samples of product 6 4 lot#(6) (4) I failing 6 4 6 4 ) (o) (4) D (4), reported result [illegible]. The scope of this investigation did not extend to all batches of (b)(4) known to have a failure reported fo1 (0) (4) (o) (4) (namely, lot f!(o) (4) ). This investigation did not identify a root cause. E. (6)(4) queries are not always ran with appropriate scopes. Examples include, but are not limited to, the following: 1. Deviation #730619, opened 29 JUN 2023, included review of the lots of stoppers and syringes used to produce product lot 6 4. (6) 4 queries for the lots of stoppers 6 4 ( ) and lots of syringes 6 4 ) were limited to supplier complaints, as opposed to any b 4 record (devia
Shortened here; the PDF has the full text.
Show observations 3 to 6
- Observation 3
Inadequate aseptic process validation
Procedures designed to prevent microbiological contamination of drug products purporting to be sterile did not include adequate validation of the aseptic process. Specifically, A. The designs of Vial Line [illegible] systems (6 4 ), Vial Line (4) system, Flexible Fill Line (6) (4), Syringe Line (6)(4) (b) (4), and Syringe (6) (4) system do not allow all (6) (4) to be (6) (with [illegible]). These include, but are not limited to, the following: • (6 4) (Vial Line [illegible] (6..,.....) (4..,.}__________ ance' .j □):..,_,,_,_________________ • (Vial Linee Qualification of the (b)(4) decontamination cycles for C-1) lines did not always include biological indicators (BIs) placed in the (6) 4 furthest away from airflow to demonstrate that the (15f) that are not 6 4 during (b) are sufficiently decontaminated. B. The design of the systems for Vial Line (15) (4) systems (15) (4), Vial Line ( ] ) 6) 4 system (>) (4)), Flexible Fill Line (6) 4 system (6J (4J, Syringe Line 6 4 system (b)(4)), and Syringe Line I6 4 C4D all require all environmental monitoring (EM) plates to be transported to a central location. To do this, operators enter numerous (15) (4) (t>) (4) to form a (o) (4) and proceed to pass EM plates through the (ti) (4) to the centralized location. Airflow visualization studies do not include visualization of these interventions in their entirety; only the action of changing plates is visualized, and not the transport of new plates from and old plates to the central location. C. Written procedures designed to prevent microbiological contamination of products purporting to be sterile during aseptic processing are not adequate. For example, the following was observed: 1. On 31 OCT 2023, we (Investigators Tan and Kuo) observed operators' bare face/skin and gowns touching (b) (4) while performing interventions (b) (4) start-up and at end of fill 6 4 system (b) (4) (Grade A) and Grade C areas in Room 820). One operator was observed blowing air from his mouth (not masked) into the surrounding environment while his face was touching 6 4 during an intervention that he was performing. 2. [Operators'] apparel appeared soiled. For example, (6) (4) (• • l:>) (4) (6) (4) (b) on Syringe Line, the Flexible Fill (Flexi) line were observed with apparent ink over [illegible]. Additionally, 6 on Flexi and (o)(4) on Syringe Line were observed to have an apparent brown residue. 3. On 31 OCT 2023, we (Investigators Tan and Kuo) observed operators performing aseptic connection of the product storage bag to the filling line. We observed that after connection of the connectors to the fill line outside the (6) (4) the unused backup connector outlets were hanging down towards the floor and contacting a (15) (4) Barrier that was used and stepped over by operators during the aseptic connection process. 4. On 01 NOV 2023, during aseptic set-up of the stopper bowl on Vial Line prior to b) < , we (Investigators Tan and Kuo) observed that the operators appeared to have difficulty in removing the 6 cover off the installed stopper bowl, potentially causing unnecessary perturbations of the unidirectional air flow in the (b) (4). 5. On Oct 31, 2023, Investigator Kuo observed that on one of the staging carts in Filling Room 820 Grade C area, one of the wheels of the cart was observed to have materials lodged within and around the (b) (4) wheel, which may prevent adequate sanitization.
- Observation 4
Insufficient control and validation procedures
Control procedures are not established which monitor the output and validate the performance of those manufacturing processes that may be responsible for causing variability in the characteristics of in process material and the drug product. Specifically, For visual inspection of drug products and biological drug products purporting to be sterile, operators are qualified on Kt>) (4I vials; doing so allows operators to inspect [6) (4) I, and t>f {:iJ vials. Justification provided for why this is allowable (document A-VPPQ-00316-S-VL-016AUG23, b) (4 Visual Inspection Process [1)] Vials) Test Method Validation (TMV) Summary Report, VL-016AUG23) is inadequate in that the justification appears to attempt to validate a [(6f (4)] process and the document's qualification methodology was not designed to mimic the conditions of the individual inspector qualifications (§3.2).
