Skip to content

Front Door Pharmacy, LLC dba Pure Pharmacy (Pure Pharmaceuticals)

8980 Kirby Dr, Houston, Texas, United States

FEI 3012937475

FDA has 2 inspections on file for Front Door Pharmacy, LLC dba Pure Pharmacy (Pure Pharmaceuticals) in Houston, Texas (FEI 3012937475), from 2019 to 2021. The latest ended on and was classified OAI, official action indicated.

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA posted the Form 483 from that inspection. FDAgent read its 7 observations from the PDF; they are listed below.

FDA issued a warning letter on (CDER). No court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Form 483 PDF ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
2, 2019–2021
2 OAI · 0 VAI · 0 NAI
Inspections with 483 observations shown
2 of 2
Form 483s posted by FDA
↗
3 posted
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDrugs · Drug Quality AssurancePosted 483 · 7 observations

    Read this Form 483 on its own page →Form 483 PDF on fda.gov

    Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.

    1. Observation 1

      Blocking first-pass HEPA air flow

      Personnel were observed conducting aseptic manipulations or placing equipment/supplies in an area that blocked the movement of first pass air around an open unit, whether before or after it is filled with sterile product. Specifically, during the visual sterile production observation of the finished sterile drug product, P105 (b) (4) (PAP30MG/ PHE1MG/ PGE10MCG), Lot # 051321, Expiry 7/16/21 being performed by the pharmacy PIC, the following conditions were observed: A. Reaching over filled sterile (b) (4) unstopped vials to fill empty vials. B. Placed a tray of empty vials awaiting filling in between the HEPA filter and the vials undergoing (b) (4) - (b) (4)/filling, resulting in an interruption of 1st air for the vials being processed the ISO 5 LAFU.

    2. Observation 2

      Inadequate media fill program

      Media fills were not performed that closely simulate aseptic production operations incorporating, as appropriate, worst-case activities and conditions that prove a challenge to aseptic operations. Specifically, your pharmacy fail to provide an adequate justification for your facility's media fill program in order to simulate aseptic production operations for the worst-case activities and conditions. For example, your firm’s media fill program documented within procedure, Conduct High Risk Media Fill Test, PURE_SOP 3.9.6 Conduct High Risk Media Fill Test fail to represent actual aseptic production practices, which include the stopper placement and affixing the (b) (4) to the vial. Repeat Observation

    Show observations 3 to 7
    1. Observation 3

      Inadequate environmental monitoring

      Environmental monitoring was not performed in your aseptic processing areas. Specifically, Environmental Monitoring and measurement frequency of microbiological contamination (viable/non-viable monitoring) of the air in the ISO 5 zone is inadequate. Your pharmacy fail to establish a viable monitoring program during sterile drug production within the ISO 5 LAFU. Additionally, during a review of your pharmacy's cleanroom re-certification for VIABLE Report 1.27.2021, which documents the (b) (4) assessment of air, and Surface contaminates within ISO 7 and ISO 8 Ante-room, fail to include the ISO 5 LAFU.

    2. Observation 4

      Sporicidal agent not validated

      The use of sporicidal agents in the cleanrooms and ISO 5 classified aseptic processing area was inadequate. Specifically, your firm failed to validate the sporicidal agent, (b) (4) Sporicidal Disinfectant Cleaner for its intended use within the cleanroom and aseptic production processing area. Repeat Observation

    3. Observation 5

      Poor airflow design / air returns

      Your facility design allowed the influx of poor-quality air into a higher classified area. Specifically, your firm air returns are adjacent to your HEPA filters in the ceiling of the ISO (b) (4) 8 Anteroom and ISO 7 Cleanroom used in the production of sterile drugs. Your pharmacy fail to provide records and/or documents (smoke study) in support of adequate air circulation in the prevention of an influx of poor quality air into a higher classified area. Repeat Observation

    4. Observation 6

      Inadequate hazardous drug cleaning/segregation

      You produced hazardous drugs without providing adequate segregation, cleaning of work surfaces and cleaning of utensils to prevent cross-contamination. Specifically, your pharmacy produces the non-sterile hazardous drug, Testosterone Cream 100 mg/mL and Estradiol 0.1MG/GM Cream without adequate segregation, deactivation, and adequate cleaning program to prevent cross-contamination. All non-sterile compounded drugs are compounded in the same designated drug production area without adequate cleaning requirements. Your firm's hazardous material SOP, Hazardous Material_4.9.6, Handling Hazardous Material - Pharmacy procedure failed to include cleaning requirements to deactivate hazardous drug products and components. Your pharmacy technician report your firm uses dish (b) (4) detergent to wash and allow to [illegible]. Equipment is wiped using (b) (4) Repeat Observation

    5. Observation 7

      ISO-5 not certified under dynamic conditions

      ISO-5 classified areas were not certified under dynamic conditions. Specifically, unidirectional airflow was not verified under operational conditions. Your firm’s clean room re-certification reports tests dated 1/27/2021, 8/5/2020, 2/12/2020, and 8/15/2019 were documented as being performed in dynamic conditions. On 5/24/2021, PIC reported no aseptic production simulation was performed at the time of any of the documented cleanroom and ISO 5 aseptic production processing areas during re-certifications. Repeat Observation

  2. OAISignificant objectionable conditions; FDA recommended regulatory actionDrugs · Drug Quality AssurancePosted 483 · 11 observations

    Read this Form 483 on its own page →Form 483 PDF on fda.gov

    Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.

