FDA issued this Form 483 to Raptim Research Private Limited in Navi Mumbai, India (FEI 3007267856) at the end of a drug inspection that ended July 25, 2025. It lists 5 observations. FDA classified the inspection Official Action Indicated (OAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Nov 12, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Jul 25, 2025
Drugs · Bioresearch Monitoring
FDA's outcome
OAI
Official Action Indicated
Posted by FDA
Nov 12, 2026
475 days after the inspection
Facility
FEI 3007267856
Navi Mumbai, India

The 5 observations on this Form 483

Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.

  1. 1

    Failure to follow SOPs and STPs

    The firm did not follow practices and procedures established in standard operating procedures (SOPs) or standard test procedures (STPs). Specifically: a. eSOP/ANA/006 (version 05, Amendment 01) states that analytical runs should not be considered final until QA has reviewed and approved the status of the run. In study (b) (4), run 251024 (b) (6) was finalized as rejected by QA on 29/Oct/2024. However, samples from subject (b) were reanalyzed on 28/Oct/2024 (run 281024REP01 (b) (6)), which was prior to final QA approval of run 251024 (b) (6). b. eSOP/ANA/001 states that biological sample disposal details should be recorded on the biological sample disposal record form, ANA-eFM-179. However, for multiple studies including (b) (4), (b) (4), (b) (4), and (b) (4), the form did not contain complete disposal details. c. The STP/ANA/946 for quantitation of (b) (4) and (b) (4) requires that samples should be discarded after completion of an analytical batch. However, in project (b) (4), samples from a repeat run completed on 03/Jul/2025, were retained in the autosampler until at least 14/Jul/2025, despite batch processing records indicating sample disposal on 04/July/2025. d. eSOP/ANA/021 states that QC audit findings reported in the QC report will be reviewed and addressed by the respective departments, including any corrections as needed. The corrections will be subsequently verified by QC. However, QC review reports found in a binder labeled "Miscellaneous data" in instrument room 02 were incomplete because they did not contain departmental responses, required corrective actions and subsequent QC verification. Impacted projects include (b) (4) and RRL-ANA-MV-896. e. eSOP/ANA/001 states that biological matrix (e.g., plasma) should be used within one year after collection. The firm had multiple lots of pooled plasma samples exceeding the one-year expiration date that were not discarded. Examples include plasma lots: i. 210324 (b) (4) - stored on 21/Mar/2024; 46 mL not discarded ii. 210324 - stored on 21/Mar/2024; 135 mL not discarded iii. 180324 - stored on 18/Mar/2024; 144 mL not discarded iv. 080523 - stored on 05/Aug/2023; 169 mL not discarded

