Equipment cleaning and cleaning validation in FDA Form 483 observations
Cleaning procedures for manufacturing equipment, the studies that show they work, and the residues, swabs and logs that prove each cleaning. FDAgent finds 197 such observations in 144 Form 483s issued to 138 facilities, for inspections that ended between October 2008 and September 2026. Each is quoted below from the 483 FDA posted.
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.
- Observations
- 197
- Each quoted below
- Form 483s
- 144
- Posted by FDA
- Facilities
- 138
- Distinct FEIs
- Most recent
- Sep 28, 2026
- Inspection end date
Observations by year
These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).
Peak: 63 in 2026
| Year | Observations |
|---|---|
| 2008 | 1 |
| 2009 | 0 |
| 2010 | 0 |
| 2011 | 0 |
| 2012 | 0 |
| 2013 | 0 |
| 2014 | 1 |
| 2015 | 1 |
| 2016 | 10 |
| 2017 | 15 |
| 2018 | 20 |
| 2019 | 5 |
| 2020 | 5 |
| 2021 | 10 |
| 2022 | 7 |
| 2023 | 22 |
| 2024 | 13 |
| 2025 | 24 |
| 2026 | 63 |
How FDAgent sorts observations into this theme
Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.
It is tied to one of these 21 CFR entries
The same entries the equipment cleaning topic counts. A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.
Or its text contains one of these phrases
“cleaning validat*” · “validat* … cleaning” · “cleaning procedure*” · “cleaning process*” · “cleaning record*” · “cleaning log*” · “cleaning agent*” · “cleaning verification” · “cleaning stud*” · “cleaning method*” · “cleaning effectiveness” · “not cleaned” · “not adequately cleaned” · “not properly cleaned” · “residue*” · “swab*” · “carry over” · “carryover”
* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.
The 197 observations, newest first
Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.
Solaris Pharmacy, Inc.Inspection ended Feb 26, 2020 · FEI 3015826783 · Observation 1
Cleaning agents used to disinfect, clean, sanitize equipment and/or production areas are not suitable for use. Specifically, a) On February 19, 2020, I observed the firm's non-sterile technician [illegible] to clean and disinfect the [(Df) (il)]; the label states that it is "intended for use" on "non-food surfaces". The firm produces capsules for oral use. Approximately, November 1, 2019 for oral administration. b) On February 19, 2020, I …
Contains “cleaning agent*”Read the Form 483 →
Blount Discount Pharmacy, Inc.Inspection ended Feb 13, 2020 · FEI 3015826061 · Observation 1
Hazardous and highly potent drugs are produced without providing adequate cleaning of work surfaces, utensils, equipment and/or personnel to prevent cross-contamination. Specifically, I observed built-up residue on the scale, stained and nicked spatulas, scratched and cloudy glassware used to produce non-sterile drug products. You use these same utensils and equipment to manufacture hazardous drugs and non-hazardous types of drugs including but …
Contains “residue*”Read the Form 483 →
Blount Discount Pharmacy, Inc.Inspection ended Feb 13, 2020 · FEI 3015826061 · Observation 2
Agents used for cleaning the laminar flow hood and other equipment between products do not include a deactivating agent (e.g. oxidizing agent) to prevent cross-contamination. Specifically, the cleaning agents I observed used between batches for equipment and the hood were [(b)(4)] and dish soap only.
Contains “cleaning agent*”Read the Form 483 →
A Chem Rx, LLC dba Cortez DrugsInspection ended Feb 7, 2020 · FEI 3011103365 · Observation 1
Observation 1: Non-microbial contamination was observed. Specifically, on 2/3/20, within the firm's Non-Sterile Suite hood/vent directly over the working surface was observed visually unclean, shedding an unknown brown residue. According to your firm's Pharmacist in Charge/Owner, the vent has not been replaced in approximately two years (all non-sterile drug products are produced within this hood).
