Raptim Research Pvt. Ltd.

A - 239/A - 242, Navi Mumbai, India

FEI 3007267856

FDA has 6 inspections on file for Raptim Research Pvt. Ltd. in Navi Mumbai, India (FEI 3007267856), from 2019 to 2025. The latest ended on and was classified OAI, official action indicated.

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA posted the Form 483 from that inspection. FDAgent read its 5 observations from the PDF; they are listed below. Two of the three FOIA requests for this site have asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Form 483 PDF ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
6, 2019–2025
2 OAI · 1 VAI · 3 NAI
Inspections with 483 observations shown
1 of 6
Form 483s posted by FDA
↗
Requested under FOIA
3 requests, latest May 5, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDrugs · Bioresearch MonitoringPosted 483 · 5 observations

    Read this Form 483 on its own page →Form 483 PDF on fda.gov

    Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.

    1. Observation 1

      Failure to follow SOPs and STPs

      The firm did not follow practices and procedures established in standard operating procedures (SOPs) or standard test procedures (STPs). Specifically: a. eSOP/ANA/006 (version 05, Amendment 01) states that analytical runs should not be considered final until QA has reviewed and approved the status of the run. In study (b) (4), run 251024 (b) (6) was finalized as rejected by QA on 29/Oct/2024. However, samples from subject (b) were reanalyzed on 28/Oct/2024 (run 281024REP01 (b) (6)), which was prior to final QA approval of run 251024 (b) (6). b. eSOP/ANA/001 states that biological sample disposal details should be recorded on the biological sample disposal record form, ANA-eFM-179. However, for multiple studies including (b) (4), (b) (4), (b) (4), and (b) (4), the form did not contain complete disposal details. c. The STP/ANA/946 for quantitation of (b) (4) and (b) (4) requires that samples should be discarded after completion of an analytical batch. However, in project (b) (4), samples from a repeat run completed on 03/Jul/2025, were retained in the autosampler until at least 14/Jul/2025, despite batch processing records indicating sample disposal on 04/July/2025. d. eSOP/ANA/021 states that QC audit findings reported in the QC report will be reviewed and addressed by the respective departments, including any corrections as needed. The corrections will be subsequently verified by QC. However, QC review reports found in a binder labeled "Miscellaneous data" in instrument room 02 were incomplete because they did not contain departmental responses, required corrective actions and subsequent QC verification. Impacted projects include (b) (4) and RRL-ANA-MV-896. e. eSOP/ANA/001 states that biological matrix (e.g., plasma) should be used within one year after collection. The firm had multiple lots of pooled plasma samples exceeding the one-year expiration date that were not discarded. Examples include plasma lots: i. 210324 (b) (4) - stored on 21/Mar/2024; 46 mL not discarded ii. 210324 - stored on 21/Mar/2024; 135 mL not discarded iii. 180324 - stored on 18/Mar/2024; 144 mL not discarded iv. 080523 - stored on 05/Aug/2023; 169 mL not discarded

    2. Observation 2

      Inadequate source records prevent study reconstruction

      The firm did not have adequate information in their source documents for complete reconstruction of study activities. a. Analytical staff could not locate the current storage location of lot 012024004103 of plasma. The plasma was transferred from the clinical department to the analytical department on 8/Nov/2024 for use as blank biological matrix for method development/validation. The Biological Matrix Consumption Record documented that the plasma was stored in Freezer INS/ANA/505 on 8/Nov/2024. Freezer INS/ANA/505 is no longer present at the facility. b. Disposal of samples from study (b) (4) (Analytical aliquot) was documented on 07/Aug/2023 in the sample disposal record. However, the samples were documented as disposed on 17/Oct/2024 in CREDBio. c. The unique identifier of a lot of whole blood used as blank biological matrix for method development RRL-ANA-MD-1336 was documented incorrectly. Lot No. 012025000575 of whole blood (300 ml) was transferred from the clinical department to the analytical department on 2/Mar/2025. The volunteer consent form for blank biological matrix donation lists the unique identifier as 012025000575. However, the Biological Matrix Consumption Record and the documentation for RRL-ANA-MD-1336 document the lot No. as 012025000515. d. CredBio and the logbook for freezer INS/ANA/507 documented that period 1, period 2, and period 3 samples from study (b) (4) were stored on shelf A4 of the freezer. However, when sample custodians attempted to retrieve the samples, period 2 samples were not located on shelf A4, but were instead on shelf A5. e. The Sample Disposal Records from in vitro study (b) (4) contained multiple discrepancies. i. Samples from Donor (b) (6) cell 10, aliquot 1 were documented as disposed of on a Biological Sample Disposal Record dated 14/May/2025. However, an additional form for in vitro sample disposal dated 18/Jun/2025 documents the disposal of samples from Donors (b) (6), aliquot 1, including Donor (b) (6) cell 10, aliquot 1. The freezer logbook for INS/ANA/259 also documented disposal of samples from Donor (b) (6) cells 1-22, aliquot 1 on 18/Jun/2025. ii. The in vitro sample disposal record dated 14/May/2025 documents that the samples were retrieved for disposal from freezer number INS/ANA/269. The samples are also documented as being retrieved and discarded of from freezer INS/ANA/259 on 14/May/2025. INS/ANA/269 is not a valid Raptim Research freezer identification number. iii. There were 308 samples collected per donor (14 samples per cell, 22 cells) in study (b) (4). The Biological Sample Disposal Record dated 14/May/2025 documents that 14 samples were discarded for Donor (b) (6) cell 10, aliquot 1. The INS/ANA/259 freezer log documents that the total number of samples discarded for Donor (b) (6) is 292 which indicates that 16 samples were previously discarded instead of 14 samples. iv. The In vitro Sample Disposal Record dated 17/Jun/2025 documents that 2380 samples were discarded for Donors (b) (6), aliquot 1. The logbook for freezer INS/ANA/259 documents that 2378 samples were discarded for Donors (b) (6), aliquot 1. f. There was a discrepancy in the biological matrix consumption record for the pooled plasma from study (b) (4) (lot 120324 (b) (4)). Pooled plasma was prepared on 12/Mar/2024, but the lot ID, matrix type, anticoagulant, number of containers, and freezer ID for that lot were documented on 12/Mar/2023. g. The biological matrix consumption record and CREDBio entry for pooled plasma lot 120324 (b) (4) from study (b) (4) documented the presence of 3 vials (vial 01 containing 100 mL; vial 02, containing 35 mL; and vial 03 containing 35 mL) in freezer INS/ANA/835 on shelf A3. However, vial 01 could not be located in the freezer. h. There was a discrepancy in the pooled plasma lot 100124 (b) (4) retrieval information between the biological matrix consumption record and CREDBio. Consumption records document storage of 6 pooled plasma containers in freezer ANA/INS/506,

