Batch records and documentation in FDA Form 483 observations
Master and batch production records, device history records, logbooks and good documentation practice. FDAgent finds 241 such observations in 143 Form 483s issued to 137 facilities, for inspections that ended between October 2008 and September 2026. Each is quoted below from the 483 FDA posted.
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.
- Observations
- 241
- Each quoted below
- Form 483s
- 143
- Posted by FDA
- Facilities
- 137
- Distinct FEIs
- Most recent
- Sep 28, 2026
- Inspection end date
Observations by year
These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).
Peak: 80 in 2026
| Year | Observations |
|---|---|
| 2008 | 3 |
| 2009 | 0 |
| 2010 | 1 |
| 2011 | 0 |
| 2012 | 4 |
| 2013 | 4 |
| 2014 | 0 |
| 2015 | 3 |
| 2016 | 16 |
| 2017 | 12 |
| 2018 | 15 |
| 2019 | 13 |
| 2020 | 0 |
| 2021 | 7 |
| 2022 | 18 |
| 2023 | 22 |
| 2024 | 15 |
| 2025 | 28 |
| 2026 | 80 |
How FDAgent sorts observations into this theme
Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.
It is tied to one of these 21 CFR entries
21 CFR 211.18021 CFR 211.18421 CFR 211.18621 CFR 211.18821 CFR 820.18121 CFR 820.184
A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.
Or its text contains one of these phrases
“batch record*” · “batch production … record*” · “master production … record*” · “master batch record*” · “device history record*” · “device master record*” · “documentation practice*” · “logbook*” · “log book*” · “BPR*” · “MPR*” · “BMR*”
* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.
The 241 observations, newest first
Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.
Strive Pharmacy LLCInspection ended Sep 28, 2026 · FEI 3021042485 · Observation 1
… The batch Production Record Checklist inaccurately lists the (b)(4) result as <1 CFU, and personnel did not complete the required evaluation for (b)(4) growth. The entire lot was dispensed to fill more than prescriptions. Semaglutide/Glycine/B12 5/5/1 mg/mL, 2 mL, batch LG364281132, sample ID #GD071926LH, result = 3 cfu Bacillus species, coagulase-negative Staphylococcus species, and Micrococcus species. This lot was dispensed. …
Contains “batch production … record*”Read the Form 483 →
Victory Medical Center PharmacyInspection ended Sep 10, 2026 · FEI 3010485868 · Observation 6
… BICARBONATE (105MG/28MG/B-COMPLEX,12MG/ML) STOCK SOLUTION, Lot 2026504, BUD October 15, 2026, Qty (b) (4) - No documented (b) (4) test was found in the batch record for this lot. ELECTRO B COMPLEX (MDV) MAGNESIUM CHLORIDE/CALCIUM CHLORIDE/B COMPLEX/SODIUM BICARBONATE (105MG/28MG/B-COMPLEX,12MG/ML) STOCK SOLUTION, Lot 2026504, BUD October 15, 2026, was used to fill (b) (4) - 49ml vials of ELECTRO B COMPLEX (MDV) MAGNESIUM CHLORIDE/CALCIUM …
Contains “batch record*”Read the Form 483 →
RC Outsourcing LLCInspection ended Sep 3, 2026 · FEI 3011967886 · Observation 1
… Review of the past two media fill batch records revealed that the records contained no directions specifying the required number of air samples an operator must collect or the prescribed time intervals at which sampling must occur. Consequently, all media fill batch records reviewed contained (b)(4) and none of the records included lot numbers for the associated media plates used during the runs. This is in contrast to commercial production, …
Contains “batch record*”Read the Form 483 →
OPS International Inc dba Olympia PharmaceuticalsInspection ended Sep 3, 2026 · FEI 3009724085 · Observation 1
… (2) According to the batch record for Myer's Cocktail (Lot #G27A05-26, Exp. 26 Nov 2026), equipment set-up of the (b)(4) auto filler, located inside the ISO (b)(4) in Suite (b)(4) began at 08:01. The first passive air sample was not opened until 08:21 on 30 July 2026. (B) The firm's personnel monitoring program does not include microbiological monitoring of gown locations for Sterile Techs who perform set-up of the (b)(4) filling line or set-up …
Contains “batch record*”Read the Form 483 →
OPS International Inc dba Olympia PharmaceuticalsInspection ended Sep 3, 2026 · FEI 3009724085 · Observation 3
Written production and process control procedures are not followed in the execution of production and process control functions and documented at the time of performance. Specifically, (B) Batch record review of Myer's Cocktail (Lot #G27A05-26, Exp. 26 Nov 2026) revealed that equipment set-up of the (b)(4) filling line, located inside the ISO 5 (b)(4) in Suite (b)(4) began on 30 July 2026 at 08:01 and was completed by Sterile Tech at 08:06 on 30 …
Contains “batch record*”Read the Form 483 →
OPS International Inc dba Olympia PharmaceuticalsInspection ended Sep 3, 2026 · FEI 3009724085 · Observation 8
