Corrective and preventive action (CAPA) in FDA Form 483 observations
Corrective and preventive actions: whether firms open them, carry them out on time and check that they worked. FDAgent finds 141 such observations in 111 Form 483s issued to 107 facilities, for inspections that ended between November 2008 and September 2026. Each is quoted below from the 483 FDA posted.
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.
- Observations
- 141
- Each quoted below
- Form 483s
- 111
- Posted by FDA
- Facilities
- 107
- Distinct FEIs
- Most recent
- Sep 28, 2026
- Inspection end date
Observations by year
These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).
Peak: 51 in 2026
| Year | Observations |
|---|---|
| 2008 | 1 |
| 2009 | 0 |
| 2010 | 1 |
| 2011 | 0 |
| 2012 | 1 |
| 2013 | 1 |
| 2014 | 0 |
| 2015 | 2 |
| 2016 | 2 |
| 2017 | 5 |
| 2018 | 9 |
| 2019 | 4 |
| 2020 | 1 |
| 2021 | 7 |
| 2022 | 8 |
| 2023 | 16 |
| 2024 | 7 |
| 2025 | 25 |
| 2026 | 51 |
How FDAgent sorts observations into this theme
Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.
It is tied to one of these 21 CFR entries
A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.
Or its text contains one of these phrases
“CAPA*” · “corrective and preventive action*” · “corrective and preventative action*” · “corrective action*” · “preventive action*” · “preventative action*” · “effectiveness check*” · “effectiveness verification*”
* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.
The 141 observations, newest first
Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.
Strive Pharmacy LLCInspection ended Sep 28, 2026 · FEI 3021042485 · Observation 1
… There was no additional investigation or corrective actions. Technicians do not document what specific activity was occurring in the LHF during sample collection. Technicians made about (b)(4) sterile drug lots inside LFH on 11/26/25. Examples of batches produced and dispensed include Semaglutide 2.5 mg/mL, 2 mL solution for injection, batches LG364186933 and LG364186939. Post-Batch (b)(4) and Surface Sampling B. About post-batch (b)(4) and …
Contains “corrective action*”Read the Form 483 →
IntegraDose Compounding Services LLCInspection ended Aug 12, 2026 · FEI 3014483112 · Observation 1
… No investigation or CAPA has been opened to assess the impact of this specific HEPA Filter on your operations. B. During the inspection walk through of your classified spaces on August 7, 2026, it was observed that HEPA filters in your ISO 7 production areas are equipped with indicator lights. Of the ten (10) HEPA filters observed, nine (9) displayed red indicator lights and one (1) displayed a green indicator light across the following areas: …
Contains “CAPA*”Read the Form 483 →
Brookfield Medical Surgical Supplies IncInspection ended Jul 10, 2026 · FEI 3011430551 · Observation 1
… DV26-062), and Triamcinolone Acetonide exhibited four (4) events (DV24-051, DV25-056, DV26-065, DV26-066, DV26-067), suggesting that root cause analyses and corrective and preventive actions (CAPAs) were either inadequate or ineffectively implemented at the contract testing laboratory. For DV25-058 (Betamethasone lot 091225-1), the root cause was undetermined. However, the contract lab did not request a second sample to verify the results when no …
Contains “corrective and preventive action*”Read the Form 483 →
Brookfield Medical Surgical Supplies IncInspection ended Jul 10, 2026 · FEI 3011430551 · Observation 2
… Your firm released the batch without any assignable cause identified or initiating any CAPA. (2) Failure to question repeated OOS results from the same contract laboratory Betamethasone Sodium Phosphate had 5 separate OOS events (DV25-055, DV25-058, DV26-060, DV26-061, DV26-062) - all tested by the same contract lab, and all apparently invalidated except as noted. QCU did not question the lab's analytical capability after recurring failures with …
Contains “CAPA*”Read the Form 483 →
Brookfield Medical Surgical Supplies IncInspection ended Jul 10, 2026 · FEI 3011430551 · Observation 3
… No root cause investigation, corrective action, or re-sampling results were available for these excursions. Bacillus species (Excursion No. 39, 2026-05-21).