- Observation 5
Endotoxin sample storage not validated
Laboratory controls do not include the establishment of scientifically sound and appropriate test procedures designed to assure that drug products conform to appropriate standards of identity, strength, quality and purity. Specifically, There is a failure to validate the endotoxin sample storage of drug products under project code: (6)(4) I a((o) (4) I up tq (6) (4) I prior to the testing. A [(o)(4) I hold study or low endotoxin recovery data demonstrates that the hold period at [(o) (4) I does not negatively affect the endotoxin recovery from samples to ensure the microbial contamination is under control.
- Observation 6
Cleaning and maintenance procedures not followed
Written procedures are not followed for the cleaning and maintenance of equipment, including utensils, used in the manufacture, processing, packing or holding of a drug product. Specifically, Work Orders (WOs) 1004043 and 1004053 were opened on 26 OCT 2023 and 27 OCT 2023, respectively, pertaining to incidences of the [(t> -(4) ] malfunctioning on System [(o)e(4 (Syringe) (4) Visual Inspection) in Building D, Room 141, used for visual inspection of Project Code ff<lL Pre-Filled Syringes. Both WOs failed to contain any information documenting the incidences nor log statements as required by the written procedure, A-SOP-07-01-015: Performing Maintenance and Calibration, Version 44, Effective 2023-08-28. Corrective actions were not performed to address these WOs and subsequently, on 31 OCT 2023, recurrences of this same issue of (e) (4) malfunctioning were apparent. During walkthroughs on 31 OCT 2023, this malfunction was observed in real time during the visual inspection of Project Code (b) (4) Lot [(b) (4)].
- Not classified yetFDA posted this Form 483; FDAgent has not matched it to a Data Dashboard inspectionNo inspection matchedForm 483 pagePDF
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssurancePosted 483 · 12 observations
Read this Form 483 on its own page →Form 483 PDF on fda.gov
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.
- Observation 1
Failure to thoroughly investigate batch discrepancies
Your firm failed to thoroughly investigate any unexplained discrepancy or failure of a batch or any of its components to meet any of its specifications, whether or not the batch has already been distributed. Specifically, A. Per various (b)(4) batch records, the Visible Particles USP <790> testing list a specification of "(b)(4) particles observed**If (b)(4) particle is observed during initial testing, test an additional (b)(4) vials with a cumulative specification of no more than (b)(4) particle observed". Between 8/11/21 and 4/7/22, twelve (b)(4) batches were released after exceeding the supplemental visible particles testing. B. Deviation 523897 regarded an atypical particle found in the (b)(4) on the (b)(4) motor in Zone (b) which lies above the filling line where filling takes place and units are unstoppered during aseptic filling of (b)(4) lot (b)(4) dated 4/28/22. Particle was identified as (b)(4) with (b)(4) material. (b)(4) is not used in the manufacturing of lot (b)(4). (b)(4) is within the formulation of (b)(4) lot (b)(4) which was manufactured directly before this batch. The investigation states, "Client (b)(4) product has been noted to be difficult to clean because of (b)(4) cleaning is performed with (b)(4) WFI and the best cleaning has been noted with (b)(4) WFI". The investigation is isolated to lot (b)(4) and does not extend to other batches previously manufactured after the aseptic filling of (b)(4) which could have been potentially affected by the inadequate cleaning procedure. The investigation did not question how (b)(4) could have potentially contaminated non-product contact equipment which lies above the filling line. Action REC 546084 was initiated on 6/15/22 to change cleaning method. A-SOP-21-01-056 Filling Line Disassembly and (b)(4) Cleaning has yet to be updated to implement (b) WFI to be used during (b)(4) cleaning following a client (b) batch on vial line (b). This change control is still open and is not due to be closed until (b)(4). Deviation 523897 is still open. C. Batches that exceeded internal control limits were not always adequately investigated. For example: The following lots exceeded internal control limit for various categories of visual defects such as Major A defect, defined as Fibers or Intrinsic Particulate (Elongated undissolved matter or particulate matter originating from a normal manufacturing process that is floating freely within the product or that is attached to the product contact area of the stopper), or Major B defect, defined as non-conformities that could lead to serious impairments of the container or potential impact to product quality. Investigation into these limit breaches to determine potential root causes did not occur as appropriate. 1. Product (b)(4) lot #(b)(4) 2. Product (b)(4) lot # (b)(4) 3. Product (b)(4) lot # (b)(4) D. OOS 520278 regarded out of specification result for (b)(4) Particulates Matter for (b)(4) product testing for (b)(4) lot (b)(4) dated 4/21/22. (b)(4) particle was observed, exceeding the specification of (b) foreign particles. Identification determined (b) particles were present, (b)(4) was (b)(4) (potentially from a hair/beard cover) and (b)(4) (potentially from a (b)(4) puncture). No obvious errors were determined to have contributed to the OOS result. Adequate justification could not be provided for only retesting (b)(4) vials during a retest of the (b)(4) Particulates Matter for (b)(4) Product instead of the (b)(4) vials required per the OOS procedure. A-SOP-09-01-019, Investigation of Out of Specification Quality Control Analytical Results, procedure requires a retesting of (b)(4) times the original independent test samples ((b)(4) vials). Deviation 542608, dated 6/8/22, was initiated to address the retesting (b)(4) particulate matter for (b)(4) product testing for (b)(4) lot (b)(4). Sr. Manager of QC stated they wanted to retest (b)(4) samples, but the client felt that since (b)(4) was identified, the initial results were contamina
Shortened here; the PDF has the full text.