Each component is not added to the batch by one person and verified by a second person. Specifically, a review of the batch record for Myer's Cocktail (Lot #(b)(4), Exp. (b)(4)) revealed that ingredients were added between (b)(4) on July 29, 2026, with no documented contemporaneous verification recorded after each individual addition. Instead, a single verification timestamp of (b)(4) was documented for this manufacturing step, more than two …
Contains “batch record*”Read the Form 483 →
TAILSTORM HEALTH INCInspection ended Aug 28, 2026 · FEI 3015929581 · Observation 2
… We reviewed the batch record for Media Fill, batch CHMF/26/001, completed (b) (4), approved 06/11/26. E. Technicians do not use BIs for routine or periodic monitoring of the (b) (4) cycle used for (b) (4) of ISO 5 Filling Line and the ISO 7 buffer room before filling. F. The (b) (4) surface remains folded inside the open (b) (4) pouch during the (b) (4) cycle for ISO 5 Filling Line and the ISO 7 buffer room. There are no BIs or (b) (4) placed …
Contains “batch record*”Read the Form 483 →
STAQ Pharma of Ohio LLCInspection ended Aug 19, 2026 · FEI 3025336457 · Observation 1
… qualify visual inspectors for products compounded and packaged in [illegible] The challenge syringes does not include known defect types listed in (b)(4) the batch production records, such as precipitation (cloudiness), particles adhered to the container, and particles adhered to the stopper. In addition, training records do not document whether visual inspection operators are trained on all known defect types prior to qualification. C. Visual …
Contains “batch production … record*”Read the Form 483 →
STAQ Pharma of Ohio LLCInspection ended Aug 19, 2026 · FEI 3025336457 · Observation 2
Batch production and control records do not include in-process results for each batch of drug product produced. Specifically, Actual light intensity values measured prior to starting visual inspection are not recorded in the production batch records; instead, the records document only whether the established light intensity acceptance criteria was met using a "Yes/No" check box. For example, for Succinylcholine Chloride 20 mg/mL, 10 mL in 10 mL …
Opens with FDA's citation wording for 21 CFR 211.188(b)(5)Read the Form 483 →
IntegraDose Compounding Services LLCInspection ended Aug 12, 2026 · FEI 3014483112 · Observation 2
… Your batch records do not document critical pieces of information related to the production and release of your drug products. Examples include but are not limited to the following: 1. Your firm uses a (b)(4) (Model (b)(4)) to (b)(4) incoming materials with a (b)(4) prior to introduction into the ISO 7 classified space. Per your firm, materials are (b)(4) and brought into the ISO 7 space. This (b)(4) process occurs in either the "Wipe" or "Wipe" …
Contains “batch record*”Read the Form 483 →
OurPharma LLCInspection ended Aug 12, 2026 · FEI 3014064135 · Observation 1
Separate or defined areas to prevent contamination or mix-ups are deficient regarding operations related to aseptic processing of drug products. Specifically, • (b)(4) Sterilization Performed in an ISO 8 Classified Room Batch records collected during the inspection confirm that (b)(4) sterilization was performed in Compounding Room (b)(4) which is classified as ISO 8. Specifically: Batch production records for two lots of Fentanyl …
Contains “batch record*”Read the Form 483 →
OurPharma LLCInspection ended Aug 12, 2026 · FEI 3014064135 · Observation 2
Records are not maintained so that data therein can be reviewed at least annually to evaluate the quality standards of each drug product to determine the need for changes in specifications or manufacturing or control procedures. Specifically, Annual product review (APR) records for six drug product families — Fentanyl Citrate, Fentanyl Citrate + Bupivacaine HCl, Fentanyl Citrate + Ropivacaine HCl, Hydromorphone HCl, Morphine HCl, and Ropivacaine …
Opens with FDA's citation wording for 21 CFR 211.180(e)Read the Form 483 →
Wanbury LimitedInspection ended Jul 17, 2026 · FEI 3002980818 · Observation 1
… B) For (b)(4) batch (b)(4) is needed to less than (b)(4) of (b)(4) However, at step number (b)(4) batches, reports (b)(4), your batch production record specified that no (b)(4) (b)(4) of a minimum of (b)(4) This is approximately (b)(4), your batch production record specified that no (b)(4) (b)(4) (b)(4), only (b)(4) was needed, for a total of (b)(4) of a minimum of (b)(4). This is approximately (b)(4) less than the validated process requires. …
Contains “batch production … record*”Read the Form 483 →
Wanbury LimitedInspection ended Jul 17, 2026 · FEI 3002980818 · Observation 2
… Despite the presence of these significant residues, the equipment usage and cleaning logbooks reflected that cleaning had been completed, and equipment status labels indicated the equipment was in a "Cleaned" state. Physical inspection of the affected equipment and interviews with the operators confirmed that the required batch-to-batch cleaning procedure, as specified in SOP No. W3/PR(b)(4)/041, had not been performed. (b)(4) had no designated …