Contains “residue*”Read the Form 483 →
GPC Medical LimitedInspection ended Oct 24, 2019 · FEI 3006513362 · Observation 1
Process Control Procedures that describe any process controls necessary to ensure conformance to the specifications have not been adequately established. For example: The Procedure for [illegible] does not identify the temperature that the process is operated at. The Cleaning process validation was performed at a temperature of [illegible]. During a walk through of the manufacturing facility on Tuesday October 22, I observed the (b)(4) Cleaning …
Contains “cleaning process*”Read the Form 483 →
GPC Medical LimitedInspection ended Oct 24, 2019 · FEI 3006513362 · Observation 4
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated. Specifically, the process validation for cleaning did not demonstrate that the acceptance was satisfied. The testing after criteria for cleaning only reported a (b)(4) (b)(4) difference to [illegible] and the acceptance criteria established that there would be no "neat" water. Also cleaning validation did not define or describe …
Contains “validat* … cleaning”Read the Form 483 →
Rapid Equine Solutions, LLCInspection ended Aug 9, 2019 · FEI 3006708880 · Observation 3
Aseptic processing areas are deficient regarding the system for cleaning and disinfecting the room and equipment to product aseptic conditions. You have not established if the current cleaning method is capable of removing residuals of API, cleaning and sanitizing agents and objectionable organisms from multi-use equipment and utensils used to produce sterile products.
Contains “cleaning method*”Read the Form 483 →
People’s Custom Rx and Clinical Care, LLCInspection ended Jul 24, 2019 · FEI 3003780900 · Observation 4
… Your firm has no evidence to show that the use of (b)(4) will effectively remove residues which might be present on the hood surface after production.
Contains “residue*”Read the Form 483 →
Front Door Pharmacy, LLC dba Pure PharmacyInspection ended Mar 18, 2019 · FEI 3012937475 · Observation 4
… The [redacted] mixer was observed with foreign residue on its outer surface. Your firm's pharmacist placed the mixer into the aseptic compounding area without disinfecting it. C. Your firm's stool, in the ISO 7 cleanroom, is covered with a material which cannot be adequately disinfected.
Contains “residue*”Read the Form 483 →
Hallettsville Pharmacy, LLC.Inspection ended Nov 8, 2018 · FEI 3014538927 · Observation 1
… Your firm has no evidence to show that the use of (b) (4) will effectively remove residues which might be present on the hood surface after production. B. Your firm uses household dish liquid detergent and potable water for the cleaning of utensils such as rubber spatulas used in the processing of highly potent drug substances such as Estriol or Progesterone. All utensils are then washed in a dishwasher before subsequent use. I noted that several …
Contains “residue*”Read the Form 483 →
Lupin Limited Unit IIIInspection ended Oct 18, 2018 · FEI 3009107538 · Observation 2
Written procedures for cleaning and maintenance fail to include description of methods, equipment and materials used, description of disassembling and reassembling equipment as necessary to assure removal or obliteration of previous batch identification. Specifically, a) cleaning validation and verification [b)(4) / (b)(4)] did not include or did not perform full cleaning studies [illegible]; production room [illegible]; such as solutions, …
Contains “cleaning validat*”Read the Form 483 →
Inventive Infusion Solutions, LPInspection ended Aug 31, 2018 · FEI 3014498447 · Observation 5
Personnel did not disinfect to prevent contamination. Specifically, On 8/22/2018 and 8/28/2018, operator (b)(4) was observed with exposed facial skin while cleaning the surfaces inside the ISO 5 area of (b)(4) Hood. The cleaning processes within the ISO 5 area are the same for the (b)(4) and (b)(4) operations of (b)(4).