    1. Observation 1

      Exposed skin during aseptic processing

      Personnel engaged in aseptic processing were observed with exposed skin. Specifically, on 3/6/2019, your pharmacist was observed with exposed skin while compounding the sterile produced drug, P-105 Lot# 030719@1 BUD 09/07/19, Qty. [illegible] vials. Your pharmacist's sterile gloves failed to extend over the sleeves of the gown worn.

    2. Observation 2

      Media-fill program inadequate

      Media fills were not performed that closely simulate, as appropriate, worst-case activities and conditions that prove a challenge to aseptic operations. Specifically, your media fill program fails to simulate activities and conditions within your firm's aseptic processing operations. For example, your firm's media fill aseptically simulates manipulating [illegible] compound batch size is [illegible] vials for the compounded [illegible].

    Show observations 3 to 11
    1. Observation 3

      Gloved hands touched non-ISO surfaces

      Personnel touched equipment or other surfaces located outside of the ISO 5 classified aseptic processing area with gloved hands and then engaged in aseptic processing. Specifically, on 3/7/2019 during the aseptic processing of P-105 Lot# 030719@1 BUD 09/07/19, Qty. [illegible] vials, personnel performed a task in the ISO 7 cleanroom/processing area multiple times without re-sanitizing gloves. The skin of the pharmacist's hands was completely exposed while working in the ISO 5 hood.

    2. Observation 4

      Equipment/materials not disinfected

      Equipment and materials or supplies were not disinfected prior to entering the aseptic process. Specifically, on 3/6/2019 the following conditions were observed: A. While compounding the sterile drug, your firm's pharmacist failed to disinfect non-sterile drug components, Phentolamine Mesylate USP, Mannitol USP, and Papaverine HCL, prior to removal from where weighed and staged on the stainless-steel table in ISO 8 anteroom and prior to entry into the ISO 5 aseptic processing area. B. The [redacted] mixer was observed with foreign residue on its outer surface. Your firm's pharmacist placed the mixer into the aseptic compounding area without disinfecting it. C. Your firm's stool, in the ISO 7 cleanroom, is covered with a material which cannot be adequately disinfected.

    3. Observation 5

      Non-sterile / unlabeled disinfectants used

      Disinfecting agents used in the ISO 5 classified aseptic processing areas were not sterile. Specifically, your firm pharmacist was observed using an unlabeled bottle of non-sterile [redacted] to disinfect the ISO 5 hood aseptic processing surfaces. Unlabeled bottles of non-sterile [redacted] were found in the pharmacy's ISO 8 anteroom and ISO 7 cleanroom. Your firm failed to provide records and/or surface disinfectant/decontaminant documentation for use in the clean room and aseptic processing area.

    4. Observation 6

      Unvalidated sporicidal use

      The use of sporicidal agents in the cleanrooms and ISO 5 classified aseptic processing area was inadequate. Specifically, your firm failed to validate the use of [redacted] cleaner as a sporicidal agent used in your firm's cleaning program.

    5. Observation 7

      Poor quality air influx from design

      Your facility design allowed the influx of poor quality air into a higher classified area. Specifically, your firm air returns are adjacent to your HEPA filters in the ceiling and ISO 7 cleanroom used in the production. Your firm failed to provide documents (smoke study) in support of adequate air circulation in the prevention of an influx of poor quality air into a higher classified area.

    6. Observation 8

      Incubator/environmental control inadequate

      Laboratory test equipment used in the preparation and testing of collected samples are inadequate. Specifically, your firm uses a [redacted model] for the incubation of collected media-fills. The unit fails to have adequate temperature and related humidity controls to perform and control these environmental conditions.

    7. Observation 9

      Hazardous drugs without proper segregation

      You produced hazardous drugs without providing adequate segregation, deactivation, and cleaning of utensils to prevent cross-contamination. Specifically, your firm produces the non-sterile hazardous drug, Testosterone Cream 100 mg/g, without adequate segregation, deactivation, and adequate cleaning to prevent cross-contamination. All non-sterile compounded drugs are produced in the same designated drug production area without adequate cleaning requirements.

    8. Observation 10

      ISO-5 not certified dynamically

      ISO-5 classified areas were not certified under dynamic conditions. Specifically, unidirectional airflow was not verified under operational conditions. Specifically, your firm's clean room re-certification tests dated [redacted] were documented as being performed while the pharmacy contract clean room certifier reported that he does not recertify the clean room while pharmacy is performing media fills and/or compounding actual sterile drug products. He reported no, he simulates motion in the clean room compounding area and documents as being in a dynamic state. The firm reported no media fills and/or drug compounding during certifications.

    9. Observation 11

      Particle-generating, hard-to-clean surfaces

      The ISO-classified areas have difficult-to-clean, particle-generating surfaces. Specifically, during a walk-through on 2/25/2019, foreign particulate matter was observed on top of [redacted] which was not in service. Foreign particulate matter was observed in the ISO 7 production area adjacent to the ISO 5 hood. The firm failed to prevent particle generation or provide adequate cleaning for visibly dirty equipment or surfaces.

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDER · DrugsFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3012937475 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3012937475. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.