Synergy Health Ede BV Form 483, October 2023
FDA issued this Form 483 to Synergy Health Ede BV in Venlo, Netherlands (FEI 3020976354) at the end of a device inspection that ended October 12, 2023. It lists 3 observations. FDA classified the inspection Voluntary Action Indicated (VAI).
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: FDA's Electronic Reading Room, posted Nov 12, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.
- Inspection ended
- Oct 12, 2023
- Devices · Compliance: Devices
- FDA's outcome
- VAI
- Voluntary Action Indicated
- Posted by FDA
- Nov 12, 2026
- 1,127 days after the inspection
The 3 observations on this Form 483
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.
- 1
Inadequate process change control procedures
Procedures for changes to a process have not been adequately established. Specifically, the process treatment [illegible] factor for xray sterilization in the Venlo facility was changed from [illegible] reportedly on September 11, 2023. The change was not documented and approved in the change control system.
- 2
Conveyor speed parameters not adequately defined
Procedures for monitoring and control of process parameters for a validated process have not been adequately established. Specifically, the conveyor speed in the Xray treatment area has not been adequately defined. Process monitoring is performed at [illegible] According to the operating technician the system allows speed settings between 0 and 4.5 m/minute. The described working range (undocumented) for the conveyor speed is between [illegible]
- 3
DHR missing X-ray treatment laps verification
The device history record does not demonstrate that the device was processed in accordance with the device master record. Specifically, the batch record for the processing of UNO 03 GentleCath Glide urethral catheters batch run [illegible] does not include the process treatment (number of laps) for xray processing. The routine irradiation specification shows that the xray process should include 14 process event laps. The customer identification report describes that the product should get [illegible] ps and [illegible] pass. The company was able to generate a run report that showed that the product was processed with [illegible] ps and [illegible] passes (not found in the DHR).
Ask FDAgent about this Form 483
FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.
Related
About this data
- The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
- The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
- FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
- FDAgent is independent of FDA.