Equipment cleaning and cleaning validation in FDA Form 483 observations
Cleaning procedures for manufacturing equipment, the studies that show they work, and the residues, swabs and logs that prove each cleaning. FDAgent finds 197 such observations in 144 Form 483s issued to 138 facilities, for inspections that ended between October 2008 and September 2026. Each is quoted below from the 483 FDA posted.
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.
- Observations
- 197
- Each quoted below
- Form 483s
- 144
- Posted by FDA
- Facilities
- 138
- Distinct FEIs
- Most recent
- Sep 28, 2026
- Inspection end date
Observations by year
These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).
Peak: 63 in 2026
| Year | Observations |
|---|---|
| 2008 | 1 |
| 2009 | 0 |
| 2010 | 0 |
| 2011 | 0 |
| 2012 | 0 |
| 2013 | 0 |
| 2014 | 1 |
| 2015 | 1 |
| 2016 | 10 |
| 2017 | 15 |
| 2018 | 20 |
| 2019 | 5 |
| 2020 | 5 |
| 2021 | 10 |
| 2022 | 7 |
| 2023 | 22 |
| 2024 | 13 |
| 2025 | 24 |
| 2026 | 63 |
How FDAgent sorts observations into this theme
Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.
It is tied to one of these 21 CFR entries
The same entries the equipment cleaning topic counts. A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.
Or its text contains one of these phrases
“cleaning validat*” · “validat* … cleaning” · “cleaning procedure*” · “cleaning process*” · “cleaning record*” · “cleaning log*” · “cleaning agent*” · “cleaning verification” · “cleaning stud*” · “cleaning method*” · “cleaning effectiveness” · “not cleaned” · “not adequately cleaned” · “not properly cleaned” · “residue*” · “swab*” · “carry over” · “carryover”
* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.
The 197 observations, newest first
Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.
Meditech Laboratories, Inc.Inspection ended May 11, 2017 · FEI 3012200488 · Observation 1
Drugs were produced without providing adequate containment and cleaning of utensils to prevent cross-contamination. Specifically, On 05/09/2017, a plastic scoop used to scoop out bulk drug substances for production was observed in the drawer in the production area containing visible white powder residue. Your firm's pharmacist in charge was unable to identify what the substances are and unable to determine when it was last cleaned. Your pharmacy …
Contains “residue*”Read the Form 483 →
Park Infusioncare LP dba Preferred HomecareInspection ended Feb 17, 2017 · FEI 3012669716 · Observation 1
… One of the ISO 5 hoods ("HOOD J") in the ISO 7 cleanroom has a work surface made of wood covered with what appears to be formica, with visible residues and abrasions on the surface. The Nurse Manager acknowledged that this surface is not amenable to suitable cleaning. b. One of the ISO 5 hoods ("HOOD H") exhibited a crack in the plexi-glass like side panel. This same ISO 5 hood exhibited an accumulation of dust above the hood. c. One of the ISO 5 …
Contains “residue*”Read the Form 483 →
Bayer AG PharmaInspection ended Jan 20, 2017 · FEI 3002806462 · Observation 3
Procedures for cleaning of the equipment and facility are not established and followed. 1. The (b) (4) building has (b) (4) (b) (4). They are not product dedicated, including (b) (4) (b) (4) that are shared use between (b) (4) products and non-(b) (4) products. Cleaning validation of the equipment failed to evaluate hard to clean and hard to visually inspect places, such as the (b) (4). Inspection of the (b) (4) (b) (4) revealed: a. The "clean" …
Contains “cleaning validat*”Read the Form 483 →
Bayer AG PharmaInspection ended Jan 20, 2017 · FEI 3002806462 · Observation 4
… Two cleaning records related to (b) (4) FPL (b) (4) on 12 January 2017. It was reported a mistake was made on the form where the original data was recorded. The operator re-wrote the data onto a new form and discarded the original. c. Numerous partially filled set-up forms for mounting of equipment associated with the start of (b) (4) campaigns were found in the waste bin. These included forms that were completely missing from batch records where …
Contains “cleaning record*”Read the Form 483 →
Aurobindo Pharma Unit 1, TelanganaInspection ended Oct 25, 2016 · FEI 3004021253 · Observation 1
Equipment cleaning procedures are not adequately validated. Specifically, a) Equipment cleaning validation studies do not support a validated time period for holding dirty equipment before initiating the cleaning process. The equipment cleaning procedure, SOP AQA-020, effective date January 20, 2016, specifies the establishment of a holding period of (b) (4) before cleaning dirty equipment through the completion of three acceptable cleaning runs; …
Contains “cleaning procedure*”Read the Form 483 →
Claris Injectables LimitedInspection ended Aug 26, 2016 · FEI 3004610460 · Observation 4
PRODUCTION SYSTEM OBSERVATION 4 Procedures designed to prevent microbiological contamination of drug products purporting to be sterile are not established, written and followed. Specifically, a) SOP PRD/004 Clean Room Entry - Exit Procedure requires use of sterile garments prior to entry into the clean room area. However, you have not validated the number of cleaning and sterilization cycles through which the garments can be processed and are …
Contains “validat* … cleaning”Read the Form 483 →
Sekisui MedicalInspection ended Jun 17, 2016 · FEI 3002806840 · Observation 1
… No raw data could be provided to support the recovery method of the cleaning validation sampling method for b) 4) 3. Complete raw data relating to sample preparation times, weights, media lot numbers, and incubators used were not maintained for the microbial testing and endotoxin testing performed during cleaning validation of(s) b)#)
Contains “cleaning validat*”Read the Form 483 →
Sekisui MedicalInspection ended Jun 17, 2016 · FEI 3002806840 · Observation 4
Failure to ensure cleaning procedures are effective. 1. Cleaning validation for the equipment used in Plant [illegible] did not include microbiological or endotoxin evaluation. The equipment is used to manufacture for the US market, which is a non-sterile API used for the preparation of sterile finished drug products. The cleaning process includes rinsing with purified water, but no sanitization steps. 2. Maximum campaign lengths for [illegible] …
Contains “cleaning procedure*”Read the Form 483 →
Absolute Pharmacy LLCInspection ended Apr 11, 2016 · FEI 3011116100 · Observation 3
… c) The pharmacist was observed swabbing the surface. Swab with the technique of [illegible] Using those and another swab to swab the (b)(4) and plated it. Your firm would recover organisms on the surfaces. Your firm would not be able to identify the location of the positive sample.