Contains “cleaning process*”Read the Form 483 →
Sun Pharmaceutical Industries LtdInspection ended Aug 31, 2018 · FEI 3002809586 · Observation 5
Written procedures are not for the cleaning and maintenance of equipment, including utensils, used in the manufacture, processing, packing or holding of a drug product. Specifically, A) You performed equipment cleaning verification activities for a specific number of batches after you identified as a high risk product. During the cleaning verification you did not perform analyses for the detergent residues nor did you perform bio-burden testing. …
Opens with FDA's citation wording for 21 CFR 211.67(b)Read the Form 483 →
Dolan Central Illinois Compounding LLC dba Preckshot ProfessionalInspection ended Aug 10, 2018 · FEI 3006572203 · Observation 5
Equipment was not disinfected prior to entering the aseptic processing areas. Specifically, Prior to sterile processing, your firm attached a pre-moistened sterile [illegible] mophead to sanitize the ISO 5 hood. I observed the foam mop head to be discolored with black colored residues on 07/30/18. I observed sterile production of Papaverine, Phentolamine, Alprostadil (15: 0.5: 25) in ML injectable, lot 07302018@18 and Papaverine, Phentolamine, …
Contains “residue*”Read the Form 483 →
The Pet Apothecary LLCInspection ended Aug 7, 2018 · FEI 3005946041 · Observation 1
You produced beta-lactam drugs without providing adequate containment, cleaning of work surfaces, and cleaning of utensils to prevent cross-contamination. Specifically, Beta-lactam drugs are produced in non-designated areas sharing the same work surfaces and equipment that are also used to produce non-beta-lactam drugs, of which are not cleaned/sanitized with bleach and isopropyl alcohol (70%) to provide adequate inactivation of the drug. For …
Contains “not cleaned”Read the Form 483 →
The Pet Apothecary LLCInspection ended Aug 7, 2018 · FEI 3005946041 · Observation 2
Cleaning methods are not established in a written procedure nor are appropriate cleaning agents specified. Specifically, a.) Drug product production equipment such as capsule filling trays and spatulas/mortars/pestles/spinners are not rinsed with purified water and/or sanitized with appropriate sanitizing agents after cleaning with hot water and dish soap. b.) There are no defined/labeled areas for storage of "clean" and "unclean" drug production …
Contains “cleaning method*”Read the Form 483 →
Auro PharmaciesInspection ended Aug 2, 2018 · FEI 3011893599 · Observation 1
OBSERVATION 1 The ISO 5 classified aseptic processing areas had visibly dirty equipment or surface. Specifically, we observed apparent brown and white residue in the ISO 5 classified hoods that are used in the production of sterile drugs. You stated the hoods are cleaned r(b) (4) production day. However, your firm was unaware of these observations until they were discussed on 07/09/2018 and subsequently on 07/18/2018. For example: A. On …
Contains “residue*”Read the Form 483 →
Unichem Laboratories Ltd.Inspection ended Jul 27, 2018 · FEI 3006456888 · Observation 1
1. Cleaning procedures are not validated. Specifically, The Unichem Laboratories Ltd. procedure PPQA029/04 is not validated to assure API manufacturing equipment are cleaned to a level that does not impact product quality when employed for the multiple batch campaign process. The cleaning validation protocol did not completely describe the equipment to be cleaned, procedures, materials, acceptable cleaning levels, parameters to be monitored and …
Contains “cleaning procedure*”Read the Form 483 →
FH Investments Inc. dba Asteria HealthInspection ended Jul 10, 2018 · FEI 3010348724 · Observation 3
Equipment and utensils are not cleaned at appropriate intervals to prevent contamination that would alter the safety, identity, strength, quality, or purity of the drug product. Specifically, the equipment cleaning process has not been validated to ensure there is no cross-contamination between the hormone active pharmaceutical ingredients (APIs), testosterone and estradiol. The (b) (4) Pellet Press, (b) (4), is used for both API products and …
Opens with FDA's citation wording for 21 CFR 211.67(a)Read the Form 483 →
Glenmark Pharmaceuticals Ltd.Inspection ended May 24, 2018 · FEI 3008565058 · Observation 1
The accuracy, sensitivity, specificity and reproducibility of test methods have not been established and documented. Specifically, Your firm's QC laboratory has not validated analytical methods used in the determination of active residue from equipment's used for processing of the following products: HP® Capsules USP? mg &P® mg, 2)? Tablets USP me.§ mg {§ mg &[H me. 3)@@ Tablets, 4)0@ Tablets, 5)0@ Capsules USP™® mg, and 6)P@ Tablets USPP® mg & …
Contains “residue*”Read the Form 483 →
21 CFR sections these observations open with
The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.
- 21 CFR 211.67Equipment cleaning and maintenanceIn this theme's rule31
- 21 CFR 211.113Control of microbiological contamination10
- 21 CFR 211.42Design and construction features6
- 21 CFR 211.22Responsibilities of quality control unit3
- 21 CFR 211.100Written procedures; deviations2
- 21 CFR 211.192Production record review2
- 21 CFR 211.28Personnel responsibilities1
- 21 CFR 211.56Sanitation1
- 21 CFR 211.58Maintenance1
- 21 CFR 211.94Drug product containers and closures1
Ask FDAgent about equipment cleaning in Form 483s
FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.
Related
About this data
- The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
- Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
- A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.
- FDAgent is independent of FDA.