- Observation 2
Inadequate drug complaint investigations and follow-up
Written records of investigation of a drug complaint do not include the findings of the investigation and the follow-up. Specifically, A. As provided by your firm, a total of 179 complaints were received for particles (foreign matter, particulate matter, black specks, black particles, foreign particle, (b)(4) impurities, black dots and other particulates, dark particles, foreign material, foreign object, (b)(4) particulate, or pieces of (b)(4) in vial) between 03 Sep 2020 to 01 Aug 2022. Complaints were not adequately investigated to include an identification of the particle in the complaint sample. For example: 1. On 23 Mar 2021, a Customer Complaint #333544 reported black specks in four vials of (b)(4), lot# (b)(4). Four vials and four (b)(4) were returned to your firm for evaluation. The black specks were not analyzed in the lab. Retain samples review did not include an inspection for particles. 2. On 01 Feb 2021, a client reported particles in a vial of (b)(4) lot# (b)(4) Customer Complaint #308230. The vial was returned to your firm for evaluation. The particles were not analyzed in the lab. Retain samples review did not include an inspection for particles. B. As provided by your firm, a total of 52 supplier complaints for stopper related issues were submitted between 01 Jul 2019 to 05 Aug 2022. Out of these, 17 supplier complaints were related to Stopper item# 402832. Three of these investigations are still open (SCR 555104, SCR 540975, SCR 534591 opened on 04 Jul 2022, 05 June 2022, and 22 May 2022, respectively). Your firm did not perform an adequate investigation into the repeat stopper defect reported. Your investigation of the 14 supplier complaints related to stopper item #402832 relied in large part on the supplier investigation. For the 14 SCR, the supplier investigations concluded that that stopper item #402832 either "met agreed qualification" or "a root cause cannot be determined". C. Your firm did not adequately extend your complaint investigation to other batches of the same lot or product when warranted. For example, Per the firm's procedure, A-SOP-03-01-024, titled Customer Complaints, Recurring Complaint is defined as "A complaint that has occurred previously and had a corrective or preventative action(s) implemented, then occurred after implementation of the action(s)." Your firm did not conduct a thorough investigation for the multiple number of complaints reported for the same lot and the high number of complaints reported for the same defect for the same product manufactured at the firm. For example: 1. Complaint #413054 on 07 Sep 2021, product (b)(4) lot# (b)(4) 2. Complaint #339767 on 06 Apr 2021, product (b)(4) lot# (b)(4) D. Complaints were not always classified according to your SOP-03-01-024, titled Customer Complaints, for classification of complaints: Critical, Major, and Minor. For example, On 30 Oct 2020, a Customer Complaint #272397 reported an empty (b)(4) of product (b)(4). Your firm classified this customer complaint as a Major complaint rather than a Critical complaint.
Show observations 3 to 12
- Observation 3
Missing internal defect control limits
Control procedures are not established which monitor the output and validate the performance of those manufacturing processes that may be responsible for causing variability in the characteristics of in-process material and the drug product. Specifically, Internal defect control limits were not established for the (b)(4) Measurement System (b)(4) inspection process. This was noted during the review of the following batches: Product (b)(4) lot# (b)(4) inspection on 11 Dec 2019 Product (b)(4) lot# (b)(4) inspection on 12 Jul 2019 and 12 Aug 2019.