Contains “corrective action*”Read the Form 483 →
Auroactive Pharma Private LimitedInspection ended Jun 26, 2026 · FEI 3030708811 · Observation 1
… Specifically: The firm's corrective actions states as 'not applicable', root cause analyses were not supported with adequate scientific evidence, and incorrect sample weight was used in calculation at one instance. Several of the cited LIRs - specifically LIR-25-00038, LIR-25-00042, and LIR-25-00045 - lack documented rationale for why new sample preparation was selected over reinjection of the existing prepared solution as the corrective path. …
Contains “corrective action*”Read the Form 483 →
Auroactive Pharma Private LimitedInspection ended Jun 26, 2026 · FEI 3030708811 · Observation 2
… 2026 % ~75 ~25% ~40 ~48% ~85 ~29% ~12 ~14% ~38 ~13% ~30 ~36% ~40 ~13% ~10 ~12% ~22 ~7% ~8 ~10% ~18 ~6% ~3 ~4% ~5 ~2% ~0 ~0% ~15 ~5% ~0 ~0% Variants Included CAPA Category Count Performed, Re Performed, Re Perfomed, Re Performed, Re Analysis, Re analysis, Re analysis performed, Re Analysis performed, Reanalysis ~180 Re Processed, Re processed, Reprocessed, Re proceessed, Re processed, Reprocessed ~85 Re Processed Re Processed, Re processed, …
Contains “CAPA*”Read the Form 483 →
Eugia Steriles Private LimitedInspection ended Jun 19, 2026 · FEI 3027357163 · Observation 3
… The firm's CAPA report further states that some of the interventions, which are performed from filling to (b)(4) loading/unloading are not clearly visible from CCTV camera and there is no feasibility of view panel to verify aseptic intervention execution (behaviour and technique). However, activity is now monitored by real time reviewer from aseptic area. Based on the above evaluation, below are the actions to be taken for clear visibility of …
Contains “CAPA*”Read the Form 483 →
Eugia Steriles Private LimitedInspection ended Jun 19, 2026 · FEI 3027357163 · Observation 4
… In addition, CAPA-EP5-PR-24-0029-OA-0003 was opened on May 27, 2024, to extend the investigation to identify the exact root cause for gross content failure. However, a trial run for Line (b)(4) was conducted to [illegible] pumps with (b)(4) pumps instead. Issues concerning the filling line machine pumps was not documented as part of the initial investigation root cause analysis. In addition, there was no hypothesis testing or scientific …
Contains “CAPA*”Read the Form 483 →
Cytovance Biologics Inc.Inspection ended Jun 17, 2026 · FEI 3006341547 · Observation 4
Failure to adequately investigate deviations and implement appropriate corrective actions. Specifically, your firm has not conducted a thorough investigation to find and remediate the source/root cause of the microbiological contamination in the (b) (4) system in Building (b) (4). DEV-2026-0012 initiated April 23, 2026, noted a trend in Total Aerobic Microbial Count (TAMC) testing of several (b) (4) drops located in Building (b) (4). Organisms …
Contains “corrective action*”Read the Form 483 →
Jubilant HollisterStier LLCInspection ended Jun 17, 2026 · FEI 3010477 · Observation 3
… PR248495, and PR252381 2025: PR268501, PR288414, and PR296936 The Quality Unit was unable to identify the source of these recurring failures and implemented CAPAs proved ineffective in preventing recurrence of failures. The source of endotoxin contamination has not been identified. The Quality Unit attributed the failures to biofilm growth and (b)(4) degradation. However, no root cause for the contamination source or the reason for (b)(4) failure …
Contains “CAPA*”Read the Form 483 →
Fayetteville VA Medical CenterInspection ended Jun 15, 2026 · FEI 1073803 · Observation 1
… Your firm failed to investigate the cause of the OOC result and implement a corrective action plan as required by your Policy Number 797-2022, entitled, Microbiological Air and Surface Monitoring Program, version 2. Additionally, your firm was unable to demonstrate that the (b)(4) was not used for the production of sterile drug products during this period.
