Batch records and documentation in FDA Form 483 observations

Master and batch production records, device history records, logbooks and good documentation practice. FDAgent finds 241 such observations in 143 Form 483s issued to 137 facilities, for inspections that ended between October 2008 and September 2026. Each is quoted below from the 483 FDA posted.

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.

Observations
241
Each quoted below
Form 483s
143
Posted by FDA
Facilities
137
Distinct FEIs
Most recent
Sep 28, 2026
Inspection end date

Observations by year

These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).

Peak: 80 in 2026

Batch records and documentation observations in Form 483s by inspection year
YearObservations
20083
20090
20101
20110
20124
20134
20140
20153
201616
201712
201815
201913
20200
20217
202218
202322
202415
202528
202680

How FDAgent sorts observations into this theme

Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.

It is tied to one of these 21 CFR entries

21 CFR 211.18021 CFR 211.18421 CFR 211.18621 CFR 211.18821 CFR 820.18121 CFR 820.184

A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.

Or its text contains one of these phrases

“batch record*” · “batch production … record*” · “master production … record*” · “master batch record*” · “device history record*” · “device master record*” · “documentation practice*” · “logbook*” · “log book*” · “BPR*” · “MPR*” · “BMR*”

* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.

The 241 observations, newest first

Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.

  1. Smithfield Biosciencs, Inc.Inspection ended Dec 23, 2025 · FEI 1527761 · Observation 1

    Process Validation performed does not demonstrate that critical process parameters produce consistent APIs meeting their specifications and quality. Specifically, Your firm's process validation for (b) (4) USP, (b) (4) product codes, does not include validation of processing times listed in the batch record and in procedure SOP# 149. For example, (b) (4) (b) (4) (b) (4) cycle minimum times in the batch record were not tested, function was not …

    Contains “batch record*”Read the Form 483 →

  2. Smithfield Biosciencs, Inc.Inspection ended Dec 23, 2025 · FEI 1527761 · Observation 3

    Batch records and laboratory records of critical steps are not reviewed and approved by the quality unit appropriately. Specifically, A.) Your analytical laboratory paper records do not include signature and date of the second person reviewing (b) (4) (b) (4) impurity data on HPLC-1 for batches (b) (4) These batches were later released by QA without investigating whether they were reviewed or not. In addition, your firm does not include an …

    Contains “batch record*”Read the Form 483 →

  3. Macau-Union Pharmaceutical LimitedInspection ended Dec 17, 2025 · FEI 3011219786 · Observation 8

    Batch production and control records do not include complete labeling control records, including specimens or copies of all labeling used for each batch of drug product produced. Specifically, Your firm did not include specimens or copies of all completed labeling, including the label adhered to packaging bottle and labeling on the (b)(4) packaging carton, in your batch production the (b)(4) records for packaging finished over-the-counter human …

    Opens with FDA's citation wording for 21 CFR 211.188(b)(8)Read the Form 483 →

  4. Wizcure Pharmaa Private LimitedInspection ended Dec 10, 2025 · FEI 3030304532 · Observation 1

    … Associated logbooks and analytical worksheets containing testing data were also incomplete or missing. Examples included: 1. There were samples in the incubators for approximately (b)(4) lots of products in varying stages of in-process sterility testing but no in-process test records. According to Microbiology Manager (b)(6), sterility (b)(4) are examined (b)(4). Results of these examinations are not recorded (b)(4) but (b)(4) record is created. …

    Contains “logbook*”Read the Form 483 →

  5. Wizcure Pharmaa Private LimitedInspection ended Dec 10, 2025 · FEI 3030304532 · Observation 6

    … You lack recordation of critical process parameter(s) in your BPRs regarding your (b)(4) process such that (b)(4) is not recorded. Your production personnel reported (b)(4) was required for (b)(4). 1. Enumeration of rejects is absent regarding your in-process visual inspection of filled (b)(4) for your (b)(4). Furthermore, critical attribute of (b)(4) integrity is apparently incorporated into your process, instead, your BPRs read in part, "OK". …

    Contains “BPR*”Read the Form 483 →

  6. Wizcure Pharmaa Private LimitedInspection ended Dec 10, 2025 · FEI 3030304532 · Observation 7

    … Logbooks (b)(4) (Testing Log Book and Bulk Sample Testing Log Book) for year 2025 were observed with missing sample received entries. 2. Log Book for gross QC calculations was observed and your QA Manager stated it was not a GMP logbook and reported the logbook is not controlled via any written SOP per your quality system.

