FDA issued this Form 483 to Creative Testing Solutions in Tempe, Arizona (FEI 3000204058) at the end of a biologics inspection that ended August 29, 2025. It lists 3 observations. FDA classified the inspection Voluntary Action Indicated (VAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Jul 15, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Aug 29, 2025
Biologics · Human Cellular, Tissue, and Gene Therapies
FDA's outcome
VAI
Voluntary Action Indicated
Posted by FDA
Jul 15, 2026
320 days after the inspection
Facility
FEI 3000204058
Tempe, Arizona

The 3 observations on this Form 483

Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.

  1. 1

    Improper reagent reservoir use/cleaning

    Testing for communicable disease agents was not performed in accordance with the manufacturer's instructions. Specifically, During testing of HCT/P donors for relevant communicable disease agents and diseases (RCDADs), you failed to follow the manufacturer’s instructions. In the (b) (4), you utilize (b) (4) reservoir to store reagents briefly during confirmatory testing to allow technicians to utilize a (b) (4). The manufacturer’s instructions state to use (Db) (4) reservoir, not (b) (4) reservoir. Your technicians utilized (b) (4) reagent reservoirs in the testing of the following tests: 1. Bio-Rad HIV-1/HIV-2 PLUS O EIA 2. Bio-Rad GS HBsAg EIA 3. Avioq HTLV-1/II In addition to utilizing (Db) (4), you are cleaning the (Db) (4) reagent reservoir using an unvalidated cleaning process that is required by the manufacturer to reuse the reservoir. Per your firm’s representatives, these reagent reservoirs are washed using (Db) (4) but no standard operating procedure has been established and no record of cleaning have been documented. From 07/15/2023 to present, you have performed (b) (4) HIV-1/HIV-2 Plus O EIA, (b) (4) GS HBsAg, and (b) (4) HTLV-II

  2. 2

    Inadequate discrepancy investigations and records

    Failure to perform a thorough investigation and make a record of the conclusions and follow-up of an unexplained discrepancy. Specifically, A. None of your firm’s standard operating procedures require investigation into all unexplained discrepancies. Your SOP titled, Error Management, only requires investigation into customer complaints and errors involving released test results. B. Your firm failed to thoroughly investigate and document improper quality control (QC) lots being maintained for reference ranges on the (b) (4) instrument. During review of donor serum protein electrophoresis results, I observed that incorrect reference percentages were used for instrument QC due to the incorrect lot number being recorded. Further review of records determined that this QC lot error was documented at least three times, on 05/15/2024, 05/16/2024, and 05/17/2024. A handwritten note dated 06/02/24 was included in batch records from 05/15/2024, 05/16/2024 and 05/17/2024. This documentation did not include when the issue was discovered, how many daily QCs were affected by this error, follow-up actions, who was involved in review, and what was considered for acceptance of released affected test results.

  3. Thorough investigations

    Failure to perform a thorough investigation and make a record of the conclusions and follow-up of an unexplained discrepancy.

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.