Contains “swab*”Read the Form 483 →
Drug Depot Inc. dba APS PharmacyInspection ended Jan 22, 2016 · FEI 3010621916 · Observation 9
Drug product containers and closures were not clean and sterilized and processed to remove pyrogenic properties to assure that they are suitable for their intended use. Specifically, A. Your firm's practice of (b)(4) lab glassware, vials and stoppers with non-sterile (b)(4) has not been verified that this process does not leave behind a chemical residue or introduce particles inside these containers and closures. B. Your firm does not …
Contains “residue*”Read the Form 483 →
Cosma SpAInspection ended Jan 15, 2016 · FEI 3002806754 · Observation 5
… Cleaning logs for production equipment or for the sampling/dispensing room are not controlled, issued or tracked by Quality. Logbook pages can be copied from the procedures by production personnel as needed. 4. Electronic chromatograms are not reviewed during the release of analytical data. In addition, not all analytical data sheets used by QC analysts during the testing of APIs are documented as reviewed.
Contains “cleaning log*”Read the Form 483 →
Cosma SpAInspection ended Jan 15, 2016 · FEI 3002806754 · Observation 6
Observation 6 The following cleaning discrepancies were noted during the review of equipment used in the manufacturing of APIs: 1. Spot checking is performed in support of cleaning validations. The spot checking of equipment used in the manufacture of non-sterile APIs intended to be further manufactured into sterile drug product do not address microbiological and endotoxin contamination. This is a repeat observation from the 2013 FDA inspection. …
Contains “cleaning validat*”Read the Form 483 →
Cosma SpAInspection ended Jan 15, 2016 · FEI 3002806754 · Observation 11
… of the procedure being trained on is not documented and training is not provided to operators on revisions to master production batch records and master cleaning records.
Contains “cleaning record*”Read the Form 483 →
Cosma SpAInspection ended Jan 15, 2016 · FEI 3002806754 · Observation 12
Observation 12 Working copies of cleaning batch records are not exact copies of the master records. Specifically, the header templates used in the printing of master and working copies of cleaning batch records contain different information. The header template used in the working copies is not approved during the approval of the master cleaning record.
Contains “cleaning record*”Read the Form 483 →
Custom Compounding CenterInspection ended Apr 9, 2015 · FEI 3002401385 · Observation 4
… Your firm's procedure, Room Cleaning Procedures-Compounding Labs-Sterile & Non Sterile (SOP# 5.005, Implemented date 9/19/14), reports shelving cleaning occurs only [illegible]. Your firm reuses a single gown an entire drug production day.
Contains “cleaning procedure*”Read the Form 483 →
DNA Pharmacy Services Inc dba Palm Beach CompoundingInspection ended Aug 13, 2014 · FEI 3004962739 · Observation 3
Equipment and utensils are not maintained at appropriate intervals to prevent contamination that would alter the safety, identity, strength, quality or purity of the drug product. Specifically, during the walkthrough observed white residue and scratches inside the [illegible] (a component used as part of the high-torque [illegible] blender) used to [illegible] inactive pharmaceutical ingredients before encapsulation. Additionally, materials …
Opens with FDA's citation wording for 21 CFR 211.67(a)Read the Form 483 →
Genzyme CorporationInspection ended Oct 10, 2008 · FEI 1000305672 · Observation 12
The firms overall assessment of the cleaning validation of Myozyme and Fabrazyme [illegible] was inadequate as the [illegible] to determine which product is the most difficult to remove was unknown until September 5, 2008.
Contains “cleaning validat*”Read the Form 483 →
21 CFR sections these observations open with
The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.
- 21 CFR 211.67Equipment cleaning and maintenanceIn this theme's rule31
- 21 CFR 211.113Control of microbiological contamination10
- 21 CFR 211.42Design and construction features6
- 21 CFR 211.22Responsibilities of quality control unit3
- 21 CFR 211.100Written procedures; deviations2
- 21 CFR 211.192Production record review2
- 21 CFR 211.28Personnel responsibilities1
- 21 CFR 211.56Sanitation1
- 21 CFR 211.58Maintenance1
- 21 CFR 211.94Drug product containers and closures1
Ask FDAgent about equipment cleaning in Form 483s
FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.
Related
About this data
- The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
- Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
- A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.