Skip to content

DyAnsys India Pvt. Ltd. Form 483, December 2022

FDA issued this Form 483 to DyAnsys India Pvt. Ltd. in Chennai, India (FEI 3010350335) at the end of a device inspection that ended December 15, 2022. It lists 1 observation. FDA classified the inspection Voluntary Action Indicated (VAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Oct 19, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Dec 15, 2022
Devices · Bioresearch Monitoring
FDA's outcome
VAI
Voluntary Action Indicated
Posted by FDA
Oct 19, 2026
1,404 days after the inspection
Facility
FEI 3010350335
Chennai, India

The 1 observation on this Form 483

Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.

  1. 1

    Failure to assure GCP compliance as sponsor

    As the sponsor of this single-site clinical trial, conducted wholly outside the United States in support of a pre-market notification submission, I found you failed to assure the study was conducted in accordance with Good Clinical Practice. Specifically, A. You could not provide documentation confirming that the clinical investigator had obtained informed consent from any of the [illegible] subjects screened for the study. [illegible] of those subjects were enrolled and received at least one device application. B. You did not assure that the investigational plan was followed. Including but not limited to: The inclusion criteria listed in the protocol requires subjects to have an average pain score of (b)(4) prior to randomization as assessed by the Brief Pain Inventory (BPI) and Visual Analog Scale for Pain (VAS). The inclusion criteria listed in the case report forms required an average pain score of [illegible] for study enrollment. The protocol (b)(4) says that subjects will be asked to keep a patient diary noting the day and date they place the study device and any adverse events experienced. Subject diaries were not used for the study. The protocol (b)(4) says that blood will be obtained for fasting blood glucose (FBG), postprandial blood glucose (PLBS), glycated hemoglobin (HbA1C) and thyroid profile. Thyroid profiles were not performed for any of the subject files reviewed. The protocol (b)(4) says that the following objective and subjective tests will be performed at all study visits: Brief Pain Inventory (BPI) - Pain Intensity Visual Analog Scale (VAS) Neuropathic Pain Score Inventory (NPSI) Neuropathic Disability Score (NDS) Semmes-Weinstein 10g Monofilament Test Vibration Perception Threshold via biothesiometer (VPT) Pinprick Sensory Threshold (PPT) Patellar Reflex Response Test (Standard Patellar Hammer) The following tests were not performed at any of the study visits after Enrollment (Baseline): Brief Pain Inventory (BPI) Neuropathic Pain Score Inventory (NPSI) Neuropathic Disability Score (NDS) Semmes-Weinstein 10g Monofilament Test Pinprick Sensory Threshold (PPT) Patellar Reflex Response Test (Standard Patellar Hammer) The Overall Neuropathy Limitations (ONLS) test was performed at all study visits, and is listed in the Statistical Analysis Plan as a secondary outcome measurement. This assessment is not listed in the protocol (b)(4) as part of any study visit. The protocol (b)(4) says all data will be entered into a validated database, and at critical junctures, data will be locked and cleaned per established procedures (b)(4) C. You did not monitor the progress of the clinical trial to ensure that the investigation was carried out consistently with the protocol, to prevent or mitigate sources of error in the conduct, collection, and reporting of critical study data, and to assure the safety of the subjects. D. You did not ensure that the Principal Investigator and Study Director were trained to comply with Good Clinical Practices or to conduct the study in accordance with the protocol. E. You did not obtain a signed investigator agreement from the Principal Investigator of the study. F. You did not maintain records of shipment and disposition of study devices including: The name and address of the study staff and study site where devices were shipped The type and quantity of the devices shipped The dates the devices were shipped The serial number of the devices shipped According to the batch records and discussions with you - (b)(4) devices were manufactured for use in the study and were shipped to the clinical site. (b)(4) devices were not used and were returned to you at the end of the study. G. The study data to be used to support the primary objective of the study, evaluating the effectiveness of (b)(4) (b)(4) The protocol lists the primary efficacy endpoints as the change from Baseline to [illegible] for the following tests: Brief Pain Inventory (BPI) Pain Intensity Visual Analog Scale (VAS) Biochemical

    Shortened here; the PDF has the full text.

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.