Batch records and documentation in FDA Form 483 observations
Master and batch production records, device history records, logbooks and good documentation practice. FDAgent finds 241 such observations in 143 Form 483s issued to 137 facilities, for inspections that ended between October 2008 and September 2026. Each is quoted below from the 483 FDA posted.
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.
- Observations
- 241
- Each quoted below
- Form 483s
- 143
- Posted by FDA
- Facilities
- 137
- Distinct FEIs
- Most recent
- Sep 28, 2026
- Inspection end date
Observations by year
These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).
Peak: 80 in 2026
| Year | Observations |
|---|---|
| 2008 | 3 |
| 2009 | 0 |
| 2010 | 1 |
| 2011 | 0 |
| 2012 | 4 |
| 2013 | 4 |
| 2014 | 0 |
| 2015 | 3 |
| 2016 | 16 |
| 2017 | 12 |
| 2018 | 15 |
| 2019 | 13 |
| 2020 | 0 |
| 2021 | 7 |
| 2022 | 18 |
| 2023 | 22 |
| 2024 | 15 |
| 2025 | 28 |
| 2026 | 80 |
How FDAgent sorts observations into this theme
Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.
It is tied to one of these 21 CFR entries
21 CFR 211.18021 CFR 211.18421 CFR 211.18621 CFR 211.18821 CFR 820.18121 CFR 820.184
A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.
Or its text contains one of these phrases
“batch record*” · “batch production … record*” · “master production … record*” · “master batch record*” · “device history record*” · “device master record*” · “documentation practice*” · “logbook*” · “log book*” · “BPR*” · “MPR*” · “BMR*”
* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.
The 241 observations, newest first
Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.
Jiangsu Hengrui Pharmceuticals Co., LtdInspection ended Jan 16, 2024 · FEI 3007373503 · Observation 4
The quality control unit lacks responsibility to approve all procedures or specifications impacting on the identity, strength, quality and purity of drug products. Specifically, A) The Quality Unit lacks adequate control over the issuance of controlled documents, such as production batch records, validation protocols and reports, change controls, and standard operating procedures. During the walkthrough of our inspection on 01/08/2024, we …
Contains “batch record*”Read the Form 483 →
Jiangsu Hengrui Pharmceuticals Co., LtdInspection ended Jan 16, 2024 · FEI 3007373503 · Observation 5
Records associated with drug product production and within the retention period for such records, were not made readily available for authorized inspection. Specifically, During our walk-through inspection of your firm's discard waste building, we requested to be taken to the area where office waste is discarded, an approximate (b)(4) walk from our location. As soon as we requested to be taken to this area, we observed at least one individual …
Opens with FDA's citation wording for 21 CFR 211.180(c)Read the Form 483 →
Jiangsu Hengrui Pharmceuticals Co., LtdInspection ended Jan 16, 2024 · FEI 3007373503 · Observation 6
Appropriate controls are not exercised over computers or related systems to assure that changes in master production and control records or other records are instituted only by authorized personnel. Specifically, Your firm's GMP related computerized systems and equipment spread across manufacturing workshops are not 21 CFR part 11 compliant. For example: A) Your firm operates portable non-viable particle monitoring equipment used to perform and …
Contains “master production … record*”Read the Form 483 →
Laurus Synthesis Private LimitedInspection ended Dec 12, 2023 · FEI 3011524794 · Observation 4
Batch production records do not contain complete information related to the production and control of the batch. Specifically, A. Portable gauges, such as the (b) (4) used to (b) (4) process water, and product and non-product contact transfer hoses are not identified or recorded within the batch production record for traceability and to ensure they are adequate for use. B. The use of process water in the manufacture of (b) (4) is not adequately …
Contains “batch production … record*”Read the Form 483 →
Laurus Synthesis Private LimitedInspection ended Dec 12, 2023 · FEI 3011524794 · Observation 5
… in procedure QA/086, "Procedure for Operation of Document Management System (DMS) Software", and can be used for any purpose as part of the execution of the batch record. These pages are not issued, controlled, or reconciled in a way to ensure that the data and records are complete and that all pages printed are accounted for as part of batch record execution and documentation.
Contains “batch record*”Read the Form 483 →
Antaria Pty Ltd.Inspection ended Nov 17, 2023 · FEI 3025978151 · Observation 9
… Your firm has no temperature or humidity data from October 6, 2022 to present except for a (b) (4) note in a logbook. There are several recent entries noting the temperature and/or humidity was out of specification however, no investigation was initiated.
