PETNET Solutions Inc. Form 483, May 2026
FDA issued this Form 483 to PETNET Solutions Inc. in California (FEI 3036193770) at the end of an inspection that ended May 22, 2026. It lists 5 observations.
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: FDA's Electronic Reading Room, posted Nov 17, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.
- Inspection ended
- May 22, 2026
- —
- FDA's outcome
- Not classified
- Not classified yet
- Posted by FDA
- Nov 17, 2026
- 179 days after the inspection
The 5 observations on this Form 483
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.
- 1
Unqualified supplier for critical precursor component
You did not use a reliable supplier as a source of each lot of component. Specifically, Your firm failed to follow procedure SOP # D0011306 titled "Supplier Qualification, Approval, and Maintenance," Revision H (Effective Date: 11/1/2025), which establishes requirements for vendor selection, approval, and oversight. For example: A. Your firm failed to show (b) (4) as an approved supplier of the precursor (b) (4) mg/vial), a critical component used in the manufacture of (b) (4) Injection. There are no supplier qualification activities performed prior to introducing the precursor (b) (4) into the manufacture of (b) (4) process validation (PV) batches of (b) (4) Injection (Batch #s (b) (4) In addition, the HPLC identity test performed to receive the precursor raw material at central facility is not specific to confirm the identity of precursor that yields as an active ingredient. B. Your firm accepted transfer tubing — a material used during product transfer in the manufacture of (b) (4) Injection — without obtaining a specification sheet to verify material specifications. In addition, there is no documentation to show that this supplier is qualified.
- 2
Inadequate visual inspection procedure and qualification
An appropriate laboratory determination to ensure that each batch of a PET drug product conforms to specifications was not conducted before final release. Specifically, A. Your procedure D0001978, Product Test: Appearance, effective 10/23/24 is not adequate. The procedure does not include the use of contrast background for visual inspection of final product to determine visible particle matter, color and clarity. B. The qualification (b) (4) used for qualification of visual inspection of the operators is not adequate. The (b) (4) of (b) (4) vials in which (b) (4) vials only contain (b) (4) of (b) (4) ((b) (4) particle). The (b) (4) is not representative for identifying all particulate matters and defects.
- 3
Missing radiochemical yield action limits
Your master production and control records did not include a statement of action limits on radiochemical yield. Specifically, A. The manufacturing batch records for (b) (4) Injection do not include action limits on radiochemical yield. For example, Manufacturing batch record for PPQ batch # (b) (4) on page 3 of 26 Section, Final Product Vial Assay and Sterility Inoculations, the radiochemical yield is only recorded, however, there is no statement on action limits on radiochemical yield beyond which investigation and corrective action are required. B. The document (D0009777) "Investigations for Low Radiochemical Yields and Failed Batches in the Production (b) (4) Labeled Products," states that the action limit for radiochemical yield for (b) (4) (b) (4) labelled product as (b) (4). The firm cannot provide any scientific rationale for the (b) (4) action limit, considering the firm manufactured around (b) (4) batches of (b) (4) injection and the obtained yield for these batches was around 65%.
- 4
Inadequate aseptic technique and personnel training
You lack personnel with the necessary background and training to perform their assigned functions. Specifically, On 05/19/2026, during a production mock preparation of final product vial subassembly conducted under the ISO 5 (b) (4) laminar airflow hood (LFH SN: (b) (4), the operator (b) (6), (b) (7) did not follow techniques described in SOP D0011940, Preparation of Final Product Vial Subassembly – (b) (4) (b) (4) Revision: Q, Effective Date: 03/13/2025. The operator failed to follow aseptic techniques as specified in the instructions of SOP D0011940. For example: A) The operator did not apply (b) (4) to their gloved hands prior to re-entering the ISO 5 critical zone. B) The operator placed the environmental monitoring (EM) settle plate cover face up of the in the critical zone during active aseptic operations. C) Materials and components were positioned in front of one another, obstructing first-air flow. D) During the cleaning, the operator did not employ (b) (4) wipe strokes. In addition, the operator used wipes from an already opened package. E) The operator did not (b) (4) and (b) (4) and (b) (4) on to the (b) (4) when performing personnel EM. We observed the operator conducted the personal EM with approximately (b) (4) motion touch for (b) (4) finger.
- 5
Failure to monitor laminar flow hood pressure
You did not implement procedures to ensure that all equipment is cleaned and suitable for its intended purposes that could reasonably be expected to adversely affect the identity, strength, quality, or purity of a PET drug, or give erroneous or invalid test results when improperly used or maintained. Specifically, You failed to monitor the pressure differential of ISO 5 Laminar Flow Hood (#(b) (4)), to ensure the pressure differential across the filter face to indicate the HEPA filters function appropriately during the final product vials assembly.
Ask FDAgent about this Form 483
FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.
Related
About this data
- The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
- The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
- FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
- FDAgent is independent of FDA.