Skip to content

Batch records and documentation in FDA Form 483 observations

Master and batch production records, device history records, logbooks and good documentation practice. FDAgent finds 241 such observations in 143 Form 483s issued to 137 facilities, for inspections that ended between October 2008 and September 2026. Each is quoted below from the 483 FDA posted.

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.

Observations
241
Each quoted below
Form 483s
143
Posted by FDA
Facilities
137
Distinct FEIs
Most recent
Sep 28, 2026
Inspection end date

Observations by year

These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).

Peak: 80 in 2026

Batch records and documentation observations in Form 483s by inspection year
YearObservations
20083
20090
20101
20110
20124
20134
20140
20153
201616
201712
201815
201913
20200
20217
202218
202322
202415
202528
202680

How FDAgent sorts observations into this theme

Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.

It is tied to one of these 21 CFR entries

21 CFR 211.18021 CFR 211.18421 CFR 211.18621 CFR 211.18821 CFR 820.18121 CFR 820.184

A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.

Or its text contains one of these phrases

“batch record*” · “batch production … record*” · “master production … record*” · “master batch record*” · “device history record*” · “device master record*” · “documentation practice*” · “logbook*” · “log book*” · “BPR*” · “MPR*” · “BMR*”

* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.

The 241 observations, newest first

Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.

  1. Qilu Antibiotics Pharmaceutical Co., LTD.Inspection ended May 30, 2023 · FEI 3009831352 · Observation 5

    … Review of the media fill batch record (b) (4) dated 06/01/2022 and batch record 01/12/23 showed an aseptic operator not receiving sampling training for either media fill conducted within (b) (4) The simulation protocol and report for unit (b) (4) workshop (b) (4) dated 01/18/2023 indicates that the operator is required to complete the sampling training but does not show as completed over the last two media fills. The operator is still …

    Contains “batch record*”Read the Form 483 →

  2. Apollo Care LLCInspection ended Apr 5, 2023 · FEI 3013927023 · Observation 1

    … -The batch records affected by deviation D-22-067 list "N/A" on the environmental monitoring sections with no further explanation of whether or not microbial growths were obtained.

    Contains “batch record*”Read the Form 483 →

  3. FRUEH VERPACKUNGSTECHNIK AGInspection ended Mar 24, 2023 · FEI 3010363503 · Observation 3

    The batch production and control records are deficient in that they do not include documentation of the accomplishment of each significant step in manufacturing, processing, and packing. Specifically, during packaging operations of drug product, we observed (b)(4) This practice of (b)(4) is not detailed in your batch record, nor is it described in a procedure.

    Opens with FDA's citation wording for 21 CFR 211.188(b)Read the Form 483 →

  4. FRUEH VERPACKUNGSTECHNIK AGInspection ended Mar 24, 2023 · FEI 3010363503 · Observation 4

    The responsibilities and procedures applicable to the quality unit are not in writing and fully followed. Specifically, Supervisory oversight over quality/production unit operations and laboratory/production electronic system and data is deficient. For Example A. The Quality Unit lacks adequate control over the issuance of production batch records purported to be controlled under your batch issuance procedures. During the inspection, it was …

    Contains “batch record*”Read the Form 483 →

  5. Genlbet Biopbarmaceuticals SAInspection ended Jan 31, 2023 · FEI 3011160795 · Observation 1

    … The (b)(4) do not represent the maximum target batch size of [illegible] as [illegible] per the current approved master batch record, (b)(4). b. There is insufficient data to ensure that the uniformity of the formulated (b)(4) containing (b)(4) [illegible] is maintained throughout the hold time and during the [illegible] within the (b)(4). (b)(4) c. The operating ranges specified during manufacturing are not optimized. For example, an (b)(4) …

    Contains “master batch record*”Read the Form 483 →

  6. Octapharma Pharmazeutika Produktionsgesellschaft m.b.H. OPGInspection ended Dec 16, 2022 · FEI 3002809097 · Observation 7

    … procedures are lack instructions to routinely perform annual data audit checks of the archived (b)(4) and (b)(4) cGMP analytical QC data and production batch records.