Contains “logbook*”Read the Form 483 →
Jiangsu Intco Medical Products Co., LtdInspection ended Nov 17, 2023 · FEI 3005260401 · Observation 1
Your firm failed to establish adequate written procedures for production and process controls designed to assure that the drug products have the identity, strength, purity, and quality that they are purported or represented to possess. Specifically, Your firm failed to retain records to support your process validation for OTC manufacturing. You did not maintain batch production records used as part of your process performance qualification for …
Contains “batch production … record*”Read the Form 483 →
Jiangsu Intco Medical Products Co., LtdInspection ended Nov 17, 2023 · FEI 3005260401 · Observation 2
… The following batches were used for your accelerated and long-term stability study, however your firm did not maintain the batch records and destroyed them on February 10, 2023. Product [Batch 1] [Batch 2] [Batch 3] [illegible 11/19/2022] (b) Your firm failed to evaluate the stability of OTC [illegible] in the container-closure system in which they are being marketed. You manufacture liquid products in the following configuration [illegible]. …
Contains “batch record*”Read the Form 483 →
Synergy Health Ede BVInspection ended Oct 12, 2023 · FEI 3020976354 · Observation 3
The device history record does not demonstrate that the device was processed in accordance with the device master record. Specifically, the batch record for the processing of UNO 03 GentleCath Glide urethral catheters batch run [illegible] does not include the process treatment (number of laps) for xray processing. The routine irradiation specification shows that the xray process should include 14 process event laps. The customer identification …
Contains “device history record*”Read the Form 483 →
Sanofi PasteurInspection ended Sep 29, 2023 · FEI 3002972083 · Observation 1
The master production and control records are deficient in that they do not include complete instructions. Specifically, there are no defined (b)(4) time frames and/or instructions for the (b)(4) preparation of the Concentrated Tetanus Protein (CTP) bulk in the Haemophilus b Conjugate Vaccine (Tetanus Toxoid Conjugate) manufacturing process. A. The CTP bulk requires the (b)(4) of the bulk after the [illegible] step in Bldg. (b)(4) (Room (b)(4)). …
Opens with FDA's citation wording for 21 CFR 211.186(b)(9)Read the Form 483 →
Linhai Tianyu Pharmaceutical Co., Ltd.Inspection ended Sep 15, 2023 · FEI 3013345425 · Observation 1
… For example, we reviewed Equipment Operational Logbooks FOM-GDL-035-01 for 2022.05.09~2022.08.10 and 2022.08.18~2022.08.28 for (b) (4) (b) (4) 215 in Workshop (b) (4) On July 4, 2022, your firm completed a (b) (4) clean after a production campaign of (b) (4) (b) (4) later, your firm completed a (b) (4) clean on (b) (4) per your firm's Cleaning Management Procedure (SMP-PR-011). On August 16, 2022, your firm began a (b) (4) batch production …
Contains “logbook*”Read the Form 483 →
Linhai Tianyu Pharmaceutical Co., Ltd.Inspection ended Sep 15, 2023 · FEI 3013345425 · Observation 4
… review room on 09/11/23, I observed your QC analyst/reviewer checking off boxes on a form titled "Checklist of Electronic Data" for approximately 5 analytical batch records for (b) (4) testing of (b) (4) API via Ion Chromatography test, one after another, without any software opened for review. In all, we counted approximately 18 analytical test records for six batches which the analyst confirmed to have performed during the observed review …
Contains “batch record*”Read the Form 483 →
Linhai Tianyu Pharmaceutical Co., Ltd.Inspection ended Sep 15, 2023 · FEI 3013345425 · Observation 7
The responsibilities and procedures applicable to the quality control unit are not in writing and fully followed. Specifically, Supervisory oversight over the laboratory electronic systems and data is deficient. Per your firm's data review procedure, Quality Control (QC) personnel perform review electronic test data for all analytical batch records using an approved review procedure and checklist for HPLC/GC network equipment. This review process …
Contains “batch record*”Read the Form 483 →
Central Glass Co., Ltd.Inspection ended Sep 15, 2023 · FEI 3002806649 · Observation 5
Your validation protocols for the execution of revalidation studies do not specify critical process steps. Your process validation studies for (b) (4) do not show the critical process parameters that were going to be studied, review of batch records shows that the validation study did not include variation of parameters within the established ranges. In addition parameters such as the timer used to control the flow in the (b) (4) are not defined …
