Batch records and documentation in FDA Form 483 observations

Master and batch production records, device history records, logbooks and good documentation practice. FDAgent finds 241 such observations in 143 Form 483s issued to 137 facilities, for inspections that ended between October 2008 and September 2026. Each is quoted below from the 483 FDA posted.

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: Form 483 PDFs in FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC. Every quote is from a 483 whose reading passed FDAgent's checks against the scan. FDAgent is independent of FDA.

Observations
241
Each quoted below
Form 483s
143
Posted by FDA
Facilities
137
Distinct FEIs
Most recent
Sep 28, 2026
Inspection end date

Observations by year

These observations by the year the inspection ended. They follow which 483s FDA has posted and FDAgent has read, so they show what is on record, not a trend in FDA's findings. 2026 runs to the latest sync (lighter bar).

Peak: 80 in 2026

Batch records and documentation observations in Form 483s by inspection year
YearObservations
20083
20090
20101
20110
20124
20134
20140
20153
201616
201712
201815
201913
20200
20217
202218
202322
202415
202528
202680

How FDAgent sorts observations into this theme

Two fixed rules, applied by code to every observation FDAgent can quote. An observation counts when either holds. No model and no similarity score decides.

It is tied to one of these 21 CFR entries

21 CFR 211.18021 CFR 211.18421 CFR 211.18621 CFR 211.18821 CFR 820.18121 CFR 820.184

A 483 prints no regulation, so FDAgent ties an observation to a section only when it opens with FDA's own citation wording for that section, word for word.

Or its text contains one of these phrases

“batch record*” · “batch production … record*” · “master production … record*” · “master batch record*” · “device history record*” · “device master record*” · “documentation practice*” · “logbook*” · “log book*” · “BPR*” · “MPR*” · “BMR*”

* matches any ending and … up to three words in between. Case is ignored unless a phrase has capitals.

The 241 observations, newest first

Quoted word for word from each Form 483, with only line breaks joined; … marks where the quote is cut. The highlight shows why the observation is in this theme. Each links to FDAgent's page for the 483.

  1. Torrent Pharmaceuticals LimitedInspection ended Apr 16, 2019 · FEI 3005029956 · Observation 2

    … During review of the batch record for PV batch # BFR5D002, it was noted that a deviation (D/I/F/2017/0477) was initiated for the film coating (b) process. This deviation addressed the differences in (b)(4) between the two validation batches. The impact of notable differences in the (b)(4) was not thoroughly understood at the time of batch execution. The impacted batch (BFR5D002) failed for Dissolution on stability at the 3M and 6M time points and …

    Contains “batch record*”Read the Form 483 →

  2. Lupin LimitedInspection ended Feb 8, 2019 · FEI 3004819820 · Observation 1

    … The batch yield inclusive of manufacturing loss (276) did not meet the batch record specification limit of RE: to! %. Your investigation concluded that the LED spots were possibly from thel®@ due to improper ®) ©) stage. However, no qualitative test was performed to confirm that the spots were from the® 4) to rule out other contaminants. The batch was released to US market.

    Contains “batch record*”Read the Form 483 →

  3. Lupin LimitedInspection ended Feb 8, 2019 · FEI 3004819820 · Observation 2

    There are no written procedures for production and process controls designed to assure that the drug products have the identity, strength, quality, and purity they purport or are represented to possess. Specifically, As per the approved batch records, the ®/) procedure is required to be continued oe ©) 4) without a discharge process of the” @. However, during manufacturing, the®@ is discharged into smaller bins © @) procedure. Currently, the …

    Contains “batch record*”Read the Form 483 →

  4. Lupin Limited Unit IIIInspection ended Oct 18, 2018 · FEI 3009107538 · Observation 2

    … c) your batch record for [illegible] cleaning and cleaning instructions on [b)(4)] room for the operator to access.