    Contains “logbook*”Read the Form 483 →

  7. Huons Co., LtdInspection ended Nov 21, 2025 · FEI 3007163376 · Observation 1

    … The Incubator Usage Record logbook (INC-3127) was manipulated to remove two entries of bioburden (b)(4) samples collected for deviation investigation regarding (b)(4) sterilization in the Vial Line. The samples of concern were for Huons (b)(4) Injection (b)(4) Batch (domestic market). Specifically, on 10/31/2025 pages 60 to 62 of the logbook containing (b)(4) information of the deviation samples were removed by the QC Microbiology Team Lead. She …

    Contains “logbook*”Read the Form 483 →

  8. Huons Co., LtdInspection ended Nov 21, 2025 · FEI 3007163376 · Observation 5

    Batch production and control records do not include complete information relating to the production and control of each batch. Electronic records are used, but they do not meet audit trail requirements to ensure that they are trustworthy, reliable and generally equivalent to paper records. Specifically, Your Audit Trail Review procedure and practice is inadequate. The checklist currently utilized for (b)(4) audit trail reviews do not specifically …

    Opens with FDA's citation wording for 21 CFR 211.188Read the Form 483 →

  9. Implantica CE Reflux LimitedInspection ended Oct 24, 2025 · FEI 3015217460 · Observation 1

    … Corrective Action - Incomplete Batch-Record Amendment The corrective-action portion of the CAPA fails to provide documented proof that all affected batch records, including previously released batch records (job-router forms, batch-release sheets, and any downstream documentation) have been updated to include the tumbler model-number (e.g., (b) (4) as the equipment used for the tumbling operation. 3. Scope & Preventive Action - Insufficient …

    Contains “batch record*”Read the Form 483 →

  10. UBI Pharma Inc.Inspection ended Oct 14, 2025 · FEI 3004713402 · Observation 5

    … Your batch record suggested that you sample [illegible] filling rate of [illegible].

    Contains “batch record*”Read the Form 483 →

  11. Hans Biomed CorporationInspection ended Sep 12, 2025 · FEI 3003828812 · Observation 1

    … The batch record documents Step (b) (4) as completed between (b) (4) on 09/09/25. I was present during that time and did not observe use of the (b) (4). Since the (b) (4) is part of Step (b) (4), the completion time for Step (b) (4) should occur (b) (4) (b) (4) - therefore, it must be (b) (4) as currently documented. Additionally, cleaning records for the (b) (4) and (b) (4) (referenced in Step (b) (4) and Step (b) (4) of other batch records) …

    Contains “batch record*”Read the Form 483 →

  12. Farmakeio Outsourcing LLCInspection ended Sep 5, 2025 · FEI 3014982757 · Observation 2

    Laboratory controls do not include the establishment of scientifically sound and appropriate specifications and test procedures designed to assure that drug products conform to appropriate standards of identity, strength, quality and purity. Specifically, A. During a review of your firm's batch record for Product ID#364, Testosterone/Triamcinolone Acetonide Pellet (200mg/40mcg), Lot# 3131, including laboratory test chromatograms, …

    Contains “batch record*”Read the Form 483 →

  13. Farmakeio Outsourcing LLCInspection ended Sep 5, 2025 · FEI 3014982757 · Observation 5

    Routine calibration and checking of automatic and electronic equipment is not performed according to a written program designed to assure proper performance. Specifically, during a review of your firm's batch record for Testosterone/Triamcinolone Acetonide Pellet 200mg/40mcg, Lot 3468, Expiry 7/23/2026, Sections 2.3.1.3 - 2.3.1.5, document pellets being processed through the [(b)(4)] Systems, [(6)(4)] Metal detection System. Section 2.3.1.5, …

