Skip to content

Corrective and preventive action (CAPA) in FDA Form 483 observations

Corrective and preventive actions: whether firms open them, carry them out on time and check that they worked. FDAgent finds 141 such observations in 111 Form 483s issued to 107 facilities, for inspections that ended between November 2008 and September 2026. Each is quoted below from the 483 FDA posted.

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.

Observations
141
Each quoted below
Form 483s
111
Posted by FDA
Facilities
107
Distinct FEIs
Most recent
Sep 28, 2026
Inspection end date

Observations by year

These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).

Peak: 51 in 2026

CAPA observations in Form 483s by inspection year
YearObservations
20081
20090
20101
20110
20121
20131
20140
20152
20162
20175
20189
20194
20201
20217
20228
202316
20247
202525
202651

How FDAgent sorts observations into this theme

Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.

It is tied to one of these 21 CFR entries

21 CFR 820.100

A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.

Or its text contains one of these phrases

“CAPA*” · “corrective and preventive action*” · “corrective and preventative action*” · “corrective action*” · “preventive action*” · “preventative action*” · “effectiveness check*” · “effectiveness verification*”

* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.

The 141 observations, newest first

Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.

  1. Farmakeio Outsourcing LLCInspection ended Sep 5, 2025 · FEI 3014982757 · Observation 1

    There is a failure to thoroughly review any unexplained discrepancy and the failure of a batch or any of its components to meet any of its specifications whether or not the batch has been already distributed. Specifically, A. Your firm failed to adequately investigate and initiate corrective and preventive actions for CAPA #, CAPA-2025-0002 dated 03 Apr 2025, which documents metal shavings being found during Testosterone/Triamcinolone Acetonide …

    Contains “corrective and preventive action*”Read the Form 483 →

  2. Farmakeio Outsourcing LLCInspection ended Sep 5, 2025 · FEI 3014982757 · Observation 4

    … During a review of your firm initiated CAPA-2025-002, Finding foreign material - granulation process, your firm documented the root cause for the origin of the foreign material resulted from the granulator impeller shaft grinding against the [(b)(4)]. A preventive maintenance (PM) was added for the [(b)(4)] - Lab Grinder to include an inspection the [(b)(4)] before running the granulator. Your firm failed to include in the PM an inspection of the …

    Contains “CAPA*”Read the Form 483 →

  3. Aptuit LtdInspection ended Aug 22, 2025 · FEI 3003444315 · Observation 2

    … The quality unit neither performed testing to the APIs to detect [illegible] nor established a corrective and preventive action to avoid recurrence. These APIs were sent to the US market and other markets.

    Contains “corrective and preventive action*”Read the Form 483 →

  4. Raptim Research Private LimitedInspection ended Jul 25, 2025 · FEI 3007267856 · Observation 1

    … reports found in a binder labeled "Miscellaneous data" in instrument room 02 were incomplete because they did not contain departmental responses, required corrective actions and subsequent QC verification. Impacted projects include (b) (4) and RRL-ANA-MV-896. e. eSOP/ANA/001 states that biological matrix (e.g., plasma) should be used within one year after collection. The firm had multiple lots of pooled plasma samples exceeding the one-year …

    Contains “corrective action*”Read the Form 483 →

  5. Elite Orthopedics, LLCInspection ended Jul 15, 2025 · FEI 3011803392 · Observation 5

    … • Packing List • Tissue Tracking Form • Incident Report Form • Corrective Action Request Form • Weekly Freezer Maintenance Form • Tissue Tracking Log

    Contains “corrective action*”Read the Form 483 →

  6. Foremount Enterprise Co., LtdInspection ended Jun 19, 2025 · FEI 3010402691 · Observation 5

    Procedures for corrective and preventive action have not been adequately established. Specifically, your firm's corrective and preventative action (CAPA) procedure Problem Improvement Procedure, QP851, Ver.20, dated December 25, 2024, has not been adequately established. Your firm was not able to provide evidence of an established CAPA system.

    Opens with FDA's citation wording for 21 CFR 820.100(a)Read the Form 483 →

  7. Medi-Fare Drug Pharmaceutical Compounding LLCInspection ended Jun 6, 2025 · FEI 3009925820 · Observation 1

    … After this batch rejection, you documented CAPA-24-005 which includes the use of a new vendor cap. While CAPA-24-005 requires container closure integrity testing of approximately [illegible] units, it does not require such actions as process validation, stability testing and / or media fills.

    Contains “CAPA*”Read the Form 483 →

  8. Medi-Fare Drug Pharmaceutical Compounding LLCInspection ended Jun 6, 2025 · FEI 3009925820 · Observation 6

    … Your root cause was inconclusive, and you did not document additional CAPA actions to address this failure.

    Contains “CAPA*”Read the Form 483 →

  9. MSD International GmbHInspection ended Apr 11, 2025 · FEI 3002808087 · Observation 1

    … In response to the deviations listed above, six CAPAs were opened to assess and remediate (b) (4) design errors that were not identified during (b) (4) validation. The status for the six CAPAs are as follows: • CAPA QN201232350 – Complete, waiting for effectiveness verification • CAPA QN201232351 – Open (Due Date: 27APR2025) • CAPA QN201232353 – Complete, waiting for effectiveness verification • CAPA QN201232354 – Complete, waiting for …

    Contains “CAPA*”Read the Form 483 →

  10. Guangxi Yulin Pharmaceutical Group Co. Ltd.Inspection ended Mar 21, 2025 · FEI 1000589993 · Observation 5

    … In addition, no preventive actions were assessed or proposed for deviation PC-20250201 in which an unplanned power outage led to a loss of power to your firm's stability chambers.