    Contains “batch record*”Read the Form 483 →

  5. Lupin Limited Unit IIIInspection ended Oct 18, 2018 · FEI 3009107538 · Observation 3

    Appropriate controls are not exercised over computerized systems, master production and control records or other records. Specifically, a) control over the computerized system in [illegible] permits modification to the sequences in HPLC and GC. The reviewer does not review the modification to the sequence. Your QC reviewer even stated that they do not investigate modifications to the sequence during normal review and accepts the audit trail entry …

    Contains “master production … record*”Read the Form 483 →

  6. Cipla LimitedInspection ended Sep 28, 2018 · FEI 3004081307 · Observation 1

    The batch production and control records are deficient in that they do not include documentation of the accomplishment of each significant step in manufacturing. Specifically, the proposed commercial master batch record for the production of [illegible] Tablets, [illegible], version 01, does not provide written instructions for the removal of the [illegible] from a [illegible] bag inside your IPCs (In Process Containers) and the subsequent …

    Opens with FDA's citation wording for 21 CFR 211.188(b)Read the Form 483 →

  7. New Vitalis Pharmacy LLC dba New Vitalis PharmacyInspection ended Sep 7, 2018 · FEI 3014391500 · Observation 1

    … apparent food-grade Grapeseed oil supplied by [(b)(4)] [illegible], and labeled by the manufacturer as "Caution: Dietary Supplement." According to your firm's batch record, Your firm produced (b)(4) finished vials of lot AL-07192018@906 using Grapeseed Oil, lot 1801260072. For example, the shipping record shows that a vial of lot AL-07192018@906 was delivered to [illegible] on 08/22/18. Your firm produced (b)(4) finished vials of lot …

    Contains “batch record*”Read the Form 483 →

  8. Celltrion Inc.Inspection ended Jul 17, 2018 · FEI 3005241015 · Observation 1

    … The batch record states, "The debris and vials around the affected areas were removed in filling room [illegible]." The record does not state how many vials were removed or any additional details. Your firm has no written procedure to ensure the vials that have the potential to be compromised with glass particles are removed from the filling line. [illegible] In addition, the batch record for [illegible] states on March 25, 2018, at [illegible], …

    Contains “batch record*”Read the Form 483 →

  9. Celltrion Inc.Inspection ended Jul 17, 2018 · FEI 3005241015 · Observation 3

    … In addition, your firm does not document the [illegible] the sensor is set to in the batch record.

    Contains “batch record*”Read the Form 483 →

  10. Biotika ASInspection ended Jun 1, 2018 · FEI 3003370212 · Observation 1

    Completion of significant steps in batch production and control records are not documented. Specifically, A. Fermentation production records do not include complete documentation of in-process test results. For example, i. in-process test results for batch (b) (4) at the fermentation [illegible]. ii. (b) (4) in-process test result for batch (b) (4) in the fermenter at cultivation hour(s) [illegible]. ii. (b) (4) in-process test results for batch …

    Contains “batch production … record*”Read the Form 483 →

  11. Glenmark Pharmaceuticals Ltd.Inspection ended May 24, 2018 · FEI 3008565058 · Observation 2

    … ¢ sample of ®® |, subsection 5.5.2 indicates "The ®@ sample shall be collected as gram unless and otherwise mentioned in the sampling [ / process validation / batch production record. [+ bags / Glass Bottles with stoppers shall be used for collecting the | samples". For example, There is no mention of the sampling amount collected during unloading of the ®* from @@ locations. IPQA employee responsible for sample collection was unable to explain …

    Contains “batch production … record*”Read the Form 483 →

  12. Glenmark Pharmaceuticals Ltd.Inspection ended May 24, 2018 · FEI 3008565058 · Observation 4

    Appropriate controls are not exercised over computers or related systems to assure that changes in master production and control records or other records are instituted only by authorized personnel. Specifically, Your firm's manufacturing equipment are not 21 CFR part 11 compliant. For example, A) There are total standalone manufacturing equipment which are not equipped with HMI/PLC/SCADA system. There is no time stamped audit trail, data …

    Contains “master production … record*”Read the Form 483 →

  13. Galaxy Surfactants Ltd.Inspection ended Mar 30, 2018 · FEI 3004879807 · Observation 8

    8. You do not ensure product quality uniformity from batch to batch such that your Batch Manufacturing Records (BMR) does not conform to the Master Formula Record (MFR). You revised your BMR (BMR-[illegible] 001-01, Effective date 11/25/2017) for Crude [illegible] outside the MFR (MFR 001-00, Effective date 8/4/2017) and did not update your MFR. This BMR was used to manufacture an intermediate for multiple APIs that were shipped to US including; …

    Contains “BMR*”Read the Form 483 →

  14. Plasma Industries Belgium SCRLInspection ended Mar 16, 2018 · FEI 3003208230 · Observation 6

    There is no written justification for the clarifying (b)(4) limits of (b)(4) as stated in Document (b)(4) 2261 - Appendix 1: Production File 2262 / 14.0 entitled (b)(4) which is a portion of the (b)(4) batch record.

