Skip to content

Creative Testing Solutions

2424 W Erie Dr, Tempe, Arizona, United States

FEI 3000204058

FDA has 7 inspections on file for Creative Testing Solutions in Tempe, Arizona (FEI 3000204058), from 2009 to 2025. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA posted the Form 483 from that inspection. FDAgent read its 2 observations from the PDF; they are listed below. One FOIA request has asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Form 483 PDF ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
7, 2009–2025
0 OAI · 1 VAI · 6 NAI
Inspections with 483 observations shown
1 of 7
Form 483s posted by FDA
↗
Requested under FOIA
1 request, latest Nov 5, 2025
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Human Cellular, Tissue, and Gene TherapiesPosted 483 · 3 observations

    Read this Form 483 on its own page →Form 483 PDF on fda.gov

    Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Observation 1

      Improper reagent reservoir use/cleaning

      Testing for communicable disease agents was not performed in accordance with the manufacturer's instructions. Specifically, During testing of HCT/P donors for relevant communicable disease agents and diseases (RCDADs), you failed to follow the manufacturer’s instructions. In the (b) (4), you utilize (b) (4) reservoir to store reagents briefly during confirmatory testing to allow technicians to utilize a (b) (4). The manufacturer’s instructions state to use (Db) (4) reservoir, not (b) (4) reservoir. Your technicians utilized (b) (4) reagent reservoirs in the testing of the following tests: 1. Bio-Rad HIV-1/HIV-2 PLUS O EIA 2. Bio-Rad GS HBsAg EIA 3. Avioq HTLV-1/II In addition to utilizing (Db) (4), you are cleaning the (Db) (4) reagent reservoir using an unvalidated cleaning process that is required by the manufacturer to reuse the reservoir. Per your firm’s representatives, these reagent reservoirs are washed using (Db) (4) but no standard operating procedure has been established and no record of cleaning have been documented. From 07/15/2023 to present, you have performed (b) (4) HIV-1/HIV-2 Plus O EIA, (b) (4) GS HBsAg, and (b) (4) HTLV-II

    2. Observation 2

      Inadequate discrepancy investigations and records

      Failure to perform a thorough investigation and make a record of the conclusions and follow-up of an unexplained discrepancy. Specifically, A. None of your firm’s standard operating procedures require investigation into all unexplained discrepancies. Your SOP titled, Error Management, only requires investigation into customer complaints and errors involving released test results. B. Your firm failed to thoroughly investigate and document improper quality control (QC) lots being maintained for reference ranges on the (b) (4) instrument. During review of donor serum protein electrophoresis results, I observed that incorrect reference percentages were used for instrument QC due to the incorrect lot number being recorded. Further review of records determined that this QC lot error was documented at least three times, on 05/15/2024, 05/16/2024, and 05/17/2024. A handwritten note dated 06/02/24 was included in batch records from 05/15/2024, 05/16/2024 and 05/17/2024. This documentation did not include when the issue was discovered, how many daily QCs were affected by this error, follow-up actions, who was involved in review, and what was considered for acceptance of released affected test results.

    Show 1 more observation
    1. Thorough investigations

      Failure to perform a thorough investigation and make a record of the conclusions and follow-up of an unexplained discrepancy.

  2. NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene TherapiesNone
  3. NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
  4. NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
  5. NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
  6. NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
  7. NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene TherapiesNone

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Nov 5, 2025. It asked for a Form 483.

  • Inspection ending Aug 29, 2025VAI

    1 request. In the inspection history above. FDA has posted its Form 483.

    1. Nov 5, 2025FDA-FOIA-2025-12603Form 483· 1 of 5 sitesRecords released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 1 section of 21 CFR in FDA’s published citations for 1 of this facility’s 7 inspections.

Observations read from posted 483 PDFs carry no CFR section of FDA’s, so they are not counted here.

Other FEIs recorded as Creative Testing Solutions

Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3000204058 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3000204058. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.