- Observation 4
Inadequate AQL sampling and acceptance criteria
Acceptance criteria for the sampling and testing conducted by the quality control unit is not adequate to assure that batches of drug products meet appropriate statistical quality control criteria as a condition for their approval and release. Specifically, A-SOP-03-04-001, Performance of AQL for Parenteral Inspections, states AQL sampling should be a representation of the entire batch. "During (b)(4) Inspection with (b)(4) (System 2440), if AQL fails for an equipment related issue that can be corrected, stopping to segregate materials and then restarting the AQL is the proper course. A Tightened AQL will be performed on segregated materials. If the issue is not able to be corrected by an adjustment to the (b)(4), the AQL will be completed in its entirety to satisfy this requirement, and a Focused Tightened AQL on the batch will be completed when appropriate." Your firm does not take AQL samples that are reflective of the entire batch. Your firm has had approximately (b)(4) drug products which exceeded AQL limits during visual inspection since 9/1/2020. It is in your procedure, A-SOP-03-04-001, to segregate product that has exceeded AQL at the time of inspection and reinspect the segregated portion under a tightened inspection. The remaining, uninspected, portion is then continued to be visually inspected under normal AQL inspection levels with a new batch size created. AQL inspection is performed while visual inspection is in progress, however, there is no documentation for when AQL samples are inspected. There is also no documentation for which (b)(4) have been sampled. According to the AQL procedure, the samples should represent "good saturation" throughout the batch, however this is not verifiable based on current AQL documentation. Segregation of the AQL portion that exceeded limits relies on immediate action from the QA inspector to hold and segregate those AQL portions. In addition, the uninspected portion is not included in the portion exceeding the AQL limit. The uninspected portion should be considered a part of the total AQL population. For example: Lot (b)(4) for product (b)(4) exceeded AQL for critical defects. This lot was run on the (b)(4) and segregated at the time of exceeding AQL limits, according to DEV 483844. The defect was determined to be a crack in the neck of the vial. The segregated portion was reinspected on a tightened AQL on the (b)(4). The remaining, uninspected, portion was allowed to continue normal AQL inspection levels on the (b)(4) with a new batch size determined from the remaining population.
- Observation 5
Inadequate automated inspection equipment qualification
Equipment used in the manufacturing of a drug product is not of appropriate design to facilitate operations for its intended use. A. 1. Your firm did not demonstrate that the (b)(4) is better than or equivalent to (b)(4) inspection. Your firm did not consider individual defect vial performance when establishing acceptance criteria during (b)(4) ((b)(4) System 2414) Performance Qualification (PQ) (b)(4) mL, VPQ-258-000-0002-P, recipe (b)(4) Rev 1.0. 2. Additionally, instances where defects were not caught by the (b)(4) were not evaluated because they were averaged out based on the acceptance criteria in part A. Your firm runs each defect type and size on the (b)(4) in (b)(4). According to your Senior Manager of Engineering, the particulates could no longer be free floating in the vial for a number of reasons such as being stuck in the stopper or on the vial wall. The (b)(4) qualification documents also did not demonstrate that the (b)(4) inspection method could resuspend the particulates in the product once they became undetectable/stuck on the vial wall. 3. Your firm did not evaluate extrinsic particulate, such as hair, detectability during qualification of the (b)(4) (System 2440) for recipe (b)(4)-304 V.9. Extrinsic Particulate Matter is a (b)(4) Critical Defect that is checked during visual inspection. B. 1. The product used for your (b)(4) inspection base line study (Knapp study) did not have the same properties as the product used during (b)(4) qualification. Your firm used a surrogate product for particulate analysis performed during the Knapp study A-DSR-18-01-091 as a baseline to qualify recipe (b)(4) Rev 1.1 on (b)(4) (System 2405). The surrogate product attributes are not detailed in the summary or executed report for the Knapp study. 2. Your firm does not have a program to re-evaluate performance qualifications based on continued process improvements. For example, your firm has implemented a test case for gray zone evaluation for new recipe performance qualifications. Your firm also now performs confidence interval calculations for acceptance criteria of (b)(4) Comparison to (b)(4) Inspection Baseline. These recently implemented test cases have not been retroactively applied to previously qualified recipes to assure they are still in a qualified state. C. Your firm has an ongoing issue of (b)(4) integrity issues that are involved in low and high-risk aseptic interventions during the aseptic filling of sterile drugs. Your firm performed a Deviation Trending Investigation for Building (b)(4) on February 9, 2022 and determined the root cause to be the (b)(4) material was not robust enough to maintain integrity during use. Your firm initiated corrective action to remediate the issue to related (b)(4) integrity by opening a change control to switch to a (b)(4) mils thicker (b)(4) from the same (b)(4) manufacture. Your firm has since cancelled that change control and as of the current establishment inspection your firm is still using the original (b)(4) that continue to experience many integrity issues.