  2. 2

    Inadequate source records prevent study reconstruction

    The firm did not have adequate information in their source documents for complete reconstruction of study activities. a. Analytical staff could not locate the current storage location of lot 012024004103 of plasma. The plasma was transferred from the clinical department to the analytical department on 8/Nov/2024 for use as blank biological matrix for method development/validation. The Biological Matrix Consumption Record documented that the plasma was stored in Freezer INS/ANA/505 on 8/Nov/2024. Freezer INS/ANA/505 is no longer present at the facility. b. Disposal of samples from study (b) (4) (Analytical aliquot) was documented on 07/Aug/2023 in the sample disposal record. However, the samples were documented as disposed on 17/Oct/2024 in CREDBio. c. The unique identifier of a lot of whole blood used as blank biological matrix for method development RRL-ANA-MD-1336 was documented incorrectly. Lot No. 012025000575 of whole blood (300 ml) was transferred from the clinical department to the analytical department on 2/Mar/2025. The volunteer consent form for blank biological matrix donation lists the unique identifier as 012025000575. However, the Biological Matrix Consumption Record and the documentation for RRL-ANA-MD-1336 document the lot No. as 012025000515. d. CredBio and the logbook for freezer INS/ANA/507 documented that period 1, period 2, and period 3 samples from study (b) (4) were stored on shelf A4 of the freezer. However, when sample custodians attempted to retrieve the samples, period 2 samples were not located on shelf A4, but were instead on shelf A5. e. The Sample Disposal Records from in vitro study (b) (4) contained multiple discrepancies. i. Samples from Donor (b) (6) cell 10, aliquot 1 were documented as disposed of on a Biological Sample Disposal Record dated 14/May/2025. However, an additional form for in vitro sample disposal dated 18/Jun/2025 documents the disposal of samples from Donors (b) (6), aliquot 1, including Donor (b) (6) cell 10, aliquot 1. The freezer logbook for INS/ANA/259 also documented disposal of samples from Donor (b) (6) cells 1-22, aliquot 1 on 18/Jun/2025. ii. The in vitro sample disposal record dated 14/May/2025 documents that the samples were retrieved for disposal from freezer number INS/ANA/269. The samples are also documented as being retrieved and discarded of from freezer INS/ANA/259 on 14/May/2025. INS/ANA/269 is not a valid Raptim Research freezer identification number. iii. There were 308 samples collected per donor (14 samples per cell, 22 cells) in study (b) (4). The Biological Sample Disposal Record dated 14/May/2025 documents that 14 samples were discarded for Donor (b) (6) cell 10, aliquot 1. The INS/ANA/259 freezer log documents that the total number of samples discarded for Donor (b) (6) is 292 which indicates that 16 samples were previously discarded instead of 14 samples. iv. The In vitro Sample Disposal Record dated 17/Jun/2025 documents that 2380 samples were discarded for Donors (b) (6), aliquot 1. The logbook for freezer INS/ANA/259 documents that 2378 samples were discarded for Donors (b) (6), aliquot 1. f. There was a discrepancy in the biological matrix consumption record for the pooled plasma from study (b) (4) (lot 120324 (b) (4)). Pooled plasma was prepared on 12/Mar/2024, but the lot ID, matrix type, anticoagulant, number of containers, and freezer ID for that lot were documented on 12/Mar/2023. g. The biological matrix consumption record and CREDBio entry for pooled plasma lot 120324 (b) (4) from study (b) (4) documented the presence of 3 vials (vial 01 containing 100 mL; vial 02, containing 35 mL; and vial 03 containing 35 mL) in freezer INS/ANA/835 on shelf A3. However, vial 01 could not be located in the freezer. h. There was a discrepancy in the pooled plasma lot 100124 (b) (4) retrieval information between the biological matrix consumption record and CREDBio. Consumption records document storage of 6 pooled plasma containers in freezer ANA/INS/506,

    Shortened here; the PDF has the full text.

  3. 3

    Inconsistent practices and procedures

    The firm did not demonstrate consistent practices and procedures. Specifically: a. In study (b) (4), segregation activity for ISR experiment 02 (samples from subjects (b) (6) and (b) (6)) was documented in CREDBio. However, this was not done for ISR experiment 01 (samples from subjects (b) (6)). b. Replicate aliquot samples were not consistently documented in CREDBio for all studies. For example, in study (b) (4), information from replicate aliquot samples was entered in CREDBio on the date the samples were received (24/Feb/2024). However, in study (b) (4), samples were received on 27/Sep/2024, but sample information was not entered in CREDBio until 13/Apr/2025. Similarly, in study (b) (4) samples were received on 03/Apr/2024, but sample information was not entered in CREDBio until 26/Feb/2025. c. Sample transfer records were not consistently uploaded into CREDBio. For example, for studies (b) (4) and (b) (4), sample transfer records for both analytical and replicate aliquots were uploaded into CREDBio. However, for studies (b) (4) and (b) (4), only the analytical sample transfer records were uploaded to CREDBio.

  4. 4

    Poor documentation practices for method validation

    The firm did not consistently follow good documentation practices in their raw data management during conduct of method validation of (b) (4) (RRL-ANA-MVR1409). For example, in raw data forms (17242, 7215, 0597) several corrections were made to address QC findings. However, the changes were made without documenting when the change was made and who made the change.

  5. 5

    Failure to follow computer backup SOP

    The firm did not consistently follow the standard operating procedure eSOP/IT/006 v 01 for back up for computer system. For example, for a 90-day period (22/April-20/July/2025) that was reviewed for instrument ID API2000-01, the backup plan failed on July 14, June 27, June 26, June 16, June 03, June 1, May 1, and April 30, 2025. The eSOP/IT/006 v 01 required a full and incremental back up of the computer systems. Although the backup activity failed, backup activity was indicated to have been verified by the IT administrator and QA in verification of backup logbook (RRL-IT-LG-007-001/25).

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.