    Contains “batch record*”Read the Form 483 →

  7. DyAnsys India Pvt. Ltd.Inspection ended Dec 15, 2022 · FEI 3010350335 · Observation 1

    … devices were shipped The type and quantity of the devices shipped The dates the devices were shipped The serial number of the devices shipped According to the batch records and discussions with you - (b)(4) devices were manufactured for use in the study and were shipped to the clinical site. (b)(4) devices were not used and were returned to you at the end of the study. G. The study data to be used to support the primary objective of the study, …

    Contains “batch record*”Read the Form 483 →

  8. Lupin Limited Unit 1Inspection ended Nov 23, 2022 · FEI 3002807511 · Observation 1

    … This batch recorded OOS results for assay (stratified samples) at three points including (b) (4) %, (b) (4) %, (b) (4) (b) (4) (Specification (b) (4) - (b) (4) %). On 11/16/2022, your Investigation Manager reviewed the respective investigation report OOS/C/21/MDP/IPDP/004. In Phase I investigation you noted the test failing for system suitability and through hypothesis test, you concluded instrument malfunction was the root cause for the recorded …

    Contains “batch record*”Read the Form 483 →

  9. Lupin Limited Unit 1Inspection ended Nov 23, 2022 · FEI 3002807511 · Observation 2

    Appropriate controls are not exercised over computer or related systems to assure that changes in master production and control records or other records are instituted only by authorized personnel. Specially, a. A review of the Sequence Audit trail data for Chromeleon, revealed several examples where your analyst deleted samples and/or blanks from the sequence lists. Your procedure SOP_MDP_QC_013135 (5.0) Assigning User/System Policies and …

    Contains “master production … record*”Read the Form 483 →

  10. Lupin Limited Unit 1Inspection ended Nov 23, 2022 · FEI 3002807511 · Observation 3

    Written procedures are not established for evaluations conducted at least annually to review records associated with a representative number of batches, whether approved or rejected. Specifically, Your Annual Product Quality Review for (b) (4) USP (b) (4) mg/(b) (4) mL and (b) (4) mg/(b) (4) mL for February 2021 to January 2022 does not include (b) (4) batches that were manufactured in year 2021 including: (b) (4). Your General Manager of Quality …

    Opens with FDA's citation wording for 21 CFR 211.180(e)(1)Read the Form 483 →

  11. Lupin Limited Unit 1Inspection ended Nov 23, 2022 · FEI 3002807511 · Observation 4

    … (Note: this is not intended to be an all-inclusive list), i.e., draft documents regarding specification/general test procedures, batch manufacturing records (BMR), proposals for change controls & supporting data, APQR meant for review purposes, Draft investigation report of OOS, OOT, Market complaints, deviations, incidents etc., document copy as Change Control, Deviation, Temporary Change Control, Complaint, CAPA used for reference to …

    Contains “BMR*”Read the Form 483 →

  12. Lupin Limited Unit 1Inspection ended Nov 23, 2022 · FEI 3002807511 · Observation 9

    … Review of your Master Batch Record of (b) (4) USP suggested that the MBR did not have provisions to continuously monitor critical process parameters (including but not limited to temperatures). In most cases, (b) (4) temperature data is recorded for the duration of the critical process steps. The above implicated (b) (4) USP Batch No. (b) (4) was used in the manufacturing of (b) (4) CAP USP (b) (4) mg - (US) Batch No. (b) (4) Manufactured date …

    Contains “master batch record*”Read the Form 483 →

  13. Lupin Limited Unit 1Inspection ended Nov 23, 2022 · FEI 3002807511 · Observation 10

    … the particular stages of manufacturing by having the production operator read the temperature on DTI (digital temperature indicator) display and record in the batch record. Your MBRs (including but not limited to) for (b) (4) USP (Doc. No. 20000001078/32), (b) (4) USP (Doc. No. 20000001495/30), and (b) (4) USP (Doc. No. 20000001614/28) have provisions to record the start and end (b) (4) processing time and (b) (4) data of temperature for the …

