Skip to content

Genzyme Corporation Form 483, October 2008

FDA issued this Form 483 to Genzyme Corporation in Allston, Massachusetts (FEI 1000305672) at the end of a drug inspection that ended October 10, 2008. It lists 16 observations. FDA classified the inspection Official Action Indicated (OAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Jan 17, 2024. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Oct 10, 2008
Drugs · Drug Quality Assurance
FDA's outcome
OAI
Official Action Indicated
Posted by FDA
Jan 17, 2024
5,577 days after the inspection
Facility
FEI 1000305672
Allston, Massachusetts

The 16 observations on this Form 483

Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.

  1. 1

    Bioburden monitoring inadequate

    Bioburden is not adequately monitored during [illegible] Cerezyme, Fabrazyme and Myozyme 160 L process [illegible] the purification of Myozyme 2000 L process [illegible]

  2. 2

    Buffers not adequately controlled

    [illegible] Buffers are not adequately controlled for [illegible] composition nor monitored for bioburden [illegible] is not adequately controlled.

  3. 3

    Batch production record control deficient

    The batch production record for drug substance is not adequately controlled. For example, the production record is printed on green paper and is issued by Quality Assurance. Attached to the production record are white manufacturing records. Operators frequently record the performance of manufacturing activities, test results or meter readings in the manufacturing records. The manufacturing records are not issued by Quality Assurance. There is no method to control the production of manufacturing records or to account for manufacturing records.

  4. 4

    Manufacturing activities not adequately documented

    A. Activities performed during drug substance manufacture are not adequately documented. For example: When performance of an activity is optional [illegible] there is frequently no place in the batch record to record whether the activity was performed. Dated and signed crossing-out of optional activities that are not performed is not used consistently in the batch record. Additional activities may be performed during manufacture of some commercial batches as part of a study. These activities are not always reflected in the batch record. Data are recorded in incorrect units. B. Activities performed during drug product manufacture are not documented. For example: a. Myozyme [illegible] was filled into vials on 9/16/08, The filling batch record does not include the number of vials that were not stoppered by the automated [illegible] stoppering machine and stoppered manually by operators prior to lyophilization. b. The forceps used to stopper the vials are not listed in the production record.

  5. 5

    Unauthorized manufacturing activities performed

    Activities are performed during drug substance manufacture that are not authorized by [illegible] During fermentation of Fabrazyme [illegible] to try to [illegible] Although the adjustments were within the approved range the study was not appropriately documented.

  6. 6

    HVAC qualification inadequate for aseptic operations

    [illegible] studies executed August 2007 during the operational qualification of the HVAC system for fill suite FF2-16, did not demonstrate the following: Critical aseptic connections [illegible] Routine functions of aseptic core operators, for example: a. manually placing stoppers or reorienting stoppers using forceps for filled vials b. withdrawing unfilled vials from the filling line for weight checks c. redirecting filled vials [illegible] with stoppers on the exit feed wheel [illegible] Unidirectional air flow surrounding the [illegible] rotary in-feed table [illegible] Opening the lyophilizer door or the automated double doors, as typically operated, into the aseptic preparation area and the effects on unidirectional airflow [illegible] Active viable air sampling and the effects on unidirectional airflow

  7. 7

    Aseptic filling line not qualified

    The aseptic filling of all drug products into vials at higher speeds on the [illegible] filling line has not been adequately qualified for its current use. [illegible] per [illegible] line is currently run [illegible] for the 20 ml vials. The data reported in validation studies documents that above [illegible] per minute, the stoppers clog in the stoppering machine and performance of the machine is affected.

  8. 8

    Surface monitoring procedure deficient

    The procedure "Surface Monitoring [illegible]" QC-019-02, version 03/11/2008 does not require personnel exiting the aseptic fill suite to [illegible] on the [illegible] Additionally, personnel monitoring observed on September 16, 2008 exhibited [illegible]

  9. 9

    Chromatography columns not maintained

    On 9/17/08 rouging was seen on the outside of the chromatography column. Internal surfaces and manual valves on the stainless steel chromatography columns used during drug substance purification are not adequately maintained. In January 2008 the facilities [illegible] chromatography skid maintenance SOP was updated [illegible] with instructions for such maintenance. Performance of this maintenance is dependent on manufacturing requesting the maintenance, [illegible] There is no procedure in manufacturing instructing them to notify facilities that maintenance is needed.

  10. 10

    Cryoshippers not validated or maintained

    For the Cryoshippers which are used to transport master cell banks and working cell banks between manufacturing facilities: The use of these cryoshippers has not been validated. The manual for these shippers lists preventative maintenance steps for both continuous maintenance and annual maintenance. The firm has not conducted any maintenance on any of [illegible] shippers currently in use. The manual for these shippers states that the life expectancy of the shippers is 5 years. [illegible] of these shippers have been in use since 2002/2003.

  11. 11

    QA approved failing validation report

    Quality assurance reviewed and approved a final validation report for The Use of Bagged Serum Stoppers in Liquid Fill Finish Processes study [illegible] which did not meet the protocol requirements.

  12. 12

    Cleaning validation assessment inadequate

    The firms overall assessment of the cleaning validation of Myozyme and Fabrazyme [illegible] was inadequate as the [illegible] to determine which product is the most difficult to remove was unknown until September 5, 2008.

  13. 13

    Document control procedure not followed

    The procedure entitled: Document System and Control, GN-001-01, is used for the review and approval of documents. This procedure was not followed in the approval of [illegible] documents.

  14. 14

    Deviation files incomplete

    Deviation files are incomplete. For example: For those deviations resulting from operator error, the deviation file did not contain documentation that the operators were re-trained or counseled as a corrective measure. For deviations where informal investigations were conducted, there was no documentation of the investigation.

  15. 15

    Incorrect buffer gravity entry

    Portions of formulating the Fabrazyme Elution Buffer are automated. The formulation for the Elution buffer [illegible] was entered into the [illegible] console which is an automated system. Since 1999, the specific gravity for the buffer was incorrectly entered into the computer program as [illegible] The correct specific gravity of the Fabrazyme Elution Buffer is [illegible] as indicated in the current production record.

  16. 16

    Master records deficient

    Master production and control records for drug product are deficient in that they: Do not include a description of the drug product containers, closures and packaging materials, a specimen or copy of each label and all other labeling, and the signatures [illegible] responsible for the approval of labeling. B. Master production and control records for drug substance are deficient in that they: Do not include an accurate statement of weight of each component [illegible] Do not include a statement of theoretical yield including the maximum and minimum percentages of theoretical yield beyond which investigation is required.

After this inspection

FDA's compliance actions for this FEI dated after this inspection and before its next one. FDAgent links a letter when FDA's case number ties it to this facility.

  • Warning LetterFeb 27, 2009CDER · 140 days after this inspection ended

Other Form 483s FDA posted for this facility

Every Form 483 FDA posts under FEI 1000305672, newest first.

All inspections at this facility →

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.