Skip to content

FRUEH VERPACKUNGSTECHNIK AG Form 483, March 2023

FDA issued this Form 483 to FRUEH VERPACKUNGSTECHNIK AG in Fehraltorf, Switzerland (FEI 3010363503) at the end of a device inspection that ended March 24, 2023. It lists 6 observations. FDA classified the inspection Voluntary Action Indicated (VAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Oct 23, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Mar 24, 2023
Devices · Compliance: Devices
FDA's outcome
VAI
Voluntary Action Indicated
Posted by FDA
Oct 23, 2026
1,309 days after the inspection
Facility
FEI 3010363503
Fehraltorf, Switzerland

The 6 observations on this Form 483

Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.

  1. 1

    Missing cleaning validation and disinfectant efficacy

    Procedures for the cleaning and maintenance of equipment are deficient regarding the protection of clean equipment from contamination prior to use. Specifically, your firm has not performed any cleaning validation studies to assess the effectiveness of your cleaning procedures for equipment and rooms used in the packaging operations for (b)(4) drug products. Additionally, no disinfectant efficacy studies have been conducted to ensure the effectiveness of your disinfectants used during cleaning operations.

  2. 2

    Supplier requirements lack bioburden controls

    Requirements that must be met by suppliers have not been adequately established. Specifically, components of (b)(4) drug product are not adequately controlled for bioburden levels. Specifications for (b)(4) do not include a bioburden requirement and incoming certificates of conformity do not certify that components meet a bioburden requirement.

  3. 3

    Batch records missing significant packaging steps

    The batch production and control records are deficient in that they do not include documentation of the accomplishment of each significant step in manufacturing, processing, and packing. Specifically, during packaging operations of drug product, we observed (b)(4) This practice of (b)(4) is not detailed in your batch record, nor is it described in a procedure.

  4. 4

    Deficient Quality Unit oversight and data integrity

    The responsibilities and procedures applicable to the quality unit are not in writing and fully followed. Specifically, Supervisory oversight over quality/production unit operations and laboratory/production electronic system and data is deficient. For Example A. The Quality Unit lacks adequate control over the issuance of production batch records purported to be controlled under your batch issuance procedures. During the inspection, it was observed that copies of official batch records for drug products (b)(4) without adequately controlling each page. For example, (b)(4) B. There are no requirements for the Quality Unit oversight of Master Batch Record storage and copying for use in manufacturing to ensure executed record instructions are accurate reproductions of the master records. Your firm does not maintain adequate master batch records for the above-mentioned drug products. During the inspection, we observed at least two recent deviations where incorrect batch record documents were issued and used by the production unit. C. Your firm does not have adequate Quality Unit oversight of visual inspection qualification test kits and the qualification process for operators performing visual inspection operations during packaging of drug products. For example, per your firm's procedure, the (b)(4) D. There is no adequate data integrity program in place to include an adequate review of all electronic data by the Quality Assurance Unit for standalone and network systems, to ensure completeness, consistency, and accuracy of all laboratory and production equipment electronic data generated by your firm. Specifically, per your firm's procedure, (b)(4) This review process does not appear to be adequate, in that all potential data generated during that specific review period for that product are not reviewed. There is no adequate reconciliation of all generated electronic test and production data during the review process for standalone and network equipment. E. Your firm does not have a written procedure, which include, but not limited to, establishing the requirements for quality agreements, the responsibilities of each party, and the review and approval for the agreements.

  5. 5

    Incoming lots lack distinctive identification codes

    Each lot in shipment received was not identified with a distinctive code for each container or grouping of containers for components. Specifically, The firm does not assign unique lot number for each lot in each shipment received for all drug components. Currently, the firm uses the (b)(4)

  6. 6

    Sterilization process not validated for dryness

    A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated. Specifically, the sterilization process (as reported in Document VB 15254) for (b)(4) (b)(4) The validation does not demonstrate that the process can consistently produce dry product.

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.