- Observation 6
Inadequate aseptic procedures airflow and monitoring
Procedures designed to prevent microbiological contamination of drug products purporting to be sterile are not established, written, or followed. A. The airflow studies demonstrate a number of instances that document a lack of successfully meeting the acceptance criteria. For example, A-VPQ-00178-S-VL-070MAR21 Vial Filling (b)(4) (System 3400) Static and Dynamic Airflow Visualization Study Summary Report, v0, dated 5/10/21 establishes the acceptance criteria for the airflow visualization studies as in part, (b)(4) (b)(4). The March 2021 airflow studies show discrepancies which do not meet the acceptance criteria, for example, * there is no smoke over many of the manual activities to determine that the manual operations do not create/generate air eddies or air turbulence; * there are numerous instances where there has been no smoke evaluations performed for the operators gloved hands and forearms; * the smoke nozzle is not positioned in such a manner to capture the airflow when personnel perform the manual operations; * the videos do not include an evaluation of the fill equipment & fill (b)(4) to determine that the equipment configuration does not negatively impact the unidirectional airflow; * due to the angle of the camera unable to observe smoke over the manual operations; * no visible smoke can be observed during various manual operations; there are numerous instances when equipment and/or personnel hands & arms block the view, which precludes the company from performing an evaluation of the unidirectional airflow. These discrepancies potentially affect the other vial and filling lines used to aseptically fill drug and biologic products. B. Adequate justification could not be provided for not performing contact plating of all (b)(4) (b)(4) used during the aseptic filling of sterile product. Instead, environmental monitoring of (b)(4) is based on the frequency of typical batch usage and what can be performed at that (b)(4) port. Usage is based on A-SOP-21-01-054, Drug Product (b)(4) Program, v16, dated 6/20/22 which specifies typical batch usage as low, medium or high. Only high usage (b)(4) and (b)(4) used to perform high risk interventions are monitored at the (b)(4) of an aseptic fill. No supporting data could be provided to justify the typical batch usage categories. Only the use of (b)(4) used to perform high risk interventions is documented. The firm has (b)(4) vial filling lines, (b)(4) filling lines and (b)(4) flexi-filling line. C. The rational to only require Quality oversight of high-risk interventions during aseptic filling is inadequate. Documentation only supports Quality oversite during high-risk interventions.
- Observation 7
Inadequate routine equipment checking program
Routine checking of equipment is not performed according to a written program designed to assure proper performance. Specifically, A. Your firm does not assure that the (b)(4) is working within expected parameters during set up for a run. Your firm uses a challenge set with gross defects only prior to running in order to check (b)(4) operability. The challenge set does not include vials that are at a known visible range of detection. According to A-SOP-22-03-014, (b)(4) Vial Inspection Test Set Composition and Handling, "The intention of the challenge is to verify (b)(4) operability ((b)(4) mechanism, and general (b)(4) functionality) is correct." B. Your firm did not qualify the (b)(4) Vial Inspection (b)(4) System CL 145) (b)(4) (b)(4) speed during any qualification activities. Your procedure, A-SOP-22-05-009, (b)(4) Process Operational Flow, states "Verify vials have at least (b)(4) in inspection zone" during set up. This verification is not documented, and no data has been generated during qualification to show that all vial sizes will perform (b)(4) during inspection. (b)(4) speeds are the only parameter verified during set up. The firm uses vial sizes ranging from (b)(4) mL on the (b)(4) inspection system.
- Observation 8
Inadequate visual inspector qualification procedures
Your firm failed to establish and follow adequate control procedures to monitor the output and to validate the performance of those manufacturing processes that may be responsible for causing variability in the characteristics of in-process material and the drug product. The following discrepancies were noted during the review of visual inspector qualification for aseptic filling: A. Although included in the emulsion visual inspection kit, visual inspectors for emulsion products are not qualified to detect Major A defects. This category contains fibers or intrinsic particulate matter. B. The emulsion visual inspection kits do not represent extrinsic particulate matter. C. A visual inspection kit does not exist for (b)(4) vials. Instead, the inspectors train on the aqueous glass kit. The aqueous kit does not contain the following defects specific to (b)(4) vials: (b)(4). D. Per A-SOP-22-01-009, titled (b)(4) Inspection Test Set Composition and Handling, version 13, dated 25 Jan 2021, "all defects will be (b)(4) during the (b)(4) qualification PBA so that over the course of the (b) required passes the inspector has had an example of all defects listed in A-WI-03-04-001-002 Parenteral Inspection Defect Criteria List". This procedure was not followed during the initial qualification of operator (b)(6), performed on 20 and 21 Jul 2021 and the emulsion qualification of (b)(6) on March 8, 2021. E. You cannot confirm through documentation the inspection times during the execution of visual inspection qualification tests is representative of routine visual inspections.
