Auroactive Pharma Private Limited Form 483, June 2026
FDA issued this Form 483 to Auroactive Pharma Private Limited in India (FEI 3030708811) at the end of an inspection that ended June 26, 2026. It lists 2 observations.
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: FDA's Electronic Reading Room, posted Nov 17, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.
- Inspection ended
- Jun 26, 2026
- —
- FDA's outcome
- Not classified
- Not classified yet
- Posted by FDA
- Nov 17, 2026
- 144 days after the inspection
The 2 observations on this Form 483
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.
- 1
Inadequate HPLC variability incident investigations
Laboratory Incident Investigations for High-Performance Liquid Chromatography (HPLC) Injection-to-Injection Area Variability Failures Were Inadequate. Specifically, The firm's Quality Control (QC) laboratory failed to conduct thorough investigations in accordance with SOP QC042 - "Handling of Laboratory Incidents" for the following Laboratory Incident Reports (LIRs): LIR-25-00008, LIR-25-00013, LIR-25-00014, LIR-25-00015, LIR-25-00038, LIR-25-00042, and LIR-25-00045 involving injection-to-injection area variability attributed to possible vial septa damage. Specifically: The firm's corrective actions states as 'not applicable', root cause analyses were not supported with adequate scientific evidence, and incorrect sample weight was used in calculation at one instance. Several of the cited LIRs - specifically LIR-25-00038, LIR-25-00042, and LIR-25-00045 - lack documented rationale for why new sample preparation was selected over reinjection of the existing prepared solution as the corrective path. For all cited LIRs, initial results were invalidated without scientific justification, and the reinjected or reanalyzed results were accepted and used for batch release. The specific findings for each LIR are detailed in the table below: LIR-25-00008 Batch (b)(4): The average results from both injections met the product specification (b)(4) to (b)(4)%w/w; however, failed to meet variability ratio specification (b)(4) to (b)(4); The original samples were reinjected stating that the failure was due to 'vial septa damaged' error without enough scientific evidence; The batch was released based on the one reinjected passing result. The API batch was utilized in manufacturing (b)(4) batch that was dispatched to the domestic market. LIR-25-00013 Batch (b)(4): The average results from both injections met the product specification (b)(4) to (b)(4)%w/w; however, failed to meet variability ratio specification (b)(4) to (b)(4); The original samples were reinjected stating that the failure was due to 'vial septa damaged' error without enough scientific evidence; The batch was released based on the one reinjected passing result. The Indian Pharmacopoeia (IP) grade API batch was dispatched to the domestic market. LIR-25-00014 Batch (b)(4): The average results from both injections met the product specification (b)(4) to (b)(4)%w/w; however, failed to meet variability ratio specification (b)(4) to (b)(4); The original samples were reinjected stating that the failure was due to 'vial septa damaged' error without enough scientific evidence; The batch was released based on the one reinjected passing result. The API batch was dispatched to the domestic market. LIR-25-00015 Batch (b)(4): The average results from both injections met the product specification (b)(4) to (b)(4)%w/w; however, failed to meet variability ratio specification (b)(4) to (b)(4); The original samples were reinjected stating that the failure was due to 'vial septa damaged' error without enough scientific evidence. Also sample weight discrepancy was observed - the balance printout recorded (b)(4) mg, whereas (b)(4) mg was used in the result calculation; results were calculated and printed on an external Excel spreadsheet and not in the official analytical data worksheet; The batch was released based on the one reinjected passing result. The API was utilized in manufacturing (b)(4) batch that was shipped to (b)(4) of (b)(4) the US market. LIR-25-00038 Batch (b)(4): The average results from both injections met the product specification (b)(4) to (b)(4)%w/w; however, failed to meet variability ratio specification (b)(4) to (b)(4); original samples were not reinjected - instead, new samples were prepared and injected with no documented scientific rationale to support new sample preparation; a 4-day gap existed between initial sample preparation (October 15, 2025) and retesting (October 19, 2025), which was not documented or justified in the investigation report; The batch was released based
Shortened here; the PDF has the full text.
- 2
Failure to investigate lab discrepancies thoroughly
The firm failed to thoroughly investigate unexplained discrepancies. Specifically, SOP QC044 - Handling of Laboratory Occurrences, Version 01, effective January 1, 2025, does not explicitly define what constitutes "reanalysis." The SOP fails to distinguish between: Reinjection of an existing prepared solution, Fresh sample preparation from the original material, or Repeat of the entire analytical sequence. Sections 5.1 and 5.2 of SOP QC044 define laboratory occurrences only in broad terms, and Section 5.8.13 references "re-analysis" without defining its scope. A review of Laboratory Occurrence Reports (LORs) conducted by the firm from February 2024 through June 20, 2026, reveals the following root cause trends, which further illustrate the systemic nature of these deficiencies: 2024 Count (~180) Root Cause Category 2024 % Analyst oversight / lack of attention ~47 ~26% Improper column washing / system preparation ~22 ~12% Instrument failure / technical issues ~22 ~12% Air bubbles / (b)(4) issues ~14 ~8% spectrum issues ~9 ~5% Improper integration / processing method ~17 ~9% Unidentified / "might be" root causes ~8 ~4% Other ~41 ~23% 2025 (~298) Count 2025 % 2026 (~83, partial) Count 2026 % ~75 ~25% ~40 ~48% ~85 ~29% ~12 ~14% ~38 ~13% ~30 ~36% ~40 ~13% ~10 ~12% ~22 ~7% ~8 ~10% ~18 ~6% ~3 ~4% ~5 ~2% ~0 ~0% ~15 ~5% ~0 ~0% Variants Included CAPA Category Count Performed, Re Performed, Re Perfomed, Re Performed, Re Analysis, Re analysis, Re analysis performed, Re Analysis performed, Reanalysis ~180 Re Processed, Re processed, Reprocessed, Re proceessed, Re processed, Reprocessed ~85 Re Processed Re Processed, Re processed, Reprocessed, Re proceessed, Re ~46 Training Training provided, Trainng provided, Training shall be provided to analyst, Training shall be provided to the analyst ~45 Other Sensor leakage, Change another system, Re calculated, Re Print, re printing, etc. ~21 Grand Total ~549 ~148 entries (approximately 27%) have CAPA recorded as "NA", meaning no corrective action was taken or documented whereas Section 5.8.20 of the SOP requires corrective and/or preventive actions to be proposed and implemented based on investigation findings.
Ask FDAgent about this Form 483
FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.
Related
About this data
- The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
- The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
- FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
- FDAgent is independent of FDA.