Skip to content

DyAnsys India Pvt. Ltd.

No 33 Meerlan Towers, 1st Floor; Hanumantha Road, Chennai, India

FEI 3010350335

FDA has 3 inspections on file for DyAnsys India Pvt. Ltd. in Chennai, India (FEI 3010350335), from 2015 to 2022. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA posted the Form 483 from that inspection. FDAgent read its one observation from the PDF; it is listed below. One FOIA request has asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Form 483 PDF ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
3, 2015–2022
0 OAI · 3 VAI · 0 NAI
Inspections with 483 observations shown
1 of 3
Form 483s posted by FDA
↗
Requested under FOIA
1 request, latest Sep 29, 2025
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Bioresearch MonitoringPosted 483 · 1 observation

    Read this Form 483 on its own page →Form 483 PDF on fda.gov

    Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.

    1. Observation 1

      Failure to assure GCP compliance as sponsor

      As the sponsor of this single-site clinical trial, conducted wholly outside the United States in support of a pre-market notification submission, I found you failed to assure the study was conducted in accordance with Good Clinical Practice. Specifically, A. You could not provide documentation confirming that the clinical investigator had obtained informed consent from any of the [illegible] subjects screened for the study. [illegible] of those subjects were enrolled and received at least one device application. B. You did not assure that the investigational plan was followed. Including but not limited to: The inclusion criteria listed in the protocol requires subjects to have an average pain score of (b)(4) prior to randomization as assessed by the Brief Pain Inventory (BPI) and Visual Analog Scale for Pain (VAS). The inclusion criteria listed in the case report forms required an average pain score of [illegible] for study enrollment. The protocol (b)(4) says that subjects will be asked to keep a patient diary noting the day and date they place the study device and any adverse events experienced. Subject diaries were not used for the study. The protocol (b)(4) says that blood will be obtained for fasting blood glucose (FBG), postprandial blood glucose (PLBS), glycated hemoglobin (HbA1C) and thyroid profile. Thyroid profiles were not performed for any of the subject files reviewed. The protocol (b)(4) says that the following objective and subjective tests will be performed at all study visits: Brief Pain Inventory (BPI) - Pain Intensity Visual Analog Scale (VAS) Neuropathic Pain Score Inventory (NPSI) Neuropathic Disability Score (NDS) Semmes-Weinstein 10g Monofilament Test Vibration Perception Threshold via biothesiometer (VPT) Pinprick Sensory Threshold (PPT) Patellar Reflex Response Test (Standard Patellar Hammer) The following tests were not performed at any of the study visits after Enrollment (Baseline): Brief Pain Inventory (BPI) Neuropathic Pain Score Inventory (NPSI) Neuropathic Disability Score (NDS) Semmes-Weinstein 10g Monofilament Test Pinprick Sensory Threshold (PPT) Patellar Reflex Response Test (Standard Patellar Hammer) The Overall Neuropathy Limitations (ONLS) test was performed at all study visits, and is listed in the Statistical Analysis Plan as a secondary outcome measurement. This assessment is not listed in the protocol (b)(4) as part of any study visit. The protocol (b)(4) says all data will be entered into a validated database, and at critical junctures, data will be locked and cleaned per established procedures (b)(4) C. You did not monitor the progress of the clinical trial to ensure that the investigation was carried out consistently with the protocol, to prevent or mitigate sources of error in the conduct, collection, and reporting of critical study data, and to assure the safety of the subjects. D. You did not ensure that the Principal Investigator and Study Director were trained to comply with Good Clinical Practices or to conduct the study in accordance with the protocol. E. You did not obtain a signed investigator agreement from the Principal Investigator of the study. F. You did not maintain records of shipment and disposition of study devices including: The name and address of the study staff and study site where devices were shipped The type and quantity of the devices shipped The dates the devices were shipped The serial number of the devices shipped According to the batch records and discussions with you - (b)(4) devices were manufactured for use in the study and were shipped to the clinical site. (b)(4) devices were not used and were returned to you at the end of the study. G. The study data to be used to support the primary objective of the study, evaluating the effectiveness of (b)(4) (b)(4) The protocol lists the primary efficacy endpoints as the change from Baseline to [illegible] for the following tests: Brief Pain Inventory (BPI) Pain Intensity Visual Analog Scale (VAS) Biochemical

      Shortened here; the PDF has the full text.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Sep 29, 2025. It asked for a Form 483.

  • Inspection ending Dec 15, 2022VAI

    1 request. In the inspection history above. FDA has posted its Form 483.

    1. Sep 29, 2025FDA-FOIA-2025-11368Form 483· 1 of 3 sitesRecords released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3010350335 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3010350335. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.