- Observation 9
Non-representative in-process particle sampling
In-process samples are not representative. Specifically, There is no scientific justification for not performing particles testing on routinely inspected product. Batches of aseptically filled (b)(4) are inspected by (b)(4). The particulate defects are rejected and the "ejects" are run through the (b)(4) (b)(4) for additional inspection. Instead of performing destructive particles testing on the good vials remaining after the (b)(4) inspection, (b)(4) vials are sampled from the batch prior to being visually inspected by this process and rather are inspected by ISC (Inspection Sample Control) personnel. The samples undergoing destructive particles testing are not fully representative of routine production.
- Observation 10
Inadequate aseptic process validation and media fills
Procedures designed to prevent microbiological contamination of drug products purporting to be sterile did not include adequate validation of the aseptic process. A. The material (b)(4) for vial filler (b)(4) has not been qualified for LIMS labels. Deviation 493708 dated 2/25/22 was initiated due to action levels being obtained on an (b)(4) in vial line (b) between the (b)(4) and the conveyor. The root cause determined the validation requirements for layout, configuration and/or label orientation for LIMS labels in the (b)(4) are not qualified. The risk assessment for batch impact determined the labels achieve a (b)(4) log reduction (decontamination), and not the preferred (b)(4) log reduction. Action 552193 was initiated 6/28/22, to update A-WI-21-06-005-02 to label contact plates in the (b)(4) with marker instead of LIMS labels. The (b)(4) is still used to pass through LIMS labels. This fill line is not dedicated and can be used to aseptically fill both biologic and drug products. B. There is no assurance that media fills are representative of routine production as media fills are not performed under the most challenging or stressful conditions and do not simulate your production process. 1. Trending of interventions during sterile manufacturing does not take into consideration packaging configuration (i.e., vial size and run speed). Instead, trending of interventions is performed by averaging all interventions from all batches run on a particular line. Vial sizes could range from (b)(4) to (b)(4) mls. 2. Adequate scientific justification was not provided for grouping interventions. TEC-000-300-068 Drug Product Manufacturing Media Fill Interventions Rational and Risk Assessment, v7, dated 2/22/22, states all interventions have been grouped based on the required steps to perform each intervention. The intervention with the highest risk was chosen to represent all interventions in that grouping during the media fills. The document equates clearing stopper jam in bowl or (b)(4) as the same risk level as stopper/seal addition with or without (b)(4) port). 3. Multiple line speeds are addressed within a single media fill batch (i.e., (b)(4) (b)(4)%). This is significant as there is no assurance that the media fills performed are representative of routine manufacturing. The Senior Manager of Value Steam stated the intention during production is to run at (b)(4)%.
- Observation 11
Failure to review quality standards and supplier complaints
Records are not maintained so that data therein can be reviewed at least annually to evaluate the quality standards of each drug product to determine the need for changes in specifications or manufacturing or control procedures. Your firm did not establish criteria during supplier qualification and requalification based on recurring supplier complaints in A-SOP-03-03-003, Supplier Qualification. In addition, your firm has not been tracking and trending supplier complaint reports when a complaint is filed to the supplier. Records are opened for many months before tracking is completed. For example: Record# 533727, dated 5/15/2022, regarding (b)(4) assembly. Record# 565406, dated 5/20/2022, regarding (b)(4) assembly. Record# 566111, dated 05/12/2022, regarding sterile bag used for sterile filling operations.