Contains “corrective action*”Read the Form 483 →
Lyophilization Services of New England, Inc. dba PCI Pharma ServicesInspection ended Jun 10, 2026 · FEI 3008546445 · Observation 1
… As part of CAPA-2024-1332, the firm stopped [applying sporicidal] onto the (b)(4) during installation and applying the sporicidal agent (b)(4) cleaning. The firm continues to use (b)(4), a non-sporicidal agent, to sanitize the (b)(4) during these steps. However, the firm does not apply any sporicidal agent to the (b)(4). Since 2/5/25, approximately [illegible] lots of the (b)(4) drug product have been manufactured on this (b)(4) line, of which …
Contains “CAPA*”Read the Form 483 →
Lyophilization Services of New England, Inc. dba PCI Pharma ServicesInspection ended Jun 10, 2026 · FEI 3008546445 · Observation 2
… CAPA-2025-0563 was initiated on 05/05/2025 to address the increased PM recoveries at the Bedford, NH facility, however, the CAPA remains open with no documented investigation into the probable aseptic breach in gowning, nor implementation of corrective actions to address the recoveries.
Contains “CAPA*”Read the Form 483 →
PQ Pharmacy LLCInspection ended Jun 1, 2026 · FEI 3017374013 · Observation 2
… HVAC failures have recurred without effective corrective action and oversight by the quality unit. For example, A) Your firm has not performed installation and operational qualification (IQ/OQ) for the HVAC systems and HEPA filtered air supply in the hazardous and non-hazardous ISO 7 and 8 cleanrooms, as confirmed by you, the CEO. Additionally, your firm has not established a written procedure governing routine qualification of your facility used …
Contains “corrective action*”Read the Form 483 →
PQ Pharmacy LLCInspection ended Jun 1, 2026 · FEI 3017374013 · Observation 3
… Per your QA Manager, the investigative efforts have been completed, however, this deviation is currently open pending closure of associated corrective/preventive actions. Its original due date was 03/20/2026 and an extension was issued without a new target due date or approval by Quality Assurance. Deviation remains open, but batch was destroyed. D) Three (3) deviations in 2025 and two (2) deviations in 2026 were initiated to document HVAC …
Contains “preventive action*”Read the Form 483 →
Janssen Vaccines, Zweigniederlassung der Cilag GmbH InternationalInspection ended May 29, 2026 · FEI 3012991344 · Observation 1
Discrepancy and failure investigations related to (b)(4) were not complete and expanded as necessary to assess product impact. For example: Between January 2025 and October 2025, you confirmed OOS results for (b)(4) in [illegible] batches of (b)(4) [illegible] Your firm's investigation (CAPA-006675) did not include an impact assessment of all batches manufactured during the subject timeframe. In addition to the rejected [illegible] batches, …
Contains “CAPA*”Read the Form 483 →
PETNET Solutions Inc.Inspection ended May 22, 2026 · FEI 3036193770 · Observation 3
… Inoculations, the radiochemical yield is only recorded, however, there is no statement on action limits on radiochemical yield beyond which investigation and corrective action are required. B. The document (D0009777) "Investigations for Low Radiochemical Yields and Failed Batches in the Production (b) (4) Labeled Products," states that the action limit for radiochemical yield for (b) (4) (b) (4) labelled product as (b) (4). The firm cannot …
Contains “corrective action*”Read the Form 483 →
Jiangsu Hansoh Pharmaceutical Group Co Ltd.Inspection ended Apr 24, 2026 · FEI 3008316144 · Observation 2
… Additionally, the CAPA did not adequately address the root cause.
Contains “CAPA*”Read the Form 483 →
Cliantha Research LimitedInspection ended Apr 17, 2026 · FEI 3017682080 · Observation 5
5. Your Quality unit failed to perform thorough and/or adequate laboratory investigations (identified by firm as event or incident) as required by the control procedure 100AB014, Version 4, Laboratory Event Management, Effective date: 23Jun2025; to ensure the appropriate root cause, propose adequate corrective / preventive actions, scientific justification to invalidate original result and management of events for any bioanalytical phase of …
Contains “preventive action*”Read the Form 483 →
21 CFR sections these observations open with
The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.
- 21 CFR 211.192Production record review24
- 21 CFR 211.22Responsibilities of quality control unit11
- 21 CFR 820.100Corrective and preventive actionIn this theme's rule6
- 21 CFR 211.42Design and construction features5
- 21 CFR 211.113Control of microbiological contamination5
- 21 CFR 211.63Equipment design, size, and location2
- 21 CFR 211.67Equipment cleaning and maintenance2
- 21 CFR 211.198Complaint files2
- 21 CFR 211.100Written procedures; deviations1
- 21 CFR 211.110Sampling and testing of in-process materials and drug products1
Ask FDAgent about CAPA in Form 483s
FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.