Contains “batch record*”Read the Form 483 →
Wickliffe LLCInspection ended Sep 15, 2023 · FEI 3002992930 · Observation 2
… batch 520880, your technician changed out a glove on the ISO-5 (b) (4), however, glove changes are not documented during media fills or reflected in the batch record. According to your technician these glove changes can occur frequently ((b) (4)). It is unknown if your firm has ever changed a glove during media fills. b) Your technician overfilled a vial of Pentosan Glucosamine batch 520880 and proceeded to pour that vial's contents into another …
Contains “batch record*”Read the Form 483 →
Samsung Biologics Co. Ltd.Inspection ended Sep 1, 2023 · FEI 3010479596 · Observation 1
… of the MSAT Laboratory was (b) (4) as an ad-hoc audit (SIR(b) (4) 05), and while no data integrity issues were identified, a recommendation to implement good documentation practice and data verification procedures was identified. At the request of Senior Quality Management, a subsequent audit was performed (b) (4) (SIR-(b) -05), based on laboratory areas of work that included GMP requirements (e.g., support of regulatory submissions), with a …
Contains “documentation practice*”Read the Form 483 →
Jiangsu Hengrui Pharmaceuticals CO. LTD.Inspection ended Aug 29, 2023 · FEI 3003404148 · Observation 3
Batch production and control records for each batch of drug product produced do not include an accurate reproduction of the appropriate master production or control record which was checked for accuracy, dated and signed. Specifically, logbook for (b) (4) Tube Replacement Records in Workshop (b) (4) were replaced, with the information in logbook rewritten in the new logbook. a) The logbook used to document the number of (b) (4) tubing uses …
Opens with FDA's citation wording for 21 CFR 211.188(a)Read the Form 483 →
Jiangsu Hengrui Pharmaceuticals CO. LTD.Inspection ended Aug 29, 2023 · FEI 3003404148 · Observation 9
The batch production and control records are deficient in that they do not include documentation of the accomplishment of each significant step in manufacturing, processing, and holding. Specifically, a) Workshop (b) (4) tanks used for holding in-process sterile drug solutions (b) (4) have (b) (4) used for (b) (4) into the tanks. The (b) (4) had labels that documented number of uses, (b) (4) installation, (b) (4) model, batch number, replacement …
Opens with FDA's citation wording for 21 CFR 211.188(b)Read the Form 483 →
UCB Farchim SAInspection ended Jul 4, 2023 · FEI 3005023799 · Observation 1
… During the walkthrough of the inspection on 06/27/2023, I observed a QA personnel delete a recently issued manufacturing batch records on their desktop following your firm's procedure, after which the recycle bin continued to be clear, showing no recent deletes. After further review, your firm's IT team was able to determine that name of files deleted are available dating back 6 months, however documents are not able to be restored. D. The …
Contains “batch record*”Read the Form 483 →
Qilu Antibiotics Pharmaceutical Co., LTD.Inspection ended May 30, 2023 · FEI 3009831352 · Observation 1
… There were no instructions and no documentation in the batch records for the ad c. For Media fill #(b) (4) two separate preparations of non-sterile (b) (4) were dissolved in (b) (4) and used for contact with (b) (4) 33105 and (b) (4) 33106 (without (b) (4) After contact with equipment, all of the (b) (4) were (b) (4) (b) (4) and emersed into (b) (4) canisters of (b) (4) media for media simulation incubation. There were no instructions and no …
Contains “batch record*”Read the Form 483 →
21 CFR sections these observations open with
The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.
- 21 CFR 211.188Batch production and control recordsIn this theme's rule17
- 21 CFR 211.100Written procedures; deviations14
- 21 CFR 211.22Responsibilities of quality control unit13
- 21 CFR 211.68Automatic, mechanical, and electronic equipment13
- 21 CFR 211.113Control of microbiological contamination11
- 21 CFR 211.180General requirementsIn this theme's rule8
- 21 CFR 211.186Master production and control recordsIn this theme's rule5
- 21 CFR 211.42Design and construction features4
- 21 CFR 211.160General requirements4
- 21 CFR 211.192Production record review4
Ask FDAgent about batch records and documentation in Form 483s
FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.
Related
About this data
- The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
- Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
- A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.
- FDAgent is independent of FDA.