Genlbet Biopbarmaceuticals SA Form 483, January 2023
FDA issued this Form 483 to Genlbet Biopbarmaceuticals SA in Oeiras, Portugal (FEI 3011160795) at the end of a biologics inspection that ended January 31, 2023. It lists 4 observations. FDA classified the inspection Voluntary Action Indicated (VAI).
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: FDA's Electronic Reading Room, posted Oct 20, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.
- Inspection ended
- Jan 31, 2023
- Biologics · Vaccines and Allergenic Products
- FDA's outcome
- VAI
- Voluntary Action Indicated
- Posted by FDA
- Oct 20, 2026
- 1,358 days after the inspection
The 4 observations on this Form 483
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.
- 1
No written production and process controls
There are no written procedures for production and process controls designed to assure that the [illegible] have (b)(4) the identity, strength, quality, and purity they purport or are represented to possess. Specifically, there is insufficient data to support the maximum target batch size, hold time, and process control parameters for the manufacture of (b)(4). [illegible] executed batch sizes for the process performance qualification a. The (b)(4) do not represent the maximum target batch size of [illegible] as [illegible] per the current approved master batch record, (b)(4). b. There is insufficient data to ensure that the uniformity of the formulated (b)(4) containing (b)(4) [illegible] is maintained throughout the hold time and during the [illegible] within the (b)(4). (b)(4) c. The operating ranges specified during manufacturing are not optimized. For example, an (b)(4) [illegible] specification for (b)(4) stages are not specified in the manufacturing batch record. Current operating ranges for (b)(4) respectively. (b)(4) detection device to ensure that the d. There is no (b)(4) is free of contamination. e. The supplier for (b)(4) respectively), used allows the [illegible] for the (b)(4) steps during the manufacturing of [illegible]. A toxicological assessment was not performed to evaluate the safety of the presence of particles which are not particulate-free. [illegible] while using the (b)(4). f. There are no equipment logs for the (b)(4) and the usage and cleaning history of these pieces of equipment are not readily available.
- 2
Inadequate equipment cleaning validation
Equipment cleaning processes are inadequate. Specifically: a. The cleaning validation study (Cleaning Validation Summary Report) performed on (b)(4) contact equipment does not ensure removal of the [illegible]. Specifically: i. (b)(4) was not performed to ensure that the product is removed from the difficult to clean areas. (b)(4) was not assessed. ii. (b)(4) iii. The cleaning validation did not include the maximum lot size of (b)(4). b. [illegible] hold time is not representative of the manufacturing process, which includes (b)(4) (b)(4) that equipment are exposed to the Grade [illegible] environment after cleaning (b)(4). c. [illegible] hold time is not representative of the manufacturing process, which includes a hold [illegible] (b)(4) decontamination. manufacturing [illegible].
- 3
Incomplete decontamination cross-contamination prevention
Decontamination is incomplete or not adequately demonstrated to ensure prevention of cross-contamination. Specifically: a. A decontamination study was not performed to evaluate the effectiveness of [illegible] which is used to (b)(4) decontaminate [illegible] parts, against (b)(4). b. (b)(4), Report on the Cleaning Validation Strategy and Results did not evaluate the material of construction of all equipment or materials including (b)(4) (b)(4) (b)(4) [illegible] and a risk assessment was not performed to justify not evaluating these materials of construction, which includes equipment or materials.
- 4
Deficient equipment cleaning maintenance procedures
Written procedures for decontamination, cleaning, and maintenance of equipment are deficient regarding sufficient details to protect equipment from contamination prior to use. For example: a. (b)(4) (b)(4) do not specify the duration that equipment may be exposed to the environment after cleaning (b)(4). b. (b)(4), Cleaning Validation Summary Report, does not specify how the hold time is determined. c. Protocols governing decontamination of containing equipment that [illegible] the manufacturing suites are [illegible] locations in order to prevent contamination. [illegible] deficient regarding sufficient details such as the [illegible].
Ask FDAgent about this Form 483
FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.
Related
About this data
- The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
- The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
- FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
- FDAgent is independent of FDA.