    Contains “preventive action*”Read the Form 483 →

  11. Hubei JXBio Pharmaceutical Co., Ltd.Inspection ended Mar 14, 2025 · FEI 3016784508 · Observation 1

    … The CAPA for replacement of all the electronic scales in the workshop with those having a printing function was not implemented using the established change control process. B. An investigation for deviation PC-C005A-24002 was initiated on 04-04-2024 for a single impurity test result (b) (4) (b) (4) (b) (4) that did not meet requirements during the of Batch (b) (4) The deviation was opened following the closing of OOS-C005A-24001 on 04-05-2024 …

    Contains “CAPA*”Read the Form 483 →

  12. Bioclinica Inc. / Clario (eResearch Technology, Inc.)Inspection ended Feb 27, 2025 · FEI 3032745961 · Observation 1

    … accuracy, and reliability of clinical trial data for study (b)(4) Specifically, your firm did not implement effective validation controls, or adequate corrective and preventive actions. For Example: A) Quality Event DEV-06296 documents that correction factors were not properly applied to data in the final data transfer on June 26, 2024, affecting 31 subjects. "There was an IQC correction in prior transfer. In the Final/End of Study transfer, the …

    Contains “corrective and preventive action*”Read the Form 483 →

  13. Emcure Pharmaceuticals LimitedInspection ended Feb 25, 2025 · FEI 3005782879 · Observation 1

    … Dec 2024", yet exhibited visible residues [illegible] (b)(4) particulates (b)(4) liquid [illegible] material, and (b)(4) [illegible] Dirty Hold Time without corrective action [illegible] and [illegible] ng was performed beyond the [illegible] (b)(4) swab tests confirmed contamination [illegible] There is no supporting data exists for DHT or Clean Hold Time (CHT) limits for Type (b)(4) or [illegible] An operator performed cleaning without …

    Contains “corrective action*”Read the Form 483 →

  14. GE HealthCareInspection ended Feb 7, 2025 · FEI 3002807778 · Observation 1

    Deviation investigations are not conducted thoroughly to identify root cause(s) and insufficient CAPA actions were implemented to prevent reoccurrences. Specifically, multiple deviation investigations were initiated for endotoxins detected in the API process intermediates, during the manufacturing of (b) (4) API (b) (4) Endotoxins were detected from multiple batches of API products manufactured between Feb 2022 to May-Sept 2024, initiating at …

    Contains “CAPA*”Read the Form 483 →

  15. GlaxoSmithKline Biologicals Kft.Inspection ended Jan 31, 2025 · FEI 3008322822 · Observation 2

    Procedures for the acceptance and rejection of components and materials utilized within drug substance manufacturing are not adequate. Procedure LSOP 9000087267 (ver. 01 – approved 20MAR2024), titled "FILT.UF (b)(4)" was implemented as a corrective action in response to deviation 201042740 and prior to use within manufacturing. The procedure requires visual inspection of the (b)(4). The procedure allows for the acceptance and use of (b)(4) …

    Contains “corrective action*”Read the Form 483 →

  16. Zhejiang Huahai Pharmaceutical Co., Ltd.Inspection ended Jan 24, 2025 · FEI 3003999190 · Observation 5

    … No investigation and subsequent CAPA plan into the cause of the gap has been made. 2. There is no record documenting the approval and use of the non-sterile tape to seal the gap within the Grade A filling line, and there is no work order or documentation detailing the installation, and subsequent replacement (if performed) of the tape. 3. An assessment was not performed to evaluate how the gap between the HEPA filter/ (b114? in the aseptic …

    Contains “CAPA*”Read the Form 483 →

  17. Janssen Vaccines CorporationInspection ended Nov 19, 2024 · FEI 3012637764 · Observation 1

    … Compliance (ET) stated that the investigation concluded the (b) (4) was not related to manufacturing process and hence the defect was not confirmed and no CAPA was implemented. Your Site Quality Head (SO) stated that the quality issue is related to end user and not with the drug product manufacturing.

    Contains “CAPA*”Read the Form 483 →

  18. Staska Pharmaceuticals IncInspection ended Sep 18, 2024 · FEI 3017473850 · Observation 1

    … CAPAs included (b) (4) vials to prevent rattling and post-cleaning inspections by the vial manufacturer. The batch was rejected but no CAPA's were implemented by your firm, and no effectiveness check was conducted. A second deviation occurred on July 19, 2024. During your firm's visual inspection of Ascorbic Acid 500mg/mL 50mL Vials for sterile injection Lot (b) (4), BUD 12/31/2024, you initially identified (b) (4) vials containing glass …

    Contains “CAPA*”Read the Form 483 →

  19. Protherics UK LimitedInspection ended Aug 13, 2024 · FEI 3001417487 · Observation 1

    … These investigations were generated in accordance with procedure (b)(4)-SOP-()-000015, Control of Quality Events and Corrective and Preventive Actions, Ver 21, 23 & 23, which allow for such actions as additional monitoring activities (e.g., (b)(4)) and requires the use of a root cause investigation (which would allow for the expansion of the scope of investigation). Although your firm initiated various corrective actions (CAPAs), the following …

    Contains “corrective and preventive action*”Read the Form 483 →

  20. Serum Institute of India Pvt. Limited Manjari SiteInspection ended Jul 2, 2024 · FEI 3020223359 · Observation 5

    DEV-003-2024-0049 detected on January 29, 2024 indicated a critical defect, (b) (4) (b) (4) was observed during AQL of (b) (4), batch (b) (4). Your firm failed to fully investigate and implement an effective corrective action/preventative action after identifying (b) (4) of (b) (4) for (b) (4) contributed to this critical defect.

    Contains “corrective action*”Read the Form 483 →

21 CFR sections these observations open with

The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.

Ask FDAgent about CAPA in Form 483s

FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
  • Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
  • A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.