Contains “logbook*”Read the Form 483 →
Wanbury LimitedInspection ended Jul 17, 2026 · FEI 3002980818 · Observation 3
A mandatory process step specified in the approved Batch Production Record (BPR) was not performed, and manufacturing activities were not contemporaneously documented. Specifically, (b)(4) On July 13, 2026, during the manufacture of, Batch No. (b)(4), the mandatory process requirement of (b)(4) (b)(4) (Operation No. (b)(4)) per BPR No. (b)(4) was not performed. Following to the required (b)(4), the process was advanced directly to completion of …
Contains “batch production … record*”Read the Form 483 →
Cascade Specialty Pharmacy LLCInspection ended Jul 17, 2026 · FEI 3015133983 · Observation 2
Batch production and control records do not include complete information relating to the production and control of each batch. Specifically, A. Amoxicillin/Clavulanate Pot 250 mg/mL Suspension Lot 657973 (RX# F wxc1) Your firm's compounding record does not specify the method to be used for crushing or dissolving tablets during production. The instructions for tablet pulverization do not indicate the required duration of pulverization needed to …
Opens with FDA's citation wording for 21 CFR 211.188Read the Form 483 →
Glenmark Pharmaceuticals LimitedInspection ended Jun 30, 2026 · FEI 3004672766 · Observation 2
Appropriate controls were not exercised over computer or related systems to assure that changes in master production and control records or other records are instituted only by authorized personnel. Specifically, A. (b)(4) machines - the (b)(4) are all documented as having no electronic data storage or audit trail capability. These machines are used to manufacture multiple tablet products for distribution in the United States, including (b)(4). …
Contains “master production … record*”Read the Form 483 →
Glenmark Pharmaceuticals LimitedInspection ended Jun 30, 2026 · FEI 3004672766 · Observation 4
Control systems for the firm's operations are not adequate to prevent contamination or mix-ups. Specifically, A. Differential pressure in the manufacturing areas of the (b)(4) facility (Unit (b)(4)) is not adequately monitored and recorded, and no alarm system is in place to detect differential pressure excursions. A review of the Batch Production Record (BPR) for Batch (b)(4) ((b)(4) USP (b)(4) mcg), confirmed that the differential pressure is …
Contains “batch production … record*”Read the Form 483 →
Dr. Reddy's Laboratories Ltd.Inspection ended Jun 25, 2026 · FEI 3014250111 · Observation 7
… shows that individual skill training sessions for critical aseptic manufacturing operations were documented as lasting as few as 29 minutes (Participate in BPR execution and closure activity, documented November 10, 2025, (b)(4)) and 9 minutes (Procedure for operation and cleaning of (b)(4) LAF, documented November 10, 2025, (b)(4)). These session durations do not provide reasonable assurance that the employee received sufficient training to …
Contains “BPR*”Read the Form 483 →
Eugia Steriles Private LimitedInspection ended Jun 19, 2026 · FEI 3027357163 · Observation 2
… There is no detailed justification by the quality unit as to why planned media fill interventions dictated by the Aseptic Process Simulation (APS) master batch record are not uniformly performed for all (b)(4) campaign batches. For example: (b)(4) (b)(4). In addition, one (1) out of the [illegible] mL media fill vials did not include media fill batches for the following intervention: Intervention No. NI-4-H: Transfer of forceps, forceps holder …
Contains “master batch record*”Read the Form 483 →
21 CFR sections these observations open with
The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.
- 21 CFR 211.188Batch production and control recordsIn this theme's rule17
- 21 CFR 211.100Written procedures; deviations14
- 21 CFR 211.22Responsibilities of quality control unit13
- 21 CFR 211.68Automatic, mechanical, and electronic equipment13
- 21 CFR 211.113Control of microbiological contamination11
- 21 CFR 211.180General requirementsIn this theme's rule8
- 21 CFR 211.186Master production and control recordsIn this theme's rule5
- 21 CFR 211.42Design and construction features4
- 21 CFR 211.160General requirements4
- 21 CFR 211.192Production record review4
Ask FDAgent about batch records and documentation in Form 483s
FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.
Related
About this data
- The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
- Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
- A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.
- FDAgent is independent of FDA.