Auroactive Pharma Private Limited
India
FEI 3030708811DUNS 644177158
FDA has one inspection on file for Auroactive Pharma Private Limited in India (FEI 3030708811). It ended on ; FDA has not posted its classification yet.
FDA posted the Form 483 from that inspection. FDAgent read its 2 observations from the PDF; they are listed below.
No warning letter, court action or import refusal is filed under this FEI.
As of Nov 21, 2026, the site is registered with FDA as a drug establishment for API manufacturing and analysis, through Dec 31, 2027. Its registrant is Auroactive Pharma Private Limited (D-U-N-S 861995263).
Sources:FDA firm profile ↗Form 483 PDF ↗FDA drug establishment registrations ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- Inspections on file
- 1, in 2026
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- ↗
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
- Drug establishment registration
- Through
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- Not classified yetClassification not posted yetPosted 483 · 2 observations
Read this Form 483 on its own page →Form 483 PDF on fda.gov
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.
- Observation 1
Inadequate HPLC variability incident investigations
Laboratory Incident Investigations for High-Performance Liquid Chromatography (HPLC) Injection-to-Injection Area Variability Failures Were Inadequate. Specifically, The firm's Quality Control (QC) laboratory failed to conduct thorough investigations in accordance with SOP QC042 - "Handling of Laboratory Incidents" for the following Laboratory Incident Reports (LIRs): LIR-25-00008, LIR-25-00013, LIR-25-00014, LIR-25-00015, LIR-25-00038, LIR-25-00042, and LIR-25-00045 involving injection-to-injection area variability attributed to possible vial septa damage. Specifically: The firm's corrective actions states as 'not applicable', root cause analyses were not supported with adequate scientific evidence, and incorrect sample weight was used in calculation at one instance. Several of the cited LIRs - specifically LIR-25-00038, LIR-25-00042, and LIR-25-00045 - lack documented rationale for why new sample preparation was selected over reinjection of the existing prepared solution as the corrective path. For all cited LIRs, initial results were invalidated without scientific justification, and the reinjected or reanalyzed results were accepted and used for batch release. The specific findings for each LIR are detailed in the table below: LIR-25-00008 Batch (b)(4): The average results from both injections met the product specification (b)(4) to (b)(4)%w/w; however, failed to meet variability ratio specification (b)(4) to (b)(4); The original samples were reinjected stating that the failure was due to 'vial septa damaged' error without enough scientific evidence; The batch was released based on the one reinjected passing result. The API batch was utilized in manufacturing (b)(4) batch that was dispatched to the domestic market. LIR-25-00013 Batch (b)(4): The average results from both injections met the product specification (b)(4) to (b)(4)%w/w; however, failed to meet variability ratio specification (b)(4) to (b)(4); The original samples were reinjected stating that the failure was due to 'vial septa damaged' error without enough scientific evidence; The batch was released based on the one reinjected passing result. The Indian Pharmacopoeia (IP) grade API batch was dispatched to the domestic market. LIR-25-00014 Batch (b)(4): The average results from both injections met the product specification (b)(4) to (b)(4)%w/w; however, failed to meet variability ratio specification (b)(4) to (b)(4); The original samples were reinjected stating that the failure was due to 'vial septa damaged' error without enough scientific evidence; The batch was released based on the one reinjected passing result. The API batch was dispatched to the domestic market. LIR-25-00015 Batch (b)(4): The average results from both injections met the product specification (b)(4) to (b)(4)%w/w; however, failed to meet variability ratio specification (b)(4) to (b)(4); The original samples were reinjected stating that the failure was due to 'vial septa damaged' error without enough scientific evidence. Also sample weight discrepancy was observed - the balance printout recorded (b)(4) mg, whereas (b)(4) mg was used in the result calculation; results were calculated and printed on an external Excel spreadsheet and not in the official analytical data worksheet; The batch was released based on the one reinjected passing result. The API was utilized in manufacturing (b)(4) batch that was shipped to (b)(4) of (b)(4) the US market. LIR-25-00038 Batch (b)(4): The average results from both injections met the product specification (b)(4) to (b)(4)%w/w; however, failed to meet variability ratio specification (b)(4) to (b)(4); original samples were not reinjected - instead, new samples were prepared and injected with no documented scientific rationale to support new sample preparation; a 4-day gap existed between initial sample preparation (October 15, 2025) and retesting (October 19, 2025), which was not documented or justified in the investigation report; The batch was released based
Shortened here; the PDF has the full text.
- Observation 2
Failure to investigate lab discrepancies thoroughly
The firm failed to thoroughly investigate unexplained discrepancies. Specifically, SOP QC044 - Handling of Laboratory Occurrences, Version 01, effective January 1, 2025, does not explicitly define what constitutes "reanalysis." The SOP fails to distinguish between: Reinjection of an existing prepared solution, Fresh sample preparation from the original material, or Repeat of the entire analytical sequence. Sections 5.1 and 5.2 of SOP QC044 define laboratory occurrences only in broad terms, and Section 5.8.13 references "re-analysis" without defining its scope. A review of Laboratory Occurrence Reports (LORs) conducted by the firm from February 2024 through June 20, 2026, reveals the following root cause trends, which further illustrate the systemic nature of these deficiencies: 2024 Count (~180) Root Cause Category 2024 % Analyst oversight / lack of attention ~47 ~26% Improper column washing / system preparation ~22 ~12% Instrument failure / technical issues ~22 ~12% Air bubbles / (b)(4) issues ~14 ~8% spectrum issues ~9 ~5% Improper integration / processing method ~17 ~9% Unidentified / "might be" root causes ~8 ~4% Other ~41 ~23% 2025 (~298) Count 2025 % 2026 (~83, partial) Count 2026 % ~75 ~25% ~40 ~48% ~85 ~29% ~12 ~14% ~38 ~13% ~30 ~36% ~40 ~13% ~10 ~12% ~22 ~7% ~8 ~10% ~18 ~6% ~3 ~4% ~5 ~2% ~0 ~0% ~15 ~5% ~0 ~0% Variants Included CAPA Category Count Performed, Re Performed, Re Perfomed, Re Performed, Re Analysis, Re analysis, Re analysis performed, Re Analysis performed, Reanalysis ~180 Re Processed, Re processed, Reprocessed, Re proceessed, Re processed, Reprocessed ~85 Re Processed Re Processed, Re processed, Reprocessed, Re proceessed, Re ~46 Training Training provided, Trainng provided, Training shall be provided to analyst, Training shall be provided to the analyst ~45 Other Sensor leakage, Change another system, Re calculated, Re Print, re printing, etc. ~21 Grand Total ~549 ~148 entries (approximately 27%) have CAPA recorded as "NA", meaning no corrective action was taken or documented whereas Section 5.8.20 of the SOP requires corrective and/or preventive actions to be proposed and implemented based on investigation findings.
FDA drug establishment registration
FDA registration search- Registered operations
- API manufacturing, analysis
- Registration runs through
- Dec 31, 2027
- Registrant
- Auroactive Pharma Private LimitedD-U-N-S 861995263
From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3030708811 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3030708811. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.