- Observation 12
Inadequate laboratory controls and test procedures
Your firm failed to establish laboratory controls that include scientifically sound and appropriate specifications, standards, sampling plans, and test procedures designed to assure that components, drug product containers, closures, in-process materials, labeling, and drug products conform to appropriate standards of identity, strength, quality, and purity. A. The verification of method AM-(b)(4) 100-018 Supplemental Destructive Visual Inspection for Visible Particulates of Client (b)(4) Drug Product Samples, v1, on 7/12/22, demonstrated the analyst could execute the method. The verification did not show the method or the analyst's ability to detect particles. B. A-SOP-09-01-061, (b)(4) Inspection of Media Filled Vials, v9, dated 12/13/21 states to (b)(4) (b)(4) the containers for approximately (b)(4) prior to inspection to suspend any potential microbial growth. (b)(4) the vial prior to initial review could potentially disperse the microbial film on the surface of the media if it existed. C. On 8/1/22 during the walk-through inspection of the Microbial Testing Laboratory, air bubbles were observed between the (b)(4) filter and the (b)(4) plate. Four out of (b)(4) water system sampling plates corresponding to samples collected on 7/19-20/22 were observed having bubbles. Inadequate contact between the filter and the (b)(4) might compromise microbial recovery. The impact of the bubbles on any microbial film on the surface of the media was not provided. D. During the walk-through inspection of the Microbial Testing Laboratory, a discrepancy on CFU counts was noted ((b)(4) CFUs on 7/25/22, (b)(4) CFUs on 8/1/22). There was no evidence that the reviewer contacted management when the discrepancy was noted. A-SOP-09-01-039, Counting and Reporting Counts of Microbial Plates, v7, effective 12/27/21 states in section 2.1, "When reviewing microbial plates, if the reviewer count differs from that of the original count, then the review must take place within (b)(4) of the original count to change the count or be changed at management discretion". This may impact the original count, and the discrepancy is not documented, and management may not be notified. E. Full reconciliation of injections performed in the quality control laboratory is not performed. The analysts are trained to choose between normal, don't process or don't process or report. A-SOP-09-06-180, Use of BLM Empower 3 Chromatography Data System, does not address these choices. If 'don't process or report' is chosen, these chromatograms are not processed and are not reviewed.
- Not classified yetFDA posted this Form 483; FDAgent has not matched it to a Data Dashboard inspectionNo inspection matchedForm 483 pagePDF
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
- NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
- Not classified yetFDA posted this Form 483; FDAgent has not matched it to a Data Dashboard inspectionNo inspection matchedForm 483 pagePDF
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood ProductsCitations not published
- NAINo objectionable conditions that need follow-upDrugs · Drug Quality AssuranceNone
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
- NAINo objectionable conditions that need follow-upDrugs · Drug Quality AssuranceNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
53 requests to FDA under the Freedom of Information Act named this site, most recently on FY2023.
Inspection ending May 22, 2026
1 request. Not listed in FDA’s Data Dashboard yet. None of FDA’s log entries for these requests say records were released.
- Sep 8, 2026FDA-FOIA-2026-8715Form 483No records found
Inspection ending Apr 24, 2026OAI
1 request. In the inspection history above. FDA has posted its Form 483.
- Jun 11, 2026FDA-FOIA-2026-5739Form 483· 1 of 5 sitesRecords released
Inspection ending Feb 1, 2026
1 request. Not listed in FDA’s Data Dashboard yet. FDA released records in answer to a request for its Form 483 + EIR on May 21, 2026. FDAgent holds no copy that FDA posted.
- May 12, 2026FDA-FOIA-2026-4866Form 483 + EIRRecords released
Inspection ending Jan 14, 2026
1 request. Not listed in FDA’s Data Dashboard yet. FDA released records in answer to a request for its Form 483 on Apr 29, 2026. FDAgent holds no copy that FDA posted.
- Mar 31, 2026FDA-FOIA-2026-3097Form 483Records released
Inspection ending Jan 1, 2026
1 request. Not listed in FDA’s Data Dashboard yet. FDA released records in answer to a request for its Form 483 on May 21, 2026. FDAgent holds no copy that FDA posted.
- Date not givenFDA-FOIA-2026-4307Form 483Records released
Inspection ending Jul 14, 2025OAI
7 requests. In the inspection history above. FDA has posted its Form 483.
- Feb 10, 2026FDA-FOIA-2026-1422Form 483· 1 of 4 sitesRecords released
- Jan 2, 2026FDA-FOIA-2026-23Form 483Records released
- Oct 15, 2025FDA-FOIA-2025-12008Inspection recordsRecords released
- Oct 15, 2025FDA-FOIA-2025-11914Form 483 + EIRRecords released
- Sep 23, 2025FDA-FOIA-2025-11211Form 483· 1 of 6 sitesClosed
- Aug 29, 2025FDA-FOIA-2025-10412Form 483Directed Requester to Publically Available Information
- Aug 22, 2025FDA-FOIA-2025-10102Form 483Records released
Inspection ending Jan 1, 2025
1 request. Not listed in FDA’s Data Dashboard yet. FDA released records in answer to a request for its Form 483 on Oct 3, 2025. FDAgent holds no copy that FDA posted.
- Aug 1, 2025FDA-FOIA-2025-9338Form 483Records released
Inspection ending Jan 1, 2024
1 request. Not listed in FDA’s Data Dashboard yet. FDA released records in answer to a request for its Form 483 + EIR on Mar 19, 2026. FDAgent holds no copy that FDA posted.
- Oct 15, 2025FDA-FOIA-2025-11922Form 483 + EIRRecords released
Inspection ending Nov 15, 2023VAI
4 requests. In the inspection history above. FDA has posted its Form 483.
