Central Glass Co., Ltd.
5253,Okiube, Ube, Japan
FEI 3002806649DUNS 706277696
FDA has 5 inspections on file for Central Glass Co., Ltd. in Ube, Japan (FEI 3002806649), from 2009 to 2023. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA posted the Form 483 from that inspection. FDAgent read its 6 observations from the PDF; they are listed below.
No warning letter, court action or import refusal is filed under this FEI.
As of Nov 21, 2026, the site is registered with FDA as a drug establishment for API manufacturing, through Dec 31, 2027. Its registrant is Central Glass Co., Ltd. (D-U-N-S 706277696).
Sources:FDA firm profile ↗Form 483 PDF ↗FDA drug establishment registrations ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 5, 2009–2023
- 0 OAI · 5 VAI · 0 NAI
- Inspections with 483 observations shown
- 1 of 5
- Form 483s posted by FDA
- ↗
- Requested under FOIA
- 2 requests, latest Jun 29, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
- Drug establishment registration
- Through
Inspection history
5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssurancePosted 483 · 6 observations
Read this Form 483 on its own page →Form 483 PDF on fda.gov
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.
- Observation 1
Inadequate laboratory controls and standards
Laboratory Controls are inadequate. Specifically, a) The working standards used during release testing of (b) (4) for assay and volatile analogues not always elude peaks for the corresponding volatile analogues. During qualification of your working standards one of the requirements is to verify the presence of all peaks. However, injection of standard lot (b) (4) during analysis of commercial (b) (4) API lots (b) (4) does not contain peaks for all volatile analogues at their respective retention time. b) Access control to samples that have not been assigned and those for which testing has been completed is not restricted. In addition on 9/11/23 completed samples of released (b) (4) batches (b) (4) [illegible] Your firm noted verbally that the completed samples were not discarded because they will be used to qualify analysts, however there was no documentation to support this. c) Your site does not conduct growth promotion of media used to perform environmental monitoring of the filling room used for (b) (4) API. d) On 9/11/23 I observed a filled GC vial labeled as "extra" unattended in the laboratory near to GC 23 stand alone equipment, when requested for the origin of the vial and content your site could not establish if the vial contained System Suitability (SST) solution, standard, or samples. Your site explained that they believed the vial was filled with System Suitability solution however, there was no documentation for this. e) During the update of software and hardware for system GC-24 the number of theoretical plates was not calculated prior to placing the system back in use.
- Observation 2
Water specs lack B. Cepacia testing
Specifications of Water used in the manufacture of APIs do not demonstrate the water is suitable for its intended use. Specifically, specifications for water used for cleaning and processing after (b) (4) water and (b) (4) (b) (4) do not include testing to confirm the absence of B. Cepacia from your water system.
Show observations 3 to 6
- Observation 3
Cleaning procedures not validated
Your cleaning procedures are not validated. Specifically, a) Your firm has not performed validation of the cleaning procedures used to clean equipment that is used for (b) (4) filling API into containers. (b) (4) varies. For example the (b) (4) however the number of consecutive batches manufactured on the equipment in (b) (4) [illegible] b) (b) (4) batches, however the (b) (4) campaign executed to asses the cleaning re-validation study of 10/15/20 consisted of (b) (4) c) Cleaning validation studies for bulk containers including clean hold time do not include samples to determine appropriate microbial levels. d) Removal of (b) (4) (b) (4) is verified using (b) (4) this method is not used during (b) (4) [illegible] f) (b) (4) sensitive to indicate if (b) (4) is reduced to levels that indicate it has been removed. e) Your test method used from 2015-2020 to evaluate cleaning validation samples had an LOQ of (b) (4) ppm. This test was not suitable for the evaluation of cleaning validation samples (b) (4) specifications for residue were (b) (4) ppm during this time.
- Observation 4
Change control system not followed
Changes that could affect the production and control of the intermediate or API are not always handled using the formal change control system. Specifically, a) The following changes were not handled using the change control system (list is not exhaustive): Change of AC system in filling room, your firm noted that the reason to not execute a change control for this (b) (4) change was because the AC was being replaced with a unit that met the same specifications. However at the time of the change the comparison between the old and new unit was not documented to justify the lack of change control. On 8/6/23 UQSE20-19 test method used to evaluate rinse samples from cleaning validation was changed to obtain a lower LOD and LOQ the residue limit on equipment was also changed from (b) (4) ppm to (b) (4) ppm. The reason for not performing this change using the change control program was not documented. b) Change control CCQ22-17 enacted to address change in storage of (b) (4) medium used by QC from refrigeration to room temperature states that the firm will verify if the storage conditions change has any impact on tests. However there were no steps, protocol or instructions on how your firm would perform the noted verification. In addition on 9/15/23 I observed that there is no calibrated device to monitor the temperature of the room where the media is stored.
- Observation 5
Validation lacks critical process parameters
Your validation protocols for the execution of revalidation studies do not specify critical process steps. Your process validation studies for (b) (4) do not show the critical process parameters that were going to be studied, review of batch records shows that the validation study did not include variation of parameters within the established ranges. In addition parameters such as the timer used to control the flow in the (b) (4) are not defined as critical process parameters.
- Observation 6
Inadequate raw material receiving procedures
Your receiving procedures for raw materials are inadequate. Specifically, You do not have procedures that require proof from your vendor that the tanks used to deliver (b) (4) (b) (4) are properly cleaned between each delivery.
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on Jun 29, 2026.
Inspection date not given
2 requests. FDA released records in answer to a request for its Inspection records on Jul 6, 2026. FDAgent holds no copy that FDA posted.
- Jun 29, 2026FDA-FOIA-2026-6198Inspection records· 1 of 3 sitesRecords released
- Nov 1, 2023FDA-FOIA-2023-9578Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
FDA drug establishment registration
FDA registration search- Registered operations
- API manufacturing
- Registration runs through
- Dec 31, 2027
- Registrant
- Central Glass Co., Ltd.D-U-N-S 706277696
From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3002806649 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3002806649. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.