    Contains “batch record*”Read the Form 483 →

  15. Aurobindo Pharma Ltd.Inspection ended Feb 20, 2018 · FEI 3008461619 · Observation 9

    Appropriate controls are not exercised over computers or related systems to assure that changes in master production and control records or other records are instituted only by authorized personnel. Specifically, master batch records that are submitted in“ for | injection, have different batch codes (document number) than the batch records that are currently in use for commercial production. There are no traceability for the changes made to the …

    Contains “master production … record*”Read the Form 483 →

  16. Suven Life Sciences Ltd.Inspection ended Feb 15, 2018 · FEI 1000623515 · Observation 1

    Appropriate controls are not exercised over computer or related systems to assure that changes in master production and control records or other records are instituted only by authorized personnel. Specifically, A. Your firm has a total of (b) (4) HPLCs and (b) (4) UPLCs used to analyze raw material, finished product, and related substances for products marketed and distributed to the US. The HPLCs and UPLCs use either Empower 3 or EZ CHROM …

    Contains “master production … record*”Read the Form 483 →

  17. Suven Life Sciences Ltd.Inspection ended Feb 15, 2018 · FEI 1000623515 · Observation 2

    … The status of this equipment was identified as cleaned and a visual inspection was documented in the Equipment Use Logbook. C. (b) (4)-3137: I observed what appears to be (b) (4) residue and black residue alongside the sides of this (b) (4). In addition, I observed the (b) (4) bolts located on the bottom of the (b) (4) that comes in direct contact with drug products to have a (b) (4) discoloration. I observed a hole at the bottom of the (b) (4) …

    Contains “logbook*”Read the Form 483 →

  18. Suven Life Sciences Ltd.Inspection ended Feb 15, 2018 · FEI 1000623515 · Observation 3

    Manufacturing Batch Records are deficient. Specifically, A. On 13 February 2018, your firm failed to document in-process checks during the manufacturing of the drug product, (b) (4) Tablets, Batch #(b) (4) on Form, F/CQO-F/320, as outlined in Step (b) (4) of the Batch Manufacturing Record. However, this step was verified by your firm's Quality Assurance. B. Recorded weights are not verified by a second verifier.

    Contains “batch record*”Read the Form 483 →

  19. Divi's Laboratories Ltd. Unit IIInspection ended Sep 19, 2017 · FEI 3004149463 · Observation 2

    … Your documented manufacturing investigation only verified that there was no deviation being reported in the batch records and operations were followed. There was no thorough investigation into manufacturing, process, maintenance, and engineering activities and involved the relevant departments that may have attributed to batch contamination. Original results were invalidated without clear identification of laboratory error and adequate …

    Contains “batch record*”Read the Form 483 →

  20. Divi's Laboratories Ltd. Unit IIInspection ended Sep 19, 2017 · FEI 3004149463 · Observation 6

    OBSERVATION 6 Batch records are not always recorded contemporaneously at the completion of an operation. For example: On 9/12/2017 during the production of [illegible] batch the in-charge production employee had signed the "CHECKED BY" column at step [illegible] in the manufacturing record to increase the batch [illegible] to [illegible]. However the "DONE BY" column was not signed by the chemist and the operation was not completed. On 9/12/2017 …

    Contains “batch record*”Read the Form 483 →

21 CFR sections these observations open with

The sections FDAgent ties the most of these observations to by FDA's wording, with how many of this theme's observations each covers. Each links to its page of FDA citations, warning letters and the regulation text.

Ask FDAgent about batch records and documentation in Form 483s

FDAgent reads FDA's Form 483s, inspections, citations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The observations are the text FDAgent read by OCR from Form 483 PDFs in FDA's Electronic Reading Room, shown only from 483s whose every printed observation passed FDAgent's checks against the scanned text, as on each 483's own page.
  • Which theme an observation belongs to is decided by the fixed rules above; every count on this page is computed from them.
  • A 483 addressed to an individual is listed without the name, and an observation that names a person is left out.
  • FDAgent is independent of FDA.