- Jun 3, 2024FDA-FOIA-2024-4818Inspection recordsClosed
- Feb 15, 2024FDA-FOIA-2024-1722Form 483Closed
- Jan 3, 2024FDA-FOIA-2024-67Form 483Closed
- Dec 20, 2023FDA-FOIA-2023-11343Form 483Closed
Inspection ending Oct 1, 2023
2 requests. Not listed in FDA’s Data Dashboard yet. None of FDA’s log entries for these requests say records were released.
- Dec 27, 2023FDA-FOIA-2023-11501EIRClosed
- Dec 27, 2023FDA-FOIA-2023-11500Form 483Closed
Inspection ending May 12, 2023
4 requests. Not listed in FDA’s Data Dashboard yet. FDA has posted its Form 483.
- FY2023FDA-FOIA-2023-5783Form 483 + EIRClosed
- FY2023FDA-FOIA-2023-5591Form 483Closed
- FY2023FDA-FOIA-2023-5574Form 483Closed over fees
- FY2023FDA-FOIA-2023-5564Form 483Closed
Inspection ending May 1, 2023
1 request. Not listed in FDA’s Data Dashboard yet. None of FDA’s log entries for these requests say records were released.
- FY2023FDA-FOIA-2023-6400Form 483Closed
Inspection ending Sep 1, 2022VAI
1 request. In the inspection history above. FDA has posted its Form 483.
- Sep 22, 2022FDA-FOIA-2022-6792Form 483Withdrawn
Inspection ending Jan 1, 2018
1 request. Not listed in FDA’s Data Dashboard yet. None of FDA’s log entries for these requests say records were released.
- Aug 8, 2025FDA-FOIA-2025-9565Form 483 + EIRClosed, no reply from requester
Inspection ending May 19, 2017VAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Sep 15, 2022FDA-FOIA-2022-6618Form 483 + EIRClosed
Inspection date not given
25 requests. FDA released records in answer to a request for its Form 483 on Oct 3, 2025. FDAgent holds no copy that FDA posted.
- FY2023FDA-FOIA-2023-7808EIRClosed
- FY2023FDA-FOIA-2023-6299Form 483 + EIRClosed
- FY2023FDA-FOIA-2023-5568Form 483Closed
- FY2023FDA-FOIA-2023-5552Form 483Closed
- FY2023FDA-FOIA-2023-5486Form 483Closed
- FY2023FDA-FOIA-2022-8045EIRWithdrawn
- Jul 5, 2026FDA-FOIA-2026-6379Form 483Directed Requester to Publicly Available Information
- Jun 8, 2026FDA-FOIA-2026-5625Form 483Records released
- Jun 4, 2026FDA-FOIA-2026-5563Form 483Records released
- Jun 4, 2026FDA-FOIA-2026-5562Form 483 + EIRRecords released
- Jun 2, 2026FDA-FOIA-2026-5470Form 483· 1 of 4 sitesRecords released
- May 8, 2026FDA-FOIA-2026-4726Inspection recordsRecords released
- Apr 1, 2026FDA-FOIA-2026-3181Form 483 + EIRWithdrawn
- Feb 5, 2026FDA-FOIA-2026-1250EIRClosed, no reply from requester
- Dec 9, 2025FDA-FOIA-2025-13771Form 483 + EIRWithdrawn
- Dec 1, 2025FDA-FOIA-2025-13404Form 483 + EIRClosed over fees
- Nov 16, 2025FDA-FOIA-2025-13003Inspection recordsWithdrawn
- Nov 14, 2025FDA-FOIA-2025-12972EIRWithdrawn
- Oct 15, 2025FDA-FOIA-2025-11928Form 483 + EIRRecords released
- Oct 15, 2025FDA-FOIA-2025-11913Form 483 + EIRRecords released
- Sep 23, 2025FDA-FOIA-2025-11188Form 483Records released
- Sep 15, 2025FDA-FOIA-2025-10929Form 483 + EIRRecords released
- Sep 23, 2024FDA-FOIA-2024-8378Form 483 + EIRNo records found
- Jan 2, 2024FDA-FOIA-2024-26Form 483Closed
- Sep 7, 2022FDA-FOIA-2022-6451EIRNo records found
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDER · Office of Manufacturing Quality · DrugsLetter on fda.govFDA record
FDA drug establishment registration
FDA registration search- Registered operations
- Manufacturing, packing, labeling
- Registration runs through
- Dec 31, 2027
- Registrant
- Catalent Indiana, LLCD-U-N-S 172209277
From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3005949964 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3005949964. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.