    Contains “batch record*”Read the Form 483 →

  14. Creative Testing SolutionsInspection ended Aug 29, 2025 · FEI 3000204058 · Observation 2

    … A handwritten note dated 06/02/24 was included in batch records from 05/15/2024, 05/16/2024 and 05/17/2024. This documentation did not include when the issue was discovered, how many daily QCs were affected by this error, follow-up actions, who was involved in review, and what was considered for acceptance of released affected test results.

    Contains “batch record*”Read the Form 483 →

  15. Tripak Pharmaceuticals Pty Ltd.Inspection ended Jul 25, 2025 · FEI 3008549263 · Observation 8

    … According to the cleaning logbooks, these tanks had been cleaned; however, the status was not clearly indicated through proper labeling. According to your written procedure no. 30022.04: Status Labeling in Production, new green cleaned labels should be attached after equipment cleaning. The absence of clear status labeling creates potential for cross-contamination and makes it difficult to verify equipment readiness for use. These tanks are being …

    Contains “logbook*”Read the Form 483 →

  16. Raptim Research Private LimitedInspection ended Jul 25, 2025 · FEI 3007267856 · Observation 2

    … CredBio and the logbook for freezer INS/ANA/507 documented that period 1, period 2, and period 3 samples from study (b) (4) were stored on shelf A4 of the freezer. However, when sample custodians attempted to retrieve the samples, period 2 samples were not located on shelf A4, but were instead on shelf A5. e. The Sample Disposal Records from in vitro study (b) (4) contained multiple discrepancies. i. Samples from Donor (b) (6) cell 10, aliquot 1 …

    Contains “logbook*”Read the Form 483 →

  17. Raptim Research Private LimitedInspection ended Jul 25, 2025 · FEI 3007267856 · Observation 4

    The firm did not consistently follow good documentation practices in their raw data management during conduct of method validation of (b) (4) (RRL-ANA-MVR1409). For example, in raw data forms (17242, 7215, 0597) several corrections were made to address QC findings. However, the changes were made without documenting when the change was made and who made the change.

    Contains “documentation practice*”Read the Form 483 →

  18. Raptim Research Private LimitedInspection ended Jul 25, 2025 · FEI 3007267856 · Observation 5

    … Although the backup activity failed, backup activity was indicated to have been verified by the IT administrator and QA in verification of backup logbook (RRL-IT-LG-007-001/25).

    Contains “logbook*”Read the Form 483 →

  19. Veganic SKN LimitedInspection ended Jul 21, 2025 · FEI 3024609429 · Observation 2

    … Additionally, there is insufficient QA involvement in batch record review during active production phases. The facility's current practice involves production employees fulfilling QA responsibilities, as confirmed by your Head of QA during our inspection. This arrangement compromises the independence required for effective quality oversight, as production personnel cannot provide the objective review necessary for proper QA functions. From …

    Contains “batch record*”Read the Form 483 →

  20. Foremount Enterprise Co., LtdInspection ended Jun 19, 2025 · FEI 3010402691 · Observation 8

    Procedures for device history records have not been adequately established. Your firm's Device History Record (DHR) procedure Product Identification and Traceability Procedure, QP751, V.18, dated July 12, 2024, has not been adequality established. Your firm was not able to provide DHRs for your Class II Non-Rebreathing Valves, Product Code CBP, for medical devices exported and intended for the US market, which demonstrated the devices were …

    Opens with FDA's citation wording for 21 CFR 820.184Read the Form 483 →

21 CFR sections these observations open with

The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.

Ask FDAgent about batch records and documentation in Form 483s

FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
  • Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
  • A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.
  • FDAgent is independent of FDA.