    Contains “batch record*”Read the Form 483 →

  14. Lupin Limited Unit 1Inspection ended Nov 23, 2022 · FEI 3002807511 · Observation 12

    … The Batch Production Records (BPR) do not provide instructions to turn on the (b) (4) prior to, or during, the (b) (4) process. There is no record to document the actual mm of water column and to assure that the filter does not "choked" when the water column reaches (b) (4) mm. c. The Corrective Action and Preventive Action (CAPA), document number SOP MUM CQA 016074, dated 31 Mar 2021, establishes the following i.e., "Target date of closure shall …

    Contains “batch production … record*”Read the Form 483 →

  15. Lupin Limited Unit 1Inspection ended Nov 23, 2022 · FEI 3002807511 · Observation 13

    The validation batch production records #(b) (4) dated 03/07/20, for the (b) (4) API Process Performance Qualification (PPQ) instruct for the (b) (4) process to "Load the (b) (4) material uniformly into the (b) (4) with the help of clean (b) (4) of approx. (b) (4) kg capacity, (b) (4) approx. (b) (4) in each (b) (4)." The General Manager of API Manufacturing confirmed that the (b) (4) (b) (4) mm (d) x (b) (4) mm (h)] does not provide any form of …

    Contains “batch production … record*”Read the Form 483 →

  16. QuVa Pharma , Inc.Inspection ended Oct 14, 2022 · FEI 3013931875 · Observation 3

    … B) The ILP section of the Batch Production Record (BPR) for syringe products such as (b) (4) 50 ml in Sodium Chloride 50ml syringe (Lot# (b) (4), Expiration Date: 11/02/2022), Phenylephrine HCl 1 mg (100 mcg/ml) 10 ml in 0.9% Sodium Chloride solution 10ml syringe KCT (Lot# (b) (4), Expiration Date: 11/07/2022) and Lidocaine HCl PF 1%, 100 mg/10 mL, 10ml syringe (Lot# (b) (4), Expiration Date: 09/29/2022) does not include a section for operators …

    Contains “batch production … record*”Read the Form 483 →

  17. QuVa Pharma , Inc.Inspection ended Oct 14, 2022 · FEI 3013931875 · Observation 4

    Appropriate controls are not exercised over computers or related systems to assure that changes in master production and control records or other records are instituted only by authorized personnel. Specifically, A) During review of user access rights for the (b) (4) (Equipment ID: (b) (4) (b) (4)), in Building (b) (4) which are used to test the (b) (4) of the (b) (4) used to sterilize in process bulk, it was observed that all operators access …

    Contains “master production … record*”Read the Form 483 →

  18. Catalent Indiana LLCInspection ended Sep 1, 2022 · FEI 3005949964 · Observation 1

    Your firm failed to thoroughly investigate any unexplained discrepancy or failure of a batch or any of its components to meet any of its specifications, whether or not the batch has already been distributed. Specifically, A. Per various (b)(4) batch records, the Visible Particles USP <790> testing list a specification of "(b)(4) particles observed**If (b)(4) particle is observed during initial testing, test an additional (b)(4) vials with a …

    Contains “batch record*”Read the Form 483 →

  19. Catalent Indiana LLCInspection ended Sep 1, 2022 · FEI 3005949964 · Observation 11

    Records are not maintained so that data therein can be reviewed at least annually to evaluate the quality standards of each drug product to determine the need for changes in specifications or manufacturing or control procedures. Your firm did not establish criteria during supplier qualification and requalification based on recurring supplier complaints in A-SOP-03-03-003, Supplier Qualification. In addition, your firm has not been tracking and …

    Opens with FDA's citation wording for 21 CFR 211.180(e)Read the Form 483 →

  20. Serum Institute of India Pvt. Ltd.Inspection ended Apr 19, 2022 · FEI 3020223359 · Observation 1

    Good Documentation Practices (GDP) and data integrity practices were deficient. Specifically, a) Deviation Report No. M26/06/21/03, dated 6/19/2021, (b)(4) reported that the limit of conductivity for [illegible] was documented as [illegible] instead of [illegible] in the following records: 1) Master Formula Record (MFR) for [illegible] 2) Batch Manufacturing Record (BMR) for [illegible] (BMR no. MBR/ [illegible] 3) SOP for Preparation of …

    Contains “documentation practice*”Read the Form 483 →

21 CFR sections these observations open with

The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.

Ask FDAgent about batch records and documentation in Form 483s

FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
  • Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
  • A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.