      Shortened here; the PDF has the full text.

    Show observations 3 to 5
    1. Observation 3

      Inconsistent practices and procedures

      The firm did not demonstrate consistent practices and procedures. Specifically: a. In study (b) (4), segregation activity for ISR experiment 02 (samples from subjects (b) (6) and (b) (6)) was documented in CREDBio. However, this was not done for ISR experiment 01 (samples from subjects (b) (6)). b. Replicate aliquot samples were not consistently documented in CREDBio for all studies. For example, in study (b) (4), information from replicate aliquot samples was entered in CREDBio on the date the samples were received (24/Feb/2024). However, in study (b) (4), samples were received on 27/Sep/2024, but sample information was not entered in CREDBio until 13/Apr/2025. Similarly, in study (b) (4) samples were received on 03/Apr/2024, but sample information was not entered in CREDBio until 26/Feb/2025. c. Sample transfer records were not consistently uploaded into CREDBio. For example, for studies (b) (4) and (b) (4), sample transfer records for both analytical and replicate aliquots were uploaded into CREDBio. However, for studies (b) (4) and (b) (4), only the analytical sample transfer records were uploaded to CREDBio.

    2. Observation 4

      Poor documentation practices for method validation

      The firm did not consistently follow good documentation practices in their raw data management during conduct of method validation of (b) (4) (RRL-ANA-MVR1409). For example, in raw data forms (17242, 7215, 0597) several corrections were made to address QC findings. However, the changes were made without documenting when the change was made and who made the change.

    3. Observation 5

      Failure to follow computer backup SOP

      The firm did not consistently follow the standard operating procedure eSOP/IT/006 v 01 for back up for computer system. For example, for a 90-day period (22/April-20/July/2025) that was reviewed for instrument ID API2000-01, the backup plan failed on July 14, June 27, June 26, June 16, June 03, June 1, May 1, and April 30, 2025. The eSOP/IT/006 v 01 required a full and incremental back up of the computer systems. Although the backup activity failed, backup activity was indicated to have been verified by the IT administrator and QA in verification of backup logbook (RRL-IT-LG-007-001/25).

  2. OAISignificant objectionable conditions; FDA recommended regulatory actionDrugs · Bioresearch MonitoringCitations not published
  3. NAINo objectionable conditions that need follow-upDrugs · Bioresearch MonitoringNone
  4. NAINo objectionable conditions that need follow-upDrugs · Bioresearch MonitoringNone
  5. NAINo objectionable conditions that need follow-upDrugs · Bioresearch MonitoringNone
  6. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Bioresearch MonitoringCitations not published

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Three requests to FDA under the Freedom of Information Act named this site, most recently on May 5, 2026. Each one asked for a Form 483.

  • Inspection ending Jul 25, 2025OAI

    2 requests. In the inspection history above. FDA has posted its Form 483.

    1. May 5, 2026FDA-FOIA-2026-4425Form 483· 1 of 5 sitesWithdrawn
    2. Nov 11, 2025FDA-FOIA-2025-12824Form 483· 1 of 2 sitesRecords released
  • Inspection date not given

    1 request. None of FDA’s log entries for these requests say records were released.

    1. Nov 14, 2024FDA-FOIA-2024-10078Form 483No records found

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3007267856 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3007267856. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.