Skip to content

AVEM Labs LLC Form 483, May 2025

FDA issued this Form 483 to AVEM Labs LLC in Las Vegas, Nevada (FEI 3017596057) at the end of a biologics inspection that ended May 21, 2025. It lists 5 observations. FDA classified the inspection Voluntary Action Indicated (VAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Oct 20, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
May 21, 2025
Biologics · Human Cellular, Tissue, and Gene Therapies
FDA's outcome
VAI
Voluntary Action Indicated
Posted by FDA
Oct 20, 2026
517 days after the inspection
Facility
FEI 3017596057
Las Vegas, Nevada

The 5 observations on this Form 483

Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.

  1. 1

    Unvalidated amniotic membrane manufacturing process

    Processes with results which could not be fully verified by inspection and tests, were not validated and approved according to established procedures. Specifically, your firm has not validated the manufacturing process for amniotic membrane (b) (4) derived from human amniotic membrane. Since January 2025, you have manufactured and distributed (b) (4) amniotic membrane (b) (4)

  2. 2

    Label accuracy verification procedures inadequate

    Labeling procedures did not include verification of label accuracy. Specifically, your quality assurance personnel are not reviewing and approving all appropriate finished product labels as they relate to the amniotic membrane (b) (4) of BioVair DL that are distributed as (b) (4) , and (b) (4) (b) (4).

  3. 3

    Temperature monitoring devices not calibrated

    Equipment used for measuring was not calibrated according to established schedules. Specifically, you have not calibrated your temperature monitoring devices, (b) (4) , S/N (b) (4) and S/N (b) (4) . These temperature monitoring devices are used to monitor temperatures of finished and quarantined lots of amniotic membrane (b) (4) BioVair DL.

  4. 4

    Supply receipt documentation not maintained

    Documentation of the receipt of supplies was not maintained. Specifically, you did not maintain records of receipt that includes the date of receipt, quantity obtained, and quality release for use of your (b) (4) part ID: (b) (4) and (b) (4) part ID: (b) (4) that were manufactured by your supplier on 03/19/2019 and are currently being used as primary packaging for your BioVair DL (b) (4) and (b) (4) amniotic membrane (b) (4).

  5. 5

    HCT/P records not retained ten years

    Records pertaining to HCT/Ps were not retained at least 10 years after the date of distribution, disposition, or expiration, whichever was latest when the date of administration of the HCT/P was unknown. Specifically, records related to manufacturing and donor eligibility were not available for review for products manufactured prior to January 2025. For example, we requested to review lot (b) (4) (b) (4) of commercially distributed (b) (4); however, the requested documentation was not made available during our inspection.

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.